SUN Helium Analyzer with Alarm
Instruction Manual
Discharge of Liability
Before operating and installing please read this manual carefully. This manual
is an integral part of the DE-OX® SUN helium analyzer. The use of this
instrument implicates total acceptance of the present section.
All warranties relevant to this instrument are void if is not operated and
maintained in accordance with this manual.
TEMC® disclaims every responsibility for the improper, wrong or incorrect use
of this measurement instrument. Before use, it is necessary to receive a
proper training on mixed gas diving and gas mixtures analysis for diving
applications. It is dangerous to breathe mixed gas with any percentage of
oxygen different from 21% and containing helium. Sport and mixed gas diving is
a potentially hazardous and dangerous activity that may lead to death if
practise incorrectly. Manipulating mixed gas with oxygen is dangerous and
requires special training and oxygen dedicated equipment. The present
instrument does not replace the necessary notional and practice training for
these activities. The data displayed are a mathematical elaboration based on
physical laws but they are not representative of the complexity of human
physiology and of the different psycho-physical situation of use by customers.
TEMC® is not liable for any damage or injury including death which may result
from DE-OX® SUN utilization.
The operator of this instrument is responsible for any damage or injury
resulting from improper use, unauthorized repair, improper maintenance, or
damage by anyone other than TEMC®. Use always another instrument for data
cross checking.
The N.O.A.A. (National Oceanic and Atmospheric Administration) oxygen partial
pressure and exposure time limits table are available on world-wide
literature.
TEMC®, whose policy is one of continuous quality improvement, reserves the
right to modify the technical characteristics of the instrument and manual
without prior notice.
General Specifications
Thank you for buying DE-OX ® SUN analyzer. It represents the simplest and
up-to-date mobile measurement instrument for the analysis of helium in mixed
gas. Unique features include two level alarms, standard analog 4-20 mA and
Open Collector outputs.
DE-OX ® SUN has been especially designed and manufactured for the
analysis of binary and ternary gas mix like Heliox and Trimix.
DE-OX ® SUN comprises in just one integrate device the helium sensor, the
electronic board and the battery. The instrument does not need any
calibration.
DE-OX ® SUN can be used for continuously reading during mixed gas
filling. It can alert and help in switching off the compressor if the helium
value goes out the set alarms range (for instance 45% helium).
DE-OX ® SUN is based on an innovative sensor developed by TEMC® for
respiratory use and for monitoring the respiratory parameters in saturation
dives.
Main features:
- Helium indication in any Oxygen-Nitrogen mix in the range 0.0 – 100.0% of volume. 0,1 % resolution.
- Full digital measurement instrument.
- Thermal conductivity stainless steel helium sensor developed for marine sport ambient.
- Extra long file sensor life
- Auto OFF with switching delay time setting.
- No calibration needed.
- Sensor and analyzer auto setting.
- Two custom audible and visible alarms.
- Analog 4-20 mA output for external devices (available on request).
- Open Collector output (available on request).
- Battery low indicator.
- milli Volt sensor output reading.
- Standard 9 Volt transistor battery.
- Simple battery replacement.
- Direct connection to every kind of tank valve with TEMC® rubber cup.
- Soft protection and transport bag.
- Numeric display of 50x35mm.
- Dimensions: max 7,5×4,2 cm weight 350 grams ca.
Powering On and Warming Up
Press the On Off key for more than three seconds, until the display shows
“On”, after that the Firmware version and the “Att” (WAIT) message will be
displayed for about 3 seconds.
Reading Page
At the end of the warming up, the instrument is now ready to read helium
concentration into the gas mix. Connect the sensors to the mixed gas flow with
a flow of 0,5-1,0 litres/minute at ambient pressure (typically 1 bar) and
temperature (typically 20°C). Now on the display the helium content will be
continuously displayed. While the instrument is in oxygen percentage display
mode, keeping the <Calib key pressed for more than 1 seconds causes the device
to display the helium sensor output in mV. It is periodically necessary to
read this value in order to know if the sensor is correctly working or it is
exhausted or faulty.
