COREMORROW P60.X250S Piezo Nanopositioning Stage User Manual
- June 9, 2024
- COREMORROW
Table of Contents
- COREMORROW P60.X250S Piezo Nanopositioning Stage User Manual
- This document describes the following products:
- DECLARATION
- 1.1 Purpose and content
- 1.2 Symbol and meaning
- 1.3 Expected reader
- 1.4 Manual lost replacement
- 2.Safety Guide
- 3.Product Overview
- 3.1 Product Introduction
- 3.2 Product Features
- 3.3 Motion direction
- 4 . Operating environment
- 5. Installation and operation
- 5.1 Installation and connection
- 5.1 Load
- 5.1 Operating
- 6.1 Transportation
- 6.2 Unpacking inspection
- 7.1 Problem Processing
- P60.X250 Piezo Nanopositioning Stage User Manual
- 7.2 Daily maintenance
- Daily cleaning:
- 8.1 Driving and Sensor cable
- 9.Customer Service
- 10.Contact us
- Harbin Core Tomorrow Science & Technology Co., Ltd.
- Read More About This Manual & Download PDF:
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
COREMORROW P60.X250S Piezo Nanopositioning Stage User Manual
This document describes the following products:
- X250S
- X250k
DECLARATION
Thank you for choosing Coremorrow’s products!
This user manual is applicable to P60.X250 series piezo nanopositioning stage.
For your smooth and correct use of this product, please read this manual
carefully before use. Please follow the instructions in the manual
installation and use procedures. Improper operation may cause injury to the
operator or damage to the product. Therefore, the entire installation and
operation process should be completed by a professional with a certain
foundation and understanding of the product principle, or under the guidance
of a professional.
If the product is disassembled or modified without permission, our company
will not be responsible for any consequences arising therefrom.
It is illegal to change the product model and other false sales of our
products. Users are advised to be vigilant. Once you find out or be deceived,
contact us directly, we will crack down on illegal activities and prevent
deception and economic losses. Coremorrow will certainly investigate legal
responsibility of the illegal subject and related parties.
With the continuous development and innovation of technology, we will add the
latest information to the manual as needed without notice. Please contact us
if you need any information.
1.1 Purpose and content
The manual is mainly for your smooth and correct use of piezo stage.
The manual contains all necessary information for P60.X250 series piezo stage.
The manual describes the precautions during installation and use.
1.2 Symbol and meaning
DANGER: Improper handling may could cause injury to the operator. Attention: Improper handling can cause damage to the equipment. No order between items.
- Please operating in
1.3 Expected reader
The manual default user has knowledge of piezo stages and controllers, and has a basis for its installation and use.
1.4 Manual lost replacement
If the manual is accidentally lost, please contact our customer service department, we will send you the latest version of user manual.
2.Safety Guide
P60.X250 piezo nanopositioning stage is designed with advanced technology and
safety standards, for safety and proper use of the product, please pay
attention to following points when using:
Temperature changed and pressure will
charge the piezo stage. Piezo stage will remain charged for a period of time
disconnected from the controller, so the
operator should pay attention when using it:
Do not disassemble P60.X250 piezo stage;
Discharge P60.X250 piezo stage before installation, it could be achieved by
connecting to the controller;
Do not pull out the controller during operation;
Before using, please check whether the connecting wire of the product is in
good condition, whether the product and the controller are effectively
protected from grounding, in case of danger, please strictly regulate the
operation during operation, do not touch it (up to 150V voltage) after power
on.
In order to prevent damage to the product caused by improper operation, please pay attention to following points when using:
P60.X250 piezo stage is used as much as possible in dust-free, oil-free,
lubricant-free environments;
Since the product adopts flexible structure design, it is recommended that the
load not exceed the bearing capacity of the product. Pay attention to the
torque when loading. Do not twist or twist the load surface to avoid damage to
the structure; Recommended voltage 0~120V for long-term operation to extend
lifetime;
Do not disassemble the product to avoid product damage;
Avoid stretching and bending the cable interface to prevent damage to the
cable; Please use the dedicated cable provided by us to connect the piezo
stage and piezo controller;
Please do not use cable extension wire, if need longer cable, please contact
us.
