Trulife GS102 Gunslinger User Guide
- June 13, 2024
- Trulife
Table of Contents
Trulife GS102 Gunslinger
Product Information: GS102 Gunslinger
The GS102 Gunslinger is a medical device designed for various applications related to shoulder injuries and surgeries. It is suitable for reconstruction surgery/shoulder replacement, traumatic injuries/fractures to the proximal humerus and shoulder, reduction of shoulder luxation, postoperative repair of rotator cuff, andBankart lesions.
The Gunslinger comes in two models: GS102L (Left) and GS102R (Right). It
offers adjustable features to ensure proper fitting and comfort for the
patient.
Manufacturer: Trulife
Model Numbers: GS102L (Left) and GS102R (Right)
Website: www.trulife.com
Product Usage Instructions: GS102 Gunslinger
- Familiarize yourself with the adjustment capabilities of the device before applying it to the patient. Refer to Figure 2 and Figure 3 for a diagram of adjustments.
- Using the provided Allen wrench, set the outrigger in a straight (neutral) position.
- With the patient supine or standing, apply the pelvic portion of the brace. Align the iliac crest with the plastic contour and secure the lower (hip) strap to ensure comfort.
- Loosen wing nuts (1), (2), and (3). Note that the forearm support bar and the attachment block slide independently of each other.
- With the patient’s elbow flexed, place the forearm in the support. Position the elbow about 1 posterior of the plastic cuff.
- Position the attachment block so that the support shaft (7) is 1 proximal of the midpoint between the wrist and elbow.
- Adjust the forearm length and loosely hand tighten wing nut(3). Remove the hand grip by squeezing the tips of the quick release fork (4) to unlock. Check the wrist to ensure unrestricted flexion.
- Accommodate flexion, rotation, and abduction adjustments as needed and hand tighten wing nuts (1), (2), and (3).
- Secure the contact closures over the forearm and readjust the posterior plastic overlap for comfort if necessary. Attach the hand grip by sliding the tips of the fork through the quick release bracket until a ‘click’ is heard.
- While supporting the forearm, loosen the socket head cap screw on the outrigger bar to allow angulation. Avoid damaging the serrations when making positional adjustments.
- Angle the shaft upward and slide the forearm outward, dependingon the required amount of abduction. The attachment at the pelvic section B (6) will compensate for elevation to maintain appropriate arc required in abduction.
- Using the Allen wrench, securely tighten the socket head cap screw to lock the angle of the outrigger. Take care to avoid mismatching the serrations when securely tightening.
- Hand tighten wing nuts (2) and (6).
- Loosen wing nuts (5) for anterior/posterior positioning and finer elevation adjustments, then hand tighten. To minimize shoulder adduction, remove the second extension post and reattach the plastic end cap.
- To maximize shoulder adduction, add the provided extension post and reattach the plastic end cap. See Figure 1 for reference.
INTRODUCTION
The Gunslinger is a sturdy, lightweight, fully adjustable orthosis which immobilizes the shoulder, elbow, wrist, and hand. This shoulder “airplane design” orthosis provides 90° of shoulder abduction from the torso.
INDICATIONS
Reconstruction surgery/shoulder replacement, traumatic injuries/fractures to the proximal humerus and shoulder, reduction of shoulder luxation, post- operative repair of rotator cuff, Bankart lesions.
Selection
- Model Number: Description
- GS102L: Gunslinger, Left
- GS102R: Gunslinger, Right
FITTING
-
Familiarize yourself with the adjustment capabilities before applying device to the patient. See FIGURE 2 and FIGURE 3 for a diagram of adjustments.
Note: Each quick release clasp is number coded with a round dot on the male and female parts for correct mating. Begin with hip (“1”), followed by the shoulder (“2”), then thoracic (“3”) FIGURE 1. Sit patient or flex patient hip if supine to ensure pelvic portion does not interfere with hip flexion. Re-adjust straps if necessary for proper length. Secure straps to a snug fit. Cut off excess strap if necessary and melt the end to prevent fraying. Refer to adjustments on the next page for continuing steps. A – Forearm Section- 1&3: Forearm AP position
- 2: Elbow flexion/extension humeral abduction (minimal)
- 3&2: Humeral flexion/extension
- 4: Hand rest quick release
B – Pelvic Section
* 5: Anterior/posterior positioning
* 6&5: Shoulder elevation
* 7: Forearm support shaft
-
2. Using Allen wrench provided set the outrigger with no angle or straight (neutral position).
-
With patient supine or standing, apply the pelvic portion of the brace. Assure comfort by aligning the iliac crest with the plastic contour and secure the lower (hip) strap.
-
Loosen wing nuts (1), (2), and (3). Note that the forearm support bar and the attachment block slide independently of each other.
