DE-OX ZIP PRO Oxygen Analyzer Instruction Manual
- June 12, 2024
- DE-OX
Table of Contents
- Discharge of Liability
- General Specifications
- Powering On and Warming Up
- Reading Page
- Programming Procedures
- Threshold Alarms Setting
- Full Scale Value Setting
- Calibration Oxygen value
- Threshold Alarms
- Oxygen Sensor Calibration
- Powering Off
- Factory Reset
- Note For Correct Oxygen Analysis
- Replacement of the Battery and of the Oxygen Sensor
- Central Unit Specifications
- General Oxygen Sensor Specifications
- Warranty
- Disposal of Old Electrical and Electronic Equipment
- Documents / Resources
Instruction Manual
Draft Issue n. 1.2 of 27/02/11 DE-OX® ZIP PRO
OXYGEN ANALYZER
2009 All Rights reserved
[ www.temc.it](http://www.temc.it)
Discharge of Liability
Before operating and installing please read this manual carefully. This manual
is an integral part of the DE-OX ® ZIP PRO oxygen analyzer. The use of this
instrument implicates total acceptance of the present section.
All warranties relevant to this instrument are void if is not operated and
maintained in accordance with this manual.
TEMC ® disclaims every responsibility for the improper, wrong or incorrect use
of this measurement instrument. Before use, it is necessary to receive a
proper training on mixed gas diving and gas mixtures analysis for diving
applications. It is dangerous to breathe mixed gas with any percentage of
oxygen different from 21% and containing helium. Sport and mixed gas diving is
a potentially hazardous and dangerous activity that may lead to death if
practise incorrectly. Manipulating mixed gas with oxygen is dangerous and
requires special training and oxygen dedicated equipment. The present
instrument does not replace the necessary notional and practice training for
these activities. The data displayed are a mathematical elaboration based on
physical laws but they are not representative of the complexity of human
physiology and of the different psycho-physical situation of use by customers.
TEMC ® is not liable for any damage or injury including death which may
result from DE -OX ® ZIP PRO utilization.
The operator of this instrument is responsible for any damage or injury
resulting from improper use, unauthorized repair, improper maintenance, or
damage by anyone other than TEMC ® .
Use always another instrument for data cross checking.
The N.O.A.A. (National Oceanic and Atmospheric Administration) oxygen partial
pressure and exposure time limits table are available on world-wide
literature.
TEMC ® , whose policy is one of continuous quality improvement, reserves the
right to modify the technical characteristics of the instrument and manual
without prior notice.
General Specifications
Thank you for buying DE -OX ® ZIP PRO analyzer. It represents the simplest and
up-to-date measurement instrument for the analysis of oxygenate mixed gas for
mixed gas diving. It has been especially designed and manufactured for the
analysis of binary and ternary gas mixtures like Nitrox, Helix and Trimix.
DE-OX ® ZIP PRO comprises in just one integrate device the oxygen sensor, the
electronic board and the battery. The electrochemical sensor gives a
proportional electrical signal with the oxygen percentage in the mixed gas. On
the unit’s display, this signal is transformed in easy-to-read oxygen
percentage information given in real time.
Main features:
-
* Display of oxygen percentage in the range 0.0 – 100.0%.
- Full digital measurement instrument
- Simple battery and sensor replacement.
- Very innovative cover design.
- 0,1 % of volume resolution
- Sensor and analyzer autosetting.
- Electrochemical sensor.
- Two custom audible and visible alarms.
- Analog 4-20 mA output for external devices.
- Numeric display of 50x35mm.
- Battery low indicator.
- mV sensor output reading
- Standard 9 Volt transistor battery.
- Simple battery and sensor replacement.
- Direct connection to every kind of tank valve with TEMC ® rubber cup.
- Soft protection and transport bag.
- Dimensions: max 7,5×4,2 cm weight 250 grams.
