SAMCON ExCam-XI80 Explosion Proof Thermographic Camera User Manual
- June 9, 2024
- SAMCON
Table of Contents
ExCam-XI80 Explosion Proof Thermographic Camera
ExCam® XI80
User Manual
Table of contents
1 Introduction ……………………………………………………………………………………………….. 4
2 Technical data ……………………………………………………………………………………………. 4 2.1 Explosion
protection ……………………………………………………………………………….. 4 2.2 Illustration of the model
key………………………………………………………………………5 2.3 Electrical parameters of the
camera…………………………………………………………..6 2.4 System cable
SKDP03-T………………………………………………………………………….6 2.5 Optical
specifications……………………………………………………………………………….6 2.6 Video-technical
characteristics …………………………………………………………………. 8 2.7 Other technical data
……………………………………………………………………………….. 8
3 Safety Instructions ……………………………………………………………………………………… 9
4 Installation ……………………………………………………………………………………………….. 11
5 Electrical connection ………………………………………………………………………………… 13 5.1 Potential
equalization ……………………………………………………………………………. 13 5.2 Variants of cable
termination ………………………………………………………………….. 14 5.3 Connection work at the device
(terminal box) …………………………………………… 14 5.4 Plug-assignment (RJ45)
………………………………………………………………………… 17 5.5 Appropriate cables & cable
entries…………………………………………………………..19 5.6 Fusing
………………………………………………………………………………………………… 19 5.7 Tests prior to switching on
voltage ………………………………………………………….. 20 5.8 Optional accessory: Industrial Process
Interface (PIF)………………………………..20
5.8.1 Process Interface of the camera…………………………………………………………………………………. 20 5.8.2
Pin assignment Industriella Process Interface (PIF)………………………………………………………. 21
5.8.3 Programmable functions PIF ……………………………………………………………………………………… 21 5.8.4
In- and outputs …………………………………………………………………………………………………………. 22 5.8.5 Extension
up to 9 outputs ………………………………………………………………………………………….. 22 5.8.6 Example for a
Fail-Safe monitoring …………………………………………………………………………….. 23
6 Network access and visualization ……………………………………………………………… 24 6.1 Minimum
system requirements……………………………………………………………….. 24 6.2 Set up installation /
first use …………………………………………………………………… 24 6.3 Allocating IP
adress……………………………………………………………………………….24
7 Visualization/ Analysis Software PIX Connect……………………………………………..28 7.1
General information about PIX Connect …………………………………………………… 28 7.2 Focusing the
thermal image …………………………………………………………………… 28 7.3 Adjustment of temperature
range ……………………………………………………………. 28 7.4 User interface
………………………………………………………………………………………. 29
8 Cleaning …………………………………………………………………………………………………… 30
9 Maintenance / Modification…………………………………………………………………………30
10 Disposal / Recycling ………………………………………………………………………………. 30
11 Drawings & 3D models …………………………………………………………………………… 31
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 2 of 32
Table of Figures and Charts
Table 2-1 Model key ……………………………………………………………………………………………. 5 Figure 2-3
Sectional view of SKDP03-T ………………………………………………………………….6 Table 2-2 Measurement
field and pixel size ……………………………………………………………. 7 Table 2-3 Other technical data
……………………………………………………………………………… 8 Table 4-1 Mounting accessories
………………………………………………………………………….. 12 Figure 5-1 ExCam XI80 Potential equalization
………………………………………………………. 13 Table 5-1 Potential equalization
………………………………………………………………………….. 13 Figure 5-2 ExCam XI80
T08-VA0.4.K1.GER-N.N-xxx.N-P ………………………………………. 14 Figure 5-3 ExCam XI80
T08-VA0.4.K1.GER-N.N-xxx.N-T ………………………………………. 14 Figure 5-4 Video Tutorial
ExTB-3 ………………………………………………………………………… 15 Figure 5-5 Pin assignment of the cable
SKDP03-T ………………………………………………… 15 Table 5-2 Wire assignment of terminal box
ExTB-3(SKDP03-T) ……………………………….15 Figure 5-6 Wired terminal box ExTB-3
………………………………………………………………….16 Figure 5-7 ExTB-3 -> Safe area
………………………………………………………………………….. 17 Figure 5-8 Plug assignment RJ45
……………………………………………………………………….. 17 Figure 5-9 Plug assignment, RJ45
………………………………………………………………………. 18 Figure 5-10 Ex-d cable selection
…………………………………………………………………………. 19 Figure 5-11 PIF
connections………………………………………………………………………………..21 Figure 5-12 Connection of 3 PIFs
………………………………………………………………………… 22 Figure 5-13 Fail-Safe monitoring
states…………………………………………………………………23 Figure 7-1 Motorized focus settings in PIX
Connect Software ………………………………….. 28 Figure 7-2 Adjust temperature range
……………………………………………………………………. 28 Figure 7-3 User interface
……………………………………………………………………………………. 29
History of revisions
Product: Title: Doc. -Id. Author: Created on:
ExCam® XI80 User Manual for ExCam® XI80 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx Eva Schneider, Grad. Eng. (UAS) 07.10.2021
Rev. Index 0
1
2
Date 07.10.2021 17.10.2022 12.12.2022
Name
Comment
E. Schneider Compilation of the document
E. Schneider E. Schneider
Extension for optional accessory Industrial Process Interface Several corrections
Approved by
Digital unterschrieben von
Thiemo Thiemo Gruber DN: cn=Thiemo Gruber, o=Samcon GmbH, ou=Hardware
Entwicklung,
Gruber email=t.gruber@samcon.eu, c=DE Datum: 2022.12.15 11:57:52 +01’00’
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 3 of 32
1 Introduction
The ExCam XI80 is a powerful thermographic camera with an integrated infrared
thermometer and autofocus for use in potentially explosive areas. It delivers
excellent infrared images (80x80p) in a temperature range from -20°C to 900°C.
