NEUATION iROLL PR35 Plate Rotator User Manual
- June 12, 2024
- NEUATION
Table of Contents
- PRODUCT DESCRIPTION
- SAFETY PRECAUTIONS
- INSTALLATION
- STANDARD PAST LISTING
- USER INTERFACE & DISPLAY
- REMOVING & li4STALLING PLATE
- Operation
- REMOVAL OF TUBES FROM PLATE
- MAINTENANCE & CLEANING
- WARRANTY STATEMENT
- PRODUCT DISPOSAL
- CUSTOMER SUPPORT
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
NEUATION iROLL PR35 Plate Rotator
PRODUCT DESCRIPTION
INTRODUCTION
This digital plate rotator is equipped with a long life BLDC motor with a
small foot print.
The rotating motion provides gentle rocking and offers pulse function for
thorough mixing.
Set up to 9 user-defined programs with the digital display for the accurate
and easy setting.
The removable plate can be set to suit tubes of different diameters.
The plate can be easily removed and replaced for cleaning.
NOTE: Before using the instrument, please read this user manual
carefully.
This user manual is Intended to assist with he operation and core of the unit
only and not Its repair.
For repair contact the supplier.
INTENDED USE
This plate rotator is used in Laboratories for mixing blood samples, viscous
samples and for liquid-solid suspensions.
Symbol
| Warning
---|---
| Waste electrical
STANDARD ACCESSORIES
- Power Adaptor
- Product user manual & warranty card
FEATURES
- Gentle rotating motion for optimum mixing
- Adjustable speed ranging from 10 to 80 RPM
- Timer ranging from 1min to 99 mins and infinite mode
- Maintenance free brushless DC motor for quiet operation and long life
- Removable plate for proper cleaning & better contamination protection
- Optional plates to hold different diameter tubes
- Pulse mode function for alternating clockwise and anticlockwise motion with 2 sec stops for better and thorough mixing
- Option for auto power on in case of power failure – on resumption it will start automatically and run for the remaining time
- Digital display for easy setting of values
- Small footprint with the best use of lab space
- Suitable for mixing blood samples and viscous substances
- Closed loop operation for exact speed control under varied load conditions
- Low voltage instrument of 24VDC
- Works With 80% humidity and 40 °C
TECHNICAL SPECIFICATIONS
Motor Type | Brushless DC |
---|---|
Motion Type | Rotating |
Speed Range | 10 to 8o RPM |
Speed Accuracy | 13 RPM |
Speed steps | +1 RPM |
Timer range | 0 to 99 mins & infinite mode |
Max Loading Capacity | 1.5 kg |
Removable attached plate | Yes |
Dimensions (mm) | 540 x 274 x 186 |
Protection class | IP 21 |
Permissible Ambient Temp | 5“C to 40 “C |
Permissible Relative humidity | s80% |
Weight | 4.5 Kg |
Altitude | Use upto an altitude of 2000 rr above MSL |
Pollution Degree | 2 |
Environment | For indoor use only |
Power Adapter Details | Input: 100 – 240 VAC, 50/ 60 Hz |
Output: 24 1.5A
Power Consumption| 4.S W
SAFETY PRECAUTIONS
Read all safety & usage Information provided in this manual carefully before using the device.
-
Never use the instrument in any manner not specified in this manual.
-
Equipment used in any manner not specified in this manual or by the manufacturer might result in the lapse of warranty.
-
Do not try to stop the DISC plate by blocking/holding It.
Do not attempt to remove DISC plate during operation. -
Never move the instrument while the equipment is under operation.
-
Repairs must only be performed by an authorized service technician.
-
Only use recommended original spare parts for best result & product safety.
-
If liquids are spilled on the DISC plate, the instrument must be cleaned carefully and properly before being used again.
-
Prior to mixing, the tubes should be visually Inspected for material damage.
Damaged tubes must not be used as this can result in sample loss and the contamination of the product. -
The maximum capacityof2 kg must not be exceeded.
-
This product must be used for specified applications only.
It must not be operated in a hazardous or flammable environment and must not be used to mix explosive or highly reactive substances. -
Do not place the potentially hazardous material within the clearance area/ envelope.
-
Wear your personal protective equipment In accordance with the hazards category of the medium to be processed.
There is risk of liquid splashing. -
Place the instrument on a flat, leveled and solid surface.
Make sure the feet of the instrument are clean and undamaged. -
Load the tubes safely and evenly.
Always distribute the samples evenly on the device. -
Do not move the instrument while it is operating or connected to the main power
-
Ensure that the knob is firmly tightened before performing any operation.
-
Ensure that only closed tubes are used for mixing
-
Do not lean on the equipment. It may damage the equipment or harm the operator.
-
Do not fill tubes near the device. Liquid spillage may harm the device.
-
In the event of contamination Caused by aggressive agents, the device must be cleaned immediately using a natural cleaning liquid.