Powering On and Warming Up
Press the On/Off key for more than one second, until the display shows “On”,
After that the “Att” (WAIT) message will be displayed for about 4 seconds,
then the sensor auto-calibration procedure is started. The auto-calibration
lasts about 5 seconds; during this time the display will show the messages
“CAL” and “Att” alternatively.
Be sure that during this time the sensor is exposed to the gas mix with Oxygen
percentage as programmed in the O2c parameter (see Programming procedure). By
default, for simplicity, the O2c value is set at 21% that corresponds to the
percentage of Oxygen in free air in this case be sure that flows air on the
sensor before switching on and during the warm up time
Warning
During the warm up time if the oxygen percentage of the mixed gas flow is
different from the value of calibration set in the instrument, a failed
calibration is obtained and a consequent wrong analysis value will be read.
Reading Page
At the end of the warming up and calibration time, the display will show the
blinking value in mV read from the oxygen sensor during calibrations: the
instrument is now ready to read oxygen concentration into the gas mix.
Connect the sensors to the mixed gas flow with a flow of 0,5-2,0 litres/minute
at ambient pressure (typically 1 bar) and temperature (typically 20°C). Now on
the display the Oxygen content will be continuously displayed. While the
instrument is in oxygen percentage display mode, keeping the <Calib key
pressed for more than 1 seconds causes the device to display the Oxygen sensor
output in mV. To read this value is periodically necessary in order to know if
the sensor is correctly working or is exhausted.
Warning
During gas analysis the instrument and the sensors must be firmly kept and
must not be shaken. The gas pipe used to provide gas mix to the sensor must
allow free gas flowing.
Warning
The mixed gas flow, must be 0,5-2 litres/min at ambient pressure (typically 1
bar) and temperature (typically 20°C) or DE-OX ® ZIP PRO will yield a wrong
gas analysis. All the analysis concentration shown on the display will be
wrong. Wrong Oxygen analysis may lead to death.
Programming Procedures
Keep the ^ Prog key pressed for more than two seconds, then release the key. On the display will appear “Pr” for two second in order to advice the start of the programming procedure. It is possible to program:
-
* AL1 Minimum alarm threshold expressed in % oxygen concentration
- AL2 Maximum alarm threshold expressed in % oxygen concentration
- FSc Value expressed in % oxygen concentration corresponding to the current output full scale value (20mA). 4mA always correspond to zero % Oxygen concentration.
- O2c Oxygen sensor calibration point
- Press ON/OFF key for going to different pages.
At the end of the of the programming procedure on the display will be displayed “End” and the instrument will display the oxygen content in the gas mix.
Threshold Alarms Setting
Press the ^ Prog key for more than two seconds, then release the key. On the
display will appear “Pr” for two seconds in order to advice the start of the
programming procedure. Then will appear “AL1” for changing the value of the
minimum alarm threshold. The display will show “AL1” and the value of the
alarm threshold alternatively.
The blinking digit shows the current cursor position.
Press the ^ Prog key for increasing the value (from 0 to 9).
Press the <Calib key to move the cursor to the next digit.
Press the On/Off key for accepting the value and jumping to AL2 threshold
programming page.
Act as for AL1 to modify the AL2 maximum threshold to the desired value. If
you want to exclude the minimum alarm AL1 program it as 00.0. If you want to
exclude the maximum alarm AL2 program it as 100.0.
Full Scale Value Setting
After the alarms concentration values setting, the instrument goes to “Fsc”
page in order to change the analogic full scale value of the instrument (if
equipped with this feature). This is the Oxygen concentration at 20 mA on the
analogical output. 4 mA is the value at 0% of Oxygen concentration.
It is displayed alternatively “Fsc” and the value of the full scale. The
blinking value is the actual changing value.
Press the ^ Prog key for increasing the value (from 0 to 9).
Press the <Calib key for going to the next value.
Press the On/Off key for accepting the value and jumping to the next program
page.
Calibration Oxygen value
In this page it is possible to change the calibration point of the instrument.