In addition, the thermal imaging camera has extensive analysis functions (e.g.
thermal analysis with hot and cold spots). The ExCam XI80 cannot only see
where it is hot, it can also measure the temperature exactly; and all that in
an 8 millisecond interval (Real Time Recording). When designing the ExCam
XI80, our main focus was on safety, mechanical precision and high quality of
stainless steel. Like all cameras of the ExCam series, it is certified
according to European regulations (ATEX) as well as international ones
(IECEx). The explosion-protected housing is approved for the ATEX group II for
zones 1, 2, 21 and 22 including the explosion groups IIC / IIIC. For more
information please visit our website at www.samcon.eu/en
2 Technical data
2.1 Explosion protection
Identification marks acc. to Directive 2014/34/EU:
Explosion protection (gas):
Explosion protection (dust):
Protection class camera: Protection class terminal box:
Transport/storage temperature: Ambient temperature (EX):
Named testing laboratory: EU type approval certificate: IECEx Certificate of
Conformity: Other certificates see:
II 2G (zone 1 and 2) II 2D (zone 21 and 22)
Ex db IIC T6 Gb
Ex tb IIIC T80°C Db
IP66/68 (IEC /EN 60529) IP66 (IEC /EN 60529)
-40°C…+70°C -10°C…+50°C
TÜV Rheinland (number 0035) TÜV 18 ATEX 8218X (2021) TUR 18.0023X (2021)
https://www.samcon.eu/en/products/network/excam-xi80/
Attention! The instructions stated on the type plates have to be observed!
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 4 of 32
2.2 Illustration of the model key
-
Ex productname
ExCam XI80 -
Type
T08- -
Housingcombination
VA0.4.K1.GER- -
Temp.range N.N-
T08- VA0.4.K1.GER- N.N-
-
Cable length [m] 005.N-
005.N- -
Cable termin. P-
T- -
Optic
012/030/0 55/080 012/030/0 55/080 Table 2-1 Model key
Explanations:
ExCam XI80 =
T08 =
VA0.4.K1.GER =
VA0.4.K1.GER = VA0.4.K1.GER = VA0.4.K1.GER =
N.N =
N.N=
005.N =
005.N =
P =
TD =
0XX =
Functional camera description of the ExCam Series (technical data/
specification of the individual camera module)
SAMCON Production- Type 08
T07 ex d housing (stainless steel 1.4404) with small diameter ØVA0=48mm) T07
VA0.4 housing with maximum body length (L.R = 171.2mm) K1 cable gland flange
Germanium sight glass, suitable for thermographic applications (MIR/ FIR)
Normal ambient temperature range (Tamb > 0°C) No high temperature battery
installed (Tamb < +50°C)
Length of the connection line in meter at delivery; 5m is the standard cable
length, max. cable length is: 005…100 [m] Non armoured cable
Plug- termination (standard) CAT6, RJ-45 plug, AWG 26-22, Terminal Box Ex-e
termination (optional)
Different optics available: 12°, 30°, 55° and 80°
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 5 of 32
2.3 Electrical parameters of the camera
Power supply of the camera: Voltage supply: Reference voltage: Maximum power consumption: Typical power consumption:
24 V DC 10 – 30 V DC 2.5 W < 2.0 W
2.4 System cable SKDP03-T
Outside diameter:
12.40 ± 0.3 mm
Bending radius:
8 x Da when installed and 4 x Da after relocation
Data line:
4 x 2 x AWG23/1 CAT.6
Performance elements:
3G1.5 (BK-BU-GN/YE)
Properties:
PUR halogen-free, flame-retardant, UV-re-
sistant, chemical resistance, shielded
Quick link:
https://www.samcon.eu/fileadmin/documents/en/60-Assembling%26mounting/SKDP03-T_Datasheet.pdf
Figure 2-3 Sectional view of SKDP03-T
2.5 Optical specifications
A selection of lenses offers the possibility to precisely measure objects in
different distances; we offer lenses for close, standard and large distances.
