Delta OHM HD2103.1 Air and Wind Speed Temperature User Manual
- June 9, 2024
- Delta OHM
Table of Contents
HD2103.1 Air and Wind Speed Temperature
English
Operating manual
Anemometers / thermometers
HD2103.1 HD2103.2
Com panies / Brands of GHM
www.deltaohm.com
Keep for future reference.
CONTENTS
INTRODUCTION ………………………………………………………………………………………………………………………………………… 3
KEYBOARD AND MENU DESCRIPTION …………………………………………………………………………………………………… 8
THE PROBES …………………………………………………………………………………………………………………………………………….. 14
WIND SPEED MEASUREMENT ………………………………………………………………………………………………………………………. 14 Flow
rate measurement …………………………………………………………………………………………………………………………. 14
AP471S1 – AP471S2 – AP471S3 – AP471S4 HOT-WIRE WIND SPEED MEASUREMENT PROBES
COMPLETE WITH SICRAM
MODULE…………………………………………………………………………………………………………………………………….. 17
Operation…………………………………………………………………………………………………………………………………………….. 18 Care and
maintenance of the probes………………………………………………………………………………………………………… 18
Dimensions………………………………………………………………………………………………………………………………………….. 19 AP472S1 –
AP472S2 VANE WIND SPEED MEASUREMENT PROBES COMPLETE WITH SICRAM MODULE
………………….. 21 Calibration……………………………………………………………………………………………………………………………………………
21 Operation…………………………………………………………………………………………………………………………………………….. 21 Care
and maintenance of the probes………………………………………………………………………………………………………… 22
Dimensions………………………………………………………………………………………………………………………………………….. 23 WCT index
………………………………………………………………………………………………………………………………………….. 23 DIRECT INPUT Pt100,
Pt1000 AND Ni1000 TEMPERATURE PROBES……………………………………………………………………. 24 How to
measure……………………………………………………………………………………………………………………………………. 24 Instructions to
connect the TP47 connector for 4 wire Pt100, Pt1000 and Ni1000 probes
………………………………. 24 Direct connection of 4-wire Pt100
sensors……………………………………………………………………………………………….. 26 WARNINGS AND OPERATING
INSTRUCTIONS …………………………………………………………………………………….. 27
INSTRUMENT SIGNALS AND FAULTS …………………………………………………………………………………………………… 28
LOW BATTERY WARNING AND BATTERY REPLACEMENT ………………………………………………………………. 30
INSTRUMENT STORAGE…………………………………………………………………………………………………………………………. 31
SERIAL INTERFACE AND USB………………………………………………………………………………………………………………… 32
STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER……………………………………………….. 34
THE LOGGING FUNCTION – ONLY FOR HD2103.2 …………………………………………………………………………………………….
34 CLEARING THE MEMORY – ONLY FOR
HD2103.2…………………………………………………………………………………………….. 34 THE PRINT FUNCTION
……………………………………………………………………………………………………………………………….. 35 CONNECTION TO A PC
……………………………………………………………………………………………………………………………. 36
CONNECTION TO THE RS232C SERIAL PORT OF THE INSTRUMENT
…………………………………………………………………….. 36 CONNECTION TO THE USB 2.0 PORT OF THE
INSTRUMENT – ONLY FOR HD2103.2……………………………………………….. 36 NOTES ABOUT WORKING AND
OPERATIVE SAFETY …………………………………………………………………………. 37
INSTRUMENT TECHNICAL CHARACTERISTICS………………………………………………………………………………….. 38
TECHNICAL DATA OF PROBES AND MODULES IN LINE WITH THE INSTRUMENT
…………………………………………………….. 40 WIND SPEED MEASUREMENT
PROBES……………………………………………………………………………………………………….. 40 TEMPERATURE PROBES Pt100
SENSOR USING SICRAM MODULE ……………………………………………………………………. 42 Probes Pt100 4 wires
and Pt1000 2 wires ………………………………………………………………………………………………… 42
ORDER CODES …………………………………………………………………………………………………………………………………………. 43
HD2103
– 2 –
V2.4
INTRODUCTION
The HD2103.1 and HD2103.2 are portable instruments with a large LCD display.
They are designed for use in the fields of air conditioning, heating,
ventilation and environmental comfort. They use hot-wire or vane probes to
measure air speed, flow rate, and temperature inside pipelines and vents.
Temperature only is measured by immersion, penetration or air contact probes.
The temperature sensor used can be chosen from the Pt100, Pt1000 or Ni1000.
The probes are fitted with the SICRAM module, with the factory calibration
settings already being memorized inside.
The HD2103.2 instrument is a data logger. It memorizes up to 38,000 samples
which can be transferred from the instrument connected to a PC via the RS232C
serial port or USB 2.0 port. The memorization interval, printing, and baud
rate can be configured using the menu. The HD2103.1 and HD2103.2 models are
fitted with an RS232C serial port and can transfer the acquired measurements
in real time to a PC or to a portable printer. The Max, Min and Avg function
calculate the maximum, minimum or average values. Other functions include: the
relative measurement REL, the HOLD function, and the automatic turning off
that can also be disabled. The instruments have IP66 protection degree.
This manual describes the HD2103.1 and HD2103.2 models: if not otherwise
specified, the description is intended to be applicable to both models.
HD2103
– 3 –
V2.4
Thermo-Anemometer HD2103.1
HD2103
– 4 –
V2.4
HD2103.1
1. Input for probes, 8-pole DIN45326 connector. 2. External auxiliary power
supply connector input. 3. Battery symbol: displays the battery charge level.
4. Function indicators. 5. Secondary display line. 6. HOLD/ key: freezes the
measurement during normal operation; in the menu, increases the
current value. 7. FUNC/ENTER key: during normal operation displays the maximum
(MAX), the minimum
(MIN) and the average (AVG) of current measurements; in the menu, confirms the
current selection. 8. REL/ key: enables the relative measurement (displays the
difference between the current value and the logged value when the key is
pressed); in the menu, decreases the current value. 9. SERIAL key: starts and
ends data transfer to the serial communication port. 10. MENU key: allows
access to and exit from the menu. 11. °C/°F-ESC key: changes the unit of
measurement for temperature between degrees Celsius and Fahrenheit; in the
menu, cancels the operation in progress without making changes. 12. UNIT/DUCT
CALC key: during normal operation, selects the unit of measurement for the
main variable. When pressed together with the FUNC key, starts the calculation
procedure for the duct flow rate. 13. ON-OFF/AUTO-OFF key: turns the
instrument on and off; when pressed together with the HOLD key, disables the
automatic turn off. 14. MAX, MIN and AVG symbols. 15. Main display line. 16.
Line for symbols and comments. 17. 8-pole MiniDin connector for RS232C. For
the connection to PC (with cable HD2110CSNM or C206) or printer (with cable
HD2110CSNM).
HD2103
– 5 –
V2.4
Thermo-Anemometer HD2103.2
HD2103
– 6 –
V2.4
HD2103.2
1. Input for probes, 8-pole DIN45326 connector.
2. External auxiliary power supply connector input.
3. Battery symbol: displays the battery charge level.
4. Function indicators.
5. Secondary display line. 6. HOLD/ key: freezes the measurement during
normal operation; in the menu, increases the
current value.
7. FUNC/ENTER key: during normal operation displays the maximum (MAX), the
minimum (MIN) and the average (AVG) of current measurements; in the menu,
confirms the current selection.
8. REL/ key: enables the relative measurement (displays the difference
between the current value and the logged value when the key is pressed); in
the menu, decreases the current value.
9. SERIAL/ERASE LOG key: starts and ends data transfer to the serial
communication port. In the menu, clears the data contained in the instrument’s
memory.
10. LOG/DUMP LOG key: during normal operation, starts and ends the saving of
the data in the internal memory; in the menu, starts the data transfer from
the instrument’s memory to the PC.
11. Mini-USB type B connector for USB 2.0. For the connection to PC (with
cable CP23).
12. MENU key: allows access to and exit from the menu.
13. °C/°F-ESC key: changes the unit of measurement for temperature between
degrees Celsius and Fahrenheit; in the menu, cancels the operation in progress
without making changes.
14. UNIT/DUCT CALC key: during normal operation, selects the unit of
measurement for the main variable. When pressed together with the FUNC key,
starts the calculation procedure for the duct flow rate.
15. ON-OFF/AUTO-OFF key: turns the instrument on and off; when pressed
together with the HOLD key, disables the automatic turn off.
16. MAX, Min and AVG symbols.
17. Main display line.
18. Line for symbols and comments.
19. 8-pole MiniDin connector for RS232C. For the connection to PC (with cable
HD2110CSNM or C206) or printer (with cable HD2110CSNM).