The display will show “O2c” and the value of the oxygen calibration value
alternatively. The blinking digit shows the current cursor position.
Press the ^ Prog key for increasing the value (from 0 to 9).
Press the <Calib key to move the cursor to the next digit.
Press the On/Off key for accepting the value and ending the programming
procedures. On the display will appear for two seconds “End”, then the
instrument will show the Oxygen concentration reading.
In this page it is possible to define the oxygen percentage value that allows
calibrating the oxygen sensor.
This value depends on the oxygen mix available. Every measure will be made
referring to this calibration value.
Wrong calibration means wrong analysis.
Typically the calibration oxygen percentage value is 21,0%, because this value
is the oxygen content in air. A calibration in air is simple and do not need
other well known oxygen gas mix.
Warning
The oxygen calibration value will be the same value of calibration set when
the instrument is switched on. During the warm up time if the oxygen
percentage of the mixed gas flow is different from the value of calibration
set in the instrument, a failed calibration is obtained and a consequent wrong
analysis value will be read.
Threshold Alarms
Should the oxygen reading goes over the maximum or below the minimum threshold
alarms (AL2 or AL1) the instrument goes to alarm mode and it activates the
relays output (open collector max 100mA) and the internal buzzer. On the
display will be shown the trespassed alarm and the actual measured value.
For stopping the buzzer song it is enough to press any key. In this event the
instrument will remain in alarm mode until the oxygen reading will be out of
the thresholds values.
Oxygen Sensor Calibration
In the reading page, it is possible to calibrate the oxygen sensors, that, due
of time, temperature, gas flow, ambient pressure, could have variation of the
measurement precision.
To do the calibration it is necessary to have a gas mix with well know and
stated percentage of Oxygen.
The known percentage has to be set into the calibration Oxygen value in the
programming procedure (see above).
For calibrating the instrument press the On/Off and <Calib keys at the same
time for two seconds. On the display will alternately appair “Cal” and Att”
for 5 seconds, then the instrument, after displaying for few seconds the
calibration value in mV read during the calibration process, goes to the main
reading page. It is possible to make multiple analysis without calibrating the
instrument every time.
Warning
During the calibration procedure, pay attention not to cover the two holes
near the hose connection in the rear of the instrument. The unit must be
exposed to the known sample mixed gas (such as air or pure oxygen) and
calibrated consequently. If you calibrate with mixed gas flow, the flow must
range between 0,5 and 2 litres/min at atmospheric pressure and temperature, or
DE-OX ® ZIP PRO will yield a wrong gas analysis. The unit must be kept steady
and must not be shaken. Avoid change in temperature between calibration gas
and analysis gas.
Warning
Calibrating the sensors with wrong gas (with different values from the setting
ones), will lead to incorrect and very dangerous analyses
The closest the calibration value is to the mixed gas percentage to be
analyzed, the best and the fastest is the DE-OX ® ZIP PRO performance.
Warning
Whatever error in the calibration procedure will make mistake in the
subsequent analysis.
Powering Off
For powering off the instrument, in the reading page press the On/Off key for more than 1 second. On the display will appear “Off” and the instrument will power off.
Factory Reset
In case it is necessary to reset the instrument to the factory settings, power on the instrument pressing at the same time for more than one second the On/Off and the <Calib keys. On the display will appear “res” and the instrument will go to the reading page.
Warning
In case of reset, the instrument will delete all the alarms settings, the full
scale value, any new calibration value of the Oxygen sensor. Before using
again the instrument, it may be necessary toprogram again the alarm values,
the full scale value, calibration value of the Oxygen sensor if changed.
All the analysis concentration shown on the display would be wrong. Wrong
Oxygen analysis may lead to death.