Warning
During gas analysis the instrument and the sensors must be firmly kept and
must not be shaken. The gas pipe used to provide gas mix to the sensor must
allow free gas flowing. Never seal or stop the sensor holes. If this event
occurs the helium sensor will be irreparably damaged.
Warning
The mixed gas flow, must be 0,5-2 litres/min at ambient pressure (typically 1
bar) and temperature (typically 20°C) or DE-OX® SUN will yield a wrong gas
analysis. All the analysis concentration shown on the display will be wrong.
Wrong helium analysis may lead to death.
Programming Procedures
Keep the ^ Prog key pressed for more than two seconds, then release the key. On the display will appear “Pr” for two second in order to advice the start of the programming procedure. It is possible to program:
- AL1 First alarm threshold expressed in % helium concentration
- AL2 Second alarm threshold expressed in % helium concentration
- FSC Value expressed in % helium concentration corresponding to the current output full scale value (20mA). 4mA always correspond to zero % helium concentration.
- t00 Auto switch off delay.
Press ON/OFF key for going to different pages.
At the end of the of the programming procedure on the display will be
displayed “End” and the instrument will display the oxygen content in the gas
mix.
Threshold Alarms Setting
Press the ^ Prog key for more than two seconds, then release the key. On the
display will appear “Pr” for two seconds in order to advice the start of the
programming procedure. Then will appear “AL1” for changing the value of the
minimum alarm threshold. The display will show “AL1” and the value of the
alarm threshold alternatively.
The blinking digit shows the current cursor position.
Press the ^ Prog key for increasing the value (from 0 to 9).
Press the <Calib key to move the cursor to the next digit.
Press the On/Off key for accepting the value and jumping to AL2 threshold
programming page.
Act as for AL1 to modify the AL2 second threshold to the desired value. If you
want to exclude the first alarm
AL1 program it as 100.0. If you want to exclude the second alarm AL2 program
it as 100.0.
Full Scale Value Setting
After the alarms concentration values setting, the instrument goes to “Fsc”
page in order to change the analogic full scale value of the instrument (if
equipped with this feature). This is the helium concentration at 20 mA on the
analogical output. 4 mA is the value at 0% of helium concentration.
It is displayed alternatively “Fsc” and the value of the full scale. The
blinking value is the actual changing value.
Press the ^ Prog key for increasing the value (from 0 to 9).
Press the <Calib key for going to the next value.
Press the On/Off key for accepting the value and jumping to the next program
page.
Auto Switch Off Delay Time
In this page it is possible to change the auto switch off delay time of the
instrument. The display will show “t00” and the seconds alternatively. The
blinking digit shows the current cursor position.
Press the ^ Prog key for increasing the value (from 0 to 9).
Press the <Calib key to move the cursor to the next digit.
In this page it is possible to define the time after that the instrument will
auto switch off in the range 0 – 1999 seconds. For keeping the instrument
already switched on insert 000 sec.
Press the On/Off key for accepting the value and quitting the programming
procedures. On the display will appear for two seconds “End”, then the
instrument will show the helium concentration.
Warning
The instrument will switch off after the time set into the Auto Switch Off
Delay Time. If you are using the instrument in continuous blending system, set
this value to 000. If not, the instrument will auto switch off without giving
any control of the mixing process. Always analyze the mixed gas you are going
to dive with. Diving with wrong mixed gas may lead to death.
Threshold Alarms
Should the helium reading goes over the threshold alarms (AL1 or AL2) the
instrument goes to alarm mode and it activates the relays output if equipped
(open collector max 100mA) and the internal buzzer. On the display will be
shown the trespassed alarm and the actual measured value.
For stopping the buzzer song it is enough to press any key. In this event the
instrument will remain in alarm mode until the analyzed value will go below
the alarm.
