Mitutoyo 99MAM033A USB Input Tool Interface Box User Manual
- June 9, 2024
- Mitutoyo
Table of Contents
- Safety Precautions
- NOTICE
- Overview
- Names and Functions of Each Part
- Connecting This Product
- Inputting Measurement Data
- Applications of Options
- Troubleshooting
- Specifications
- List of Dogmatic Connecting Cables by the Tools to Be Connected
- CUSTOMRT SUPPORT
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Mitutoyo 99MAM033A USB Input Tool Interface Box
Microsoft®, Windows®, and Excel® are registered trademarks and/or trademarks of Microsoft Corporation in the United States and/or other countries.
Safety Precautions
To ensure operator safety, use this instrument in conformance with the
directions and specifications given in this User’s Manual.
Non-conformance to the instructions may compromise safety.
NOTICE
- Be sure to confirm the following to ensure this product is used correctly and safely. This product operates using power supplied from a PC. Because the PC might generate a large amount of noise depending on its manufacturer or type, check this product’s operation with the entire system before using it. To check the operation, perform the following procedure.
- Connect this product to a measuring tool and a PC, and then turn on the PC.
- Perform Zero setting for the measuring tool. (Refer to the measuring tool’s User’s Manual.)
- After measuring, return the contact point/stylus to the zero set position, and check that the display has returned to zero.
- Repeat step 3) about five times. If zero is indicated each time, there is no problem. You can use this product safely.
- If any problems are found, please contact our office.
- Do not use this product in locations where it will can be directly affected by machining oil, water, dust, etc.
- Do not use this product at sites where it will be exposed to direct sunlight or blowing hot air. Do not leave this product in a car left in the sunlight or near open flame.
- Applying excessive force to this product or dropping it may damage it. Please handle it with care.
- Do not disassemble this product or do anything that might cause it to short out.
- Clean dirt off this product by wiping it with a lint-free cloth or paper soaked in alcohol. Do not use an organic solvent such as thinner.
Overview
This product is a tool for inputting the measurement data of a measuring tool into a PC. By simply pressing the DATA button on this product or on the measuring tool, measurement data can be entered into spreadsheet software such as Excel or a text editor. The operation of entering measurement data values using the PC’s keyboard and pressing the [Enter] key can be executed with a single button press.
Names and Functions of Each Part
- USB cable (A-Micro B type: 06AFZ050) Connects to this product on one side, and the USB port of a PC or a commercially-available USB hub on the other side.
- DATA button Inputs measurement data into the PC.
- Connector for the foot switch Connects to the optional foot switch (937179T or 12AAJ088)
- USB connector (Micro-B Receptacle) Connects to the supplied USB cable.
- Connector for measuring tools Connects to an optional dogmatic connecting cable.
- Label Indicates the code number, etc., of this product.
Connecting This Product
Connect an optional dogmatic connecting cable to the dogmatic output port of
the measuring tool and to the connector for measuring tools of this product.
Also, connect the USB cable to the USB port of the PC and to the USB connector
of this product. When this product is connected, a driver is automatically
installed onto the PC.
Multiple measuring tools can be connected to one PC.
NOTICE
- When connecting the Type A/B/G to a measuring tool, be sure to tighten the mounting screws. If the screws are not tightened, water resistance is not maintained, resulting in a failure. For the Type G, use the screws supplied with the measuring tool. When attaching or removing screws, use the screwdriver supplied to the measuring tool, or use a commercially-available size 0 screwdriver. Tighten screws using a torque of about 5 to 8 N·cm. For details about connection methods, refer to each measuring tool’s User’s Manual.
- When connecting multiple measuring tools, that number of USB ports must be available. When the number of required USB ports are not available on the PC, use a commercially-available USB hub.
Inputting Measurement Data
- Start up the software that will be used for inputting measurement data.
- Set the PC’s text input to single-byte input (only when using languages that support double-byte characters). 3
- Change the cell position or move the cursor to the position where the measurement data will be input.
- Press the DATA button on this product or on the measuring tool.
NOTICE
- When multiple measuring tools are connected, measurement data can be input from one tool at a time.
- To input measurement data from multiple measuring tools at once, the special software USB-ITPAK (option) is required. (Refer to “5. Applications of Options”.)
Applications of Options
The following various functions are available when the special software USB- ITPAK (option) is used.
- Inputting measurement data from multiple measuring tools at once
- Specifying the input destination for measurement data
- Using other software (software compatible with the communication interface of this product) to input measurement data via the VCP driver (included with USB-ITPAK)
- Inputting measurement data via foot operation (optional foot switch and USB-FSW are required)
For details, refer to each option’s User’s Manual.
NOTICE
When a foot switch is directly connected to this product, measurement data is input for the measuring tool that is connected to the product.
Troubleshooting
If this product does not operate correctly, check if any of following problems
apply.
If the following solutions do not resolve the problem, please contact the
dealer or office where you obtained the product.