Please note that the ExCam XI80 has a fixed optic. A change of optics is not
possible. The following table shows a relationship between the measurement
object distance and the pixel size on the object level.
HFOV: Horizontal field of view VFOV: Vertikal field of view DFOV: Diagonal
field of view IFOV: Instantaneous field of view (size of the single pixel on
object level)
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 6 of 32
55°-optic
Enlargement of HFOV total measuring field at object VFOV
level [m]
DFOV
Size of the
single pixel
[mm]
IFOV
0,05 0,057 0,057
0,08
0,1 0,11 0,11 0,156
0,2 0,218 0,218 0,308
Distance to measurement object [m]
0,3
0,5
1
2
4
6
0,325 0,539 1,07 2,14 4,27 6,41
0,325 0,539 1,07 2,14 4,27 6,41
0,459 0,762 1,52 3,02 6,04 9,06
10 10,7 10,7 15,1
30 32 32 45,3
100 106,7 106,7 150,9
0,6
1,2
2,3
3,4
5,6 11,2 22,4 44,6 66,9 111,5 334,5 1114,8
12°-optic
Enlargement of total measuring field at object level [m] Size of single pixel [mm]
HFOV VFOV DFOV
IFOV
Distance to measurement object [m]
0,05
0,1
0,2
0,3
0,5
1
2
4
6
10
0,022 0,043 0,065 0,11 0,21 0,43 0,85 1,28 2,1
0,022 0,043 0,065 0,11 0,21 0,43 0,85 1,28 2,1
0,031 0,061 0,092 0,15
0,3
0,6
1,2 1,81
3
0,3
0,5
0,8
1,3
2,7
5,3 10,6 15,9 26,6
30 100 6,4 21,3 6,4 21,3
9 30,1
79,7 265,6
30°-optic
Enlargement of HFOV total measuring field at object VFOV
level [m]
DFOV
Size of single
pixel [mm]
IFOV
0,05 0,028 0,028 0,039
0,33
0,1 0,056 0,056 0,079
0,67
0,2 0,111 0,111 0,158
1,33
Distance to measurement object [m]
0,3
0,5
1
2
4
6
0,167 0,279 0,557 1,115 2,23 3,346
0,167 0,279 0,557 1,115 2,23 3,346
0,24 0,39 0,79 1,58 3,15 4,7
2 3,33 6,67 13,33 26,67
40
10 5,6 5,6 7,9
66,67
30 16,7 16,7 23,7
100 55,8 55,8 78,9
200 666,67
80°-optic
Enlargement of total measuring field at object level [m] Size of single pixel [mm]
HFOV VFOV DFOV
IFOV
0,05 0,09 0,09 0,127
0,08
0,1 0,174 0,174 0,246
1,6
0,2 0,343 0,343 0,483
3,2
Distance to measurement object [m]
0,3 0,5
1
2
4
6
0,509 0,884 1,682 3,357 6,708 10,058
0,509 0,884 1,682 3,357 6,708 10,058
0,72 1,19 2,38 4,75 9,49 14,2
10 16,8 16,8 23,7
30 50,3 50,3 71,1
100 167,5 167,5 236,9
4,7 7,8 15,5
31 61,9 92,8 154,6 463,7 154,5
Table 2-2 Measurement field and pixel size
For distances smaller than the minimum measuring distance (0.3m@12°;
0.2m@30°/55°/80°), the measuring accuracy can be outside the specification.
As an alternative to the table, you can use the Optris GmbH opticcalculator:
https://www.optris.de/optikkalkulator-fuer-waermebildkameras
Or the free optris Optikkalkulator App: Optris calculator Apps bei Google Play
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 7 of 32
2.6 Video-technical characteristics
We use the Optris Xi80 Spot-finder IR-Camera in a pressure-resistant
enclosure. For details, please refer to the Product Documentation, video-
technical data of optris®:
https://www.optris.com/optris-xi-80
2.7 Other technical data
Camera
Permissible ambient temperature -10°C … +50°C
Terminal box
-60°C … +55°C
Protection class as per EN 60529/IEC 529
Housing material Weight Dimensions
IP66/68 (Test conditions: 0.5h/8m water column 5°C) Stainless steel, mat. no. 1.4404 About 0.93 kg D48mm x 171.2mm
IP66
Polyester resin 1 kg 145mmx145mmx71mm
Table 2-3 Other technical data
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 8 of 32
3 Safety Instructions
Please absolutely observe the installation instruction’s safety directions of
the T08 ExCam series!