Powering Off
For powering off the instrument, in the reading page press the On/Off key for
more than 3 seconds. On the display will appear “Off” and the instrument will
power off.
Factory Reset
In case it is necessary to reset the instrument to the factory settings, power
on the instrument pressing at the same time for more than one second the
On/Off and the <Calib keys. On the display will appear “res” and the
instrument will go to the reading page.
Warning
In case of reset, the instrument will delete all the alarms settings, the full
scale value, any auto delay switching off time. Before using again the
instrument, it may be necessary to program again the alarm values, the full
scale value, the auto delay switching off time if changed. Wrong helium
analysis may lead to death.
Note For Correct Helium Analysis
Before doing mixed gas analysis, connect the mixed gas hose (not supplied) to
the flow adapter in the rear of the instrument and let the gas flow goes
through the internal helium sensor. The sensor, when coming in contact with
the mixed gas flow, gives an electrical signal. This signal is transmitted to
the central unit and processed in order to give the helium percentage
information on the display. The mixed gas flow that comes in contact with the
sensor, must range from 0,5 to 1 litres/min (best 0,5 lit/min) at the same
calibration pressure (typically ambient pressure 1 bar) and temperature
(Typically 20°C) or a wrong gas analysis will occur (preferably use a flow
regulator with meter connected to the mixed gas tank). Higher flow value or
pressure or temperature variations, will cause wrong analysis and will
irreparably damage the sensor. During the gas analysis, the unit must be kept
steady and must not be shaken or submitted to vibrations. Allow gas freely
flows into the pipes.
During gas analysis, the helium percentage of the mixed gas being analyzed is
available on the display. It is possible to make several gas analysis without
waiting the instrument goes to 0 reading.
It is possible to let DE -OX® SUN in ambient for ambient analysis of helium.
In this case the reading will be slower (up to 5 minutes).
Warning
When battery is nearly exhausted BAT appears on the display. In this case
helium analysis is not anymore reliable.
Warning
Breathing mixed gas with helium needs the right training. NOOA and medical
studies recommend not to expose to an oxygen partial pressure higher than 1.6
bar or lower than 0.18 bar because of serious personal injury including death.
Refer to proper medical literature and receive the right training.
Replacement of the Battery and of the Helium Sensor
DE-OX® SUN requires one 9 Volt transistor alkaline battery. It must be replaced when “bat” appears on the display. To replace the battery:
- Switch off DE-OX® SUN.
- Remove the three closing screws and carefully open the instrument.
- Remove the old battery and replace it with the new one.
- Do not throw anyway the old battery. Check the correct recycling procedure.
- Close the instrument screwing the three closing screws.
- Switch on the instrument with the On/Off key and check the battery level.
Warning
Don’t leave battery inside the instrument if you do not use it for a long
time.
The helium sensor must be replaced when unable to correctly read helium
concentration. The sensor life is over 24 months or more under normal
operating conditions. Always contact TEMC® for replacing the sensor.
Warning
Replace the helium sensor when the unit is unable to correctly analyze the
gas. The use of an exhausted sensors may cause dangerous analysis and lead to
death. Wrong helium analysis may lead to death. Always contact TEMC® for
sensor replacement. Do not use any other helium sensor brand.
To replace the helium sensor:
- Switch off DE-OX® SUN with the On/Off key.
- Remove the three closing screws and carefully open the instrument.
- Unplug the sensor connector.
- Remove the old sensor screwing un-clockwise the flow adapter and replace it with the new one.
- Connect again the plug to the sensor connector. Screw again the flow adapter.
- Do not throw anyway the old sensor. Check the correct recycling procedure.
- Close the instrument, screwing the three closing screws.
- Switch on the instrument with the On/Off key and check the unit.
Warning
Do not try to disassemble the sensor. Sealed unit contains delicate electronic
components that may be destroyed. If this event occurs the helium sensor will
be irreparably damaged. If disassembled or handled the sensor will not
properly work and will not correctly analyze the gas mix. Wrong helium
analysis may lead to death. Do not wet in any case. Do not touch any of the
internal part. Some sensor internal parts are heated to up 500°C and may
injure you causing serious burning. In this event get medical attention.
