MEDI-RENT Bio Compression BioPants for Legs User Guide

June 9, 2024
MEDI RENT

MEDI-RENT Bio Compression BioPants for Legs User Guide
MEDI-RENT Bio Compression BioPants for Legs

INSTRUCTION

INSTRUCTION
Place the device somewhere easily accessible.
Wear lightweight long pants under the Bio Compression leg garments.

INSTRUCTION
Plug your power lead in, ensuring that it is not a trip hazard. (Often you will need to visit the bathroom during treatment due to fluid evacuation)

INSTRUCTION
Connect cuff hose to the device, ensuring the numbers and indicators match.

INSTRUCTION
Find the position where you will be rested during treatment on a bed or couch where your legs are elevated and supported.

INSTRUCTION
Place your leg into the sleeve by unzipping it part-way.

INSTRUCTION
First Time Users unless otherwise advised by your health professional, starting pressure is set at a default of 50mmHg.

If you need to adjust settings, see overleaf.

INSTRUCTION
Press “Power On/Off” button to turn on. The settings will appear on the LED display and the pump will start.

INSTRUCTION
At the end of a timed treatment, your pump will turn off. To turn off the pump during a treatment, press the “Power On/Off” button – “Hold” will show in the LED and then the pump will turn off.

INSTRUCTION
Whilst the compression pump is running please read the ‘Settings Navigation Guide.’

Allow the device run for at least 10 minutes after any setting change you make for the full effect to be felt.

INSTRUCTION
To adjust PRESSURE whilst the pump is running:

Press and hold the up arrow until you see “HOLD” on display.

The pump will return to resting position and the current pressure setting will be on display.

Use the up/down arrows to adjust.

Press the SET button to save your changes and recommence treatment.

INSTRUCTION
To adjust SETTINGS: Begin with the pump plugged in but turned off.

Press and hold “SET” for 5 seconds (until LED indicators flash) “SETUP” will appear on display.

With PRESSURE LED lit – use up/down arrows to adjust. Press “SET”.

With TREATMENT TIME LED lit – use up/down arrows to adjust Press “SET”.

With FOCUS LED lit – use up/down arrows to adjust. Press “SET”.

The pump will turn off.

Press the power button to turn the device on and commence treatment.

Refer to the Owner Manual for more information on settings.

INSTRUCTION
If you wish to clean the garment at any time, simply use warm soapy water or Clinell wipes. These specific cleaning wipes can be purchased from Medi-Rent and do not contain alcohol like many other sanitising wipes which can damage the garments.
Do not place in washing machine or

WARNING

  • Be advised by your Health Professional on setting selection, pressure and duration.
  • Always check your body for any pressure areas after treatment. If using the pump for prolonged periods adhere to pressure area prevention techniques such as changing position regularly and sitting/lying on a supportive surface.

CONTRAINDICATIONS FOR USING A COMPRESSION PUMP

  • Known or suspected deep vein thrombosis
  • Active metastatic disease affecting oedematous region
  • Pulmonary embolism
  • Thrombophlebitis
  • Cellulitis or other infection not treated with antibiotics
  • Uncontrolled/severe cardiac failure
  • Pulmonary oedema

Have questions?

Call us on AU PH 1300 726 666 or NZ PH 0800 725 555

MEDIRENT logo

© 2022 Medi-Rent Pty Limited

Medi-Rent Pty Limited
ABN: 37 603 454 360
E: info@medirent.com.au

Au PH: 1300 726 666

HEAD OFFICE

Unit 1, 61 Beauchamp Rd
Matraville NSW 2036
Unit 3, 93 Riverside Avenue
Werribee VIC 3030
Unit 3, 8 Emplacement Crescent
Hamilton Hill WA 6163
AUSTRALIA
F: 02 9316 9998

NZ PH: 0800 725 555

PO Box 40724
Upper Hutt 5140
NEW ZEALAND
www.medirent.com.au
www.medirent.co.nz