Trulife A193RG Cervical Orthosis User Guide
- June 10, 2024
- Trulife
Table of Contents
A193RG, A193SM, A193PED
Lerma MINERVA Cervical Orthosis
Installation and User Guide
25486-001 REV H 2022-05-23
INTRODUCTION
The Lerman MINERVA is a cervical orthosis that consists of two shells, anterior and posterior, which are joined by torso and shoulder straps. The posterior segment has an attached occipital support, which may be moved vertically. A sliding mandibular support is attached to the anterior shell. This mandibular support may also be adjusted vertically. A forehead strap is included to provide additional support. The Lerma MINERVA is constructed entirely of non-ferrous materials.
Product Code | Size | *Measurement (Chin to Distal Edge of Brace)** |
---|---|---|
A193REG | Regular | 13”‐16” (33‐41 cm) |
A193SM | Small | 11”‐13” (28‐33 cm) |
A193PED | Pediatric | 9”‐11” (23‐28 cm) |
*Distal edge of the brace should be positioned at the inferior end of the sternum.
Product Code | Description |
---|---|
A193REG-CHIN-MR | CHIN ASSY REG MINERVA |
A193SM-CHIN-MR | CHIN ASSY SM MINERVA |
A193PED-CHIN-MR | CHIN ASSY PED MINERVA |
A193REG-OCPT-MR | OCCIPITAL ASSY REG MINERVA |
A193SM-OCPT-MR | OCCIPITAL ASSY SM MINERVA |
A193PED-OCPT-MR | OCCIPITAL ASSY PED MINERVA |
O10901 | MINERVA PAD SET RG |
O10902 | MINERVA PAD SET SM |
O10927 | PAD CHIN RG MINERVA |
O10928 | PAD OCCIP RG MINERVA |
O10931 | PAD CHIN SM MINERVA |
INDICATIONS
The Lerma Minerva is intended to support and stabilize C2-T3 stable fractures, muscular/ligamentous injury, degenerative joint disease, cervical pain, and nerve impingement.
CONTRAINDICATIONS
Unstable fractures, type II dens fractures.
INSTALLATION AND USE
-
Compare the posterior section with the patient to roughly determine the correct height of the occipital support with respect to the posterior shell.
-
Make height adjustments as necessary to the posterior section by removing the screws and sliding the bar. Replace the screws when adjustments are complete.
-
Contour and adjust the posterior section as needed.
-
With the help of an assistant, log‐roll the patient to his/her side to fit the posterior section.
Be sure the head and neck are supported during the log-roll process. -
When a successful posterior fit is achieved, roll the patient onto his/her back.
-
Make height adjustments as necessary to the anterior section by removing the screws and sliding the bar. Replace the screws when adjustments are complete.
-
Contour and adjust the anterior section as needed.
-
Adjust all hook and pile closures to a snug comfortable position.
-
Attach the optional forehead strap for additional support if needed.
MAINTENANCE
- Periodically check the Lerman MINERVA fasteners for looseness and tighten as appropriate.
- Qualified assistance should be obtained when cleaning the device. Homecare providers must be trained by the fitting practitioner.
- The shell of the Lerman MINERVA can be cleaned with soap and warm water or rubbing alcohol.
- The hook and loop adhered liners can be removed and washed with a gentle soap and cold water and air‐dried away from any heat sources.
- If the Lerman MINERVA becomes contaminated with biological fluids, it should be handled as medical waste upon disposal.
STORAGE AND USE
There are no identified restrictions on the temperature of storage and use.
DISPOSAL
There are no hazardous materials in the device. Comply with local and national
laws and regulations.
LEGAL INFORMATION
The use of this class 1 medical device is subject to the respective national laws of the country of use and may vary accordingly. The user of this device should report any serious incident to Trulife and the competent authority of the country of use.
Trulife
3225 Woburn St. Suite 160 Bellingham, WA 98226 USA
Phone (+1) 360 697 5656
Email supportop@trulife.com|
MDSS GmbH
Schiff graben 41
30175 Hannover Germany|
MDSS CH GmbH
Laurenzenvorstadt 61
5000 Aarau
Switzerland
---|---|---
LIMITED WARRANTY
Trulife warrants that the PRODUCT will be free from defects in material and
workmanship from the date of installation for the warranty period stated on
the PRODUCT warranty card.
This warranty will not apply if the PRODUCT has been damaged by misuse, abuse,
neglect, improper care, failure to follow instructions, abnormal wear and
tear, or in the event that the PRODUCT has been modified/repaired by persons
unauthorized by Trulife.
If a defect in material or workmanship is found during the warranty period,
Trulife will, at Trulove’s option, either repair or replace the product. If it
is not possible to repair or replace the product, Trulife will be limited to
refunding the purchase price.
Trulife will not be liable under any legal theory for any direct, indirect,
special, incidental or consequential damages arising from the use of or
inability to use this product.
The application guidelines for this Trulife product are for the use of and by
a certified, qualified practitioner only. Patients are not to attempt to apply
or adjust the item unless expressly instructed to do so by the practitioner
responsible for the prescription and/or initial fitting of the device. All
patient questions should be referred to the practitioner and not to the
manufacturer. The manufacturer warrants only that the enclosed product has
been inspected for quality and can be effective for certain indications, but
final decisions and ongoing monitoring must be made by the medical
professional(s) prescribing and/or fitting the device to determine its
effectiveness for an individual patient. Patient compliance is an integral
part of the entire protocol and must be adhered to in order to avoid potential
problems and to maximize the effectiveness of the prescribed product.
As a condition of the sale of any Trulife product, this product is restricted
to a “Single Patient Use Only” by the originally fitted patient in order to
protect the care provider and the patient against potentially adverse
consequences of infectious disease transmission, material instability in
adapting to the configuration of the original user and/or decrease in
effectivity. Any express or implied warranties are voided if the product is
reused or fitted to another patient. Additionally, a license of right to use
under any relevant patents pertaining to the product is terminated with the
cessation of use by the original patient. As with all Trulife products, this
product must be prescribed and applied by a qualified practitioner to
determine it meets the needs of the particular patient and accomplishes the
desired results.