SPECTRO-UV ML-3500S MAXIMA Ultra High Intensity UV-A Lamp Instruction Manual

October 30, 2023
SPECTRO-UV

SPECTRO-UV ML-3500S MAXIMA Ultra High-Intensity UV-A Lamp

SPECTRO-UV-ML-3500S-MAXIMA-Ultra-High-Intensity-UV-A-Lamp-
product

MAXIMATM ML-3500 Series Micro Discharge Light (MDL) UV-A Lamps

The MAXIMATM ML-3500 series ultra-high intensity UV lamps feature micro- discharge light (MDL) technology to make all inspections easier, safer, and more reliable. The MAXIMA series lamps are available in three line-voltage models: ML-3500S, with a spot bulb; range. All M model MAXIMA lamps operate in both alternating (AC) or direct current (DC). Each volt, 7 amp/hr rechargeable battery that will operate the lamp for two hours, and a battery charger and carrying case. Each comes equipped with a detachable AC cord for the country of shipment (120V, 230V, 240V, or 100V).

IMPORTANT SAFETY INSTRUCTIONS

  1. Read all instructions.

  2. Care must be taken, as burns can occur from touching hot parts.

  3. Do not operate the lamp with a damaged cord or if the lamp has been damaged.

  4. Wear UV-absorbing protective eyewear (such as the UVS-30 glasses provided) when operating this or any UV lamp.

  5. The filter lens will become hot during use. Never operate the
    lamp with the UV filter lens removed.

  6. Never use this equipment in any manner not specified in these instructions because the protection provided by the equipment may be impaired.

WARNING: UV light is emitted from this product. Avoid eye and skin exposure to unshielded products. FOR PROFESSIONAL EXAMINATION USE ONLY.

ML-3500S AC OPERATION

  1. Connect the plug to a power source that complies with the electrical requirements specified.
  2. Use UVS-30 UV-absorbing protective eyewear when operating the lamp.

ML-3500MS AC/DC MODELS FOR AC OPERATION

  1. Align the top-notch of 7-pin plug on the AC cord to the notch in the ballast outlet; turn the locking ring to the right over the plug to secure it.
  2. Press ON the BLACK ballast rocker switch.
  3. Press ON the RED Inverter rocker switch.
  4. Use UVS-30 UV-absorbing protective eyewear when operating the lamp.

ML-3500MS AC/DC MODELS FOR DC OPERATION

  1. Refer to the BP-12A operating instructions for full information.
  2. Insert the plug into the receptacle on the battery pack.
  3. Align the top-notch of 7-pin plug on the DC cord to the notch in the ballast outlet; turn the locking ring to the right over the plug to secure.
  4. Press ON the BLACK ballast rocker switch.
  5. Press ON the RED Inverter rocker switch.
  6. Use UVS-30 UV-absorbing protective eyewear when operating the lamp.
  7. IMPORTANT: Disconnect the lamp cord plug from the battery pack when the unit is not in use.

INTRODUCTION

The MAXIMA™ ML-3500 series ultra-high intensity UV lamps feature micro- discharge light (MDL) technology to make all inspections easier, safer, and more reliable. The MAXIMA series lamps are available in three line-voltage models: ML-3500S, with a spot bulb; range. All “M” model MAXIMA lamps operate in both alternating (AC) or direct current (DC). Each
volt, 7 amp/hr rechargeable battery that will operate the lamp for two hours, and a battery charger and carrying case. Each comes equipped with a detachable AC cord for the country of shipment (120V, 230V, 240V, or 100V).

WARNING:

  • UV light emitted from this product.
  • Avoid eye and skin exposure to unshielded products.
  • FOR PROFESSIONAL EXAMINATION USE ONLY.

SAFETY INSTRUCTIONS

IMPORTANT

  1. Read all instructions.
  2. Care must be taken, as burns can occur from touching hot parts.
  3. Do not operate the lamp with a damaged cord or if the lamp has been damaged.
  4. Wear UV-absorbing protective eyewear (such as the UVS-30 glasses provided) when operating this or any UV lamp.
  5. The filter lens will become hot during use. Never operate the lamp with the UV filter lens removed.
  6. Never use this equipment in any manner not specified in these instructions because the protection provided by the equipment may be impaired.