HD2103
– 7 –
V2.4
KEYBOARD AND MENU DESCRIPTION
Foreword The instrument keyboard is composed of single-function keys, like the
MENU key, and doublefunction keys such as the ON-OFF/Auto-OFF key. In the
double-keys, the function in the upper part is the “main function”, while the
one in the bottom part is the “secondary function”. When the instrument is in
standard measurement mode, the main function is active. In the menu or in
conjunction with the FUNC key, the secondary function is enabled. The pressing
of a key is accompanied by a short confirmation beep: a longer beep sounds if
the wrong key is pressed. Each key specific function is described in detail
below.
ON-OFF/Auto-OFF key
The instrument is turned on and off using the ON/OFF key. The turning on
enables all display segments for a few seconds, starts an auto-test including
the detection of the probe connected to the input, and sets the instrument
ready for normal measurement.
mm kµ
mcimn 23s
MIN
During turning on, should no probes be connected, the message “NOPRBE SER_NUM” is displayed in the line of symbols for a few seconds, and in the main line a series of dashes appears while the ERR message replaces the temperature. When the probe is inserted into a functioning instrument, the “NEW_PROB_DET” (New probe detected) message appears: as the probe’s data are captured upon turning the instrument on, it is necessary to turn the instrument off and on again. Replace the probes when the instrument is off.
Automatic turning off
The instrument has an AutoPowerOff function that automatically turns the instrument off after about 8 minutes if no key is pressed during the intervening time. The AutoPowerOff function can be disabled by holding the HOLD key pressed down when turning the instrument on: the battery symbol will blink to remind the user that the instrument can only be turned off by pressing the <ON/OFF> key.
HD2103
– 8 –
V2.4
The automatic turning off function is disabled when external power is used. On the other hand, it cannot be disabled when the batteries are discharged.
FUNC/ENTER key
During normal measurement this enables the display and logging of the maximum
(MAX), minimum (MIN) and average (AVG) value of the measurements captured by
the probe connected to the instrument, updating them with the acquisition of
new samples. The acquisition frequency is once a second. The calculation is
performed on the variable appearing on the display upon pressure of the FUNC
key: by changing the variable or unit of measurement, the Max, Min and Avg
values are cleared. The MAX, MIN and AVG measurements remain in the memory
until the instrument is on, even after exiting the calculation function. To
reset the previous values and restart with a new measurement session, press
FUNC until the message “FUNC CLR” appears, then use the arrows to select YES
and confirm using ENTER.
In the menu, the ENTER key confirms the current parameter and then goes to the
next one. Pressed together with the UNIT/DuctCalc key, enables the flow rate
calculation (please see the UNIT key description).
Attention: the data captured using the Record function cannot be transferred
to the PC.
HOLD/ key
It increases the current parameter when used in the menu; when used in
measurement mode, it freezes the measurement in progress, and upon application
of pressure on the key, the message HOLD appears in the upper side of the
display. To return to the current measurement, press the key again. Upon
turning on the instrument, the AutoPowerOff function can be disabled by
holding the HOLD key down (please see the ON-OFF key description). During
calculation of the flow rate (Duct Calc), pressing this key adds a new sample.
UNIT/DuctCalc key
In measurement mode, it allows selection of the unit of measurement of the main input quantity (shown in the central line of the display). By repeatedly pressing the function key, the different units of measurement are shown in sequence:
· wind speed: m/s – km/h – ft/min – mph (miles/hour) – knots
· calculated flow rate: l/s (litres/s) – m3/s – m3/min – m3/h – ft3/s – ft3/min
· the apparent temperature WCT (wind chill equivalent temperature)
The WCT parameter represents the apparent temperature in presence of cold wind and is determined up to an actual temperature of 10°C. Above this limit the display shows an error (ERR).
HD2103
– 9 –
V2.4
This setting changes the information displayed and the immediate print of data (SERIAL key). The data recorded using the LOG function (HD2103.2) and those sent to the printer or PC through the serial port using the SERIAL function (HD2103.1 and HD2103.2), keep the chosen unit of measurement and display it.
Spatial averaging calculation
Simultaneously pressing the FUNC and UNIT keys starts the spatial averaging calculation, using the unit of measurement selected with the UNIT key. It is possible to calculate the spatial averaging, of the wind speed or flow rate i.e. across a section of a duct. Please see the paragraph dedicated to wind speed measurement on page 14.
°C/°F – ESC key
Changes the unit of measurement for the temperature between Centigrade or
Fahrenheit. When used in the menu, it clears or cancels the active function.
It allows exiting the spatial averaging calculation (Duct Calc).
REL/ key
In measurement mode, it displays the difference between the current value and
that measured on pressing the key for both measurements – main and secondary.
The REL message appears on the display; press the key again to return to the
current measurement. When used in the menu, it decreases the current variable
value.
MENU Key
The first menu item is accessed by initially pressing on the MENU key; press
ENTER to go to the following items. To modify the item displayed, use the
arrow keys ( and ). The current value is confirmed by pressing the ENTER key
and the display moves on to the next parameter, the setting is cancelled by
pressing the ESC key. To exit the menu, press the MENU key at any time.
The menu items are listed in this order:
- Probe type: the message “>>>_PRBE_TYPE” is scrolled in the comment line. The main line in the center of the display shows the type of probe connected to the instrument. The following probes can be connected to the input: · hot- wire probes complete with SICRAM module · vane probes complete with SICRAM module
HD2103
– 10 –
V2.4
· temperature probes Pt100 complete with SICRAM module · direct 4-wire Pt100
probes complete with TP47 module · 2-wire Pt1000 probes complete with TP47
module · 2-wire Ni1000 probes complete with TP47 module
Upon turning on the instrument automatically detects the probes fitted with
SICRAM module: the Probe Type menu item is configured by the instrument and
cannot be modified by the user.
Upon turning on, the direct 4-wire Pt100, the Pt1000 and Ni1000 temperature
probes display the message “NO_PRBE_SER_NUM”. In this case the probe type must
be entered manually. Select Probe type using the MENU key and then select the
type of probe used with the arrow keys; confirm using ENTER.
2) SECT m2 – SECT inch2: this parameter defines a duct’s section area for flow
rate calculation. It is expressed in m2 or inch2. To change the unit of
measurement, press UNIT. Use the arrows and confirm with ENTER. Please see the
paragraph dedicated to wind speed measurement on page 14.
3) AVG TIME SECS: time interval according to which the moving average is
calculated, in seconds, during measurement of the flow. The value ranges from
1 (no average) to 99 seconds. Please see the paragraph dedicated to wind speed
measurement on page 14.
4) Management of memorized data (only HD2103.2): the message
“>>>_LOG_DUMP_or_ERAS” (Transfer data or erase) is scrolled in the comment
line. The center figure reports the number of free memory pages (FREE). All
memory data are permanently erased by pressing SERIAL/EraseLOG. By pressing
LOG/DumpLOG, the data transfer of the logged data on the serial port is
started: the “BAUD-RATE” must have previously been set to the maximum value
(please see the menu items described below and the paragraph “STORING AND
TRANSFERRING DATA TO A PERSONAL COMPUTER” on page 34).
5) Sleep_Mode_LOG (Automatic turning off during recording) (only HD2103.2):
this function controls the instrument’s automatic turning off during logging,
occurring between the capture of a sample and the next one. When the interval
is lower than 60 seconds, the instrument will always remain on. With intervals
greater than or equal to 60 seconds, it is possible to turn off the instrument
between loggings: it will turn on at the moment of sampling and will turn off
immediately afterwards, thus increasing the battery life. Using the arrows
select YES and confirm using ENTER in order to enable the automatic turning
off, select NO and confirm to disable it and keep the instrument on
continuously. Note: even if Sleep_Mode_LOG=YES is selected, the instrument
does not turn off for less than one minute intervals.
6) Print and log interval: sets the interval in seconds between two loggings
or data transfers to the serial port. The interval can be set at 0, 1s, 5s,
10s, 15s, 30s, 60s (1min), 120s (2min), 300s (5min), 600s (10min), 900s
(15min), 1200s (20min), 1800s (30min) and 3600s (1hour). If the value 0 is
set, SERIAL works on command: the sending of data to the serial port is
performed each time the key is pressed. Recording (LOG) is performed with one
second intervals even if the interval is set to 0. With an interval from 1 to
3600s, continuous data transfer is started when the SERIAL key is pressed. To
end the recording (LOG) and continuous data transfer operations (SERIAL with
an interval greater than 0), press the same key again.
HD2103
– 11 –
V2.4
- YEAR: to set the current year. Use the arrows to modify this parameter and confirm using ENTER.
- MNTH (month): to set the current month. Use the arrows to modify this parameter and confirm using ENTER.
- DAY: to set the current day. Use the arrows to modify this parameter and confirm using ENTER.
- HOUR: to set the current hour. Use the arrows to modify this parameter and confirm using ENTER.
