PCE Instruments PCE-CS xxxxLD Crane Scales User Manual
- June 12, 2024
- PCE Instruments
Table of Contents
PCE Instruments PCE-CS xxxxLD Crane Scales User Manual
Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.
- The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- Since lifting loads is associated with hazards, certain obligations must be fulfilled by the manufacturer as well as the user in order to prevent these hazards. The use of the measuring equipment must not constitute a hazard at any time.
- The permissible nominal loads [MAX] of the test equipment and other means used (e. g. crane, load sling, etc.) must not be exceeded under any circumstances.
- When not in use, no loads must be attached to the scales.
- Before using the scales, make sure that all brackets are firmly mounted.
- Make sure that there are no persons under the suspended load.
- The measuring device is only permitted for lifting and weighing freely moving loads. Twisting the load (torsional stress) is not permitted. Transporting persons, pulling loads at an angle, tearing them loose, dragging them etc. is prohibited.
- Installation, commissioning, maintenance and operation of the measuring equipment may only be carried out by trained persons wearing appropriate protective equipment.
- Never touch the moving parts during the weighing process. This can lead to injuries.
- Do not enter the danger zone during use. This does not only include the area directly below the load but also areas that could become a hazard, e. g. due to swinging ropes or other load suspension equipment.
- Non-observance of the safety notes can cause damage to the device and injuries to the user.
We do not assume liability for printing errors or any other mistakes in this manual.
We expressly point to our general guarantee terms which can be found in our general terms of business.
Specifications
PCE-CS 300LD
Measurement range| 0 … 300 kg / 0 … 661 lbs
Minimum load| 2 kg / 4.4 lbs
Resolution| 0.05 kg / 0.1 lbs
Accuracy| 1 % of range
PCE-CS 500LD
Measurement range| 0 … 500 kg / 0 … 1102 lbs
Minimum load| 4 kg / 8.8 lbs
Resolution| 0.1 kg / 0.2 lbs
Accuracy| 1 % of range
PCE-CS 1000LD
Measurement range| 0 … 1000 kg / 0 … 2205 lbs
Minimum load| 10 kg / 22 lbs
Resolution| 0.2 kg / 0.4 lbs
Accuracy| 1 % of range
Further specifications
Additional functions| totalising function, user-defined unit
Tare range| 100 % of range
Zero point range| 4 % of range
Units| kg, lb, un (user defined)
Digit height| 41 mm / 1.6″
Automatic power off of the backlight| after 30 seconds of inactivity
reactivation after pressing a key or when the weight is changed
Power supply crane scales| 4 x 1.5 V AA battery
Power supply remote control| 2 x 1.5 V AA battery
Operating conditions| 5 … 35 °C / 41 … 95 °F<90 % RH, non-condensing
Weight| 1300 g / 2.9 lbs
Delivery scope
- 1 x PCE-CS xxxxLD crane scales
- 1 x remote control
- 6 x 1.5 V AA battery
- 1 x user manual
Dimensions
Letter | Length in mm |
---|---|
A | 290 |
B | 23 |
Device description
No. | Description |
---|---|
1 | Shackle |
2 | Display |
3 | Hook |
No. | Description |
---|---|
1 | Battery compartments |
2 | Nameplate |
3 | Keypad |
No. | Function |
---|---|
1 | Press: cancel or return to weighing mode Press and hold: switch on and off |
2 | Press: freeze displayPress and hold: change unit |
3 | Press: tare / disable taring Press and hold: zero point |
Remote control
No. | Description |
---|---|
1 | Zero point, increase value |
2 | Totaliser function view |
3 | Totalised weight, decrease value |
4 | Change unit |
5 | Press: back to weighing mode Press and hold: switch off the scales |
6 | Secondary function level |
7 | Delete last weight, go one step to the left |
8 | Freeze display, accept entry, confirm entry |
9 | Tare / disable taring, go one step to the right |
Display description
Possible display | Meaning |
---|---|
STB | Displayed measured value is stable |
Zero | Displayed measured value is the zero point |
Tara | Displayed measured value is tared |
HOLD | Displayed measured value is frozen |
lb | Measured value in pounds |
kg | Measured value in kilogrammes |
UN | Measured value is displayed in a user-defined unit |
Indications
Possible display | Meaning |
---|---|
– – – – – | Weight detected |
Setup | Settings |
Bat99 | Battery indicator in per cent |
End | Save and return |
oFF | Shutdown |
ouEr | Overrange |
Err | Invalid action |
ACC | Totalising (accumulation) |
DEL | Delete data |
CLR | Delete all data |
Operation
Switch scales on and off
To switch on the scales, press and release the key once. The scales are powered on and a start-up process begins.
