Avery Weigh-Tronix ZM 110 BSL Bench Scale User Manual

June 9, 2024
Avery Weigh-Tronix

Avery Weigh-Tronix LOGO

Avery Weigh-Tronix ZM 110 BSL Bench Scale

Avery Weigh-Tronix ZM 110 BSL Bench Scale-PRODUCT

Cleaning of Scale

Do

  • Wipe down the outside of the product with a clean cloth, moistened with water and a small amount of mild detergent cleaning fluid.
  • Spray on to the cloth when cleaning and not directly onto the indicator area.

Do NOT

  • Use harsh abrasives, solvents, scouring cleaners or alkaline cleaning solutions.
  • Do not attempt to clean the inside of the machine.

Sharp Objects
Do not use any sharp objects such as screwdrivers or long fingernails to operate the keys.

Introduction

The ZM110 is a general-purpose indicator ideal for connecting to most bench and floor scales. This indicator can easily be desk or pole mounted and comes with a power adaptor along with a rechargeable lead acid battery to allow the scale to be used in applications where power outlets are limited. This ZM110 indicator comes standard with a range of easy-to-operate General weighing functions along with basic Check weighing, Counting and Hold modes.

Unpacking

  • Carefully take the scale out of its package, make sure it’s not damaged and all accessories are included.

  • Remove the scale from the carton

  • Remove the protective covering
    Inspect the scale and indicator for damage

  • Make sure all components are included:

    • ZM110 Indicator
    • Adaptor plus plug adaptors
    • BSL Platform, column and brackets
    • Product manual

General Installation Guidelines
To get the best performance from the scale try to place the ZM110 in a location that will not degrade its accuracy.

  • Avoid extremes of temperature.
  • Try to avoid placing the scale in direct sunlight or near air vents
  • Place the scale on a level flat surface. It is not advised to place the scale near vibrating machinery
  • Avoid unstable power sources.
  • Do not use near large users of electricity e.g. welding equipment.

Levelling the Scale
Level the scale using the four adjustable feet on the bottom of the scale, Figure 1, and the bubble level near the base of the column, Figure 2. Always check the level prior to using scale.Avery Weigh-Tronix ZM 110 BSL Bench
Scale-FIG-2

Scale Installation

  1. Attach the support column to the bracket at the back of the scale base using the three mounting screws. Use a 3mm hex (Allen) key to tighten the screws. See Figure 3.
  2. Place the indicator on the top bracket as shown in Figure 4 and slide it down until it snaps into the final position. Loosen the adjustment knob shown in Figure 4 and adjust the indicator’s angle for best viewing of the display, then tighten the knob.
  3. The cable from the base to the indicator is run through the column and out through the plastic bracket at the top when you unpack the scale. Make sure the connector is free of obstructions, i.e. packing foam, and plug the cable into the connector, as shown in Figure 5. Tighten the threaded collar.
  4. Attach the AC power adapter to the connector on the back of the indicator. See Figure 5.

Battery Operation

The ZM 110 can be operated from a 6V 4Ahrs battery located inside the indicator or plugged into an acceptable power outlet. Battery life is approximately 107 hours with the backlight off. When the battery needs charging this symbol, , appears in the upper left corner of the weight display. The scale will still operate for about 10 hours after which it will automatically switch off to protect the battery.

When the scale is turned on, the battery status is:

  •  Green LED, the battery is fully charged;
  • Red LED, the battery is low and needs to be recharged.

Powering ON/OFF the Scale
With the platform connected and the battery fully charged or the charger plugged into the indicator, press the ON key to power ON the scale. A display test will be performed on power up and the scale will finish in normal weighing mode. Press the OFF key to turn the scale off.

Front Panel
The indicator has a sealed keypad and a 6 digits LCD display 0.59″ (40mm) high with a white backlight. Only the active annunciators are shown below.Avery
Weigh-Tronix ZM 110 BSL Bench Scale-FIG-7.