If any damage is seen, contact the service technician. -
The power adaptor given with the unit is designed to be used for this particular unit.
Do not use any other power adaptor. -
Only process media which will not react dangerously to the extra energy produce by mixing.
-
Do not operate the appliance in the explosive atmosphere with hazardous substance or under water.
INSTALLATION
Gently remove the upper packaging and take out the plate rotator by holding It
from the bottom.
When this equipment is used for the first time, ensure that all the packaging
accessories are removed from the product.
Please keep all the packaging in safe storage for at least2 years for warranty
purposes.
LOCATION & MOUNTING
Place the device on a flat and leveled surface and ensure that all four feet
stand on the surface firmly.
Avoid installing on a slippery surface or surface prone to vibration.
- Ideal ambient temperature is 2O”C + S°C; avoid placing the device indirect sunlight.
- Keep clearance of at least 3 cm from all sides to guarantee cooling efficiency.
- Keep this away from heat or water to avoid sample temperature issues.
- Do not place the device in such a place that it becomes difficult to operate.
CONHECTING THE POWER ADAPTOR
- Connect one side of the power adaptor to the rear side of device and other æ the power supply as shown in the Figure below.
- ensure the power switch is OFF- while connecting the power adaptor.
STANDARD PAST LISTING
Front Side
Back Side
USER INTERFACE & DISPLAY
No. | Description |
---|---|
1. Display | The display shows the set speed as well as the time value. |
It also displays selected Program when in Prog Mode
2. Speed| Press ’Speed’ to select speed value and use increment ar
decrement button to set require speed
3. Time| Press ’Time to select time value and use increment or
decrement button to set require run time
4/5. Increment /Decrement| Press Increment / Decrement to set require
value of speed, time or program (Prog Mode)
---|---
6. Stan/5top| Press * Start/Stop ” to start the operation or to stop
the ongoing operation.
7. Pulse & LED| Press ’Pulse’ once to start pulse operation and press it
again to stop operation.
Pulse LED will glow to indicate Pulse is activated.
8. Prog| Hold down ’Prog” to enter in program mode and use other
require buttons to set parameter for selected program
REMOVING & li4STALLING PLATE
REMOVING PLATE
Plate can be removed / changed by turning the knob in counter clockwise
direction.
Turn the screw in counter clockwise direction and then remove the plate as
shown in the below representational image:
INSTALLING Plate
Follow the exact reverse process to insert back the p1ate.
Below image ‹howl the step by step procedure to insert the plate back into the
unit.
Operation
Powering The Device On
After connecting the power adaptor to back side of the device, switch the main
power supply to “ON” & then switch the device on from the rear.
Make sure that tubes are loaded properly before use.
SPEED SETTING
Press ’Speed’ to select speed value and then press ’Increment / Decrement to
set required speed.
The speed for operation ranges from 10 R PM to 80 R P M with
incremental/decremental steps of 1 RPM. Speed values are saved after blinking
5 times.
TIME SETTING
Press “Time” to select time value and use increment or decrement button to set
run time.
Run time for operation can be set between 1 minute to 99 minutes with an
Incremental/decremental step of 1minute.
It can also be set for continuous or infinite mode indicated by )( in time
display.
The value will be saved after 5 blinks.
The countdown timer will get activated as soon as the operation starts.
Time cannot be modified when the unit is in operation.
START / STOP DPERATION
Once the required speed and time are set, press “Start / Stop” to START
operation.
Speed gradually increases from lower RPM to set RPM and time takes effect as
soon as the operation starts.
Press “Stan /Stop“ again to STOP the ongoing operation at anytime. Speed
gradually decreases from set RPM to zero (0) and stops in a few seconds.
NOTE: Speed and time cannot be changed/modified during operation.
DEFAULT PULSE I\4ODE
Pulse mode is used for vigorous mixing of samples by rolling with a momentary
clockwise and counter clockwise pause every 30 seconds.
Default cycle time set for pulse mode is 3Dseconds.
Set required speed and total run time before activating pulse mode to guide
the device to start pulse mode operation for that particular speed and total
run time.
Once the speed and total run time are set, press “Pulse Button” to START pulse
mode operation for that particular speed and total run time.
In pulse mode, pulse LED will glow indicating pulse mode operation is
activated.
Example: Speed set to 50 RPM and total run time set to 2 minute.
There will be rotating motion in clockwise direction for 30 seconds (cycle
time).
After completion of 30 seconds.
disc plate mill have a 2 cycles of momentary pause in clockwise and counter
clock wise direction.
Post these 2 cycle, disc plate will have a rotating motion for 30 seconds in
counter clockwise direction.
This entire process Press “Pulse” again to stop ongoing pulse mode operation.
User can also Press “Start/Stop ” to STOP ongoing pulse mode operations.
PROGRAM MODE
The program mode operation feature allows the user to save 9 programs which
can have different speed and time parameters.
Users can use any of these programmed parameters by simply tuning on the
program mode.