NOTE: Due to the ultra-high UV-A intensity produced by this lamp, subject (penetrants, metals, plastics) fluorescence may fade after a period of exposure. Therefore, this lamp should not be left shining on any one area for an extended period.
ML-3500S AC OPERATION

  1. Connect the plug to a power source that complies with the electrical requirements specified on the lamp’s label. For best performance and durability, the power source should be regulated, surge free and noise free. In those regions where mains may experience fluctuations greater than ± 10%, an overvoltage protection device is available.
  2. Press ON the BLACK ballast rocker switch.

ELECTRICAL SPECIFICATIONS

ML-3500S MODELS

MAXIMA

Model

| ML-3500S| ML-3500S/F| ML-3500S/FB| ML-3500S/FA| ML-3500/SJ
---|---|---|---|---|---


Power Source

| 120V/60Hz/

1.00A

| 230V/50Hz/

0.75A

| 230V/50Hz/

0.75A

| 240V/50Hz/

0.75A

| 100V/50-60Hz/

1.00A

ML-3500MS AC/DC MODELS FOR AC OPERATION

  1. Align the top notch of the 7-pin plug on the AC cord to the notch in the ballast outlet; turn the locking ring to the right over the plug to secure.
  2. Press ON the BLACK ballast rocker switch.
  3. Press ON the RED Inverter rocker switch.

FOR DC OPERATION

  1. Connect the 12-volt plug to the Spectroline® BP-12A 12-volt battery pack or to a suitable DC supply. The BP-12A must be fully charged before using! Refer to the BP-12A operating instructions for full information.
  2. Insert the plug into the receptacle on the battery pack.
  3. Align the top notch of the 7-pin plug on the DC cord to the notch in the ballast outlet; turn the locking ring to the right over the plug to secure.
  4. Press ON the BLACK ballast rocker switch.
  5. Press ON the RED Inverter rocker switch.

NOTE: The BP-12A battery pack includes a 12-volt battery, 120V/60Hz- charger, the BP-12A/F has a 230V/50Hz charger, the BP-12A/FA has a 240V/50Hz charger and the BP-12A/J has a 100V/50-60Hz charger. All battery packs include a nylon carrying case. IMPORTANT: Disconnect lamp cord plug from battery pack when the unit is not in use.
ELECTRICAL SPECIFICATIONS—ML-3500MS AC/DC MODELS

MAXIMA Model ML-3500MS ML-3500MS/F ML-3500MS/FA ML-3500MS/J
Power Source 120V/60Hz/

1.00A

| 230V/50Hz/

0.75A

| 240V/50Hz/

0.75A

| 100V/50-60Hz/

1.00A

BP-12A

Power Source

| 12VDC| 12VDC| 12VDC| 12VDC

OPERATOR SERVICING

  • The following instructions are intended for limited servicing of the lamp. Do not perform any servicing that is not covered in these instructions. Return the lamp to the factory for maintenance not covered in this section.

HOUSING AND FILTER MAINTENANCE

  • CAUTION: Always shut off the switch on the ballast and unplug the ballast from the power outlet before performing any maintenance or servicing. Allow the lamp to cool before handling.
  • Clean the lamp and ballast housings with water and a mild detergent and wipe dry. Periodically clean the filter with a glass cleaner and a soft cloth.

FILTER REPLACEMENT

  • The lamp is shipped with a 2F958 filter lens. This may be used with either the
  • BLE-35RA spot bulb or the BLE-35RAF flood bulb. The 2F350 filter lens may be ordered with a BLE-35RA spot bulb (only) to produce a more diffused spot beam that is not as wide as the flood beam.
  1. Peel the silicone-rubber filter holder from around the circumference of the lamp housing. The filter will remain inside the filter holder.
  2. Press the existing filter out of the filter holder.
  3. Press the new filter gently but firmly into the filter holder.
  4. Reattach the filter holder by pressing it around the circumference of the lamp housing with the heel of your hand. Be sure the groove in the filter holder seats over the rim of the housing.

BULB REPLACEMENT

  • Use only a Spectroline UV bulb/reflector assembly.
  • Using any other bulb may damage the unit and will void the warranty.
  • The lamp is shipped with a BLE-35RA spot bulb/reflector assembly as standard.
  • The lamp may be ordered with a BLE-35RAF flood bulb/reflector assembly.
  1. Remove the filter holder.
  2. Insert the bulb/reflector tool in the two holes in the old bulb/reflector, rotate counterclockwise and pull.
  3. Hold the new bulb/reflector by its protective cover while inserting it into the socket with a clockwise quarter-turn bayonet action. Firm pressure is needed to overcome the spring in the socket. Pop off the black fasteners and remove the protective cover.
  4. Reattach the filter holder.