Quick link:
https://www.samcon.eu/fileadmin/documents/en/22-Ex-Network-Cameras/ExCam-
Series-T08-EX-Installation-Manual-2020.pdf
It is absolutely mandatory to adhere to the national safety regulations and
regulations for prevention of accidents, as well as to the safety instructions
given below in this User Manual!
Attention! Cameras of type T08 ExCam are not suitable for use in zones 0 and
20. The ambient temperature, temperature class and explosion group written on
the enclosure nameplate must be absolutely adhered to! The customer is not
allowed to make any alterations of the camera! The camera must be operated in
a proper and sound condition and only in the way intended.
Attention! Repairs may only be carried out by using original parts from the
manufacturer. Repairs which affect the explosion protection may only be
carried out in accordance with the nationally applied regulations and
exclusively by the manufacturer.
Attention! The alignment of the camera to intensive energy sources (e.g.
devices which emit laser radiation or reflections of such equipment) can cause
an irreparable defect of the infrared detector. This is also valid if the
camera is switched off.
Attention! Devices of type T08-VA0.x.K1.GER may not be used in mining/
firedamp hazard (ATEX group 1) or in areas with high risk of mechanical stress
(ATEX group 2).
Attention! Prior to installation, take external sources of heat or cold into
account! The temperature ranges prescribed for storage, transport and
operating must be adhered to!
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 9 of 32
Attention! Adhere to the warnings given on the nameplate:
“WARNING DO NOT OPEN IN HAZARD AREAS” Using the camera in explosion-
protected areas with regard to temperature and dust layers is defined in the
respective national regulations. When installing the ExCam, adhere to the
requirements of the EN/IEC 60079-14.
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 10 of 32
4 Installation
For commissioning and operating the camera, the relevant national regulations,
as well as the generally accepted rules of technology shall prevail. Before
mounting the camera, thoroughly check it for any transport damage regarding
the housing and cable. Installation, electrical connection and the first
commissioning must only be carried out by qualified specialists.
Work preparation:
Attention! Prepare your work carefully and in accordance with the relevant
regulations.
Attention! Depending on classification of hazard areas, it is imperative to
obtain a work approval first! When you open the pressure-resistant enclosure
under voltage, it is absolutely necessary to prevent potentially explosive
atmosphere!
To ensure the best image quality delivered by the network camera, plan the
installation site carefully (consider object distance or size, angle and
minimum object distance to the focus).
Use appropriate tools and aids. When working, ensure a safe stand. Make sure
that any static charge is avoided.
Attention! Please observe the national security, installation and accident
prevention regulations (e.g. DIN EN 60079-14) and the safety instructions in
this User Manual, as well as the ones in the Installation Guidelines!
Attention! Adhere to the provisions of the IECEx, ATEX and EX installation
instructions for mounting and starting up!
The ExCam® XI80 consists of a flame-proof camera housing (T08 Ex-d) and an
ex-e terminal box (VA0.4-…-T) or a Plug (VA0.4-…-P). The camera is equipped
with a systemcable (5 to 100 m). Mount the camera according to the desired
field of view. Install the terminal box so that a good accessibility is given,
in order to facilitate the electrical connection.
Drawings for drill hole patterns and further information can be viewed on our
product page:
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 11 of 32
Quick link: https://www.samcon.eu/en/products/network/excam-xi80/
Optional mounting accessories Wall bracket WMB-…
Pole adapter PMB-…
Hinge attachment SCH-…
WALL MOUNT BRACKET WMB-VA0.x/VA1.x Wall bracket for devices of T08-VA0-series
Suitable for hanging the camera on walls. Material: stainless steel 1.4404
Weight: 0.68 kg Dimensions: 80 x 100 x 205 mm
POLE MOUNT EXCAM VA1.x (-) Pole apter for VA wall mount Material: stainless
steel 1.4404 Suitable for pole diameters between 50 and 105 mm Load-bearing
capacity: 45 kg Dimensions:120x180x130 at masts of Ø 60 mm)
Hinge attachment SCH-VAx.x Hinge attachment for easy mounting on round sight
glasses acc. to DIN 28120/28121 or similar for VA Material: stainless steel
AISI 316L/1.4404 Weight: ca. 0.04 kg Dimensions WxHxD [mm]: 29.2x40x73.1
Table 4-1 Mounting accessories
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 12 of 32
5 Electrical connection
Attention! The electrical connection of the equipment may only be carried out
by qualified and skilled personnel!
