Enable Sleep

Control-IQ technology offers optional settings for a Sleep Activity that will change the treatment values when enabled.

  1. From the Options menu tap Activity.

  2. Tap START next to Sleep.
    Note: Sleep and Exercise cannot be enabled at the same time.

  3. Sleep is now enabled.
    Note: To disable Sleep Activity, repeat the previous steps and tap STOP.

  4. Tap STOP to turn Sleep off.

Using Sleep Schedules is recommended

Within the Sleep Activity, users can set up two Sleep Schedules, which will automatically turn Sleep on and off at pre-programmed times. For example, one might be set up for weekdays and the other for weekends.

  • Note: No Automatic Correction Boluses will be delivered during Sleep.
  • Note: If not using pre-programmed Sleep Schedules, Sleep must be manually turned on and off. If you have a Sleep Schedule set and go to bed earlier/later it will still turn off at the programmed time.
  • Note: If Exercise is enabled at the time Sleep is scheduled to start, Sleep will not begin. Once Exercise is turned off, the user will need to manually start Sleep or wait until the next scheduled sleep cycle
  1. From the Options menu tap Activity, and then Sleep Schedules. Then, select one of the two Sleep Schedules to set it up.
  2. Select any or all days of the week. Set the time that the Sleep function will start and end on those days. These times should reflect the time the user generally goes to sleep and wakes up.
  3. Tap to save the settings.

Enable Exercise

  1. From the Options menu tap Activity.

  2. Tap START next to Exercise.
    Note: Sleep and Exercise cannot be enabled at the same time.
    Note: If Exercise is active at the time Sleep is scheduled to start, Sleep will not begin. Once Exercise is turned off, the user will need to manually start Sleep or wait until the next scheduled sleep cycle.

  3. Exercise is now enabled.
    Note: To disable Exercise Activity, repeat the previous steps and tap STOP.

  4. Tap STOP to turn Exercise off.

Responsible Use of Control-IQ Technology
Even with advanced systems such as the t:slim X2 insulin pump with Control-IQ technology, users are still responsible for actively managing their diabetes. Control-IQ technology does not prevent all high and low blood glucose events. The system is designed to help reduce glucose variability, but it requires that users accurately input information, such as meals and periods of sleep or exercise. Control-IQ technology will not function as intended unless all system components, including CGM, infusion sets and pump cartridges, are used as instructed. Importantly, the system cannot adjust insulin dosing if the pump is not receiving CGM readings. Because there are situations and emergencies that the system may not be capable of identifying or addressing, users should always pay attention to their symptoms and treat accordingly.

Important Safety Information

The t:slim X2 insulin pump with Control-IQ technology (the System) consists of the t:slim X2 insulin pump, which contains Control-IQ technology, and a compatible continuous glucose monitor (CGM, sold separately). The t:slim X2 insulin pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in people requiring insulin. The t:slim X2 insulin pump can be used solely for continuous insulin delivery and as part of the System. When used with a compatible CGM, the System can be used to automatically increase, decrease, and suspend delivery of basal insulin based on CGM sensor readings and predicted glucose values. The System can also deliver correction boluses when the glucose value is predicted to exceed a predefined threshold. The pump and the System are indicated for use in individuals six years of age and greater. The pump and the System are intended for single user use. The pump and the System are indicated for use with NovoRapid or Humalog U-100 insulin. The System is intended for the management of Type 1 diabetes.

WARNING: Control-IQ technology should not be used by anyone under the age of six years old. It should also not be used in patients who require less than 10 units of insulin per day or who weigh less than 25 kilograms.
The System is not indicated for use in pregnant women, people on dialysis, or critically ill users. Do not use the System if using hydroxyurea. Users of the pump and the System must: be willing and able to use the insulin pump, CGM, and all other system components in accordance with their respective instructions for use; test blood glucose levels as recommended by their healthcare provider; demonstrate adequate carb-counting skills; maintain sufficient diabetes self-care skills; see healthcare provider(s) regularly; and have adequate vision and/or hearing to recognize all functions of the pump, including alerts, alarms, and reminders. The t:slim X2 pump and the CGM transmitter and sensor must be removed before MRI, CT, or diathermy treatment. Visit tandemdiabetes.com/safetyinfo for additional important safety information.

© 2022 Tandem Diabetes Care, Inc. All rights reserved. Tandem Diabetes Care, Control-IQ, and t:slim X2 are either registered trademarks or trademarks of Tandem Diabetes Care, Inc. in the United States and/or other countries. ML- 1007838_B