- MIN (minutes): to set the current minutes. In order to correctly synchronize the minute, it is possible to reset the seconds by pressing the UNIT key. Use the arrows to set the current minute plus one, and as soon as that minute is reached press UNIT: this synchronizes the time to the second. Press ENTER to go onto the next item.
- BAUD_RATE: indicates the frequency used for the serial communication with the PC. Values from 1200 to 38400 baud. Use the arrows to modify this parameter and confirm using ENTER. The communication between instrument and PC (or serial port printer) only works if the instrument and PC baud rates are the same. If the USB connection is used this parameter value is automatically set (please see the details on page 34).
LOG/DumpLOG key – only HD2103.2
In measurement mode, this function starts and stops the logging of a data
block to be saved in the instrument’s internal memory. The data logging
frequency is set in the “Print and log interval” menu parameter. The data
logged between a start and subsequent stop represent a block. When the logging
function is on, the LOG indication is on the display, the battery symbol
blinks and a beep is issued each time a logging occurs; when using an external
power supply the battery symbol does not appear. To end the logging, press
LOG.
The HD2103.2 can turn off during logging between one capture and the next: the
function is controlled by the Sleep_Mode_LOG parameter. When the logging
interval is less than one minute, the logging instrument remains on; with an
interval of at least one minute, it turns off between one capture and the next
if the parameter Sleep_Mode_LOG=YES.
Dump LOG – only HD2103.2
Press the MENU key until the “>>>_LOG_DUMP_or_ERAS” item is displayed and then press on the LOG key: the transfer of the data contained in the instrument internal memory via the serial port is started. Please see the paragraph dedicated to data transfer on page 34.
HD2103
– 12 –
V2.4
SERIAL key – only HD2103.1
SERIAL/EraseLOG key – only HD2103.2
In measurement mode, this function starts and stops the data transfer to the
RS232C serial output. According to the settings entered in the Print and log
interval menu item, a single sample can be printed if Print and log interval=0
or a continuous indefinite printing of the measured data can be set up if
Print and log interval=1…3600. The printing operation is accompanied by the
display of the RS232 symbol and the blinking of the battery symbol; when using
an external power supply the battery symbol does not appear. Press SERIAL to
end the continuous printing.
Before starting the printing with SERIAL, set the baud rate. To do so, select
the Baud Rate menu item and select the maximum value equal to 38400 baud by
using the arrows. Confirm by pressing ENTER. The DeltaLog9 software for PC
will automatically set the baud rate value during connection. If you are using
a different program than DeltaLog9, be sure the baud rate is the same for both
the instrument and the PC: the communication will only work in this way.
Erase memory – only HD2103.2
Press the MENU key until the “>>>_LOG_DUMP_or_ERAS” item is displayed and then press on the SERIAL/EraseLOG key: this clears permanently all the data contained in the instrument’s memory.
HD2103
– 13 –
V2.4
THE PROBES
The instrument works with hot-wire, vane and temperature probes fitted with
the SICRAM module. It works also with temperature probes with 4-wire Pt100,
2-wire Pt1000 or Ni1000 sensors.
Some probes are fitted with SICRAM module that acts as an interface between
the sensor on the probe and the instrument. There is a microprocessor circuit
with a permanent memory inside the module that enables the data logger to
recognize the type of probe connected and to read its calibration information.
The temperature probes without SICRAM module are not automatically detected by
the instrument and must be set up in the Probe type menu item (please see the
description of the menu on page 10).
The probe is detected during turn on, and this cannot be performed when the
instrument is already on, therefore if a probe is connected and the instrument
is on, it is necessary to turn it off and on.
WIND SPEED MEASUREMENT
The AP471 and AP472 series probes measure the incident wind speed and flow
rate; some also measure air temperature. The hot-wire measurement principle is
used for the AP471 series and the vane principle for the AP472 series. On
request, the probes of the AP471 and AP472 series can be fitted with a
telescopic rod that eases measurements in areas difficult to reach (for
example vents). The typical applications are wind speed and flow rate checks
in air conditioning, heating and cooling systems, or environmental comfort
determination, etc. Hot-wire probes are usually employed for precise
measurements with low-medium wind speeds (up to 10 m/s), the vane probes with
speeds from 5 to 50m/s. In addition, the measured fluid temperature must be
considered: the probes measure air flows at 80°C maximum temperature.
The measurements provided by the instrument using the probes are: wind speed,
flow rate, and air temperature.
Connecting the probe and turning the instrument on, it is possible to choose
the unit of measurement for the value displayed in the main line by using the
UNIT key. The following units are available:
· for wind speed: m/s – km/h – ft/min – mph (miles/hour) – knots; · for air
temperature: °C and °F; · for flow rate: l/s (litres/s) – m3/s – m3/min – m3/h
– ft3/s – ft3/min.
Also the WCT apparent temperature in the presence of air flows is calculated.
Flow rate measurement
The flow rate measurement requires knowledge of the duct or vent area
orthogonal to the flow: the menu items indicated by “SECT m2” and “SECT INC2”
define the section area m2 or inch2. To set the area value, select the “SECT
m2” menu item, and using the arrows, set the desired value. Confirm the
selection by pressing the ENTER key.
HD2103
– 14 –
V2.4
To set the measurement in inch2, select the “SECT m2” menu item, and using the
UNIT key, change the unit of measurement from m2 to inch2. Proceed to enter
the data using the arrows: confirm with ENTER.
The area comprised must be between 0.0001m2 (1cm2) and 1.9999m2.
After input of the duct’s section area, select the unit of measurement for the
flow rate by using the UNIT key: l/s – m3/s – m3/min – m3/h – ft3/s – ft3/min.
The display shows the calculated flow rate on the section set with the
parameters “SECT m2” and “SECT INC2″.
To obtain the correct flow rate measurement, the fact that the air speed is
not constant on the section but varies from point to point needs to be
considered, therefore an average speed over the entire section is required.
Furthermore, the speed varies through time at the same point. This is
particularly true when the surface considered is wide or when turbulence is
generated, near a grid or diffuser. The anemometers provide some solutions in
order to obtain correct measurements even in presence of these disturbance
phenomena.
- Spatial averaging (Duct Calc function) It is always best to perform the
measurements at different points and consider the average value as the valid
data. Using the Duct Calc function (sub-function of the UNIT key), the
HD2103.1 and HD2103.2 anemometers can capture more than one measurement and
supply the maximum, average, and minimum value. In particular, the average
value (AVG) is the most important as it supplies the average speed, and
therefore the calculated flow rate along the entire section and not on a
single measurement point.
Procedure: Using the UNIT key, select the speed or flow rate unit of measurement for which you wish to obtain the average, maximum, and minimum. Enable the calculation function by pressing simultaneously the DUCT CALC and FUNC keys: the central line of the display shows the instantaneous value of the selected variable (speed or calculated flow rate) while the lower line figure indicates the number of times the measurement has been logged. The measurement can be interrupted and there are no time limits to abide by between two subsequent measurements. The captured values are not erased, so it is possible to get some samples, turn the instrument off and later turn it back on again to proceed with the logging of more samples without losing the measurements already carried out. The maximum number of samples is 99. To reset the calculation, press FUNC until the message ” CLR FUNC ” appears, then use the arrows to select YES and confirm using ENTER. Position the probe in the first measurement point and press HOLD/ to capture the first point value. Repeat the same procedure with all the other points that you feel it is necessary to capture, by pressing HOLD/ each time: the indicator will provide the total number of acquired samples. After recording the first sample or at the end of the capture, press FUNC/ENTER. The MAX, MIN and average values of the chosen quantity, that is, the speed or calculated flow rate on the entire section will be displayed, according to the variable selected at the beginning of the measurement. To end the function, press ESC. Generally, the greater the number of captured measurements (maximum 99) the greater the accuracy of the result obtained.
HD2103
– 15 –
V2.4
- Moving average The Duct Calc function provides a spatial averaging of the
captured values and therefore compensates the speed differences between one
point and another in the duct section. There is also another source of errors
due to the variation of the flow in time: that is, the flow is not constant
but increases or decreases in the same point. To compensate for this second
source of instability, it is possible to get a temporal moving average of the
last n measurements capture: with n>1, the displayed value will not be the
single captured value but the continuously updated current average of the last
n measurements. To set the “n” value, use the AVG TIME SECS menu function: by
using the arrows select the desired n value and confirm with ENTER. “n” can be
set from 1 (no average) to 99.
Note: the presence of grids or diffusers with tilting fins introduces errors in the measurement of the flow due to the existing turbulence. These are generated because part of the flow, when meeting the obstacle (fin), is slowed down when the rest of the flow proceeds at maximum speed. In this case, to obtain a correct measurement, it is better to temporarily insert a duct after the grid with a length at least double that of the grid’s diagonal and perform the measurements at its end. The area considered for the flow rate calculation will be that of the provisional duct placed between the grid and the probe.