Attention! It is absolutely necessary to ground the ExCam® series’ housing via
the PA connection.
Attention! Please observe the national security, installation and accident
prevention regulations (e.g. DIN EN 60079-14) and the safety instructions in
this User Manual, as well as the ones in the Installation Guidelines!
The delivered ExCam® XI80 is equipped with an electrical connection cable of
the type SKDP03-T. The maximum transmission range from the camera to the next
active network interface is 100 m and can be specified by the client. The user
is NOT authorised to carry out any electrical connection work inside the
pressure-resistant enclosure (T08).
5.1 Potential equalization
Figure 5-1 ExCam XI80 Potential equalization
Potential equalization/grounding of the camera housing is absolutely necessary, in order
to avoid static charges and thus the formation of sparks. For this purpose, a screw terminal
is provided at the rear side’s bottom (see Figure 5-1). The cross-section of the potential equalization has to comply with the National Ground Rules (at least 4mm2).
Wiring table:
Potential
Colour (IEC 60757) Cross-sec-
Comment
PA
GN/YE
tion 4 mm2 (rigid)
Terminal: Slotted screw M3x0.5×5 (DIN 84) with washer Ø3,2mm (DIN 125A), Keep
1.2 Nm tightening torque!
Table 5-1 Potential equalization
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 13 of 32
5.2 Variants of cable termination Figures 5.2 and 5.3 illustrate the cable
termination of the ExCam XI80.
Figure 5-2 ExCam XI80 T08-VA0.4.K1.GER-N.N-xxx.N-P
Figure 5-3 ExCam XI80 T08-VA0.4.K1.GER-N.N-xxx.N-T
5.3 Connection work at the device (terminal box) Attention! Never open the
Ex-e terminal box under voltage! Attention! Adhere to the international
installation regulations for connection chambers with increased safety (Ex-e).
Attention! Adhere to attached separate User Manual for the Ex-e terminal box.
Video Tutorial: Observe our video tutorial: “SAMCON 01 Wiring the cable
SKDP03-T to the junction box ExTB-3″ https://go.samcon.eu/v01
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 14 of 32
Figure 5-4 Video Tutorial ExTB-3
The pin assignment of the SKDP03-T is executed in accordance with the standard
EIA/TIA568B for 24VDC, as follows:
Camera (Ex-d) (T568B)
SHIELD Tx+ TxRx+ Rx-
(GND ISO)
(AO / RS485 B)
(AI/DI/RS485 A)
(n.c.)
(n.c.)
VCC GND PE/PA
Colour SKDP03-T (IEC60757)
WH / OG OG WH / GN GN WH / BU BU BN WH / BN
BK BU YE / GN
Figure 5-5 Pin assignment of the cable SKDP03-T
Terminal
Cross-sec- Comment
ExTB-3
tional sur-
face
PE
1.5 mm2
Shield
1
0.26 mm2
Solid conductor
2
0.26 mm2
Solid conductor
3
0.26 mm2
Solid conductor
4
0.26 mm2
Solid conductor
5
0.26 mm2
Solid conductor
6
0.26 mm2
Solid conductor
7
0.26 mm2
Solid conductor
8
0.26 mm2
Solid conductor
PE
9
1.5 mm2
24VDC
10
1.5 mm2
0VDC
PE
1.5 mm2
PE
Table 5-2 Wire assignment of terminal box ExTB-3(SKDP03-T)
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 15 of 32
Figure 5-6 Wired terminal box ExTB-3
Particularly in EMC critical environments, it is necessary to make sure that
the cable shield is grounded on side of the terminal block! The maximum cable
length is 100 m.
Attention! Finally, check your network installation. Attention! Cables and
wires must comply with the requirements of the IEC 60079-0/1/7 & 14.
Attention! The supply line must have a sufficient cross-section. The cable
protection must comply with national and international regulations.
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 16 of 32
Example: Direct routing from ExTB-3 to safe area
5.4 Plug-assignment (RJ45)
24 VDC power supply camera; Optional PIF (24 VDC/IO)
Figure 5-7 ExTB-3 -> Safe area
Figure 5-8 Plug assignment RJ45
If the cable termination uses a plug, the latter should be plugged into the
RJ45 slot of the network device (PSE). Prior to connecting it to the camera,
the network device (PSE) can
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 17 of 32
already be supplied with power, hence there is no ,,power ON” priority which
has to be observed.