Display Annunciator

Avery Weigh-Tronix ZM 110 BSL Bench Scale-
FIG-8

Operation KeysAvery Weigh-Tronix ZM 110 BSL Bench Scale-
FIG-9
Error Codes

Err H Initial zero too high (over Full Scale + 10%)
Err L Initial zero too low (under Full Scale – 10%)
Err N Unstable internal count, this may indicate the load cell or

electronics are faulty
-OL-| Overload, full capacity +9d

Scale Operation

This section covers the scale operations of simple weighing, counting and hold. The steps for various kinds of weighing are explained in the following pages.
A warm-up time of 15 minutes is required to stabilize the measured values.

Gross Weighing
To perform a gross weighment, follow these steps:

  • Power up the scale. Be sure the scale is displaying weight in the correct unit of measure. Press the UNITS key, if necessary to change the units of measure the scale is working in.
  • Zero the scale if necessary by pressing the ZERO key. If the weight change is within the Zero window area the display will show 0.000. Decimal positions available may vary.
  • Place the item(s) to be weighed on the scale platform. 4. The display will now display the weight of the item that is on the weight platter

Tare Weighing

  1. If you want to do NET weighing, such as weighing objects in a container, follow these steps:
  2. Power up the scale. Be sure the scale is displaying weight in the correct unit of measure. Press the UNITS key, if necessary to change the units of measure the scale is working in.
  3. Zero the scale if necessary by pressing the ZERO key. If the weight change is within the Zero window area the display will show 0.000. Decimal positions available may vary.
  4. Place the item to be tared on the scale platform.
  5. Press the TARE key. If the container weight is outside the Zero window area the display will show 0.000 weight and the NET annunciator is illuminated. Decimal positions available may vary.
  6. Place the item to be weighed on the scale platform. The Net weight of the item is displayed. Press the SELECT key to choose the Gross or Net weight.Avery Weigh-Tronix ZM 110 BSL Bench Scale-FIG-11

Example : annunciators lit when tare weighing with 5 kg box and 10 kg of items.

Clear a Tare
Remove any load from the scale and press the TARE key until 0.000 is displayed. Decimal positions available may vary.

Example : annunciators lit when clearing a 5 kg container tare

Counting Mode

The ZM110 Counting mode has been designed to allow the operator to easily carry out basic sampling and counting routines on this scale even if the scale is mainly used for general weighing applications. By pressing the units’ key until Pcs is illuminated on the display, the scale can now be quickly used to sample parts ready for counting.

Sampling Parts Prior to Counting
Follow these steps to count items using the sampling method.

  1. Power up and if necessary Zero the scale by pressing the ZERO key. If the weight change is within the Zero window area the display will show 0.000. Decimal positions available may vary.
  2. Press the UNITS key until 0 Pcs is displayed.
  3. Press SELECT to choose the sample size from: 10, 20, 50, 100 and 200.
  4. Manually count the sample of parts onto the scale to the agreed sample size and press the PRINT key to confirm.
  5. Place the remaining parts onto the scale platform to Count Parts. The display will show the number of parts on the scale.

Example : annunciators lit at steps: 1, 2, 3 and 4 when sample size 10 Pcs

Counting can also be done while using a tare following these steps:

  •  After you have set the piece weight place the empty container on the empty scale and press the TARE key.
  •  The scale will Zero the display and illuminate the Net annunciator. Place the parts to be counted into the container and read off the net count value.

Check Counting
Make sure that the unit of measure is PCs and follow the instructions in section 3.4.1, page 14.

