Laserliner ThermoMaster Plus Thermometer User Manual
- June 10, 2024
- Laserliner
Table of Contents
- ThermoMaster Plus
- Function / Application
- General safety instructions
- Safety instructions
- Symbols
- General instructions
- [1] Inserting batteries
- [2] Connecting thermocouples
- [3] ON OFF
- [4] HOLD function
- [5] Temperature measurement (T1, T2)
- [6] Show MAX / MIN / AVG / T2-T1
- [7] Memory function / call up memory
- [8] Auto power off
- [9] Temperature alarm
- [10] Temperature alarm LO
- [11] Temperature alarm HI
- [12] Offset temperature T1 / T2
- [13] Setting unit of temperature
- [14] LCD backlight
- [15] Setting type of thermocouple
- [16] Clearing memory
- [17] Restoring factory setting
- [18] Standard view
- Data transfer
- Application (app)
- Information on maintenance and care
- Calibration
- SERVICE
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ThermoMaster Plus
DUAL INPUT SENSOR TYPE K -50°C…300°C
| Completely read through the operating instructions, the “Warranty and
Additional Information” booklet as well as the latest information under the
internet link at the end of these instructions. Follow the instructions they
contain. This document must be kept in a safe place and passed on together
with the device.
---|---
Function / Application
The digital thermometer is used to measure temperature and temperature differences with the aid of exchangeable type K/J/T/E thermcouples/sensors. The digital thermometer is predominantly used for temperature measurement in laboratories and in industrial applications. With the aid of the MAX function it is possible to determine limit temperature infringements in extended series of measurements.
General safety instructions
– The device must only be used in accordance with its intended purpose and
within the scope of the specifications.
– Not suitable for potentially explosive areas or medical diagnoses.
– The measuring tools and accessories are not toys. Keep out of reach of
children.
– Modifications or changes to the device are not permitted, this will
otherwise invalidate the approval and safety specifications.
– Do not expose the device to mechanical stress, extreme temperatures,
moisture or significant vibration.
– Do not power the thermocouple (K-type) with an external voltage.
– The device must no longer be used if one or more of its functions fail or
the battery charge is weak.
– Please ensure compliance with the safety regulations set out by local and
national authorities with regard to the correct and proper use of the device.
Safety instructions
Dealing with electromagnetic radiation
– The measuring device complies with electromagnetic compatibility regulations
and limits in accordance with the EMC Directive 2014/30/EU which is covered by
the Radio Equipment Directive 2014/53/EU.
– Local operating restrictions – for example, in hospitals, aircraft, petrol
stations or in the vicinity of people with pacemakers – may apply. Electronic
devices can potentially cause hazards or interference or be subject to hazards
or interference.
– The measuring accuracy may be affected when working close to high voltages
or high electromagnetic alternating fields.
Dealing with RF radiation
– The measuring device is equipped with a wireless interface.
– The measuring device complies with electromagnetic compatibility and
wireless radiation regulations and limits in accordance with the RED
2014/53/EU.
– Umarex GmbH & Co. KG hereby declares that the ThermoMaster Plus radio-based
system complies with the requirements and other stipulations of the European
Radio Equipment Directive 2014/53/EU (RED).
The EU Declaration of Conformity can be found in its entirety at the following
address:
http://laserliner.com/info?an=AHQ
Symbols
Warning about hazardous electrical voltage: Unprotected live components inside the device housing are capable of posing a risk of electric shock.
Danger area warning
Protection class II: The test device has reinforced or double insulation.
Earth potential
Read operating instructions
General instructions
– If the thermometer is subject to large fluctuations in the ambient
temperature, after the temperature has stabilised, wait 20 minutes before
performing a measurement.
– Always ensure adequate thermal coupling at the measuring point in order to
avoid measuring errors caused by temperature loss.
– Please note that all thermometers with a contact sensor influence the
measurement and their thermal capacity can reduce the actual temperature. More
thermal energy should therefore be applied to the thermocouple than it can
dissipate.
– Four lines will appear in line A if no thermocouple is connected.
– Lo or Hi will be displayed if the measured temperature is outside the
measuring range.
– Only use the correct type of thermocouple (type K, J, T or E) and make sure
that the device is set to the corresponding type. The wrong type can cause
significant measuring errors.