Attention! Use appropriate RJ45 plugs! Check the cable shielding, cross-
section and the outside diameter! Attention! It is imperative to ensure a
correct routing of the individual wires according to the EIA/TIA-568B”
Detailed instructions on how to connect an RJ45 plug are available in our
video tutorial: “SAMCON 03 Mounting and installing the RJ45 jack to SAMCON
cables” https://go.samcon.eu/v03
Figure 5-9 Plug assignment, RJ45
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 18 of 32
5.5 Appropriate cables & cable entries An integral part of the device safety
is the correct selection of the cables, wires and cable entries.
Attention! Cables and wires must comply with the requirements of the IEC
600790/1/7 & 14. Attention! The supply line must have a sufficient cross-
section. The cable protection must comply with national and international
regulations. For non-binding configuration and planning guidelines, please
visit our website: Perhaps our video will help you: “Cables for flameproof
devices in potentially explosive atmospheres” http://go.samcon.eu/video-cable-
ex
Figure 5-10 Ex-d cable selection
5.6 Fusing The power supply fusing depends on the cable cross-section and
length. 300 mA mT- fuse is recommended.
Attention! Please pay attention to the national and international regulations
regarding selectivity and line protection.
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 19 of 32
5.7 Tests prior to switching on voltage
Attention! Prior to starting the device, perform all tests as indicated by the
national regulations. Furthermore, check the correct function and installation
of the device in accordance with this User Manual and other applicable
regulations.
Attention! Incorrect installation or operation of the camera may lead to a
loss of warranty!
Attention! Do not switch on the camera at temperatures below 0°C!
5.8 Optional accessory: Industrial Process Interface (PIF)
An industrial process interface with 500 VACeff insulation voltage between
ExCam XI and process is available for permanent monitoring of hardware and
software during operation. The PIF is equipped with self-monitoring (fail-safe
mode) so that states such as interruptions in the cable connection,
termination of the software, etc. can be recorded and output as an alarm. The
fail-safe time constant is 1.5 seconds. The PIF is not explosion-proof and may
therefore only be used in safe areas.
5.8.1 Process Interface of the camera
ExCam XI 80/410 offers the following direct In- and Outputs:
Name
Description
Max. range / status
AI
Analog Input
0 – 10 V1)
or
Digital Input
24 V
DI
(Low-aktive = 0 … 0.6 V)
AO
Analog Output
0/4 20 mA
Alarm Output
0/4 20 mA
- the AI is designed for max. 24 V, voltage level above 10 V is not
interpreted
In addition the above direct in- and outputs the ExCam XI80/410 has a RS485- interface. This interface can be used to control the external, industrial PIF. When using the RS485 interfacethe direct out- and inputs are not available. If the process interface of the camera is directly connected to external hardware (without using the supplied PIF cable) an activation of the field ,,Support proprietary PIF
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 20 of 32
cable” in the menu Tools/ Configuration/ Device (PIF) in the PIX Connect Software is necessary.
Consider that the input of the PIF is not protected if there is a direct PIF connection! A voltage > 3 V on the INT-pin will destroy the device!
5.8.2 Pin assignment Industrial Process Interface (PIF)
GN YE WH BN SHIELD
RS485 B RS485 A 12 V GND GND
Figure 5-11 PIF connections
5.8.3 Programmable functions PIF
Analog Input (AI) Analog Output (AO) Digital Output (DO)
Emissivity, ambient temperature, reference temperature, uncommitted value, flag control, triggered recording, triggered snapshots, triggered line- scanner, triggered event grabber, reset peak-/valley-hold, switch temperature range Main measure area, measure area, internal temperature, flag status, recording status, line scan status, alarm, frame sync, fail-safe, external communication, autonomous status Flag status, recording status, line scan status, alarm, frame sync, fail-safe, external communication, autonomous status
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 21 of 32
5.8.4 In- and outputs
Name
Description
Max. range1) / status
INPUT 1/2/3
Analog- or digital input
0 – 10 V2)
AO 1/2/3 DO 1/2/3
Analog output 1, 2 and 3 Alarm output 1,2 and 3 Relay output 1,2 and 3 Alarm output
0 10 V 0/4 20 mA open/closed (red LED on)/ 0…30 V,
400 mA
FS
Fail-safe relay
0/4 20 mA open/closed (green LED on)/ 0…30
V, 400 mA
- depending on supply voltage; for 0-10 V on the AO the PIF has to be powered with min. 12 V. 2) the AI is designed for max. 24 V, the voltage level above 10 V is not interpreted
5.8.5 Extension up to 9 outputs
The industrial PIF has a maximum of 3 analog outputs. To use up to 9 analog or
alarm outputs maximum 3 PIFs can be connected. Each stackable industrial PIF
must have its own RS485 address. The address must be set directly on the
board. For the PIF which is the furthest away the 120R switch (TERM.