- “8.8.8.8.” flashes twice
- maximum capacity display
- display of the battery level in per cent
- measurement begins
If – – – is displayed during the start-up process, a weight has been detected and must be removed.
To switch off the scales, press and hold the key on the remote control for about one second or press the key on the scales. The scales are switched off with the following display:
- Display of the battery level in per cent
- oFF
- The scale is now switched off
Note : It is not possible to switch on the scales via the remote control. When the total battery voltage is 3.75 V, the scales are switched off automatically because the battery voltage is too low.
Zero point
To set the zero point, press the key on the remote control or hold down the key on the scales.
Note : If the attached load is in motion, tared or outside the manual zero range, “Err” is displayed.
Taring
To tare the weight, press the key on the remote control or press and release the key on the scales once. To delete the tared weight, press the key again.
Note: If the suspended load is in motion, negative or out of the tare range, “Err” is displayed.
Hold function
To freeze the display, press the key on the remote control or the key on the scales. The display will also show “HOLD” next to the held reading. To resume the measurement, press the key again.
Totalising weights (accumulation)
Press the key on the remote control to add up the displayed weight. Once the value has been accepted, “ACC” is briefly displayed once.
Note : If the suspended load is not stable, negative or the load has not been suspended, “Err” is displayed.
Display the total weight
To display the sum of the weights, press and release the F1 key on the remote control once. The total is displayed flashing. Press the key on the remote control to display the highest or lowest totalised value. To return to measuring mode, press and release the key once.
Note: It is not possible to add further values in this view. To do this, you must go back to measuring mode
Delete last measured value
Press the key on the remote control to delete the last totalised value. If “dEL” is displayed, the last value has been deleted.
Note : Only the last value can be deleted.
Delete all saved values
To delete all saved values, first press the key on the remote control and then
press the key.
As soon as “CLR” is shown on the display, all values are deleted.
Change unit
To set the unit, press the F2 key on the remote control or briefly press and
hold the key on the scales. You can choose kg, lborun.
Settings
To make settings to the scales, the remote control is mandatory. To open the
settings menu, first
press the key and then the key. “SEtUP” is displayed.
Now select the desired menu item with the key. Use the and keys to change the value. Press the key again to accept the change.
Menu structure
Display | Function | Possible settings |
---|---|---|
Here you can set the time after which the scales switch off automatically | ||
in case of inactivity. | 0 = off (default) 5 = 5 minutes10 = 10 minutes15 = 15 |
minutes30 = 30 minutes60 = 60 minutes
| The brightness of the backlight can be set here.| Off = off1 = dimmed2 =
normal3 = bright
****| Here you can set the time during which the backlight is active.| 0 =
off5 = 5 seconds10 = 10 seconds15 = 15 seconds30 = 30 seconds (default)60 = 60
seconds
End| You have reached the end of the menu.| No settings possible. The scales
automatically return to weighingmode.
Advanced settings
Note:
The advanced settings should only be used by advanced users as incorrect
parameterization can cause problems.
Changing the parameters does not change the given specifications.
To open the advanced settings, press the key twice. “P0000″ is then displayed. Use the arrow keys on the remote control to enter the password P0258 and confirm the entry with the key.
The scales now display “SCALE”. The operation of the advanced settings is identical to the operation of the normal settings.