3.Product Overview
3.1 Product Introduction
P60.X250 piezo nanopositioning stage has the following advantages: High
performance piezo stack:
P60.X250 piezo stage is driven by high-performance piezo stack. The piezo
stage is fully insulated, so its performance and service life are far superior
to those of conventional stage. The insulating layer could effectively prevent
the piezo stack from being damaged by moisture. The stability of piezo stack
could be ensured under extreme conditions. The piezo stage has the
characteristics of no rebound, no wear. Flexible hinge structure:
P60.X250 adopts flexible hinge mechanism, which has the advantages of no friction and high positioning accuracy. The flexible hinge is a component that is not affected by static and dynamic friction. It is based on the elastic deformation of solid (such as steel) and has no rolling and sliding parts. With high stiffness and load carrying capacity, the hinge guides are maintenance-free and wear-free. Suitable for a wide range of temperatures without the need for lubricants. SGS sensor:
SGS sensor detects position information by expanding, they are fixed at
appropriate positions in transmission portion, and the displacement is
measured.
P60.X250 piezo stage could travel up to ±165μm. It could also be customized
according to the requeirements, with high flexibility and easy operation to
meet different applications. P60.X250 is mainly used in biotechnology,
interferometry, quality inspection, optical detection, nanopositioning,
microscopy and imaging, image processing and other fields, and has very good
development prospects.
3.2 Product Features
Quick response time;
Closed-loop SGS sensor with high positioning accuracy; Mainly used for high
frequency.
3.3 Motion direction
4 . Operating environment
Operating temperature: -20~80℃
Environment free of dust, clean, and non-corrosive substances Relative
humidity: <60%
Recommended voltage between 0~120V for long-term and high-reliability
operation
5. Installation and operation
Before installation, first check the safety, such as whether the cable is damaged or broken, and whether the power supply has grounding protection.
5.1 Installation and connection
- Ensure that the installation plane is
- Secure the piezo stage with M4 screws and
- Install the load and tighten properly to ensure positioning
- Connect the cable to the corresponding interface of the controller according to the connector
P60.X250 installation diagram is as follows:
5.1 Load
When installing the load, ensure that the load mass is less than the max load
of P60.X250, and install the load as much as possible at the center of the
moving surface.
The correct placement of the load is shown below:
For protection, we don’t recommend the load is too high. If it is used a higher load, please reduce the position center of the load gravity. The error placement method is as follows:
It is forbidden to load the load on one side of the piezo stage, which would cause serious damage to the product, as shown in the following figure:
If the screws used in tightening and connecting are too long, the screws will
not be installed or some of screws will leak out, which will scratch the stage
and affect the motion accuracy. If the screws are too short, the connection
will not be compact and easy to loosen. Looseness could damage the internal
structure of the stage.
We recommend using the connection method shown below:
Please do not connect as shown below:
Note:
Slowly tighten the load when installing it, and the torque should not be too large
to avoid damage to mechanism.
Avoid stretching and bending cables.
Please connect P60.X250 piezo nanopositioning stage and controller with special cable provided by CoreMorrow.
Please do not extend the cable without authorization, please contact us if you need a longer cable.
5.1 Operating
Please confirm the following information before starting the power supply: P60.X250 piezo nanopositioning stage has been grounded.
The environment meets requirements of P60.X250 working environment. The fixing screws must be tightened.
The above information is determined to perform controller related operations:
- Please carefully read piezo controller user manual before operating, then start the controller;
- Set the piezo controller to zero;
- After checking for no abnormalities, input voltage values according to requirements (keyboard or analog control);
- It could work normally within the rated voltage range of piezo ceramic (0~120V, max not exceeding 150V);
- Please set the piezo controller to zero after
Please do not touch the product during the operation, avoid affecting the piezo stage performance or even damage the piezo stage.