Note For Correct Oxygen Analysis
Before doing mixed gas analysis, connect the mixed gas hose (not supplied) to
the adapter in the rear of the instrument and let the gas flow goes through
the internal oxygen sensor. The sensor, when coming in contact with the mixed
gas flow, gives an electrical signal. This signal is transmitted to the
central unit and processed in order to give the oxygen percentage information
on the display. The oxygen analysis is made by comparing the calibration value
to the analysis value. The mixed gas flow that comes in contact with the
sensor, must range from 0,5 to 2 litres/min (best 0,5 lit/min) at the same
calibration pressure (typically ambient pressure 1 bar) and temperature
(Typically 20°C) or a wrong gas analysis will occur (preferably use a flow
regulator with meter
connected to the mixed gas tank). Higher flow value or pressure or temperature
variations, will cause wrong analysis and may damage the sensor. Whatever
error in the calibration procedure will make mistake in the subsequent
analysis. During the gas analysis, the unit must be kept steady and must not
be shaken or submitted to vibrations. Allow gas freely flows into the pipes.
During gas analysis, the oxygen percentage of the mixed gas being analyzed is
available on the display. It is possible to make more gas analysis without
intermediary calibration.
It is possible to let DE-OX ® ZIP PRO in ambient for ambient analysis Oxygen.
In this case the reading will be slower (up to 5 minutes).
Warning
When battery is nearly exhausted BAT appears on the display. In this case
oxygen analysis is not be reliable anymore.
Warning
NOOA and medical studies recommend not to expose to an oxygen partial pressure
higher than 1.6 (Ata) because of serious personal injury including death.
Refer to proper medical literature.
Replacement of the Battery and of the Oxygen Sensor
DE-OX® ZIP PRO requires one 9 Volt transistor alkaline battery. It must be
replaced when “bat” appears on the display. To replace the battery:
– Switch off DE-OX® ZIP PRO.
– Remove the three closing screws and carefully open the instrument.
– Remove the old battery and replace it with the new one.
– Do not throw anyway the old battery. Check the correct recycling procedure.
– Close the instrument screwing the three closing screws.
– Switch on the instrument with the On/Off key and check the battery level.
Warning
Don’t leave battery inside the instrument if you do not use it for a long
time.
The Oxygen sensor must be replaced when unable to correctly read Oxygen
concentration. The sensor life is about 12-24 months or more under normal
operating conditions.
Warning
Replace the Oxygen sensor when the unit is unable to correctly analyze the
gas. The use of an exhausted sensors may cause dangerous analysis and lead to
death. Wrong Oxygen analysis may lead to death.
To replace the Oxygen sensor:
– Switch off DE-OX® ZIP PRO with the On/Off key.
– Remove the three closing screws and carefully open the instrument.
– Unplug the sensor plug connection.
– Remove the old sensor screwing un-clockwise and replace it with the new one.
– Connect again the plug to the sensor.
– Do not throw anyway the old sensor. Check the correct recycling procedure.
– Close the instrument, screwing the three closing screws.
– Switch on the instrument with the On/Off key and check the unit.
Warning
Do not try to disassemble the sensor. Sealed unit contains caustic liquid
(KOH) which can cause severe burns to skin and eyes. In case of contact, flush
15 minutes with water. For contact to eyes also get medical attention.
Central Unit Specifications
DE-OX ® ZIP PRO contains delicate electronic devices, hence it is absolutely necessary to:
– Avoid shock (greater than 2g) or vibrations
– Avoid exposure to environmental temperature higher than 30°C and lower than
5°C (41 F)
– Do not supply the instrument with voltage higher than 9 V DC. This will lead
to damaging the battery and the internal electronic carts
– Press the keys delicately with your fingers and don’t use any kind of tools
– Avoid exposure to spray liquids and to corrosive gas, don’t submerge and do
not expose to water splash.
– For cleaning, use a soft dry brush not soaked with any liquid.
– Don’t try to remove screws from DE-OX® ZIP PRO or to open it. For any
problem call TEMC ® .
– Dimensions: max 7,5×4,2 cm weight 250 grams ca.
Warning
Never wet or splash or submerge the instrument. The instrument or any of its
part are waterproof . If one of these events occur the DE-OX ® ZIP PRO will be
irreparably damaged.