Check weighing Mode

  1. Checkweighing is a function that allows the operator to quickly see if a product is within an agreed acceptable tolerance, ideal for fast repetitive work where accuracy is important. By setting upper and lower acceptable weight or count limits for a product the scale can be used to check for regular weight or count conformity. To set the limits and beeper mode, see Chapter 3.4.1. and 3.4.2. on page 14. To perform a gross weight, follow these steps:
  2. Power up the scale. Be sure the scale is displaying weight in the correct unit of measure. Press the UNITS key, if necessary to change the units of measure the scale is working in.
  3. Zero the scale if necessary by pressing the ZERO key. If the weight change is within the Zero window area the display will show 0.000. Decimal positions available may vary.
  4. Place the item to be checked weighed on the scale platform. The weight is displayed and either the LO, OK or HI annunciator will be illuminated letting the operator quickly see if the item being weighed is within an acceptable weight tolerance to proceed.
  5. Remove item from the scale platform and repeat steps 3 and 4 for each item to be check and weighed.Avery Weigh-Tronix ZM 110 BSL Bench Scale-FIG-14

Check weighing can also be done while using a tare. After you have Zeroed the scale (step 2), place the item to be tared on the scale, press the TARE key and continue with check weighing as described in steps 3 and 4.

Example : annunciators lit when acceptable limits are: LO < 1.00 kg and HI > 4.00 kg

Setting the Checkweighing Limit

  • When the scale is ON, press the PRINT and TARE keys together; UF-1 will be displayed.
  • Press TARE (►) key; UF-2 will be displayed.
  • Press PRINT(↵) key;
  • Enter or edit the LO limit value by pressing the ZERO(◄) and/or TARE (►) keys to move the flashing digit left or right and press the SELECT (▲) key to increase the numeric value;
  • Press the PRINT (↵) key to confirm;
  • Repeat points iv. and v. for the HI limit;
  • Choose the beeper mode [000; 001; 002; see section 3.4.2
  • Table] by pressing the ZERO(◄) and/or TARE (►) keys to move the flashing digit left or right and press the SELECT (▲) key to increase the numeric value;
  • Press the PRINT (↵) key to complete and leave the procedure;
  • Press the UNITS (ESC) key to exit the User Menu.

Checkweighing Beeper Mode

o 000 The beeper is disabled.
o 001 The beeper tones only when the load is within acceptable

limits.
o 002| The beeper tones only when the load is outside the acceptable limits.

Hold Mode for Animal Weighing
This function can be used to capture and hold stable weight/pcs values. The indicator temporarily freezes (Hold) the value on the screen for the user to view or record.

To enable the Hold mode, see chapter 3.5.1 To perform a Hold weighment, follow these steps:

  1.  Power up the scale and zero the display, if necessary, by pressing the ZERO key. Be sure the scale is displaying weight in your preferred unit of measure. Press the UNIT key, if necessary.
  2.  Open the User Menu and select the Hold function [UF-5] to enable the option [Hold 1]. See the instructions below.
  3.  Place the item(s)/animal to be weighed/counted on the scale platform.
    • The indicator captures the weight.
    • The indicator freezes the stable weight.
  4.  Removing the item(s)/animal from the scale platform will clear the hold weight.
  5.  Repeat the steps above for other Hold weighments.

Setting the Hold modeAvery Weigh-Tronix ZM 110 BSL Bench Scale-
FIG-20
Open the User Menu and select the Hold function [UF-5] to enable the option [Hold 1]. Settings:

  •  PCt is the stability weight window in number of divisions. PCt range = 1 to 100 divs.
  •  timE is the number of consecutive weight readings within the stability window to be classed as stable weight. The average of these weight readings is the hold weight that the indicator will display. timE range = 1 to 64.

For example, a smaller PCt value sets a tighter stability window in which the stable weight has to be captured timE number of times.
Valid modes are HOLD 0 and HOLD 1 only. Recommended starting settings: PCt 100 and timE 4. Then adjust to meet the weighing application requirements.

Hold Mode Examples
PCt = 20 divisions, stability window timE =16 weight readings to be taken from within the stability window The hold weight displayed is the last weight reading (H) taken from within the stability window. Avery Weigh-Tronix ZM
110 BSL Bench Scale-FIG-22

Note : a smaller PCt value sets a tighter stability window in which the stable weight has to be captured timE number of times.