– In addition to being subject to ageing, a thermocouple is also greatly
depending on the operating conditions and should therefore be checked
regularly.
– High pressure or mechanical deformation can change the grid structure and
therefore have an influence on the thermoelectric voltage output.
– The thermometer and thermocouples have different measuring ranges and
accuracies and therefore need to be considered separately.
[1] Inserting batteries
Open the battery compartment and insert batteries according to the symbols. Be sure to pay attention to polarity.
[2] Connecting thermocouples
| Pay attention to the polarity information on the thermocouple as well as on
the device connection.
---|---
1 Hold current measured value
2 ON/OFF
3 Reset MAX / MIN / AVG; alarm OFF
4 Change menu selection
5 Show MAX / MIN / AVG / T2-T1 for T1
6 Battery compartment (rear)
7 Memory function
8 Settings menu
9 Exit menu / Switch off alarm
10 Show MAX / MIN / AVG / T1-T2 for T2
11 Change menu selection
12 Input for thermocouple T2
13 Input for thermocouple T1
* Two connected thermocouples
One connected thermocouple
a Measured value T1* / Measured value T2
b* Measured value T2
c Battery charge
d HOLD function
e Digital Connection
f Alarm
g Auto power off
h Offset temperature T1
i Offset temperature T2
j Thermocouple type
k Memory
l Menu
m T2-T1: Value T2-T1
n T1: Thermo-couple T1
T2: Thermo-couple T2
MAX: MAX value
MIN: MAX value
AVG: Average value
T1-T2: Value T1-T2
°C K:** Units
[3] ON OFF
[4] HOLD function
The HOLD function shows the last measurement or measured value in the display.
[5] Temperature measurement (T1, T2)
With one thermocouple
When measuring with one thermocouple, the measured value of the connected thermocouple T1 or T2 is shown in line A. The display shows whether T1 or T2 is connected.
With two thermocouples
The measured value of thermocouple T1 is shown in line A and the measured value of thermocouple T2 in line B.
[6] Show MAX / MIN / AVG / T2-T1
With one thermocouple
| TIP: When line A is set to MAX and line B to MIN a direct MAX/MIN
comparison can be made with one thermocouple.
---|---
With two thermocouples (Example T1)
Press button A to show the MAX, MIN, AVG values of thermocouple T1 as well as the difference T2-T1.
Press button B to show the MAX, MIN, AVG values of thermocouple T2 as well as the difference T1-T2.
[7] Memory function / call up memory
The device has 10 storage locations.
The current measured value display is stored at the next free memory location by short-pressing the LOG button. An acoustic signals confirms successful storage.
The measured value memory is called up by long-pressing the LOG button.
The measured value view and storage location are shown alternately.
- Change storage location
The measured value memory is exited by short-pressing the ESC button.
[8] Auto power off
- Confirm / next menu item
- Change selection
- Next menu item
- Auto power off deactivated
- Auto power off activated
[9] Temperature alarm
By activating the “Temperature alarm” function, deviations from the required temperature range are indicated by the (f) symbol on the display flashing and a two-stage (10 minutes / 3 seconds) acoustic signal. The temperature alarm can be cancelled by pressing the ESC button and is set to “OFF” in the menu.
You can set the temperature range, see section 10 “Temperature alarm LO” and section 11 “Temperature alarm HI”.
- Confirm / next menu item
- Change selection
- Alarm (10 minutes)
- Next menu item
- Alarm (3 seconds)
- Alarm deactivated
[10] Temperature alarm LO
- Change selection
- Next menu item
- Factory setting: -40 °C
[11] Temperature alarm HI
- Next menu item
- Factory setting: 400 °C
[12] Offset temperature T1 / T2
- Change selection
- Confirm selection
- Offset T1 activated
- Offset T1 deactivated
- Highest value: 10 °C
- Lowest value: -10 °C
- The display shows that an offset temperature is set.
[13] Setting unit of temperature
- Confirm / next menu item
- Change selection
- Next menu item
- Kelvin
- ° Celsius
[14] LCD backlight
- Change selection
- Next menu item
[15] Setting type of thermocouple
- Change selection
- Next menu item
- Type E
- Type T
- Type J
- Type K
[16] Clearing memory
-
Next menu item
-
The display flashes while clearing the memory.