Always contact TEMC® for sensor replacement. Do not use any other helium
sensor brand.
Central Unit Specifications
DE-OX® SUN contains delicate electronic devices, hence it is absolutely necessary to:
- Avoid shock (greater than 2g) or vibrations
- Avoid exposure to environmental temperature higher than 30°C and lower than 5°C (41 F)
- Do not supply the instrument with voltage higher than 9 V DC. This will lead to damaging the battery and the internal electronic carts
- Press the keys delicately with your fingers and don’t use any kind of tools
- Avoid exposure to spray liquids and to corrosive gas, don’t submerge and do not expose to water splash.
- For cleaning, use a soft dry brush not soaked with any liquid.
- Don’t try to remove screws from DE-OX® SUN or to open it. For any problem call TEMC®.
Warning
Never wet or splash or submerge the instrument. The instrument or any of its
part are waterproof . If one of these events occur the DE-OX® SUN will be
irreparably damaged.
Helium Sensor Specifications
The Helium sensor is developed for related underwater applications. It
operates by comparing the thermal conductivity of the sample gas with that of
a reference gas (in this case air). A heated sensor is mounted so that it is
exposed to the sample gas while the reference sensor is enclosed in a sealed
compartment of the reference gas. If the sample gas has a higher thermal
conductivity than the reference gas, heat is lost from the exposed element and
its temperature decreases; if the thermal conductivity is lower than that of
the reference, the sensor heats up. The sensor is designed to only analyze
helium into nitrogen and oxygen gas mix.
Specifications:
- Measurements range 0÷100% of Helium in air or Nitrogen or oxygen.
- Resolution 0,1%
- Operating humidity from 0 to 95% non condensing. Prevent condensation on the surface of the sensing surface.
- Operating temperature range from 5°C (41F) to 40°C (104F).
- Storage temperature range from -15°C (5F) to 50°C (122F).
- Do not expose sensor to a biased voltage or to a short circuit.
- Response Time <10 sec for 90% response at 23°C.
- Accuracy ±2% over full scale.
- Linearity within ±2% over full scale.
- Due to the thermal conductivity technology, the accuracy of the sensor is no more warranted if into the gas mix there are gases like: acetylene, ammonia, argon, carbon dioxide, ethane, ethylene, freon, hydrogen, methane, neon, propane, etc.
In the interest of product improvement, TEMC® reserves the right to alter
design features and specifications without notice. Check TEMC® for the latest
helium sensor specifications.
Unless otherwise stated, all product specifications are quoted at standard
temperature and pressure.
Warning
Do not try to disassemble the sensor. Sealed unit contains delicate electronic
components that may be destroyed. If disassembled or handled the sensor will
not properly work and will not correctly analyze the gas mix. Wrong helium
analysis may lead to death. Do not wet in any case. Do not touch any of the
internal part. Some sensor internal parts are heated to up 500°C and may
injure you causing serious burning. In this event get medical attention Always
contact TEMC® for sensor replacement. Do not use any other helium sensor
brand.
4-20mA Connections (if available)
In one side of the instrument (if equipped with this feature) there is a jack
for the 4-20 mA output. Insert the plug into the jack for taking out the
signal. For connections refer to the following diagram.
Warning
The plug should be connected or disconnected when the instrument is switched
off, or the instrument will autonomously switch off.
Open Collector Connections (if available)
In one side of of the instrument (if equipped with this feature) there is a
jack for the open collector output. Insert the plug into the jack for taking
out the signal. For connections refer to the following diagram.
Warning
The plug should be connected or disconnected when the instrument is switched
off, or the instrument will autonomously switch off.