PIPE SECTION
D
PIPE
TEMPORARY PIPE PROBE
AIR VENT
L = 2 x D
HD2103
– 16 –
V2.4
AP471S1 – AP471S2 – AP471S3 – AP471S4 HOT-WIRE WIND SPEED MEASUREMENT PROBES
COMPLETE WITH SICRAM MODULE
The AP471S1 and AP471S3 probes measure incident air flows up to 40m/s. The
AP471S2, AP471S4 probes are fitted with an omnidirectional sensor allowing
measurement of speeds up to 5m/s in any direction of the air flow incident on
the probe. The AP471S4 probe is fitted with support base and sensor
protection. The wind speed measurement is compensated for according to
temperature within the range of 0…+80°C. The probes AP471S1, AP471S2, and
AP471S3 measure the environment temperature in the range of -25°C…+80°C;
AP471S4 in the range 0°C…+80°C. The AP472S… modules are calibrated in the
factory; no calibration is required by the user.
AP471 S1
AP471S4 HD2103
AP471 S3
The AP471S1, S2 and S3 probes are fitted with a cylindrical protection screen
that can slide longitudinally over a groove. The screen has two end-of-travel
positions that block it in measurement condition (completely low) or rest
condition (completely high). To reduce the space occupied when not used, the
AP471S4 is supplied with a protection cylinder that can be screwed on the
probe’s head.
LINE OF STREAM DIRECTION
SPEED SENSOR
COMPENSATION SENSOR
AMBIENT TEMPERATURE SENSOR
CYLINDRICAL SLEEVE
AP471S1 – AP471S3
AP471S2
– 17 –
V2.4
Operation Where present, extend the telescopic rod to the necessary length
paying attention to the cable so that it can slide freely and without strain.
Uncover the sensor and introduce the probe in the air flow being measured,
maintaining the arrow at the top of the probe parallel to the flow, as
indicated in the figures.
Flow direction indicator
Top of the probe
The probe should be maintained orthogonal to the flow and not tilted in
relation to it:
OK
NO
Proceed with measurement following the instructions provided in this chapter.
Care and maintenance of the probes
The speed sensor of the AP471S… probes is heated and, in the presence of gas
vapours, could trigger a fire or explosion. Do not use the probe in the
presence of inflammable gases. Ensure that no gas or explosive vapor leakage
is present in the measurement spaces. The probe is very delicate and should be
handled with extreme care. Even a simple collision, especially of the
omnidirectional probes that have an uncovered sensor, could render the probe
unusable. After measurement, the sensor set on the probe head must be
protected with the supplied metallic screen or threaded cylinder. During use,
the AP471S4 omnidirectional probe must be protected with the special metallic
grid. During transportation, the sensor must be closed into a cylinder screwed
on the end part of the probe. Do not let fingers touch the sensors. Use only
alcohol to clean the probe.
HD2103
– 18 –
V2.4
Dimensions
85
200 min – 880 max
Ø13
Ø13
200 min
85
880 max
102
Ø13
87
Ø8 AP471 S1 110
Ø8 AP471 S2 87
Ø8 AP471 S3
HD2103
– 19 –
V2.4
AP471S4
HD2103
– 20 –
V2.4
AP472S1 – AP472S2 VANE WIND SPEED MEASUREMENT PROBES COMPLETE WITH SICRAM
MODULE
The AP472S1 and AP472S2 vane probes measure the incident wind speed and flow
rate. The probe AP472S1 measures also the temperature using a thermocouple of
type K. On request, they can be fitted with a telescopic rod that eases
measurements in areas difficult to reach (for example vents). The probes’
speed and temperature measurement ranges are outlined in the table below:
AP472S1 AP472S2
Speed (m/s) 0.6…25 0.5…20
Temperature (°C)
-25…+80 -25…+80 (working temperature)
Temperature sensor
Thermocouple K
—-
Diameter (mm) 100
60
Greater diameters are suitable for flow measurements in the presence of turbulence with mediumlow air speeds (i.e. at the exit of the ducts). Lower diameters are suitable in applications where the probe surface must be much slower than the duct cross section within which the measurement is carried out, i.e. ventilation ducts.
Calibration
The AP472S1 and AP472S2 probes are calibrated in the factory; no calibration
is required by the user.
Operation
Where present, extend the telescopic rod to the necessary length paying
attention to the cable so that it can slide freely and without strain.
Introduce the probe in the air flow being measured, maintaining the arrow at
the top of the probe parallel to the flow as indicated in the following
figure.
HD2103
– 21 –
V2.4
The probe should be maintained orthogonal to the flow and not tilted in relation to it
OK
NO
The probe is correctly positioned in relation to the air-flow when the value measured is the maximum possible. Proceed with measurement following the instructions provided in this chapter.
Care and maintenance of the probes
The probe performance, mainly at low speeds, largely depends on the very slow
friction of the vane on its own axis. In order not to compromise this
characteristic, it is recommended that forcing is avoided, as well as blocking
or rotating the vane with the fingers, and if possible, avoid inserting it in
air flows that could spoil the probe.
HD2103
– 22 –
V2.4
Dimensions
1 2
3
AP472S1
AP472S2
To mount the telescopic rod on the probes, unscrew the handle (3) holding the probe body still in the point (1). Screw the rod end on the screw (2). The handle (3) is screwed on the telescopic rod.
WCT index
With particularly cold climatic conditions, the effect of wind speed on the
perceived temperature is not negligible: i.e. a temperature of -16°C in
presence of wind at 30km/h produces on the human body the same sensation of a
-37°C temperature in absence of wind.
When measuring the apparent temperature, the WCT parameter (wind chill
equivalent temperature) represents the apparent temperature in presence of
cold wind and is determined up to an actual temperature of +10°C; above this
limit the display shows an error (ERR). The WCT value is displayed in the main
line; the secondary line shows the actual temperature.
HD2103
– 23 –
V2.4
DIRECT INPUT Pt100, Pt1000 AND Ni1000 TEMPERATURE PROBES
The instrument accepts the input of Platinum temperature probes with
resistances of 100 and 1000, as well as Nickel temperature probes with a 1000
resistance. The Pt100 are connected to 4 wires, the Pt1000 and Ni1000 to 2
wires, with the excitation current chosen minimizing the sensor self-heating
effects. All probes with SICRAM module are calibrated in the factory, the 2 or
4-wire probes with direct input are checked for conformity with class A
tolerance according to norm IEC751 – BS1904 DIN43760. For the temperature
probes without SICRAM module (Pt100 4 wires, Pt1000 and Ni1000) the model
configuration is required (please see the description of the Probe Type menu
on page 10).
The °C or °F unit of measurement can be chosen for display, printing and
logging using the °C/°FESC key.
The probe is detected during turn on, and this cannot be performed when the
instrument is already on, therefore if a probe is connected and the instrument
is on, it is necessary to turn it off and on.
How to measure The response time for the measurement of the temperature in air
is greatly reduced if the air is moving. If the air is still, stir the probe.
Please remember that the response times in any case are longer than those for
liquid measurements.
The temperature measurement by immersion is carried out by inserting the probe
in the liquid for at least 60mm; the sensor is housed in the end part of the
probe.
In the temperature measurement by penetration the probe tip must be inserted
to a depth of at least 60mm, the sensor is housed in the end part of the
probe. When measuring the temperature on frozen blocks it is convenient to use
a mechanical tool to bore a cavity in which to insert the tip probe.
In order to perform a correct contact measurement, the measurement surface
must be even and smooth, and the probe must be perpendicular to the
measurement plane. So as to obtain the correct measurement, the insertion of a
drop of oil or heat-conductive paste is useful (do not use water or solvents).
This method also improves the response time.
Instructions to connect the TP47 connector for 4 wire Pt100, Pt1000 and Ni1000
probes All Delta Ohm probes are provided with a connector. The HD2103.1 and
HD2103.2 instruments also work with 4-wire direct Pt100, Pt1000 and Ni1000
probes manufactured by other producers; the TP47 connector is used for the
instrument connection to which the probe’s wires should be welded.
The instructions to connect the Platinum or Nickel probe to the module are provided below. The module is supplied complete with fairlead and gasket for 5mm maximum diameter cables.
HD2103
– 24 –
V2.4
Do the following to open the module and connect a probe: Unscrew the fairlead and extract the gasket, remove the label using a cutter, unscrew the ring on the opposite side as illustrated in the figure:
Open the two module shells: the printed circuit to which the probe must be connected is housed inside. On the left there are the 1…4 points on which the sensor wires must be welded. The JP1…JP4 jumpers are in the center of the board. These must be closed with a tin bead for some type of sensors:
4
Not Used
3
Ni1000
2
Pt1000
1
Pt100 3 wires
Before welding, pass the probe cable through the fairlead and gasket. Weld the wires as shown in the table:
Sensor
TP47 card connection
Pt100 4 wires
Pt100 4 wires
4
JP4
3
JP3
2
JP2
1
JP1
Jumper to close None
4
JP4
Pt1000 2 wires
Pt1000 2 wires
3
JP3
2
JP2
JP2
1
JP1
4
JP4
Ni1000
Ni1000 2 wires
3
JP3
2
JP2
JP3
1
JP1
HD2103
– 25 –
V2.4
Ensure the welds are clean and perfect. Once the welding operation is complete, close the two shells, insert the gasket in the module, and screw the fairlead and the ring with the O-Ring. Make sure the cable is not twisted while you are screwing the fairlead. Now the probe is ready.
