SPER SCIENTIFIC 800008 Advanced Thermocouple Datalogger Instruction Manual
- October 30, 2023
- SPER SCIENTIFIC
Table of Contents
SPER SCIENTIFIC 800008 Advanced Thermocouple Datalogger
Advanced Datalogging Thermometer – 800008 Copyright ©2015 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may be developed, including photocopying, recording, or any information storage and retrieval system without the express permission from Sper Scientific.
INTRODUCTION
Your digital thermometer accurately measures temperature in Centigrade or Fahrenheit from -200°C to 1370°C (-328°F to 2498°F). It conforms to the National Bureau of Standards and IEC584 temperature and voltage tables for K-type thermocouple probes.
Features include a data hold function, maximum and minimum value recording, memory and clock functions, and a low battery indicator. An RS232 interface allows the unit to communicate with a computer. Applications include HVAC, lab work, environmental work, process control, and the food industry.
PANEL DESCRIPTION
- K Type temperature sensor input connector
- LCD Display
- Power Button
- Relative Readout Button
- Record Button
- Tripod Connector (on back)
- Battery Cover (on back)
- Min/Max Button
- Hold Button
- °C/°F Button
- Offset Calibration Screw
- Digital Output Connector
- AC Power Adaptor Connector
MEASUREMENT PROCEDURE
Setting the Clock
- Press the green on/off button and MAX/MIN button simultaneously, release.
- The meter will show SEL.
- Press the CLOCK button to select the year, month, day, hour and minute (in that order).
- To set the year, month, day, hour and minute, press the ▲ button to increase the number and the ▼ button to decrease the number.
- Press the CLOCK button again to exit clock setup.
- To cancel the setup process at any time, press the power button to turn the meter off.
Taking Measurements
-
Press the POWER button to turn the unit on. The amount of available memory space will be displayed.
-
Plug the thermocouple plug into the meter.
-
The meter will default to the temperature scale last used. Press the °C/°F button to change the temperature scale, if necessary.
-
Begin taking temperature readings.
-
Press the HOLD button to freeze the temperature reading. H will be displayed.
Note…
When the data hold function is active, the REL, MAX/MIN, and °C/°F buttons will be disabled. If you press these buttons, two beeps will sound. -
To resume measurement, press the HOLD button again.
-
To cancel setup process at any time without saving the changes, press the power button to turn the meter off.
Recording Data
Each time you use the REC button to record data, a data set will be stored in
the memory. Up to 16241 records can be stored.
- Press the REC button to begin recording data.
- To stop recording data, press the REC button again
Clearing Data from Memory
- With the meter turned off, press and hold both the POWER button and the REC button for at least 2 seconds.
- Release the power button, then release the REC button. CLR will be displayed
Relative Operation
When the meter is in relative operation mode the °C/°F button wil be disabled. Two beeps will sound when the REL button is pressed.
- Press the REL button. The meter will now store the current reading. The difference between the stored data and the new reading will be displayed.
- When you have finished, press the REL button again to exit relative operation mode.
- To resume measurement, press the HOLD button again
Setting the Recording Interval for Datalogging
- With the meter turned off, press and hold the MAX/MIN button.
- Press the POWER button to turn the meter on. Release the buttons.
- Press the HOLD button to select minutes and seconds.
- To set minutes and seconds, press the ▲ button to increase the number and the ▼ button to decrease the number.
CALIBRATION
Ice Point Calibration
Rinse container, gloves and tolls with pure distilled water. Using shaved ice
made from pure distilled water, fill container and add distilled water. The
ice bath is ready when there is no ice floating and no excess water on the
surface. Stir the ice bath to cool the water. Immerse the probe for 2 minutes.
Without removing the probe from the ice bath, use the calibration screw to
adjust the thermometer to 0°C/32°F.
AUTOMATIC SHUTOFF
If the meter is not used for 30 minutes, it will shut off automatically. To disable automatic shutoff:
- With the meter turned off, press and hold down the HOLD button.