The resolution of piezo ceramic is not limited. It depends on the performance of piezo controller. In order to better achieve the motion accuracy, it is recommended to use CoreMorrow piezo controller. For the operation of piezo controller, please refer to piezo controller usual manual.
P60.X250 Piezo Nanopositioning Stage User Manual
Note:
Please do not remove the grounding protection during operation. If it needs to be
temporarily removed, reconnect the grounding protection before starting again.
Drive voltage should not exceed 150V;
If abnormal noise or oscillation occurs during operating, immediately turn off the power check parameter setting.
|
6.1 Transportation
Piezo stage is packed in carton. The transportation is carried out under the condition of packaging. This product can be transported by various means of transportation under normal conditions. During transportation, it should avoid direct rain and snow, avoid contact with corrosive substances, and avoid strong collision, try to avoid improper behavior such as squeezing, irregular placement.
6.2 Unpacking inspection
Please check below information:
Before opening the package, please carefully check the packaging for damage,
bruising, wetting, moisture, deformation, etc.
Please check the parts on a case-by-case basis based on the supply contract
and the packing list
Whether there is obvious damage on the surface of the product Whether the
product identification is clear and complete
Whether the connector is loose, the connection line is broken or not, etc.
If any of the above occurs, please make a detailed record and take a photo,
and please contact us immediately
Precautions:
Please handle gently when unpacking to avoid permanent damage to the stage
caused by bumps
Please do not use heavy tools or use rough methods to open them Please do not
discard all packaging materials for return shipments
7.1 Problem Processing
Common problems are shown in the following table:
Problem Description | Possible reasons | Solutions |
---|
No displacement or displacement becomes smaller
|
The cable is not connected correctly or is in faulty contact.
|
Check the cable connection
Excessive load
|
Loaded by product max load
Sensor zero drift
|
Calibrate the sensor
Reduced accuracy
|
The installation base is uneven
|
Mount the stage on flat surface
Loose connection
|
Fastening connection
Wiring is wrong
|
Check the wiring is correct
Vibration or inaccurate positioning at start-up
|
The parameters are incorrect.
|
Stop immediately, check the parameter settings
Resonance
|
operating frequency less than the resonant frequency
P60.X250 Piezo Nanopositioning Stage User Manual
If the user has problems that could not be solved. Please record the fault
situation and contact with us. The professional technical engineer would help
solve the problem.
Problems caused by improper operation are not covered by the warranty;
CoreMorrow is not responsible for any problems caused by dismantling the peizo
stage without permission.
7.2 Daily maintenance
If the piezo stage is not used for a long time, please store it in an room environment with no dust, good ventilation, clean and non-corrosive substances.
Please do not loosen the screws on the stage structure, avoiding to affect the accuracy of the stage.
Daily cleaning:
Please disconnect the piezo stage to controller, then clean the surface of it.
When cleaning, first wipe a small amount of detergent with a towel and gently
wipe the surface of the piezo stage, then wash it with alcohol. Please do not
use ultrasonic waves for cleaning.
When cleaning the surface, try not to pick up the product. If it is necessary
to pick up, please place a sponge under the piezo stage for protection, the
piezo stage should not be too high from the sponge to prevent the piezo stage
from falling off.
8.1 Driving and Sensor cable
Driving cable interface is 1-contact LEMO, model number is LEMO ERA.00.250.CTL, as shown below:
Sensor cable interface is 4-contacts LEMO, model number is LEMO
ERA.0S.304.CLL, as shown below:
LEMO Sensors Connector:
9.Customer Service
If you have questions about the products you are currently using, please let
us know the following information:
Piezo stage model number and related number Controller model for this product
Software driver version for this product
Computer operating system for installing supporting software