Direct connection of 4-wire Pt100 sensors Sensor Direct soldering to the connector
Pt100 4 wires
View of the soldering side of the free female connector
4-wire Pt100 sensors can be soldered directly to the pins of the free female connector without making use of the TP47 board. The 4 wires of the Pt100 sensors have to be soldered as indicated in the figure on the left. In order to use this type of probe it is necessary to set up the menu item “Probe Type” as described at page 10. The P100 probe is recognized upon turning on the instrument: connect the probe when the instrument is switched off and then turn it on.
HD2103
– 26 –
V2.4
WARNINGS AND OPERATING INSTRUCTIONS
1. Do not expose the probes to gases or liquids that could corrode the
material of the sensor or the probe itself. Clean the probe carefully after
each measurement.
2. Do not bend the probe connectors or force them upward or downward.
3. Do not bend or force the contacts when inserting the probe connector into
the instrument.
4. Do not bend, deform or drop the probes, as this could cause irreparable
damage.
5. Always select the most suitable probe for your application.
6. Do not use probes in presence of corrosive gases or liquids. The sensor
container is made of AISI 316 stainless steel, while the contact probe
container is made of AISI 316 stainless steel plus silver. Avoid contact
between the probe surface and any sticky surface or substance that could
corrode or damage it.
7. Above 400°C and below 40°C, avoid violent blows or thermal shocks to
Platinum temperature probes as this could cause irreparable damage.
8. To obtain reliable temperature measurements, temperature variations that
are too rapid must be avoided.
9. Temperature probes for surface measurements (contact probes) must be held
perpendicular against the surface. Apply oil or heat-conductive paste between
the surface and the probe in order to improve contact and reduce reading time.
Whatever you do, do not use water or solvent for this purpose. A contact
measurement is always very hard to perform. It has high levels of uncertainty
and depends on the ability of the operator.
10. Temperature measurements on non-metal surfaces usually require a great
deal of time due to the low heat conductivity of non-metal materials.
11. Probes are not insulated from their external casing; be very careful not
to come into contact with live parts (above 48V). This could be extremely
dangerous for the instrument as well as for the operator, who could be
electrocuted.
12. Avoid taking measurements in presence of high frequency sources,
microwave ovens or large magnetic fields; results may not be very reliable.
13. Clean the probe carefully after use.
14. The instrument is water resistant and IP66, but should not be immersed in
water without closing the free connectors using caps. The probe connectors
must be fitted with sealing gaskets. Should the instrument fall into the
water, check for any water infiltration. Gently handle the instrument in such
a way as to prevent any water infiltration from the connectors’ side.
HD2103
– 27 –
V2.4
INSTRUMENT SIGNALS AND FAULTS
The following table lists all error indications and information displayed by
the instrument and supplied to the user in different operating situations:
Display indications
Explanation
ERR
This appears if the probe, has already been detected by the instrument, but is disconnected. At the same time an intermittent beep is issued.
– – –
This appears in the display central line when a temperature only probe is connected. In the lower line the temperature is shown correctly.
This appears if the probe, has already been detected by the instrument, PROB COMM LOST but is disconnected. At the same time an intermittent beep is issued.
OVER
Measurement overflow: indicates that the probe is measuring a value exceeding the measuring range.
LOG MEM FULL
Memory full; the instrument cannot store further data, the memory space is exhausted.
NEW PROBE DET
This message appears when a new probe is inserted while the instrument is on. Turn the instrument off and then back on again.
PROB ERR
A probe with SICRAM module has been inserted when not admissible for that specific instrument.
SYS ERR
CAL LOST
Instrument management program error. Contact the instrument’s supplier and
communicate the numeric code # reported by the display.
Program error: it appears after turning on for a few seconds. Contact the
instrument’s supplier.
BATT TOO LOW CHNG NOW
Indication of insufficient battery charge appearing on turning on. The instrument issues a long beep and turns off. Replace the batteries.
HD2103
– 28 –
V2.4
The following table reports the indications provided by the instrument as they appear on the display, together with their description.
Display indications
Explanation
_LOG_DUMP_or_ERAS >>>_PRBETYPE AVG TIME SECS BATT TOO LOW – CHNG NOW BAUDRATE >>> COMM STOP COMM STRT DAY DUMP_END DUMP_In_PROG >>> ERR FUNC CLR FUNC CLRD HOUR LOG In PROG LOG MEM FULL LOG_CLRD LOG_STOP LOG_STRT MIN >>> USE_UNIT_TO_ZERO SEC MNTH NEW_PROB_DET NOPRBE SER_NUM OVER PLS_EXIT >>> FUNC RES_FOR_FACT ONLY PRBE_SER #### #### PRNT AND LOG INTV PRNT INTV >>> PROB COMM LOST PROB ERR SECT inch2 SECT m2 SLP_MODE_LOG SYS ERR # YEAR
transfer or erase data type of probe connected time interval according to
which the moving average is calculated battery discharged – replace it
immediately baud rate value printing complete printing started day data
transfer complete data transfer in progress error max, min and average values
clearing max, min and average values cleared hour logging in progress memory
full memory data cleared logging complete logging started minutes >>> use the
UNIT key to reset the seconds month new probe detected the connected probe’s
serial number is absent maximum limit exceeded please exit using ESC >>>
function reserved to factory calibration
serial number #### #### of the connected probe printing and logging intervals
printing interval lost communication with probe error – unexpected probe
section in inch2 section in m2 turning off during recording mode program error
number # year
HD2103
– 29 –
V2.4
LOW BATTERY WARNING AND BATTERY REPLACEMENT
The battery symbol on the display constantly shows the battery charge status.
To the extent that batteries have discharged, the symbol “empties”. When the
charge decreases still further it starts blinking…
In this case, batteries should be replaced as soon as possible. If you
continue to use it, the instrument can no longer ensure correct measurement.
The memory data are maintained. If the battery charge level is insufficient,
the following message appears when you turn the instrument on:
BATT TOO LOW CHNG NOW
The instrument issues a long beep and turns off. In this case, replace the
batteries in order to turn the instrument back on.
If the HD2103.2 is logging and battery voltage falls below the minimum
operating level. the logging session is ended in order to avoid data loss.
The battery symbol turns off when the external power supply is connected.
To replace the batteries, switch the instrument off and unscrew the battery
cover counter clockwise. After replacing the batteries (4 1.5V alkaline
batteries – type AA) screw the cover on clockwise.
After replacing the batteries, the date, time, baud rate, type of probe, printing interval, logging parameters must be set again: in order to simplify the operation, on insertion of the new batteries the instrument turns on automatically and requests these parameters in sequence. To go to the next item press ENTER; to return to measurement mode, press MENU.
HD2103
– 30 –
V2.4
MALFUNCTIONING UPON TURNING ON AFTER BATTERY REPLACEMENT After replacing the
batteries, the instrument may not restart correctly; in this case, repeat the
operation. After disconnecting the batteries, wait a few minutes in order to
allow circuit condensers to discharge completely; then reinsert the batteries.
WARNING ABOUT BATTERY USE · Batteries should be removed when the instrument is
not used for an extended time. · Flat batteries must be replaced immediately.
· Avoid batteries leaking. · Always use good quality leakproof alkaline
batteries. Sometimes on the market, it is possible to find new batteries with
an insufficient charge capacity.
INSTRUMENT STORAGE
Instrument storage conditions: · Temperature: -25…+65°C. · Humidity: less than
90%RH without condensation. · Do not store the instrument in places where:
Humidity is high. The instrument may be exposed to direct sunlight. The
instrument may be exposed to a source of high temperature. The instrument may
be exposed to strong vibrations. The instrument may be exposed to steam, salt
or any corrosive gas.
The instrument case is made of ABS plastic and the protections are rubber: do
not use any incompatible solvent for cleaning.
HD2103
– 31 –
V2.4
SERIAL INTERFACE AND USB
The HD2103.1 and HD2103.2 instruments are fitted with an electrically isolated
RS-232C serial interface; the HD2103.2 also has an USB 2.0 interface. The
following serial cables can be used:
· HD2110CSNM: serial connection cable with 8-pole MiniDin connector on one end
and 9-pole Sub D connector on the other end;
· C.206: serial connection cable with 8-pole MiniDin connector on one end and
USB type A connector on the other end. With integrated RS232/USB converter;
· CP23: connection cable with Mini-USB type B connector on one end and USB
type A connector on the other end (only for HD2103.2).