A signal sounds when the memory is cleared. -
| The clearing procedure cannot be stopped or cancelled.
---|---
[17] Restoring factory setting
-
Next menu item
-
The display flashes while restoring the factory setting.
A signal sounds when the factory setting is restored. -
| The clearing procedure cannot be stopped or cancelled.
---|---
[18] Standard view
The display is reset to the standard view (view after switching on with thermocouple connected) by unplugging and reconnecting a thermocouple. The standard view can also be set by pressing the ESC button.
Data transfer
This device has digital connectivity which allows wireless data transfer to mobile devices such as smart phones or tablets with a wireless interface.
The system prerequisites for a digital connection are specified at
http://laserliner.com/info?an=ble
This device can generate a wireless connection to devices which are compatible with the wireless standard IEEE 802.15.4. The wireless standard IEEE 802.15.4 is a transfer protocol for Wireless Personal Area Networks (WPAN). The range is set to a maximum distance of 10 m from the terminal device and greatly depends on the ambient conditions such as the thickness and composition of walls, sources of interference as well as the transmit / receive properties of the terminal device.
Application (app)
An app is required to use the digital connection.
You can download the app from the corresponding stores for the specific type
of terminal device:
| Make sure that the wireless interface of the mobile device is activated.
---|---
After starting the app and activating the digital connection, a connection can be set up between a mobile device and the measuring device.
If the app detects several active measuring devices, select the matching device.
This measuring device can be connected automatically the next time it is switched on.
Information on maintenance and care
Clean all components with a damp cloth and do not use cleaning agents,
scouring agents and solvents.
Remove the battery pack before storing for longer periods. Store the device in
a clean and dry place.
Do not touch the lens.
Calibration
The measuring device must be calibrated and tested on a regular basis to ensure it is accurate and working properly. We recommend carrying out calibration once a year. Contact your distributor or the UMAREX-LASERLINER service department.
Technical Data|
Subject to technical changes without notice. 21W42
---|---
Measured variable| Contact temperature
Functions| Alarm, continuous measurement, difference, hold, MIN/MAX, Mean
value
Contact temperature measuring range| Type K: -150°C … 1370°C
Type T: -150°C … 400°C
Type J: -150°C … 1200°C
Type E: -150°C … 900°C
Contact temperature accuracy| At ambient temperature 18°C … 28°C:
-150°C … -100°C (±(0.2% of measured value + 1°C))
-100°C … 1370°C (±(0,1% of measured value + 1°C))
Contact temperature resolution| 0,1°C
Thermocouple measuring range| -60°C … 300°C
Interface| Digital Connection
Connections| Thermocouple type K/J/T/E
Unit of temperature| °C (Celsius), K (Kelvin)
Memory| 10 storage locations
Auto power off| after 20 minutes
Power supply| 4 x 1.5V LR03 (AAA)
Operating time| approx. 100 hours
Radio module operating data| IEEE 802.15.4. LE ≥ 4.x (Digital Connection)
interface; Frequency band: ISM band 2400–2483.5 MHz, 40 channels; Transmission
power: max. 10 mW; Bandwidth: 2 MHz; Bit rate: 1 Mbit/s; Modulation: GFSK/FHSS
Operating conditions| 0°C … 50°C, max. humidity 80% rH, no condensation, max.
working altitude 2000 m above sea level
Storage conditions| -20°C … 60°C, max. humidity 80% rH, no condensation
Dimensions (W x H x D)| 75 x 167 x 35 mm
Weight| 216 g (incl. batteries)
EU directives and disposal
This device complies with all necessary standards for the free movement of goods within the EU.
This product is an electric device and must be collected separately for disposal according to the European Directive on waste electrical and electronic equipment.
Further safety and supplementary notices at: http://laserliner.com/info?an=AHQ
SERVICE
Umarex GmbH & Co. KG
– Laserliner –
Möhnestraße 149, 59755 Arnsberg, Germany
Tel.: +49 2932 638-300, Fax: +49 2932 638-333
info@laserliner.com
8.082.96.167.1 / Rev21W42
Umarex GmbH & Co. KG
Donnerfeld 2
59757 Arnsberg, Germany
Tel.: +49 2932 638-300, Fax: -333
www.laserliner.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>