- Press the POWER button to turn the meter on. Release the buttons. Two beeps will sound to indicate that automatic shutoff is disabled.
- The automatic shutoff feature will reactivate the next time the unit is turned on.
BATTERY REPLACEMENT
The low battery indicator appears when the battery needs replacement. Insert a fresh 9V battery
PC CONNECTION
- The Digital Output is a 9600bps N 81 serial interface
- The RX is a 5V normal high input port.
- The TX is 5V normal high output port
The command of Digital Output is listed below:
RS232 Command | Function | Remarks |
---|---|---|
K(ASC 4BH) | Ask for model No. | Return 4 bytes |
A(ASC 41H) | Inquire all encoded data | Return electrode 10 byte |
H(ASC 48H) | Hold button | |
M(ASC 4DH) | Max/Min button | |
N(ASC 3EH) | Exit Max/Min mode | |
R(ASC 52H) | REL button | |
C(ASC 43H) | °C/°F button | |
U(ASC 55H) | Dump all memory of therm. | Return 32768 bytes |
P(ASC 50H) | Load recorded data | |
Command K: | Return 4 bytes. For example, when sending command “K” to |
meter, it will return “3”,”0”,”5”, ASCII(13).
---|---
Command U:| Return 32768 bytes.
Command P:| Instead of returning all 32768 bytes, it only
returns recorded data.
Command H:| Equivalent to one pushing on the HOLD button
and no message is returned.
Command M:| Equivalent to one pushing on the MAX/MIN button and no
message is returned.
Command N:| Equivalent to one pushing and hold the MAX/MIN button for
two seconds to exit MAX/MIN mode.
Command R:| Equivalent to one pushing on the REL button
and no message is returned.
Command C:| Equivalent to one pushing on the °C/°F button and no message is returned.
Command A
1st BYTE:
The first byte is the start byte, its value is 2.
2nd BYTE:
bit7| bit6| bit5| bit4| bit3| bit2| bit1|
bit0
---|---|---|---|---|---|---|---
°C/°F| Low Bat| Hold| REL| Min/Max| REC
- bit 0 : 1→recording mode, 0→not recording
- bit 2 bit 1
- 0 0 →normal mode
- 0 1 →MAXIMUM mode
- 1 0 →MINIMUM mode
- 1 1 →calculate MAX/MIN in background mode
- bit3 : no use.
- bit4 : 1→REL
- bit5 : 1→ HOLD, 0→not HOLD
- bit6 : 1→LOW BATTERY , 0→BATTERY NORMAL
- bit7 : 1→°C 0→°F
3rd BYTE
bit7| bit6| bit5| bit4| bit3| bit2| bit1|
bit0
---|---|---|---|---|---|---|---
Auto Power Off| Memory Full| | | | Resolution| Sign| OL
- bit0: 1→main window value is OL, 0→not OL
- bit1 : 1→main window value is minus, 0→main window value is plus.
- bit2 : 1→4th byte and 5th byte represent #### , 0→4th byte and 5th byte represent ###.#
- bit6 : 1→Memory is full. 0→Memory is not full.
- bit7 : 1→Auto power off enabled. 0→Auto power off disabled
4th BYTE: | first two BCD code of main window value. |
---|---|
5th BYTE: | last two BCD code of main window value |
6th BYTE: | BCD code of month. |
7th BYTE: | BCD code of day. |
8th BYTE: | BCD code of hours. |
9th BYTE: | BCD code of minute. |
10th BYTE: | End byte, its value is 3, 1st and 10th are used |
to check frame error.
Setup ThermoLog (Thermo DataLogger)— RS232 interface software
The ThermoLog package contains:
- Two 3.5” diskettes
- Custom designed RS232 cable for THERMOLOG.