The connection via the C.206 cable requires the previous installation of the
cable USB drivers. Install the drivers before connecting the C.206 cable to
the PC. The connection via the CP23 cable does not require the installation of
USB drivers: when connecting the instrument to the PC, the Windows® operating
system automatically recognizes the device as an HID device (Human Interface
Device) and uses the drivers already included in the operating system.
Cable
HD2110CSNM C.206 CP23
Instrument port
RS232 (MiniDin) RS232 (MiniDin) USB (Mini-USB)
PC port RS232 (9-pole SubD) USB USB
Installation of USB drivers
No Yes No
The instrument standard serial transmission parameters are:
· Baud rate 38400 baud
· Parity
None
· N. bit
8
· Stop bit 1
· Protocol Xon/Xoff
It is possible to change the RS232C serial port baud rate by setting the “Baudrate” parameter in the menu (please see page 12). The possible values are: 38400, 19200, 9600, 4800, 2400, 1200. The other transmission parameters are fixed. The USB 2.0 connection does not require the setting of parameters.
The instruments are provided with a complete set of commands and data queries to be sent via the PC. The serial commands work with a standard serial communication program (e.g. Hyperterminal) only through the RS232 serial port of the instrument, using the cable HD2110CSNM or the cable C.206.
All the commands transferred to the instrument must have the following structure: XYcr where: XY is the command code and cr is the Carriage Return (ASCII 0D)
Command P0 P1 S0 G0
Response & &
3.25 22.7 Model HD2103 -2
Description Ping (locks the instrument keyboard for 70 seconds) Unlocks the instrument keyboard Captured measurements (24 characters) Instrument model
HD2103
– 32 –
V2.4
Command G1 G2 G3 G4 G5 G6 G7 G8 GB GC LN LD LE
K1
K0 K4 K5 K7 K6 KP KQ RA RP RUA RUB
Response M=Thermo-Anemometer SN=12345678 Firm.Ver.=01-00 Firm.Date=2004/06/15
cal 0000/00/00 00:00:00 Probe=Sicram hot wire Probe SN=11119999 Probe
cal.=2004/01/12 User ID=0000000000000000
&1999 PRINTOUT OF LOG & PRINTOUT IMMEDIATE MODE
& & & & & & & # & 600 U= m/s U= °C
WA#
&
WC0
&
WC1
&
Description Model description Instrument serial number Firmware version
Firmware date Calibration date and time Type of probe connected to input Probe
serial number Probe calibration date User code (set with T2xxxxxxxxxxxxxxxx)
Print instrument’s heading Number of free pages in the flash memory Print data
logged in flash Erase data in flash memory
Immediate printing of data
Stop printing data Start logging data Stop logging data Enable REL function
Disable REL function Auto-power-off function=ENABLE Auto-power-off
function=DISABLE Reading of LOG/PRINT interval set Battery level (Resolut.
0.01V) Main unit of measurement Secondary unit of measurement Setting
LOG/PRINT interval. # is a hexadecimal number 0…D that represents the position
of the interval in the list 0, 1, 5, 10, …, 3600 seconds. Setting SELF off
Setting SELF on
Command characters are exclusively upper case characters. Once a correct command is entered, the instrument responds with “&”; when any wrong combination of characters is entered, the instrument responds with “?”. The instrument response strings end with the sending of the CR command (Carriage Return). The instrument does not send the LF command (Line Feed). Before sending commands to the instrument via the serial port, locking the keyboard to avoid functioning conflicts is recommended: use the P0 command. When complete, restore the keyboard with the P1 command.
HD2103
– 33 –
V2.4
STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER
The HD2103.1 and HD2103.2 instruments can be connected to a personal computer
via an RS232C serial port or USB 2.0 port, and exchange data and information
through the DeltaLog9 software running in a Windows operating environment.
Both models can send in real time input measured values directly to a PC,
through the PRINT function; the HD2103.2 can also store the values measured by
using the Logging function (LOG key) in its internal memory. If necessary, the
data stored in the memory can be transferred to a PC later.
THE LOGGING FUNCTION – ONLY FOR HD2103.2
The Logging function allows the recording up to 38,000 measurements registered
by the probe connected to the input. The time interval between two consecutive
measurements can be set from 1 second to 1 hour. The logging starts by
pressing the LOG key and ends by pressing the same key again: the data
memorized in this way form a continuous block of data.
See the description of the menu item on page 10.
If the automatic turning off option between two recordings (menu >>
Sleep_Mode_LOG) is enabled, upon pressing the LOG key the instrument logs the
first data and turns off (the LOG ON indication periodically appears on the
display to indicate that the Logging function is active). 15 seconds before
the next logging instant, it turns on again to capture the new sample, and
then turns off.
The data stored in the memory can be transferred to a PC using the DUMP LOG
command: MENU >> LOG. During data transfer the display shows the message DUMP;
to stop the data transfer press ESC on the instrument or on the PC.
CLEARING THE MEMORY – ONLY FOR HD2103.2
To clear the memory, use the Erase Log function (MENU >> SERIAL). The
instrument starts clearing the internal memory; at the end of the operation,
it goes back to normal display.
NOTES: · Data transfer does not cause the memory to be erased; the operation
can be repeated as many
times as required. · The logged data remain in the memory apart from the
batteries charge conditions. · In order to print the data to a parallel
interface printer, you must use a parallel-serial adaptor (not
supplied). · The direct connection between instrument and printer via a USB
connector does not work. · Some keys are disabled during logging. The
following keys work: HOLD, FUNC (Max-Min-
Avg) and SERIAL. · Pressing the HOLD, REL and FUNC keys has no effect on the
logged data if these keys are
pressed after starting the recording, otherwise the following is valid. · The
recording started with the display in HOLD mode proceeds normally with the
actual
measured values (that is, not in “HOLD” mode). Only the display is frozen to
the values present when the HOLD key was pressed. · The same is true for the
Max-Min-Avg function. · If the logging is started when the display is in REL
mode, the relative values are logged. · It is possible to activate both the
logging (LOG) and direct transfer (PRINT) functions at the same time.
HD2103
– 34 –
V2.4
THE PRINT FUNCTION
The PRINT function sends the measurements taken in real time by the instrument
inputs directly to a PC or a printer. Print data units of measurements are the
same as those used on the display. The function is started by pressing SERIAL.
The time interval between two consecutive prints can be set from 1 second to 1
hour (please see the Print and log interval menu item on page 10). If the
print interval is equal to 0, by pressing SERIAL the single data is sent to
the connected device. If the print interval is higher than 0, the data
transfer continues until the operator stops it by pressing SERIAL again.
The PRINT function works with a standard serial communication program (e.g.
Hyperterminal) only through the RS232 serial port of the instrument, using the
cable HD2110CSNM or the cable C.206.
Connect the HD40.1 printer using cable HD2110CSNM.
NOTES:
· The print out is formatted across 24 columns. · Some keys are disabled
during serial transmission. The following keys work: ON/OFF, HOLD,
FUNC (Max-Min-Avg) and LOG. · Pressing the HOLD, REL and FUNC keys has no
effect on the printed data if these keys are
pressed after starting the printing, otherwise the following is valid. · The
transfer started with the display in HOLD mode proceeds normally with the
actual
measured values (that is, not in “HOLD” mode). Only the display is frozen to
the values present when the HOLD key was pressed. · The same is true for the
Max-Min-Avg function. · If the serial transfer is started when the display is
in REL mode, the relative values are transferred. · It is possible to activate
both the logging (LOG) and direct transfer (PRINT) functions at the same time.
HD2103
– 35 –
V2.4
CONNECTION TO A PC
HD2103.1 Connection to the PC with the cable:
· HD2110CSNM: 8-pole MiniDin connector on one end and 9-pole Sub D connector
on the other end;
· C.206: 8-pole MiniDin connector on one end and USB type A connector on the
other end. With integrated RS232/USB converter (requires the installation of
the USB drivers).
HD2103.2 Connection to the PC with the cable:
· CP23: Mini-USB type B connector on one end and USB type A connector on the
other end; · HD2110CSNM: 8-pole MiniDin connector on one end and 9-pole Sub D
connector on the
other end;
· C.206: 8-pole MiniDin connector on one end and USB type A connector on the
other end. With integrated RS232/USB converter (requires the installation of
the USB drivers).
The instruments are supplied with the DeltaLog9 software that manages the
connection, data transfer, graphic presentation, and printing operations of
the captured or logged measurements. The DeltaLog9 software is complete with
“On-line Help” (also in PDF format) describing its characteristics and
functions.
