pico Technology TA506 PicoBNC+ 10:1 Attenuating Lead User Guide
- June 12, 2024
- pico Technology
Table of Contents
TA506 PicoBNC+ 10:1 Attenuating Lead
User Guide
Introduction
This passive high-impedance attenuating lead is designed for use with Pico Technology PicoBNC+ automotive oscilloscopes. It can be used to measure electrical signals up to the limits specified in this guide.
Warranty
Pico Technology warrants this oscilloscope accessory for normal use and
operation within specifications for a period of one year from date of shipment
and will repair or replace any defective product which was not damaged by
negligence, misuse, improper installation, accident or unauthorized repair or
modification by the buyer. This warranty is applicable only to defects due to
material or workmanship. Pico disclaims any other implied warranties of
merchantability or fitness for a particular purpose. Pico will not be liable
for any indirect, special, incidental, or consequential damages (including
damages for loss of profits, loss of business, loss of use or data,
interruption of business and the like), even if Pico has been advised of the
possibility of such damages arising from any defect or error in this manual or
product.
Disposal
Your cooperation is required to help protect our environment. Therefore either
return this product at the end of life to the manufacturer or ensure WEEE-
compliant collection and treatment yourself. Do not dispose of as unsorted
municipal waste.
Safety information
To prevent possible electrical shock, fire, personal injury, or damage to the
product, carefully read this safety information before attempting to install
or use the product. In addition, follow all generally accepted safety
practices and procedures for working with and near electricity.
The product has been designed and tested in accordance with the European
standard publication EN 61010-031 (Hand-held probe assemblies) and left the
factory in a safe condition.
The following safety descriptions are found throughout this guide:
A WARNING identifies conditions or practices that could result in injury
or death.
A CAUTION identifies conditions or practices that could result in damage
to the product or equipment to which it is connected.
Symbols
These safety and electrical symbols may appear on the product or in this
guide:
Symbol | Description |
---|---|
Direct current | |
Earth (ground) terminal |
Terminal can be used to make a measurement ground connection. The terminal is
NOT a safety or protective earth.
| Possibility of electric shock
| Caution
Appearance on the product indicates a need to read these safety and operation
instructions.
| Do not dispose of this product as unsorted municipal waste.
WARNING
To prevent injury or death, use the product only as instructed and use only
accessories that have been supplied or recommended. Protection provided by the
product may be impaired if used in a manner not specified by the manufacturer.
Maximum input ratings
The table and frequency derating plot below indicate the full-scale
measurement range and overvoltage protection range for these probes. The full-
scale measurement ranges are the maximum voltages that can be accurately
measured by the probe. The overvoltage protection ranges are the maximum
voltages that will not damage the lead.
WARNING
To prevent electric shock, do not attempt to measure voltages outside of the
specified full-scale measurement range.
Attenuation| Full-scale measurement range| Overvoltage protection
range
---|---|---
10:01| 400 V (DC + peak AC)| 400 V (DC + peak AC)
WARNING
Signals exceeding the voltage limits in the table below are defined as
“hazardous live“ by EN 61010.
Signal voltage limits of EN 61010-031
±60 V DC| 30 V AC RMS| ± 42.4 V pk max.
To prevent electric shock, take all necessary safety precautions when working
on equipment where hazardous live voltages may be present.
WARNING
These leads do not carry a measurement category rating. To prevent electric
shock, do not connect to a utility power (line) voltage or a derived voltage
that can carry the overvoltage transients that may be present. Measurement
categories are defined in EN 61010-031 as follows:
No measurement category (not in CAT II, CAT III, or CAT IV)
Definition| For measurements performed on circuits not directly connected to a utility power supply.
WARNING
To avoid overloading the lead, note that its maximum input voltage rating
decreases as the frequency of the applied signal increases.
