SPER SCIENTIFIC 300067 Brix Refractometer Instruction Manual

June 9, 2024
SPER SCIENTIFIC

SPER SCIENTIFIC 300067 Brix Refractometer

Product Information

Refractometer
The Refractometer is a device that measures the refractive index of a liquid sample. It can measure the concentration of a solution by determining the amount of light that is refracted as it passes through the sample. This particular model has a range of 0-80% Brix and comes with a calibration screw and distilled water for calibration.

Product Usage Instructions

  1. Open the COVER PLATE and clean the PRISM with a soft cloth to avoid scratching the surface.
  2. Aim the refractometer toward a light source and rotate the EYEPIECE to obtain the clearest focus.
  3. Perform Zero Point Calibration:
  4. Open the COVER PLATE.
  5. Apply a few drops of the included Distilled Water onto the PRISM platform.
  6. Close the COVER PLATE.
  7. Turn the CALIBRATION SCREW until the dark and light boundary line coincides with the bottom line of the scale.
  8. Carefully dry the prism platform and COVER PLATE.
  9. Place a few drops of the test solution on the prism.
  10. Close the COVER PLATE.
  11. Look through the EYEPIECE and read the Brix value on the scale.
  12. After use, clean the prism with distilled water and a cloth to remove any residue.

Note: The temperature of the zero point calibration should be at the same temperature as the test solution. For variations in temperature, the zero point should be adjusted once every 30 minutes.

DESCRIPTION

SPER-SCIENTIFIC-300067-Brix-Refractometer-1

OPERATING PROCEDURE

  1. With the COVER PLATE open, clean the PRISM with a soft cloth to avoid scratching the surface.
  2. Aim the refractometer toward a light source and rotate the EYEPIECE to obtain the clearest focus.
  3. Zero Point Calibration:
    • A. Open the COVER PLATE.
    • B. Apply a few drops of the included Distilled Water onto the PRISM platform.
    • C. Close the COVER PLATE.
    • D. Turn the CALIBRATION SCREW until the dark and light boundary line coincides with the bottom line of the scale.
  4. Carefully dry the prism platform and COVER PLATE.
  5. Place a few drops of the test solution on the prism and close the COVER PLATE so the solution spreads evenly on the prism.
  6. Aim the front of the refractometer towards the light source and focus the eyepiece on the boundary line.
  7. The boundary line indicates the brix percentage of the test sample at 20°C.
  8. After use, clean the prism with distilled water and a cloth to remove any residue.

Note: The temperature of the zero point calibration should be at the same temperature as the test solution. For variations in temperature the zero point should be adjusted once every 30 minutes.

Precautions

  • Never submerge the unit, and do not let liquid seep into the unit’s body.
  • Clean the refractometer after each use with a soft cotton cloth.
  • Do not scratch the surface of the prism.
  • Store in a dry, clean, and non-corrosive environment.
  • Avoid strong impacts.

Specifications

Model Range Resolution Size Weight Accessories
300067 0-80% Brix 0.5% Brix 6” x 1 ½” (153mm x 40 mm) 7.1oz. (200 g.)

Screwdriver, Carrying Case,
Transfer Pipette, Manual.

WARRANTY

Sper Scientific warrants this product against defects in materials and workmanship for a period of one (1) year from the date of purchase, and agrees to repair or replace any defective unit without charge. If your model has since been discontinued, an equivalent Sper Scientific product will be substituted if available. This warranty does not cover probes, batteries, battery leakage, or damage resulting from accident, tampering, misuse, or abuse of the product. Opening the meter to expose its electronics will break the waterproof seal and void the warranty. To obtain warranty service, ship the unit postage prepaid to:
Sper Scientific Ltd. offers a warranty for defective units. The defective unit must be accompanied by a description of the problem and your return address. Register your product online at https://www.sperwarranty.com/register within 10 days of purchase.

SPER SCIENTIFIC LTD.
8281 E. Evans Rd., Suite #103
Scottsdale, AZ 85260
480-948-4448

The defective unit must be accompanied by a description of the problem and your return address. Register your product online at https://www.sperwarranty.com/register within 10 days of purchase.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

SPER SCIENTIFIC User Manuals

Related Manuals