Menu structure
Display | Function | Possible settings |
---|---|---|
The display resolution can be set here. | 0.001 / 0.002 / 0.005 / 0.01 / |
0.02 /0.05 / 0.1 / 0.2 / 0.5 / 1 / 2 / 5 / 10 /20 / 50
| The automatic zero point range can be set here.| 0 = off2 = ±2 % of range3 =
±3 % of range4 = ±4 % v. range10 = ±10 % of range20 = ±20 % of range (default)
100 = ±100 % of range
| The manual zero point range can be set here.| 0 = off2 = ±2 % of range3 = ±3
% of range4 = ±4 % v. range (default) 10 = ±10 % of range20 = ±20 % of
range100 = ±100 % of range
| The zero tracking range can be set here. This function can help to maintain
accuracy in the event of temperature fluctuations.| 0.0 = off0.5 = ±0.5e
(default) 1.0 = ±1.0e1.5 = ±1.5e2.0 = ±2.0e2.5 = ±2.5e3.0 = ±3.0e3.5 =
±3.5e4.0 = ±4.0e4.5 = ±4.5e5.0 = ±5.0e
| The zero point range can be set here. This function defines the range in
which the measured value must be lower than the set value before a measurement
is possible.| 0.0 = off 0.5 = ±0.5e 1.0 = ±1.0e 1.5 = ±1.5e 2.0 = ±2.0e 2.5 =
±2.5e 3.0 = ±3.0e 3.5 = ±3.5e 4.0 = ±4.0e 4.5 = ±4.5e5.0 = ±5.0e (default)
| This function measures the weight based on the last zeroing. When this
function is active, automatic zeroing is deactivated.
| On = switched on
Off = switched off (default)
---|---|---
| Here you can set the measuring rate.
If the weight is unstable, a high value should be selected. If the attached weight is stable, a low value can be
selected.
| 0 = off
1 = weakest (default) 2 = weak
3 = normal
4 = strong
5 = strongest
| Dynamic weighing can be enabled here. If the weighing accuracy is of secondary importance and the speed of a measurement is crucial, the “STB” function can be
permanently activated here.
| On = on
Off = off (default)
| Here you set the gravitational value of the place of use after a calibration.| 0000 … 9999
You can change the value with the arrow keys on the remote control.
You can move the decimal point with the F1 key.
****| Here you can store your own unit via a factor.| 0000 … 9999
You can change the value with the arrow keys on the remote control.
You can move the decimal point with the F1 key.
Adjustment
If the scales show incorrect values, it can be adjusted. An adjustment should only be carried out by an advanced user. Make sure in advance that you have calibrated weights available to achieve an accurate result.
Note : Changing parameters does not change the specifications.
To open the adjustment menu, press the key twice. “P0000″ is then displayed. Enter the password P8416 using the arrow keys on the remote control and confirm the entry with the key. The scales now display “CAL”. The operation of the advanced settings is identical to the operation of the normal settings.
Display | Function | Possible settings |
---|---|---|
Here you can set the unit with which the calibration is to be carried out. |
kg, lb
| Here you set the gravitational value of the calibration location.| 0000 …
9999You can change the value with the arrow keys on the remote control. You
can move the decimal pointwith the F1 key.
| The maximum capacity can be set here.| 0000 … 9999You can change the value
with the arrow keys on the remote control. You can move the decimal pointwith
the F1 key.
| Here you set the zero point.| Then press the key.Let the scales hang
freely and wait until the displayed digit value hasstabilised. Then press the
key.
| Here you can set the first calibration point.| 00000 … 99999You can change
the value with the arrow keys on the remote control. You can set the decimal
point with the F1 key.Then press the key to accept the value. The digit
value is now displayed. Attach the set weight and wait until the value has
stabilised.Then press the key.
| Here you can set the second calibration point.| See previous step.
| Here you can set the third calibration point| See previous step.You have
now completed thecalibration and the scales return to weighing mode.
Battery
The battery status is displayed each time the scales are switched on and off. As soon as the batteries are flat, they must be replaced with new ones. For battery replacement, there are two battery compartments on the back of the scales that can be opened with a screwdriver. Two 1.5 V AA batteries must be inserted in each battery compartment. Ensure correct polarity. After inserting the batteries, the battery compartment must be closed. The scales can then be used again.
There is an LED on the remote control. It should flash when a key is pressed. If it does not flash, the batteries are probably flat and need to be replaced. The battery compartment is on the back. Use two 1.5 V AA batteries. Ensure correct polarity when inserting them
Contact
If you have any questions, suggestions or technical problems, please do not hesitate to contact us. You will find the relevant contact information at the end of this user manual.
Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste.
They must be given to collection points designed for that purpose.
In order to comply with the EU directive 2012/19/EU we take our devices back. We either re-use them or give them to a recycling company which disposes of the devices in line with law.
For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
If you have any questions, please contact PCE Instruments.
PCE Instruments contact information
Germany
PCE Deutschland GmbH
Im Langel 26
D-59872 Meschede
Deutschland
Tel.: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29
info@pce-instruments.com
www.pce-instruments.com/deutsch