General Oxygen Sensor Specifications
Sensor type | Galvanic cell type |
---|---|
Electrical Connector: | 3.5 mm mono jack, Do not expose sensor to a biased |
voltage or to a short circuit.
Measurement Range:| 0 to 100 Vol.%
Nominal Sensor Life:| > 1, 000, 000 Vol.% h
Expected Operating Life:| > 2 years @ ambient air. Sensor must be replaced
when unable to calibrate or to analyze mixed gas correctly.
Output Signal:| 11±3 milli Volt mV @ dry ambient air at 23°C (74F) and 60% RH
and at 1 ata.
Response Time t90 :| < 12 s
Drift:| < 1% volume O2 / month @ air
Operating Temperature:| 0 to 40°C
Pressure Range:| 750 to 1250 hPa
Linearity Error:| = 2% @ 100% O2 applied for 5 min
Zero Offset Voltage:| = 200 µV in 100% N2, applied for 5 min
Repeatability:| ± 1 % volume O2 @ 100% O2, applied for 5 min
Influence of Humidity:| -0.03% rel. O2 reading / % RH
Temperature Compensation:| NTC
Interferences:| according to DIN EN 12598 and ISO 7767
Storage Conditions|
Temperature Range:| recommended: 5 to 30°C maximum: – 20 to 50°C
Humidity:| up to 100% RH
Weight:| approximately 26 g
Material in contact with media:| PA, PPS, PTFE, stainless steel
In the interest of product improvement, TEMC ® reserves the right to alter
design features and specifications without notice. Check TEMC ® for the latest
oxygen sensor specifications.
Unless otherwise stated, all product specifications are quoted at standard
temperature and pressure.
Note: The figures in this table are typical values and should not be used
as a basis for cross calibration. Cross sensitivities may not be linear and
should not be scaled. For some cross interferences break through will occur if
gas is applied for a longer time period.
Warning
Do not try to disassemble the sensor. Sealed unit contains caustic liquid
(KOH) which may cause severe burns to skin and eyes. In case of contact, flush
15 minutes with water. For contact to eyes also get medical attention. Do not
wet in any case. If not properly cured the eyes may have permanent damages.
Warning
Never wet or splash or submerge the instrument. The instrument or any of its
part are waterproof . If one of these events occur the DE-OX ® ZIP PRO will be
irreparably damaged.
4-20mA Connections (if available)
In the rear of the instrument (if equipped with this feature) there is a jack for the 4-20 mA output. Insert the plug into the jack for taking out the signal. For connections refer to the following diagram.
Warning
The plug should be connected or disconnected when the instrument is switched
off, or the instrument will autonomously switch off.
Open Collector Connections (if available)
In the rear of the instrument (if equipped with this feature) there is a
jack for the open collector output. Insert the plug into the jack for taking
out the signal. For connections refer to the following diagram.
Warning
The plug should be connected or disconnected when the instrument is switched
off, or the instrument will autonomously switch off.
Warranty
TEMC ® warrants that its DE-OX ® ZIP PRO computer will be free from defects on
material and workmanship for a period of twelve (12) months from the date of
delivery, with the exception of sensor that is not manufactured inhouse and
that is warranted for six (6) months, provided that the warranty Registration
Card is filled in and returned to TEMC ® at the time of delivery.
Warranty will be void by failure to install, use or maintain DE-OX ® ZIP PRO
according to TEMC ® instructions. To avail oneself of the warranty, send the
product with carriage prepaid to TEMC ® .
THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED,
IMPLIED, OR LEGAL. TEMC ® MAKES NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR
ANY PARTICULAR PURPOSE. Buyer’s exclusive remedy shall be the replacement of
any DE -OX ® ZIP PRO Oxygen analyzer or sensor therefore that fails to comply
with the above warranty. Under no circumstances will TEMC ® be liable for
economic, special, incidental or consequential damages of any kind whatsoever.