-
With patient’s elbow flexed, place forearm in support. The elbow should be positioned about 1” posterior of the plastic cuff.
-
Position the attachment block with respect to the forearm so that the support shaft (7) is 1” proximal of the midpoint between the wrist and elbow.
-
Adjust forearm length and loosely hand tighten wing nut (3). Remove hand grip by squeezing tips of quick release fork (4) to unlock. Check the wrist to ensure unrestricted flexion.
-
Accommodate flexion, rotation, and abduction adjustments accordingly and hand tighten wing nuts (1), (2), and (3).
-
Secure contact closures over forearm and readjust posterior plastic overlap for comfort as necessary. Attach hand grip by sliding tips of fork through quick release bracket until a ‘click’ is heard.
-
While supporting the forearm, loosen the socket head cap screw on the outrigger bar to allow angulation. Loosen the screw enough to avoid damaging the serrations when making positional adjustments.
-
With wing nuts (6) and (2) loose enough to allow motion, angle the shaft upward while sliding the forearm outward. You will notice, depending on the amount of abduction required, the attachment at the pelvic section B (6) will compensate elevation to maintain appropriate arc required in abduction.
-
Using the Allen wrench, securely tighten the socket head cap screw to lock angle of outrigger. Care should be taken to avoid mismatching the serrations when securely tightening.
-
Hand tighten wing nuts (2) and (6).
-
Loosen wing nuts (5) for anterior/posterior positioning and finer elevation adjustments then hand tighten.
Note: To minimize shoulder adduction, remove the second extension post and reattach the plastic end cap.
Note: To maximize shoulder adduction, add the provided extension post and reattach the plastic end cap. See Figure 1. -
If two or three extension posts are used, disassemble them using the provided wrench. Each extension post will have an exposed set screw that is firmly cemented into it. Do not try to remove these. Apply the provided red thread locking compound to the exposed threads. Reattach and securely tighten with the provided wrench.
Step 15 must be followed to prevent loosening during use. -
Secure all wing nuts, especially (2), with the wrench provided when all adjustments have been completed.
STORAGE AND USE
There are no identified restrictions on the temperature of storage and use.
DISPOSAL
There are no hazardous materials in the device. Comply with local and national laws and regulations.
LEGAL INFORMATION
The use of this class 1 medical device is subject to the respective national laws of the country of use and may vary accordingly. The user of this device should report any serious incident to Trulife and the competent authority of the country of use.
Trulife
3225 Woburn St. Suite 160
Bellingham, WA 98226
USA
Phone (+1) 360 697 5656
Email supportop@trulife.com
LIMITED WARRANTY
Trulife warrants that the PRODUCT will be free from defects in material and workmanship from the date of installation for the warranty period stated on the PRODUCT warranty card. This warranty will not apply if the PRODUCT has been damaged by misuse, abuse, neglect, improper care, failure to follow instructions, abnormal wear and tear, or in the event that the PRODUCT has been modified/repaired by persons unauthorized by Trulife. If a defect in material or workmanship is found during the warranty period, Trulife will, at Trulife’s option, either repair or replace the product. If it is not possible to repair or replace the product, Trulife will be limited to refunding the purchase price. Trulife will not be liable under any legal theory for any direct, indirect, special, incidental or consequential damages arising from the use of or inability to use this product. The application guidelines for this Trulife product are for the use of and by a certified, qualified practitioner only. Patients are not to attempt to apply or adjust the item unless expressly instructed to do so by the practitioner responsible for the prescription and/or initial fitting of the device. All patient questions should be referred to the practitioner and not to the manufacturer. The manufacturer warrants only that the enclosed product has been inspected for quality and can be effective for certain indications, but final decisions and ongoing monitoring must be made by the medical professional(s) prescribing and/or fitting the device to determine its effectiveness for an individual patient. Patient compliance is an integral part of the entire protocol and must be adhered to in order to avoid potential problems and to maximize the effectiveness of the prescribed product.
As a condition of the sale of any Trulife product, this product is restricted to a “Single Patient Use Only” by the originally fitted patient in order to protect the care provider and the patient against potentially adverse consequences of infectious disease transmission, material instability in adapting to the configuration of the original user and/or decrease in effectivity. Any express or implied warranties are voided if the product is reused or fitted to another patient. Additionally, a license of right to use under any relevant patents pertaining to the product is terminated with the cessation of use by the original patient. As with all Trulife products, this product must be prescribed and applied by a qualified practitioner to determine it meets the needs of the particular patient and accomplishes the desired results.
Trulife www.trulife.com
References
- Prosthetics Orthotics Pressure Care Specialists | Global | Trulife
- Prosthetics Orthotics Pressure Care Specialists | Global | Trulife
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>