UV INTENSITY VERIFICATION

  • To ensure that your MAXIMA™ ML-3500 is operating at the required UV intensity, it should be checked periodically. The AccuMAX™ XRP-3000 radiometer/photometer kit uses a dual- wavelength UV-A/VIS sensor detector to accurately measure ultraviolet irradiance or visible light. The AccuMAX XR-1000 digital readout unit with the XS-365 UV sensor or the Spectroline DM-365XA digital readout unit are also recommended for accurate UV-A measurement. The meters are specially designed to measure the UV irradiance from 320-400m with a peak at 365m.
  • For more information about these units, please contact the Customer Service Department at Spectro-UV. Call toll-free 1-866-230-7305.

ENVIRONMENTAL CONDITIONS

  • These UV lamps are designed to be safe under the following conditions:
  • Indoor use
  • Altitude up to 2,000 m (6,562 ft)
  • Temperature 5°C to 40°C (41°F to 104°F)
  • Maximum relative humidity 80% for temperatures up to 31°C (88°F decreasing linearly to 50% relative humidity at 40°C (104°F)
  • Mains supply voltage fluctuations not to exceed +10% of the nominal voltage
  • Installation Category I
  • Pollution Degree 2

LIMITED WARRANTY

  • The warranty policy for these UV lamps is provided on the Certificate of Limited Warranty enclosed separately with each unit.
  • NOTE: Please contact the Customer Service Department at Spectro-UV for assistance. Have the model and serial numbers of the unit and the date of purchase available when calling.

PARTS LIST

ITEM PART DESCRIPTION
1 122283 BULB REMOVAL TOOL
3 123604 FILTER HOLDER

5

| 2F958| FILTER, LENS
2F350| FILTER, LENS (DIFFUSING)
2F100C| FILTER, LENS (CLEAR)


7

| BLE-35RA| BULB/REFLECTOR ASSEMBLY, 35W SPOT
BLE-35RAF| BULB/REFLECTOR ASSEMBLY, 35W FLOOD
9| 125060| IGNITER ASSEMBLY
9A| 124863| ADAPTER
9B| 125061| IGNITER
11| 124736| LAMP HOUSING
13| 100310| PISTOL GRIP
15| 103976| STRAIN RELIEF
17| 125943| CORD, SECONDARY, 8-FT
19| 123666| BALLAST, HID, 35W, AC, ELECTRONIC ASSEMBLY (NOT SHOWN)
21| 123667| INVERTER, 100W, 12V DC TO 120V AC (NOT SHOWN)
23| 100253| BALLAST SWITCH, ROCKER
* 25| 119797| ML-3500S CORD, PRIMARY, 8-FT, AC, 120V, 100V (SEE 125462)
WC-0105| ML-3500S/F, AC, 230V
WC-0127| ML-3500S/FB. AC. 230V
WC-0144| ML-3500S/FA, AC 240V
**
27| 124828| ML-3500MS CORD, 7-PIN, 8-FT, AC, 120V, 100V
124827| CORD, 7-PIN 8-FT, AC, 230V
125186| CORD, 7-PIN, 8-FT, AC, 240V
29
| 124826| CORD, 7-PIN, 2-FT, DC
31**| BP-12A| 12V DC BATTERY PACK
33| 122122| INVERTER SWITCH, ROCKER

  • Each of the primary cords in the ML-3500S series lamps is fitted with the appropriate plug for the country of shipment.
  • Each of the ML-3500MS series models are equipped with the AC/DC cords and battery chargers appropriate for the country of shipment.
  • Caution (refer to accompanying documents)
  • 1.888.610.7664
  • www.calcert.com
  • [email protected]

Documents / Resources

| SPECTRO-UV ML-3500S MAXIMA Ultra High Intensity UV-A Lamp [pdf] Instruction Manual
ML-3500S MAXIMA Ultra High Intensity UV-A Lamp, ML-3500S, MAXIMA Ultra High Intensity UV-A Lamp, Ultra High Intensity UV-A Lamp, High Intensity UV-A Lamp, Intensity UV-A Lamp, UV-A Lamp, Lamp
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

SPECTRO-UV User Manuals

Related Manuals