Problem | Solution |
---|
- Even after this product is connected to the PC, it is not recognized by the PC.
- When the PC is restored from suspension (sleep mode, standby mode), this product is not recognized by the PC.
|
-
Disconnect this product from the PC, wait about 10 seconds, and then reconnect it. Do not operate this product or the PC while the PC is recognizing this product. Perform the following operations on the PC to check whether this product is recognized by the PC.
(1) Open the [Device Manager] screen.
While pressing the Windows logo key, press X, and then select [Device Manager].
(2) Open [Human Interface Devices] and disconnect and reconnect the USB connector to check that the number of devices in [Human Interface Devices] or [USB Input Devices] increases or decreases by one.
If the number of devices does not increase or decrease, try the following solutions. -
If a spare PC or USB hub is available, replace the PC or USB hub being used with the spare for the check.
-
When connecting a USB hub to a USB hub, make sure the longest chain is only two hubs long.
-
The number of this product that can be recognized by a PC is 20 or less.
Measurement data cannot be input to the PC.|
- If a spare PC or measuring tool is available, replace the PC or measuring tool being used with the spare for the check.
- If a spare of this product is available, replace the product being used with the spare. If data can be input using the replacement device, the problem device may have a damaged DATA button, broken cable, or worn connector.
The measurement data input to the PC is not displayed on separate lines, but
in a single row.| Set the PC’s text input to single-byte input (only when
using languages that support double-byte characters).
When this product is connected to your PC or USB hub, etc., interference
caused by noise occurs. (For example, the zero point of the measuring tool
shifts.)| This product is compliant with EMC directives, but operation when
connected to a PC or USB hub, etc., of any type is not guaranteed. If
interference from noise occurs, try the following noise measures.
- Attach a ferrite core to the cable of this product (near the USB connector).
- Use a USB hub that is self-powered (powered via its own power adapter).
- Ground the frame ground of the PC case. (Ground the PC case.)
NOTICE
This product supports suspension (sleep mode, standby mode) function. However,
operation with a PC or USB hub of any type is not guaranteed.
When operation is impaired by using suspension function, set the PC’s power
management settings so that suspension does not occur during operation.
Specifications
Power supply : +5 V supplied by the PC main body via the USB connector
(“A” plug)
Compatible operating systems : Windows 10 (operation using Windows 10
Mobile not guaranteed)
External dimensions : W 38 mm × D 64 mm × H 21 mm (Main body dimensions)
Cable length : 1 m
Mass : Approx. 55 g
Operating temperature : 0°C to 40°C
Storage temperature : -10°C to 60°C
Compliant standards : USB standard: USB 2.0 certified, communication
speed: 12 Mbps (Full Speed)
CE marking/UKCA marking : EMC Directive/Electromagnetic Compatibility
Regulations: EN61326-1
Immunity test requirement: Clause 6.2 Table 2
Emission limit: Class B RoHSDirective/The Restriction of the Use of
Certain Hazardous Substances in Electrical and Electronic Equipment
Regulations: EN IEC 63000
- A display value on this product may flicker or disappear temporarily due to electromagnetic interference caused by electrostatic discharge. However, this product will return to normal after removing the interference.
- External power supply models may not acquire correct measurement data due to electromagnetic interference acting on the AC or DC power line. If this is the case, check the circumference of the power line and then perform measurement again.
- External power supply models will be turned off automatically if a brownout occurs. However, this product will return to normal after the recovery from the low voltage.
List of Dogmatic Connecting Cables by the Tools to Be Connected
Select the digimatic connecting cable applicable to the measuring tool that is to be connected.
Connector type| Part No.| Form of the connector of measuring tool| Main
applicable measuring tool
---|---|---|---
1m| 2m
A| Equipped with the output switch Waterproof type| 05CZA624| 05CZA625| | ABS
coolant proof caliper CD-PMX/PM/GM
B
| Equipped with the output switch Waterproof type| 05CZA662| 05CZA663| |
Coolant proof micrometer MDC-MJ Quantumlike MDE
C| Equipped with the output switch Straight| 959149| 959150| | ABS digimatic
caliper CD-AX/CX/C
D| Flat (10-pin)| 936937| 965014| | ABS dogmatic indicator ID-H/F QM-Height
QMH
E| Round (6-pin)| 937387| 965013| | Quick micro MDQ Dogmatic Hotlist HTD ABS
Bregmatic SBM-C
F| Flat straight| 905338| 905409| | ABS dogmatic indicator ID-CX/C/S/U
G| Flat straight Waterproof type| 21EAA194| 21EAA190| | ABS dogmatic indicator
ID-N/B
SF| Digimatic S1 straight| 06AGL011| 06AGL021| | ABS dogmatic indicator ID-
F_NX
Mitutoyo Corporation 20-1, Sakado 1-Chome, Takatsu-ku, K213-8533, Japan
CUSTOMRT SUPPORT
1.888.610.7664
www.calcert.com
sales@calcert.com