System Required:
- Windows 95 or Windows 98 or Windows NT 4.0 above. Minimum Hardware Required:
- 486-100 MHz PC compatible, 16 MB RAM; At least 5 MB hard disk space available to install THERMOLOG program. Recommended display resolution is 800X600
Software Download:
With technology upgrading, some computers no longer have a disc drive for you
to install the software that comes with your meter. You can download the
software that came with your meter directly. Go to the
www.sperdirect.com/software.htm and
find your meter or search for you meter (800008) and download from product
page.
Note…
Some software may not be directly available to download due to it must be
purchased. If you find that the software you need is not available, please
contact our Customer Support at 480.948.4448 or email
info@sperscientific.com for further
assistance.
Install ThermoLog:
- We recommend closing all other application before installing ThermoLog.
- Insert setup CD.
- Choose the Start button on the Taskbar and selct Run.
- Type A :\SETUP and choose OK, then it will copy ThermoLog.exe and help file to your hard disk.
- For other operating instructions, please refer to the online help whil executing ThermoLog
10 Main Menu:
Link Test:
Open Link Test window to search for thermometer connected to PC. When you
start the THERMOLINK, this window will display first and search for
thermometer. The result will be shown in the text box.
Control Panel:
By opening the Control Panel Window, the user can control thermometer via the
button in this window.
Datalogger:
By opening the DataLogger Window, the user can load the recorded data from the
thermometer.
Tabular :
By opening the Tabular window, the present data from the thermometer will be
listed in a scrolling table. These data can be stored as a file or the table
can be copy to other software such as EXCEL for further analysis.
Graph :
Open Real-Time Graph window to show the present data in graph.
Exit :
Terminates THERMOLOG program.
Tray Icon:
When THERMOLOG is running, there will be an icon displayed on the Windows
Taskbar area, you can click this icon then it will show a pop-up menu.
SPECIFICATIONS
Range | Resolution | Accuracy |
---|---|---|
-200°C to 200°C | 0.1°C | ±(0.2% reading + 1°C) |
200°C to 400°C | 1°C | ±(0.5% reading + 1°C) |
400°C to 1370°C | 1°C | ±(0.2% reading + 1°C) |
-328°F to -200°F | 1°F | ±(0.5% reading + 2°F) |
-200°F to 200°F | 0.1°F | ±(0.2% reading + 2°F) |
200°F to 2498°F | 1°F | ±(0.3% reading + 2°F) |
Temperature Coefficient:
For ambient temperatures from 0°C to 18°C and 28°C to 50°C, for each
°C ambient below 18°C or above 28°C add the following tolerance into the
accuracy spec: 0.01% of reading + 0.03°C (0.01% of reading + 0.06°F)
Probe Temp Range| -50 to 300ºC (-58 to 572º)
Numerical Display:| 4 digital Liquid Crystal Display
---|---
Input Projection at
Thermocouple Input:
| 60V DC or 24 Vrms AC
Operating Environment:| 0°C to 50°C (32°F to 122°F); 0 to 80% RH
Storage Environment:| 10°C to 60°C (14°F to 140°F); 0 to 80% RH
Storage Altitude| ≤ 2000 meters
Power Requirement:| 9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22
Battery Life:| Approx. 100 hrs. with alkaline battery
AC Adapter:| 9V DC ±15% 100 mA
Dimensions| 7¼” × 2½” × 1¼” (184 × 64 × 32 mm)
Weight| 7.4 oz (210 g)
WARRANTY
Sper Scientific warrants this product against defects in materials and workmanship for a period of five (5) years from the date of purchase, and agrees to repair or replace any defective unit without charge. If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available. This warranty does not cover probes, batteries, battery leakage, or damage resulting from accident, tampering, misuse, or abuse of the product. Opening the meter to expose its electronics will break the waterproof seal and void the warranty. To obtain warranty service, ship the unit postage prepaid to
SPER SCIENTIFIC LTD.
8281 East Evans Road, Suite #103 Scottsdale, AZ 85260
The defective unit must be accompanied by a description of the problem and your return address. Register your product online at www.sperwarranty.com within 10 days of purchase
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>