Voltage rating (V peak max.) v frequency curve
CAUTION
Do not exceed the voltage rating marked on any accessory. If an accessory is
not marked with a voltage rating on either the connector, cable or body, or if
a protective finger guard is removed, do not exceed the EN 61010 “hazardous
live” limits above.
WARNING
To prevent injury or death, do not use the lead or an accessory if it appears
to be damaged in any way, and stop use immediately if you are concerned by any
abnormal operations.
CAUTION
Exceeding the voltage rating of any cable, connector or accessory can cause
permanent damage to the lead and other connected equipment.
WARNING
When connecting one or multiple accessories and a probe assembly together, the
lowest voltage rating in the interconnected set applies.
Grounding
WARNING
The lead‘s ground connection through the oscilloscope is for measurement
purposes only. The lead does not have a protective safety ground.
Never connect the ground input to any electrical power source. To prevent
personal injury or death, use a voltmeter to check that there is no
significant AC or DC voltage between the lead ground and the point to which
you intend to connect it.
CAUTION
Applying a voltage to the ground input is likely to cause permanent damage to
the lead and other connected equipment. It is good practice to connect the
lead output to the measurement instrument and the ground lead to earth ground
before connecting the lead to the circuit under test. Disconnect the lead
input and the ground lead from the circuit under test before disconnecting the
lead from the measurement instrument.
External connections
CAUTION
Take care to avoid mechanical stress or tight bend radii for all connected
leads, including all coaxial leads and connectors. Mishandling will cause
deformation and will degrade performance and measurement accuracy.
Environment
WARNING
To prevent injury or death, do not use in wet or damp conditions or near
explosive gas or vapor.
CAUTION
To prevent damage, always store the device in appropriate environments.
| Storage| Operating
---|---|---
Temperature| –20 to 80 °C| 0 to 80 °C
Humidity| 5 %RH to 95 %RH noncondensing| 5 %RH to 80 %RH noncondensing
Altitude range| Up to 2 000 m
Pollution degree| 2. As defined in EN 61010-031. Only non-conductive
pollution. Occasionally, however, a temporary conductivity caused by
condensation must be accepted.
Care of the product
The lead contains no user-serviceable parts. Repair, servicing and calibration
require specialized test equipment and must only be performed by Pico or an
approved service provider. There may be a charge for these services unless
covered by the Pico one-year warranty.
Inspect the lead and all connectors, cables and accessories before use for
signs of damage.
WARNING
To prevent electric shock do not tamper with or disassemble the probe, case
parts, connectors or accessories.
CAUTION
When cleaning the product, use a soft cloth and a solution of mild soap or
detergent in water. To prevent electric shock, do not allow liquids to enter
the probe casing, as this will compromise the electronics or insulation
inside.
Avoid mechanical shock to the probe in general to guarantee accurate
performance and protection.
Specifications
Probe characteristics | TA506 |
---|---|
Attenuation ratio | 10:01 |
Bandwidth | 20 MHz |
Rise time (calculated) | 17.5 ns |
Input resistance | 10 MΩ ±2% |
Input capacitance | 33 pF ±1 pF |
Max. working voltage | 400 V pk |
Total length | 2.85 m nominal |
Weight | 100 g |
United Kingdom headquarters
Pico Technology
James House
Marlborough Road
Colm worth Business Park
Eaton Socony
St. Neots
PE19 8YP
United Kingdom
Tel: +44 (0)1480 396395
sales@picotech.com
support@picotech.com| North America regional
office
Pico Technology North America
320 N Glenwood Blvd.
Tyler
TX 75702
United States
Tel: +1 800 591 2796 (toll free)
sales@picotech.com
support@picotech.com| Germany regional office
and EU authorized representative
Pico Technology GmbH
Im Rehwinkel 6
30827 Garbsen
Germany
Tel: +49 (0) 5131 907 6290
info.de@picotech.com
---|---|---
www.picotech.com DO368-4 Copyright © 2022
Pico Technology Ltd. All rights reserved.