CONNECTION TO THE RS232C SERIAL PORT OF THE INSTRUMENT
1. The measurement instrument must be switched off. 2. Using the Delta Ohm
HD2110CSNM or C.206 cable, connect the measuring instrument to the
first free RS232C (COM) or USB serial port of the PC. 3. Turn on the
instrument and set the baud rate to 38400 (MENU >> ENTER until the Baud Rate
parameter >> select 38400 using the arrows >> confirm with ENTER). The
parameter remains in the memory until replacement of the batteries. 4. Launch
the DeltaLog9 application and press CONNECT. Wait for the connection to occur
and follow the indications on the screen. For a description of the DeltaLog9
application, please refer to its on-line Help.
CONNECTION TO THE USB 2.0 PORT OF THE INSTRUMENT – ONLY FOR HD2103.2
The connection via the CP23 cable does not require the installation of USB
drivers: when connecting the instrument to the PC, the Windows® operating
system automatically recognizes the device as an HID device (Human Interface
Device) and uses the drivers already included in the operating system. To
check if the connection has been successfully completed, double-click on
“Device Manager” from the Control Panel. The following items should appear:
“Human Interface Device” >> “HID-compliant device” “Human Interface Device” >>
“USB Human Interface Device”
When the USB cable is disconnected, the items disappear and reappear when it
is connected again.
HD2103
– 36 –
V2.4
NOTES ABOUT WORKING AND OPERATIVE SAFETY
Authorized use
The technical specifications as given in chapter “TECHNICAL CHARACTERISTICS”
must be observed. Only the operation and running of the measuring instrument
according to the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.
General safety instructions
This measuring system is constructed and tested in compliance with the EN
61010-1:2010 safety regulations for electronic measuring instruments. It left
the factory in a safe and secure technical condition. The smooth functioning
and operational safety of the measuring system can only be guaranteed if the
generally applicable safety measures and the specific safety instructions in
this operating manual are followed during operation. The smooth functioning
and operational safety of the instrument can only be guaranteed under the
environmental and electrical operating conditions that are in specified in
chapter “TECHNICAL CHARACTERISTICS”.
Do not use or store the product in places such as listed below: · Rapid
changes in ambient temperature which may cause condensation. · Corrosive or
inflammable gases. · Direct vibration or shock to the instrument. · Excessive
induction noise, static electricity, magnetic fields or noise.
If the measuring system was transported from a cold environment to a warm
environment, the formation of condensate can impair the functioning of the
measuring system. In this event, wait until the temperature of the measuring
system reaches room temperature before putting the measuring system back into
operation.
Obligations of the purchaser
The purchaser of this measuring system must ensure that the following laws and
guidelines are observed when using dangerous substances:
EEC directives for protective labour legislation National protective labour
legislation Safety regulations
HD2103
– 37 –
V2.4
INSTRUMENT TECHNICAL CHARACTERISTICS
Instrument Dimensions (Length x Width x Height) Weight Materials Display
Operating conditions Working temperature Storing temperature Working relative
humidity Protection degree
Power Supply Batteries Autonomy (*) Power absorbed with instrument off Mains
(cod. SWD10)
Measuring units
Security of memorized data
Time Date and time Precision
Measured values memorization – model HD2103.2 Type Quantity Selectable storage
interval
Serial interface RS232C Type Baud rate Data bit Parity Stop bit Flow Control
Serial cable length
185x90x40mm 470g (complete with batteries) ABS, rubber 2×4½ characters plus
symbols Visible area: 52x42mm
-5…50°C -25…65°C 0…90%RH without condensation IP66
4 1.5V type AA batteries 200 hours with 1800mAh alkaline batteries 20µA Mains
adapter 100-240Vac/12Vdc-1A °C – °F – m/s – km/h – ft/min – mph – knot l/s –
m3/s – m3/min – m3/h – ft3/s – ft3/min WCT Unlimited, independently of battery
charge conditions
Schedule in real time 1min/month max drift
2000 pages of 19 samples each Total of 38000 samples 1s, 5s, 10s, 15s, 30s,
1min, 2min, 5min, 10min, 15min, 20min, 30min and 1hour
RS232C electrically isolated Can be set from 1200 to 38400 baud 8 None 1
Xon/Xoff Max 15m
(*) It’s referred to all the probes except the hot-wire ones, which autonomy is stated forwards in the relevant characteristics.
HD2103
– 38 –
V2.4
Selectable print interval
USB interface – model HD2103.2 Type
Connections Input module for the probes RS232 serial interface USB interface
(only HD2103.2) Mains adapter (cod. SWD10)
Temperature measurement Pt100 measurement range Pt1000 measurement range
Ni1000 measurement range Resolution Accuracy Drift after 1 year
immediate or 1s, 5s, 10s, 15s, 30s, 1min, 2min, 5min, 10min, 15min, 20min,
30min and 1hour
1.1 – 2.0 electrically isolated
8-pole male DIN45326 connector 8-pole MiniDin connector Mini-USB type B
connector 2-pole connector (positive at centre)
-200…+650°C -200…+650°C -50…+250°C 0.1°C ± 0.1°C (excluding probe error) 0.1°C/year (only the instrument)
HD2103
– 39 –
V2.4
TECHNICAL DATA OF PROBES AND MODULES IN LINE WITH THE INSTRUMENT
WIND SPEED MEASUREMENT PROBES Hot-wire probes: AP471S1 – AP471S2 – AP471S3 – AP471S4
AP471S1 – AP471S3
AP471S2
AP471S4
Type of measure
Wind speed, calculated flow rate, air temperature
Type of sensor
Speed Temperature Measurement range
NTC thermistor NTC thermistor
Omnidirectional NTC thermistor NTC thermistor
Speed Temperature Measurement resolution:
0.1…40 m/s -25…+80°C
0.1…5 m/s
-25…+80°C
0…80°C
Speed
Temperature Measurement accuracy:
0.01 m/s 0.1 km/h 1 ft/min 0.1 mph 0.1 knot
0.1°C
Speed
Temperature Minimum speed
±0.2 m/s (0.10…0.99 m/s) ±0.4 m/s (1.00…9.99 m/s) ±0.8 m/s (10.00…40.00 m/s)
±0.8°C (-10…+80°C)
±0.2 m/s (0.10…0.99 m/s) ±0.3 m/s (1.00…5.00 m/s)
±0.8°C (-10…+80°C)
0.1 m/s
Air temperature compensation
0…80°C
Battery life Unit of Measurement
Approx. 20 hours @ 20 m/s with Approx. 30 hours @ 5 m/s with
alkaline batteries
alkaline batteries
Speed Flow rate
m/s km/h ft/min mph knot l/s – m3/s – m3/min – m3/h – ft3/s – ft3/min
Pipeline section for flow rate calculation
0.0001…1.9999 m2
Cable length
~2 m
HD2103
– 40 –
V2.4
Vane probes: AP472S1 – AP472S2 AP472S1
AP472S2
Type of measure
Wind speed, calculated flow rate,
air temperature
Wind speed, calculated flow rate
Diameter Type of measurement
Speed Temperature Measurement range Speed Temperature Resolution Speed
Temperature Accuracy
Speed Temperature Minimum speed Unit of Measurement Speed Flow rate Pipeline
section for flow rate calculation Cable length
100mm
60mm
Vane K thermocouple
Vane —-
0.6…25 m/s -25…+80 (*)
0.5…20 m/s -25…+80 (*)
0.01 m/s 0.1 km/h 1 ft/min 0.1 mph 0.1 knot
0.1°C
—-
± (0.4 m/s + 1.5% f.s.) ±0.8°C 0.6 m/s
± (0.4 m/s + 1.5% f.s.) —-
0.5 m/s
m/s km/h ft/min mph knot l/s – m3/s – m3/min – m3/h – ft3/s – ft3/min
0.0001…1.9999 m2
~2m
(*) The indicated value refers to the vane’s working range.
HD2103
– 41 –
V2.4
TEMPERATURE PROBES Pt100 SENSOR USING SICRAM MODULE
Model TP472I TP472I.O TP473P.I TP473P.O TP474C.O TP475A.O TP472I.5 TP472I.10
TP49A.I TP49AC.I TP49AP.I TP875.I
TP876.I TP87.O
Type
Application range
Accuracy
Immersion
-196°C…+500°C
Immersion Penetration Penetration
Contact
Air
Immersion
Immersion
Immersion
Contact
Penetration Globe-thermometer
Ø 150 mm Globe-thermometer
Ø 50 mm Immersion
-50°C…+300°C -50°C…+400°C -50°C…+300°C -50°C…+300°C -50°C…+250°C -50°C…+400°C -50°C…+400°C -70°C…+250°C -70°C…+250°C -70°C…+250°C -30°C…+120°C
-30°C…+120°C -50°C…+200°C
±0.1°C (@ 0 °C) ±0.2°C (-50 °C t 250 °C) ±0.3°C (t < -50 °C; t > 250 °C)
Common characteristics Resolution
Temperature drift @ 20°C
0.01°C in the range ±199.99°C, 0.1°C in the remaining field 0.003%/°C
Probes Pt100 4 wires and Pt1000 2 wires
Model
TP47.100.O TP47.1000.O TP87.100.O TP87.1000.O
Type
Pt100 4 wires Pt1000 2 wires Pt100 4 wires Pt1000 2 wires
Application range
-50…+250°C -50…+250°C -50…+200°C -50…+200°C
Accuracy
Class A Class A Class A Class A
Common characteristics Resolution
Temperature drift @ 20°C Pt100 Pt1000
0.01°C in the range ±199.99°C, 0.1°C in the remaining field
0.003%/°C 0.005%/°C
HD2103
– 42 –
V2.4
ORDER CODES
HD2103.1
Kit including the instrument HD2103.1, 4×1.5V alkaline batteries, operating manual, case and DeltaLog9 software downloadable from Delta OHM website. The probes and the cables must be ordered separately.