To turn the program mode on, hold down “Prog” until “Prog on” displays
indicting program mode is activated.
After a few seconds, the display will. show “P”1” and “”speed & time value”
alternatively.
This shows that the current program is P1 with some RPM and time value.
Press the “Program” button to Select any program from “P1ta P9”.
Programs can also be selected by pressing “increment or decrement” buttons,
Once the desired program
is selected, press the “speed button” to select and set speed value.
Press “Time” to select and set time value.
Speed and time value for that particular program is now saved.
Press the “Start Button” to start the program mode operation.
Once set RPM is reached, the program “Px” (with x = 1to g) and its “Speed” &
“Time” value will display alternatively.
To stop operation, press “Stop”.
To leave Program mode, hold down the “Program” button again to turn OFF the
Programmed, It will display “Prog off ” indicating program mode is off,.
Note: On every restart, the device automatically comes out pf program
mode.
It shows the last run default speed and time value.
POWER DIODE
The function of this particular feature is that in case of power failure,
equipment saves the ȘPEED, TIME and. other details as set by the user.
When the ”power returns, equipment runs for the remaining time left with cycle
time of +/- 5 mins. Equipment.
Saves the run time every 5 minutes of operation.
To select this feature, long press “TIME“ button.
Unless and until this feature is not switched on, plate rotator will not
resume its operation after power failure.
Example: User has set RPM of 45,time of 35 mins.
After completion 23 mins, there was power cut and operation comes to a stand
still. just before the power cut, remaining the seen on the display was 23
mins.
when the power returns, the plate rotator will assume its operations from 25
mins with the set RPM and other operational values.
REMOVAL OF TUBES FROM PLATE
In order to remove tubes from the plate, do not try to rotate the plate by
bare hands.
Procedure to remove the tubes attached at the bottom Of the plate is
mentioned below:
- Single press PULSE button to rotate the plate so that the tubes at the bottom of the plate comes at the top.
- Now pull out the tubes from the clamps of the plate one by one.
- Similarly, follow the above process to remove the tubes In order to avoid damage to the plate & tubes.
MAINTENANCE & CLEANING
Warning: Ensure the unit is disconnected from the power supply before attempting maintenance or cleaning.
- Clean the unit by a soft damp cloth or mild non-corrosive (pHs8) detergent and do not spray the unit.
- Wear clean when cleaning and be sure that the unit completely dry before operation.
- Cleaning is made easier If spillages are attended promptly.
- In any case, spillages of acids and alkali must be removed immediately as these chemicals can attack and damage the casework finish.
- Ensure that the appropriate safety precaution is observed.
WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship for a period of two (2) years from the date of purchase. Your product will be duly repaired upon prompt notification In compliance with the following conditions:
This warranty is valid only if the product is used for its Intended purpose
and within the guidelines specified in this instruction manual.
This warranty does not cover damage caused by accident, neglect, misuse,
improper service, natural forces or other causes not arising from defects in
original material or workmanship.
This warranty does not cover any incidental or consequential damages,
commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will
immediately terminate all liabi1ities on us for the products or damages caused
by its use.
The customer shall be responsible for the product or use of products as well
as any supervision required for safety.
If requested the products must be returned to the distributor in well packed
and insured manner and all shipping charges must be paid.
NOTE: Some states donor allow limitation on the length of implied
warranties or the exclusion or limitation of incidental or consequential
damages.
This warranty is given you specific legal rights.
This warranty is given expressly in lieu of all other warranties, expressed or
implied.
The purchaser agrees that there is no warranty of merchantability or of
fitness for any intended purpose and there are no other remedies or
warranties, implied, which extend beyond the description on the face of the
agreement.
This warranty is only applicable to the original purchaser.
Product received without proper authorization will not be processed for
warranty or service.
All items returned for service should be sent with postage prepaid in the
original packaging or other suitable packaging, padded to avoid damage.
We will not be responsible for damage incurred by improper packaging.
Note: This warranty is valid only if the warranty is registered with the
supplier within 30 days from the date of purchase.
For you reference, make a note of serial number, date of purchase and supplier here.
Serial No.: Purchase Date:
Supplier:
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are
to be observed.
Information on the disposal of electrical and electronic devices in the
European Community
The disposal of electrical devices is regulated within the European Community
by national regulations based on EU Directive 2012/19/EU on waste electrical
and electronic equipment {WEEE).
According to these regulations, any devices supplied after 13.06.05 in the
business to business sphere, to which this product is assigned, may no longer
be disposed off in municipal or domestic waste.
They are marked with the following symbol to indicate this.
As disposal regulations with in the EU may vary from country to country,
please contact your supplier if necessary.
CUSTOMER SUPPORT
Accumax Lab Devices Private Limited
Plot No. 15,16 & 32 GIDC Electronic Park SEZ; Sector 26,.Gandhinagar, Gujarat,
India – 382026.
Website: www.neuation.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>