Termination) has to be set.
Figure 5-12 Connection of 3 PIFs
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 22 of 32
5.8.6 Example for a Fail-Safe monitoring
Figure 5-13 Fail-Safe monitoring states
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 23 of 32
6 Network access and visualization
The most important procedures of initially starting the camera are described
below. A detailed software description can be found on the supplied USB stick
and in the Help menu in the PIX Connect software under Help Documentation.
Alternatively, the software can also be downloaded from the Optris website at
the following link: https://www.optris.global/pix
6.1 Minimum system requirements
· Windows 7, 8 or 10 · USB port · Hard disk with min. 30 MByte free storage ·
Minimum 128 MByte RAM
6.2 Set up installation / first use
All drivers are loaded automatically by the Windows operating system. No
driver installation is necessary. The software starts automatically in the
installed language.
1. Connect the supplied USB stick to your PC 2. Start the Setup.exe. Follow
the instructions of the wizard until the installation is finished.
After installation, you will find the software on your desktop (as a program
icon) and in the start menu: Start Programs Optris GmbH PIX Connect 3. Select
the desired language in the menu Tools Language
6.3 Allocating IP adress
The ExCam XI80 is designed for use in an Ethernet network and requires an IP address for access and control.
Software configurations based on OEM operating instructions (Optris)
Ethernet setup (point-to-point connection) After you have connected the
Ethernet cable to the camera and the PC, you first have to make the network
settings on the PC.
The devices are delivered with the following factory settings:
IP-address camera: 192.168.0.101
IP-address PC: 192.168.0.100
Port-number: 50101
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 24 of 32
To do this, go to All Control Panel Items and open the Network and Sharing Center. Go to Ethernet.
Now press Properties.
Now select Internet Protocol Version 4 (TCP / IPv4) in the properties window and then go to Properties again.
In the register card General, activate the
check box “Use the following IP address”.
Now enter a user-defined IP address for
your PC (192.168.0.100). This must be identi-
cal to the address set in the PIX Connect soft-
ware.
Then close the window with OK. The network settings on the PC are complete.
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 25 of 32
Start the PIX Connect software and activate the Ethernet function. To do this,
go to Devices and Enable Ethernet in the menu.
3.
Make sure that all three networks (Domain, Private, Public) are allowed in the
Windows Firewall window to ensure a connection to the device
4.
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 26 of 32
With the option “port no. assign automatically” a port number is assigned
automatically.
If several cameras with different port numbers are connected and communication
is to take place with a specific camera, the corresponding camera can be
determined by reacting to port number.
The authorizations of programs can also be activated retrospectively in the
Windows firewall settings of the PC (Windows firewall and allow a program or
feature through Windows Firewall).
The port range used should be between 49152 … 65535. When using other ports, it is possible that these are already reserved or assigned.
When using several XI80 cameras in one network, pay attention to the data
rate:
Switch with 100Mbits/s: approx. 17 devices (XI80)
Switch with 1000Mbits/s: approx. 170 devices (XI80) In addition to the data
rate, you should also pay attention to the PC performance. A separate instance
is required for each device used.
The device is now ready for the Ethernet connection and is listed in the menu Devices. The camera is identified by a network symbol and the network address and port number. Select the device. A connection to the device is established and the temperature measurement can begin. To change the corresponding address settings, go to Devices and Ethernet Settings (TCP/IP) Settings in the menu. The address for the device is assigned under Device address. This must have a different address (last block) to the other participant (e.g. PC) (Send to address (local computer)). It is important that the network part (first three blocks) must be identical for both addresses. The address range of the individual blocks can be between 0 and 255.
5. 6.
In addition, a separate port number must be configured. The selected number can be between 1 and 65535.
If you cannot assign the IP address, it may be necessary to check the firewall settings!
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 27 of 32
7 Visualization/ Analysis Software PIX Connect
7.1 General information about PIX Connect The configuration menu of the web
surface allows an intuitive navigation and offers several configuration
possibilities. For detailed documentation and information how to use the web
interface, please see the Optris PIX Connect Operator’s Manual or visit the
following website: https://www.optris.com/optris-xi-80
The radiometric processing of the image data enables subsequent detailed image
analysis with the convenient PIX Connect software. 7.2 Focusing the thermal
image To make sure that the focus of thermal channel is adjusted correctly the
motorized focus can be adjusted in the PIX Connect software (Menu View/
Windows/ Distance or over the icon ). Adjustment to the left = near,
adjustment to the right = infinite.