HD2103.2
Kit including the HD2103.2 data logger, 4×1.5V alkaline batteries, operating manual, case and, CP23 and DeltaLog9 software downloadable from Delta OHM website. The probes and the cables must be ordered separately.
HD2110CSNM Connection cable 8-pole MiniDin Sub D 9-pole female for RS232C.
C.206
Connection cable 8-pole MiniDin USB type A. With integrated RS232/USB converter.
SWD10
Stabilized power supply at 100-240Vac/12Vdc-1A mains voltage.
HD40.1
The kit includes: 24-column portable thermal printer, serial interface, 57mm paper width, four NiMH 1.2V rechargeable batteries, SWD10 power supply, instruction manual, 5 thermal paper rolls.
BAT.40
Spare battery pack for HD40.1 printer with in-built temperature sensor.
RCT
The kit includes 4 thermal paper rolls 57mm wide and 32mm in diameter.
PROBES COMPLETE WITH SICRAM MODULE
WIND SPEED MEASUREMENT PROBES
HOT-WIRE PROBES:
AP471S1 AP471S2
AP471S3
AP471S4
Hot-wire telescopic probe, measuring range: 0.1…40 m/s. Cable length 2 metres.
Omnidirectional hot-wire telescopic probe, measuring range: 0.1…5 m/s. Cable
length 2 metres.
Hot-wire telescopic probe with terminal tip for easy position, measuring
range: 0.1…40 m/s. Cable length 2 metres.
Omnidirectional hot-wire telescopic probe with base, measuring range: 0.1…5
m/s. Cable length 2 metres.
VANE PROBES:
AP472S1 AP472S2 AST.1 AP471S1.23.6 AP471S1.23.7
Vane probe with thermocouple, Ø100 mm. Speed from 0.6 to 25 m/s; temperature
from -25 to 80°C. Cable length 2 metres.
Vane probe, Ø60 mm. Measurement range: 0.5…20 m/s. Cable length 2 metres.
Telescopic shaft (minimum length 210 mm, maximum length 870 mm) for AP472S1
and AP472S2 vane probes.
Fixed extension shaft 16 x 300 mm, M10 male thread on a side, female on the
other. For vane probes AP472S1 and AP472S2.
Fixed extension shaft 16 x 300 mm, M10 female thread on a side only. For vane
probes AP472S1 and AP472S2.
HD2103
– 43 –
V2.4
TP472I TP472I.O TP473P.I TP473P.O TP474C.O TP475A.O TP472I.5 TP472I.10 TP49A.I TP49AC.I TP49AP.I TP875.I TP876.I TP87.O
TEMPERATURE MEASUREMENT PROBES
Immersion probe, sensor Pt100. Stem Ø3 mm, length 300 mm. Cable length 2
metres.
Immersion probe, sensor Pt100. Stem Ø3 mm, length 230 mm. Cable length 2
metres.
Penetration probe, sensor Pt100. Stem Ø4 mm, length 150 mm. Cable length 2
metres.
Penetration probe, sensor Pt100. Stem Ø4 mm, length 150 mm. Cable length 2
metres.
Contact probe, sensor Pt100. Stem Ø4 mm, length 230 mm, contact surface Ø 5
mm. Cable length 2 metres.
Air probe, sensor Pt100. Stem Ø4 mm, length 230 mm. Cable length 2 metres.
Immersion probe, sensor Pt100. Stem Ø6 mm, length 500 mm. Cable length 2
metres.
Immersion probe, sensor Pt100. Stem Ø6 mm, length 1000 mm. Cable length 2
metres.
Immersion probe, sensor Pt100. Stem Ø2.7 mm, length 150 mm. Cable length 2
metres. Aluminium handle.
Contact probe, sensor Pt100. Stem Ø4 mm, length 150 mm. Cable length 2 metres.
Aluminium handle.
Penetration probe, sensor Pt100. Stem Ø2.7 mm, length 150 mm. Cable length 2
metres. Aluminium handle.
Globe-thermometer Ø150 mm with handle. Cable length 2 metres.
Globe-thermometer Ø50 mm with handle. Cable length 2 metres.
Immersion probe, sensor Pt100. Stem Ø3 mm, length 70 mm. Cable length 2
metres.
TEMPERATURE PROBES WITHOUT SICRAM MODULE
TP47.100.O
Immersion probe, sensor Pt100 direct 4 wires. Probe’s stem Ø3 mm, length 230 mm. 4-wire connection cable with connector, length 2 metres.
TP47.1000.O Immersion probe, sensor Pt1000. Probe’s stem Ø3 mm, length 230 mm. 2-wire connection cable with connector, length 2 metres.
TP87.100.O
Immersion probe, sensor Pt100 direct 4 wires. Probe’s stem Ø3 mm, length 70 mm. 4-wire connection cable with connector, length 2 metres.
TP87.1000.O Immersion probe, sensor Pt1000. Probe’s stem Ø3 mm, length 70 mm. 2-wire connection cable with connector, length 2 metres.
TP47
Only connector for probe connection: Pt100 direct 3 and 4 wires, Pt1000 and Ni1000 2 wires.
DELTA OHM metrology laboratories LAT N° 124 are accredited by ACCREDIA for Temperature, Humidity, Pressure, Photometry / Radiometry, Acoustics and Air Velocity. They can supply calibration certificates for the accredited quantities.
HD2103
– 44 –
V2.4
NOTES
HD2103
– 45 –
V2.4
GUARANTEE
TERMS OF GUARANTEE
All DELTA OHM instruments are subject to accurate testing, and are guaranteed
for 24 months from the date of purchase. DELTA OHM will repair or replace free
of charge the parts that, within the warranty period, shall be deemed non
efficient according to its own judgement. Complete replacement is excluded and
no damage claims are accepted. The DELTA OHM guarantee only covers instrument
repair. The guarantee is void in case of incidental breakage during transport,
negligence, misuse, connection to a different voltage than that required for
the appliance by the operator. Finally, a product repaired or tampered by
unauthorized third parties is excluded from the guarantee. The instrument
shall be returned FREE OF SHIPMENT CHARGES to your dealer. The jurisdiction of
Padua applies in any dispute.
The electrical and electronic equipment marked with this symbol cannot be disposed of in public landfills. According to the Directive 2011/65/EU, the european users of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new one. The illegal disposal of electrical and electronic equipment is punished with an administrative fine.
This guarantee must be sent together with the instrument to our service centre. IMPORTANT: Guarantee is valid only if coupon has been correctly filled in all details.
I nst rum ent Code:
HD2103.1
HD2103.2
Serial Number
REN EW ALS
Date I nspect or
Date I nspect or
Date I nspect or
Date I nspect or
Date I nspect or
Date I nspect or
GHM GROUP Delta OHM | Delta Ohm S.r.l. a socio unico Via Marconi 5 | 35030 Caselle di Selvazzano | Padova | ITALY Phone +39 049 8977150 | Fax +39 049 635596 www.deltaohm.com | sales@deltaohm.com
GHM GROUP Delta OHM | Delta Ohm S.r.l. a socio unico Via Marconi 5 | 35030
Caselle di Selvazzano | Padova | ITALY Phone +39 049 8977150 | Fax +39 049
635596 www.deltaohm.com | sales@deltaohm.com
The qualit y level of our inst rum ent s is t he result of t he const ant
developm ent of t he product . This m ay produce som e differences bet ween t
he inform at ion writ t en in t his m anual and t he inst rum ent you have
purchased. We cannot com plet ely exclude t he possibilit y of errors in t he
m anual, for which we apologize. The dat a, im ages and descript ions included
in t his m anual cannot be legally assert ed. We reserve t he right t o m ake
changes and correct ions wit h no prior not ice.
GHM GROUP Delta OHM | Delta Ohm S.r.l. a socio unico Via Marconi 5 | 35030
Caselle di Selvazzano | Padova | ITALY Phone +39 049 8977150 | Fax +39 049
635596 www.deltaohm.com | sales@deltaohm.com
V2 .4 21/ 01/ 2021