PR-44052-BAS Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Instruction Manual

June 4, 2024
Arrow

PR-44052-BAS Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Instruction Manual
PR-44052-BAS Pressure Injectable Arrowg+ard Blue Advance Two-Lumen
PICC

Contents

  • 1: Two Lumen TaperFree® Catheter with Arrowg+ard Blue Advance® Antimicrobial/Antithrombogenic Protection1: 5.5 Fr. (1.88 mm OD) x 40 cm, Pressure Injectable, T-Port Connector, Blue FlexTip® and Placement Wire
  • 1: GlideThru™ Peel-Away Sheath: 5.5 Fr. x 2-3/4″ (7 cm) Radiopaque over 5.5 Fr. Dilator
  • 1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm) (Straight Soft Tip on One End – Straight Stiff Tip on Other)
  • 1: Safety Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW 1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
  • 1: Syringe: 10 mL Luer-Lock 2: Dust Cap: Non-Vented 1: SecondSite™ Adjustable Hub: Fastener
  • 1: SecondSite™ Adjustable Hub: Catheter Clamp
  • 1: SharpsAway® Disposal Cup
  • 1: SharpsAway® II Locking Disposal Cup
  • 1: Catheter Trimmer
  • 1: Safety Scalpel: #11 1: Patient ID Card
  • 1: Patient Information Booklet 2: Paper Tape Measure
  • 1: Tourniquet
  • 1: Dressing: STATLOCK®2 Catheter Stabilization Device

1 Licensed under US Patent No. 7,329,412.
2 A registered trademark of C. R. Bard, Inc.
All components are CE 2797 unless otherwise noted.

Rx only

Warning: Read all package insert warnings, precautions, and instructions prior to use. Failure to do so may result in severe patient injury or death. www.teleflex.com/IFU California Prop. 65 ⚠

WARNING: Cancer and Reproductive Harm. www.P65Warnings.ca.gov Not made with natural rubber latex. Store below 25°C (77°F). Avoid excessive heat above 30°C (86°F). Fluid path components are non-pyrogenic.
Contraindications: The Pressure Injectable Arrowg+ard Blue Advance antimicrobial/ antithrombogenic catheter is contraindicated:

  • for patients with known hypersensitivity to chlorhexidine
  • in the presence of device related infection in the intended insertion vessel or catheter pathway
  • in the presence of thrombosis in the intended insertion vessel or catheter pathway.

| *Priming Volume (mL)| Gravity Flow Rate† (mL/hr)| Pump Flow Rate†† (mL/hr)| MAX Pressure Injection Flow Rate (mL/sec)
---|---|---|---|---
Distal (18 Ga.) Proximal (18 Ga.)| 0.41
0.43| 330
343| 3850
3880| 5
5

  • Priming volumes are approximate and are done without accessories.
    † Flow rate values are approximate and are determined using deionized water at 100 cm head height.
    †† Pump flow rates are determined at maximum pump pressure of 10 psig and represent approximate flow capabilities.
    ** Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.

Support

EU Authorized Representative and Importer

Teleflex Medical IDA Business and Technology Park Dublin Road, Athlone Co. Westmeath, Ireland

Arrow International LLC Subsidiary of Teleflex Incorporated 3015 Carrington Mill Blvd. Morrisville, NC 27560 USA

Logo

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Arrow User Manuals

Related Manuals