Figure 7-1 Motorized focus settings in PIX Connect Software
7.3 Adjustment of temperature range Set the temperature range that suits you
in the field at the bottom left.
Figure 7-2 Adjust temperature range
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 28 of 32
7.4 User interface
There are many possibilities within the user interface.
Figure 7-3 User interface
1. IR-Live view from the camera 2. Temperature profile: Temperature
distribution on a maximum of two lines, which can
be positioned anywhere in the image in terms of size and position. 3. Digital
display group: Possible display of all temperatures from e.g. defined
measuring
fields, cold spots, hot spots, temperature at the mouse pointer, the internal
temperature and the chip temperature. Alarm Settings: Bar with graphical
representation of a defined lower temperature threshold (blue arrow) and an
upper threshold (red arrow). The color of the digits of the displayed
temperature changes to RED when the upper alarm value is exceeded and to BLUE
when it falls below. 4. Temperature of measure area: Analyses the temperature
according to the selected shape of the field, e.g. the mean value of the
rectangle. This value is also shown in the live image (top right) and in the
digital display. 5. Reference bar: Shows a color scale with the corresponding
temperature values. 6. Temperature-time diagram: Shows the temperature curve
over time for the selected ROI (region of interest). 7. Histogram: Statistical
distribution of individual temperature values in the image. 8. Automatic /
manual scaling of the reference bar and thus the displayed temperature range:
Man., </> (min, max), 1 : 1 sigma, 3 : 3 sigma, OPT: Optimized palette 9.
Distance Function: Adjusting the motor focus to focus the image 10. Menu and
toolbar (icons) 11. Symbol for switching the individual palette views in the
reference bar. 12. Status bar: Serial number, optics, temperature range, mouse
cursor position, device frame rate/display frame rate, emissivity, ambient
temperature, flag status
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 29 of 32
There are also many helpful video tutorials for the PIX-Connect software from
Optris
https://www.youtube.com/results?search_query=optris+pix+connect+software
Configure your device according to your individual requirements.
8 Cleaning
Dirt on the viewing window must be removed as it can falsify the result of the
temperature measurement. For regular cleaning, never use cleaning agents
containing solvents (neither for the optics nor for the housing). Loose
particles can be blown away with clean compressed air (Attention in dust-
explosive environments). The sight glass can be cleaned with a soft, damp
microfibre cloth (moistened with water) or a lens cleaner (e.g. Purosol or B+W
Lens Cleaner).
9 Maintenance / Modification
The applicable regulations for the maintenance and servicing of electrical
devices in potentially explosive atmospheres must be adhered to.
The required maintenance intervals are specific to the individual devices. The
operating company has to determine these intervals depending on the
application parameters. The maintenance tasks especially include examination
of parts on which the ignition protection depends (e.g., proper condition of
the casing, seals and cable entry points). If maintenance measures are
necessary they have to be initiated and/or executed.
Repairs may only be carried out with original parts of SAMCON
Prozessleittechnik GmbH. Damaged pressure-resistant housings have to be
replaced completely. In case of doubt, send the part in question back to
SAMCON Prozessleittechnik GmbH. Reparations concerning the explosion
protection must only be carried out in accordance with nationally applied
regulations by SAMCON Prozessleittechnik GmbH or by an authorised electrical
technician authorised by SAMCON Prozessleittechnik GmbH. Rebuilding of or
alterations to the devices are not permitted.
10 Disposal / Recycling
When disposing of the device, nationally applicable regulations must be
observed. This Document is subject to alterations and additions.
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 30 of 32
11 Drawings & 3D models
All drawings, 3D models, certificates and other information are available in
the download area of the product page on our website:
https://www.samcon.eu/en/products/network/excam-xi80/
If you wish additional technical information, please contact us at:
support@samcon.eu
Doc.-ID: 211007-PT08BA-ES-ExCam XI80_en_rev.02.docx, Page 31 of 32
Schillerstrasse 17, 35102 Lohra-Altenvers, Germany
www.samcon.eu, info@samcon.eu Phone: +49 6426 9231-0, fax: – 31
References
- YouTube
- Explosion Proof Camera Housings - IECEx and ATEX CCTV Systems
- Explosion Proof Camera Housings - IECEx and ATEX CCTV Systems
- Explosion Proof Camera Housings - IECEx and ATEX CCTV Systems
- YouTube
- YouTube
- Spot finder IR camera optris Xi 80
- Optikkalkulator: Finden Sie die richtige Wärmebildkamera-Optik
- optris PIX Connect for download
- ExCam XI80 - Explosion Proof Thermographic Camera, ATEX, IECEx & IP68
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>