Warranty
TEMC® warrants that its DE -OX® SUN computer will be free from defects on
material and workmanship for a period of twelve (12) months from the date of
delivery, with the exception of sensor that is not manufactured in-house and
that is warranted for six (6) months, provided that the warranty Registration
Card is filled in and returned to TEMC® at the time of delivery.
Warranty will be void by failure to install, use or maintain DE -OX® SUN
according to TEMC® instructions. To avail oneself of the warranty, send the
product with carriage prepaid to TEMC®.
THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED,
IMPLIED, OR LEGAL. TEMC® MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR
ANY PARTICULAR PURPOSE. Buyer’s exclusive remedy shall be the replacement of
any DE-OX® SUN helium analyzer or sensor therefore that fails to comply with
the above warranty. Under no circumstances will TEMC® be liable for economic,
special, incidental or consequential damages of any kind whatsoever.
This warranty does not cover any damage due to accidental events (ex. impact
or falls) or natural events (ex. fires calamities, earthquakes, etc.).
(Warranty Registration Card)
(Fill with required data and send in a sealed envelope within 5 days from
purchase to):
TEMC ®
Via Donna Prassede 5 A
I-20142 MILANO ITALY
Phone +39 02 8463648 or +39 080 4490264
E-Mail: temc@iol.it
Website: www.temc.it
(First name) ____
(Family name) _
(Address) __
(#)_____
(Town) ____
(ZIP)__
(Country)
(Telephone) __
(E-Mail) ____
(Instrument Serial number) ____
(Purchase date)
(I have read and well understood the instruction manual and I accept all
warranty terms).
(Date) ____
(Signature)
Disposal of Old Electrical and Electronic Equipment
In the EU when you buy a new equipment you may give back the old equivalent
unit to the supplier free of charge for correct disposal.
Disposal of old Electrical & Electronic Equipment (Applicable throughout
the European Union and other European countries with separate collection
programs)
This symbol, found on your product or on its packaging, indicates that this
product should not be treated as household waste when you wish to dispose of
it. Instead, it should be handed over to an applicable collection point for
the recycling of electrical and electronic equipment. By ensuring this product
is disposed of correctly, you will help prevent potential negative
consequences to the environment and human health, which could otherwise be
caused by inappropriate disposal of this product. The recycling of materials
will help to conserve natural resources. For more detailed information about
the recycling of this product, please contact your local city office,
household waste disposal service or the retail store where you purchased this
product.
Warning
Please refer to the complete manual in electronic format before operating
the instrument. This guide is just a quick support. Misuse of this instrument
may lead to death.
For switching on the instrument, keep pressed the ON/OFF key for few seconds.
For analyzing connect the sensor to the gas source at 0,5-1 lit/min and 1 bar.
Pressing the CALIB key it is displayed the milli Volt output of the helium
sensor.
Pressing the PROG key you will go to the programming pages (Pr): Alarm1,
Alarm2, Full scale value, Auto switch off.
Alarm 1 (AL1) sets the lower alarm. At 100.0 the alarm is off.
Alarm 2 (AL2) sets the higher alarm. At 100.0 the alarm is off.
Full scale value (FSC) sets the span of the instrument for the 4-20 mA output,
if available.
Auto Switch Off Delay Time (t00) sets the switching off delay time in seconds.
At 000 the instrument will not automatically switch off.
For changing the value into any programming page press PROG^ key.
For going to the next value press CALIB< key.
For saving the value and going to the next program page press ON/OFF key. At
the end of the programming pages is displayed End.
For switching off the instrument keep pressed the ON/OFF key for few seconds.
For replacing the battery, unscrew the three screws on the rear, open the
instrument, replace the battery with a new one, close the instrument screwing
the three screws.
TEMC ®
Via Donna Prassede 5 A
I-20142 MILANO ITALY
Phone +39 02 8463648 or +39 080 4490264
E-Mail: temc@iol.it
Website: www.temc.it
HELIUM ANALYZER
2011 All Rights reserved
www.temc.it
Draft Issue n. 1.1 of 10/02/11