This warranty does not cover any damage due to accidental events (ex. impact
or falls) or natural events (ex. fires calamities , earthquakes, etc.).
Carolina di Grazia (Warranty Registration Card) Compilate invirase in busts
chiesa entry 5 gitorin dall’acquisto a:
(Fill with required data and send in a sealed envelope within 5 days from
purchase to):
TEMC ®
Via Donna Preside 5 A
I-20142 MILANO ITALY
Phone +39 02 8463648 or +39 080 4490264
E-Mail: temc@iol.it Website:
www.temc.it
Nome ____Cognome
(First name) (Family name)
Via____n
(Address) (#)
Città__CAP__
(Town) (ZIP)
Nazione____
(Country)
Telefono
(Telephon)
E-Mail
(E-Mail)
N.serie strumento_____
(Instrument Serial number)
Data Acquisto____
(Purchase date)
Dichiaro di aver letto e compreso il manuale di istruzione e di accettare le
condizioni di garanzia del presente strumento.
(I have read and well understood the instruction manual and I accept all
warranty terms).
Data__
(Date)
Firma__
(Signature)
Disposal of Old Electrical and Electronic Equipment
In the EU when you buy a new equipment you may give back the old equivalent unit to the supplier free of charge for correct disposal.
Disposal of ooid Electrical & Electronic Equipment (Applicable throughout the
European Union and other European countries with separate collection programs)
This symbol, found on your product or on its packaging, indicates that this
product should not be treated as household waste when you wish to dispose of
it. Instead, it should be handed over to an applicable collection point for
the recycling of electrical and electronic equipment.
By ensuring this product is disposed of correctly, you will help prevent
potential negative consequences to the environment and human health, which
could otherwise be caused by inappropriate disposal of this product. The
recycling of materials will help to conserve natural resources, For more
detailed information about the recycling of this product, please contact your
local city office, household waste disposal service or the retail store where
you purchased this product.
Warning
Please refer to the complete manual in electronic format before operating the
instrument. This guide is just a quick support. Misuse of this instrument may
lead to death.
For switching on the instrument, keep pressed the ON/OFF key for few seconds.
The analyser will calibrate in the set value (default is 21.0%).
For analyzing connect the sensor to the gas source at 0,5-1 lit/min and 1 bar.
Pressing the CALIB key it is displayed the milli Volt sensor output.
Pressing the PROG key you will go to the programming pages (Pr): Alarm1,
Alarm2, Full scale value, Oxygen calibration value.
Alarm 1 (AL1) sets the minimum alarm. At 000.0 the alarm is off.
Alarm 2 (AL2) sets the maximum alarm. At 100.0 the alarm is off.
Full scale value (FSC) sets the span of the instrument for the 4-20 mA output,
if available.
Oxygen calibration value (O2c) sets the oxygen value during the calibration
procedure.
For changing the value into any programming page press PROG^ key.
For going to the next value press CALIB< key.
For saving the value and going to the next program page press ON/OFF key. At
the end of the programming pages is displayed End.
For calibrating in the oxygen set value, connect the sensor to the calibration
gas and press at the same time for few seconds ON/OFF and the <CALIB keys.
For switching off the instrument keep pressed the ON/OFF key for few seconds.
For replacing the battery or the sensor, unscrew the three screws on the rear,
open the instrument, replace the battery or sensor with a new one, close the
instrument screwing the three screws.
For resetting the instrument to the original factory settings (rES), power on
the instrument pressing at the same time for few seconds the ON/OFF and the
<CALIB keys.
TEMC®
Via Donna Preside 5 A
I-20142 MILANO ITALY
Phone +39 02 8463648 or +39 080 4490264
E-Mail: temc@iol.it Website:
www.temc.it
Documents / Resources
| DE-
OX ZIP PRO Oxygen
Analyzer
[pdf] Instruction Manual
ZIP PRO Oxygen Analyzer, ZIP PRO, Oxygen Analyzer, Analyzer
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>