  • PCt = 20 divisions, stability window time
  • E =16 16 weight readings to be taken from within the stability window

The hold weight displayed is the last weight reading (H) taken from within the stability window. Avery Weigh-Tronix ZM 110 BSL Bench Scale-
FIG-21

There are three menus that allow configuring, enabling and executing specific functions or options.

User Menu
In the User Menu there are various submenus available to configure specific sections of the scale operating modes, including a diagnostic feature.

User Menu Levels [UF]
Navigate the options by pressing ZERO (), TARE () and increase a numeric value by SELECT () key. Press UNITS (ESC) to exit the option or the menu and PRINT () key to confirm.

User Functions Descriptions [UF]

Supervisor Menu
In the Supervisor Menu there are various submenus available to configure system parameters, specific sections of the operating modes and the calibration feature.
See Service Manual ZM110_s_en_501813.pdf .

Quick Calibration Menu
It is recommended to perform only the very first calibration following the function LF 1 in the Supervisor Menu. To prevent any unintentional amendment of the scale configuration, all the next Calibration should be performed by entering the Quick Calibration Menu. See Service Manual ZM110_s_en_501813.pdf

Note : perform the linearity procedure only if test weights applied to the scale between the zero and span calibration points are showing slight inaccuracies, such as ± a few divisions. If large inaccuracies are recorded, this indicates a possible mechanical problem or possible load cell failure which linearity calibration may not be able to correct.

Scale Configuration [LF 2]
Enter the function LF 2 under the Supervisor Menu to set the units, counting mode, calibration units, the decimal position of the separator and division size. See Service Manual ZM110_s_en_501813.pdf

RS232 Communications
RS232 allows the scale to communicate with a remote peripheral device such as a printer or computer. RS232 kit p/n AWT15-501905. See Service Manual ZM110_s_en_501813.pdf

Output Specifications

  • RS-232 output of weighing data ·
  • ASCII code ·
  • 8 data bits ·
  • Parity None, Stop Bit 1
  • Baud rate from 1200bps to 38400bps. If enabled the default baud rate is 9600bps.

Replacement Parts
See Service Manual ZM110_s_en_501813.pdf

Loadcell Pinouts
  • PIN 1 : Ex +
  • PIN 2 : Sense +
  • PIN 3 : Ex-
  • PIN 4 : Sense –
  • PIN 5: Sig +
  • PIN 6 : GND
  • PIN 7 : Sig –

Avery Weigh-Tronix is a trademark of the Illinois Tool Works group of companies whose ultimate parent company is Illinois Tool Works Inc (“Illinois Tool Works”). Copyright © 2018 Illinois Tool Works. All rights reserved.

No part of this publication may be reproduced, stored in an electronic retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise without the prior written consent of the copyright owner, or as permitted by law or under license. Full acknowledgment of the source must be given. Avery Weigh-Tronix is a registered trade mark of the Avery Weigh-Tronix group of companies. This publication was correct at the time of going to print however, Avery Weigh-Tronix reserves the right to alter without notice the specification, design, price or conditions of supply of any product or service at any time. All third party brands and product names used within this document are trademarks or registered trademarks of their respective holders.

General Information and Warnings

FCC

FCC Declarations of Compliance

United States
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Canada
This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications.

Training – Formation
Do not attempt to operate or complete any procedure on a machine unless you have received the appropriate training or read the instruction books.
CAUTION! Risk of electrical shock. Refer to qualified service personnel for service.

Safe Battery Disposal
Please be aware this product contains a lead acid battery which MUST be removed and disposed of safely prior to any disposal of the scale. This battery can be easily accessed by removing the battery cover found on the underside of the indicator.
CAUTION! The danger of explosion if the battery is incorrectly replaced. Replace only with the same or equivalent type recommended by the manufacturer.

Routine Maintenance

IMPORTANT : This equipment must be routinely checked for proper operation and calibration. Application and usage will determine the frequency of calibration required for safe operation.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Avery Weigh-Tronix User Manuals

Related Manuals