FLIR VS80C10-25RM Pipe Inspection Camera Probe User Manual
- October 30, 2023
- FLIR
Table of Contents
- FLIR VS80C10-25RM Pipe Inspection Camera Probe
- Product Information
- Product Usage Instructions
- Introduction
- Product Description
- Safety
- Connect the VS80
- Camera Accessories
- Distance Gauge
- Measurement Considerations
- Specifications
- Warranty and Support
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
FLIR VS80C10-25RM Pipe Inspection Camera Probe
Product Information
Introduction
The Pipe Inspection Camera Probe is an industrial product designed for
inspecting pipes and other hard-to-reach areas. It consists of a handle,
camera head, distance gauge, probe wheel with 82 ft (25 m) coiled probe, and a
VS80 connector. The digital display shows the probe travel distance, battery
status, and unit of distance measured (meters or feet).
Safety
Before using the product, please read and understand all safety cautions.
Do not insert the probe into flammable liquid or gas. Do not use abrasive
detergents, solvents, corrosives, or alcohol to clean the probe. Use a high-
quality lens cleaning fluid and a lint-free swab to clean the camera and work
light lenses. Do not disassemble the product, avoid direct sun exposure, and
store in a cool, dry, and well-ventilated area. The probe contains sensitive
circuitry that controls the optics and electronics of the camera. Do not
strike the probe or allow it to impact objects forcefully.
Product Usage Instructions
Connecting the VS80
Always switch off the VS80 power before connecting the probe. Align the
white dot on the probe connector with the white dot on the VS80 monitor
connector. Plug the probe into the monitor unit, ensuring proper alignment and
full insertion. Tighten the collar nut to secure the connection. Do not
overtighten.
Using the Distance Gauge
The backlit electronic gauge measures and displays the distance the probe
travels into the inspection area in meters and feet. The gauge tracks forward
and reverse movement. To switch on/off the gauge, press the power button for
three seconds. After one hour, the gauge will automatically switch off to
conserve battery energy. The battery status icon shows the current battery
power. To replace the batteries, switch off the meter, loosen and remove the
battery compartment screw, remove the compartment cover, remove the old
batteries, and install two new AAA batteries observing correct polarity. Close
the compartment and secure it with the compartment screw.
Changing the Measurement Units
To change the measurement units, press the unit of measure button.
Camera Accessories
The Pipe Inspection Camera Probe comes with a centring head and a side view
mirror installation kit. The distance gauge is also included.
Introduction
Thank you for selecting the FLIR pipe inspection probe. The probe includes a digital camera with mirror and centring tool accessories, a coiled 82 ft (25 m) probe, and a distance gauge. The probe is intended for use with the VS80 High-Performance Vid-endoscope. See the VS80 manual for complete instructions on image and video management, Wi-Fi connectivity, split-screen modes, and other features. Obtain the VS80 user manual and additional translations of this user manual on the support site: https://support.flir.com
Product Description
- Handle
- Camera head
- Distance gauge
- Probe wheel with 82 ft (25 m) coiled probe
- VS80 connector
- Digital display shows probe travel distance
- Power/Reset button. Press for 3 seconds to switch ON or OFF. Short press to reset/zero the display
- Battery compartment cover
- Battery compartment screw
- Display backlight button
- Unit of measure button
- Battery status
- Unit of distance measure (meters)
- Unit of distance measure (feet)
Safety
Safety Note
Please read and understand all safety cautions before use.
CAUTION
Do not insert the probe into flammable liquid or gas. This product is designed
for industrial use only. It is not intended for use in hu-man or other
biological inspection.
CAUTION
To clean the probe, wipe with a soft cloth dampened with a mild detergent
and clean water. Do not use abrasive detergents, solvents, corrosives, or
alcohol to clean the probe. Use a high quality lens cleaning fluid and a lint-
free swab to clean the camera and worklight lenses. Do not disassemble the
product, damage to the product and electrical shock may occur. Avoid direct
sun exposure. Store in a cool, dry, and well-ventilated area.
- The probe contains sensitive circuitry that controls the optics and electronics of the camera. Do not strike the probe or allow it to impact objects forcefully.
- The following liquids have been approved for sub-mersion of the camera probe: Brake and transmis-sion fluid, diesel fuel, unleaded gasoline, engine oil, and CPC hydraulic oil (46AWS).
Connect the VS80
The steps below explain how to safely connect the probe to the VS80 monitor. Always switch the VS80 power off before connecting the probe.
- Align the white dot on the probe connector with the white dot on the VS80 monitor connector.
- Plug the probe into the monitor unit, ensuring proper alignment and full insertion.
- Tighten the collar nut to secure the connection. Do not overtighten.
Camera Accessories
The kit includes a side view mirror and two centring tools (Ø13 and Ø20 mm), shown above, left to right. The mirror tool provides a side view of the inspection area, and the centring tools maintain a straight trajectory for the probe in a variety of pipe sizes.
Side View 70° Mirror Accessory
- 70° mirror
- Align the probe work light with the opening on the accessory
- Threaded probe connection
Note that the mirror’s protective film must be removed before use. Clean the mirror with a lint-free swab and good-quality lens cleaner. Remove dust and moisture droplets with compressed air when necessary.
Attach the accessory as shown in the illustration below.
Centring Head Installation
Centring Head with Side View Mirror Installation
Distance Gauge
The backlit electronic gauge (3) measures and displays the distance the probe travels into the inspection area in meters and feet. The gauge tracks forward and reverses the movement.
Powering the Distance Gauge
The distance gauge is switched ON or OFF by pressing the power button (7) for
three seconds. After one hour, the gauge will automatically switch off, to
conserve battery energy.
The battery status icon (12) shows the current battery power. To replace the batteries, follow the steps below:
- Switch the meter off by pressing the power button (7) for three seconds.
- Loosen and remove the battery compartment screw (9).
- Remove the compartment cover (8).
- Remove the old batteries and dispose of them responsibly.
- Install the two new AAA batteries, observing the correct polarity.
- Close the compartment and secure it with the compartment screw.
Change the Measurement Units
The display shows the probe travel distance in meters (m) and feet (ft), short
press the units button (11) to toggle the displayed units.
Reset (Zero) the Distance Gauge Display
- With the gauge powered, short press the power/reset button (7) to reset the display to zero.
- For the best distance accuracy, zero the display at the point of probe insertion into the inspection area.
- If you wish to set a reference point, other than the starting point, press the reset button at any time during an inspection. Now you can monitor the distance traveled from this new reference point.
Distance Gauge Backlight
Short press the backlight button (10) to switch the display backlight ON or
OFF. The backlight will automatically switch OFF after 10 seconds to conserve
battery energy.
Measurement Considerations
- Ensure that the camera lens cap is removed before inserting the probe into the pipe.
- Store the lens cap safely while it is removed.
- For protection, attach the lens cap when storing the camera.
- After connecting the probe to the VS80, verify that a clear cam-era image is displayed. If not, or if the image is flickering or otherwise compromised, remove the probe and reattach it. Contact FLIR support for assistance if issues persist.
- Turn the probe wheel (4) slowly when moving the probe through a pipe. Do not force the wheel if resistance is detected. Check the nature of the resistance and clear as necessary before continuing.
- Use the distance gauge (3) to keep track of the probe travel distance. Zero the display when you first insert the probe into the pipe for the best accuracy.
- Take still images and video, adjust the display parameters, and learn of the other features available on the VS80 as described in the VS80 manual.
- When the inspection is complete, carefully recoil the probe slowly onto the storage wheel. Clean with a damp cloth and mild detergent as necessary. Use a high-quality lens cleaning fluid and a lint-free swab to clean the camera, mirror, and work light lenses.
Specifications
Camera Specifications
Camera image resolution | 640 * 480 pixels |
---|---|
Frame rate | 30 frames per second (fps) |
Probe head and body diameter | 0.4 in (10 mm) |
Probe head length | 0.7 in (16.7 mm) |
Probe length | 82 ft (25 m) |
Probe head material | Stainless steel |
Worklights | One front facing and one side facing LED |
Field of View (FOV) | 87° |
Depth of Field (DOF) | 0.4 to 4.0 in (1 cm to 10 cm) |
Accessories | Mirror adaptor and centring tools |
Operating temperature | 32℉ to 140℉ (0℃ to 60℃) |
Storage temperature | 14℉ to 140℉ (-10℃ to 60℃) |
Water-proof, ingress rating | Probe tip and body: IP 67 (IEC 60529); |
Protected from submersion in water up to 3.3 ft (1 m) for 30 minutes, maximum. In use: rain, splashes, and accidental submersion
Liquids approved for probe submersion| Brake and transmission fluid, diesel fuel, un- leaded gasoline, engine oil, and CPC hy- draulic fluid (46AWS)
Distance Gauge Specifications
Display type | Backlit LCD |
---|---|
Display digits | 3 digits with 0.1 resolution (99.9) |
Unit of measure | Meters (m) and feet (ft); selectable |
Accuracy | 1.3 ft (0.4 m) |
Control buttons | Power/reset, unit of measure, and display backlight |
Power | Two (2) AAA batteries, included |
≥ 90 hours operation
---|---
Operating and storage temperature| 14℉ to 120℉ (-10℃ to 60℃)
Dimensions| 2.4 x 4.3 x 10.0 in (62 x 110 x 25 mm)
Warranty and Support
Three-Year Limited Warranty
This product is protected by FLIR’s 3-Year Limited Warranty. Visit
www.flir.com/testwarranty to read the
3-Year Limited Warranty document.
CUSTOMER SUPPORT
Customer Support Telephone List|
https://support.flir.com/contact
---|---
Repair, Calibration, and Technical Support|
https://support.flir.com
Wlastebsitepage
http://www.flir.com
Customer support
http://support.flir.com
Copyright
© 2022, FLIR Systems, Inc. All rights reserved worldwide.
Disclaimer
Specifications are subject to change without further notice. Models and
accessories are subject to regional market considerations. License procedures
may apply. Products described herein may be subject to US Export Regulations.
Please refer to exportquestions@flir.com
with any questions.
References
- support.flir.com
- Thermal Imaging, Night Vision and Infrared Camera Systems | Teledyne FLIR
- Test & Measurement Warranties | Teledyne FLIR
- support.flir.com
- support.flir.com/contact
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>