ACCUREX TX Series Weighing Indicators User Manual
- June 9, 2024
- ACCUREX
Table of Contents
Series Weighing Indicators
User Manual User
Manual
TX series Weighing Indicators
Applications: Electronic platform scale, weighing scale, etc.
Main Specification
1. Model : | TX |
---|---|
2.Accuracy class : | class Ⅲ |
3.Max verified counts: | n=3000 |
4.Operating temperature | -10~40℃ |
5.Power supply: | AC22 AC220V/50HZ Battery: 7.2V/2AH 0V/50HZ |
6.Sensor excitation voltage: | DC 5V |
7.Display: | 6 bits LCD/LED display |
8.Serial communication interface: | RS232 signal transmission distance ≤ 20m |
9.Keyboard: | 5 mechanical keys |
10.DC power supply: | built in maintenance free 7.2v/2ah NiMH battery |
11. Humidity: | ≤ 85% RH |
12.Storage and transportation temperature: | – 25 ~ 55 ℃ |
Keyboard function
2.1. Panel diagram 2.2. Keyboard function
key name | Function |
---|---|
Press this key for turning ON or OFF the indicator | |
keep pressing this button for more than 3 seconds in weighing mode, it will |
enter user’s parameter setting mode ;
When the serial communication mode is manual output, Short press to output
weight data.
| In weighing mode, short press this key to accumulate the weight; in the
self-checking status, press and hold this key, it will enter calibration mode;
In weighing mode, long press this key to enter counting function
| In weighing mode, press this key to tare;
In parameter setting mode, press this key to change parameter value
| In weighing mode, press this key to zero;
In calibration process, press this key to change the value
Indicator operating instruction
3.1. Tare
Put the weighing cargo on the scale platform, when the display digits is
stable, press the button to tare. The display show zero, and net weight
indicating light turns on. take off the cargo, press the button key again. In
this way you can clean the tare weight, then the net weight indicating light
turn off.
3.2. Zero
3.2.1 Manual zero
The display value deviates from zero point. but it is still in the range of
zero setting range (± 2% of zero point at starting-up) and the stable
indicating light is still lightning, press to zero, then the zero indicating
light turn on.
3.2.2 Power on auto zero
When power on auto zero point ranges within ±10% of the capacity, the
indicator will zero automatically.
3.3. Animal weighing function
Set the value into 2 in the P9 of the system setting then the animal weighing
function will be switched on. Set the value into 1,the animal weighing
function will be switched off.
3.4. Accumulating function
In weighing state, press key to accumulate the current weight, when the weight
is over 20e and the weighing data is stable. And the accumulating light is
lightning. When pressing key again, it will return the weighing mode. And the
accumulating light will be off. It will show current accumulating value when
pressing the key under the zero point state. In accumulating sate, press key
to clean the accumulating value .When press the key again, it will return to
weighing mode again also. And the accumulating light will be off.
Note: the scale must zero before next accumulation operation. otherwise it
can not perform the second time accumulating operation.
3.5. Counting function
In weighing mode, long press key and hold on it to enter the counting mode.
The counting light will be lightning. And it will display 【Count 】. When it is
stable, press key to enter sampling numbers setting and it display 【C00000】.
Press key to shift after putting some samples on the scale platform. Press
key to set the sample numbers value. After setting the value of the current
sample number, Press key to perform the sampling and it display 【—— 】. When it
show number, it means you can operate the counting now. This time sampling
date will be saved. Press key to exit the counting mode.
Note:
If the sample weight is smaller than the minimum sampling weight, the sampling
will be failed. It will grab the previous sampling data.
If the sample unit weight is smaller than the minimum sampling unit weight,
the sampling will also be failed. It will also grab the previous sampling
data.
System setting
In weighing mode, long press key to enter user setting mode. Press key to
change the value. press key to enter the next parameter; after all parameter
setting is finished, press key to exit the user setting mode.
(1) Automatically power off
Display | Parameter | Instruction |
---|---|---|
[131 1] | 1 | No automatic shutdown |
function
2| 10min
3| 20min
4| 30min
(2) Power saving mode
Display | Parameter | Instruction |
---|---|---|
[P4 1] | 1 | LCD no backlight; LED no |
power saving function
2| LCD automatically backlight;
LED have power saving function
3| LCD backlight keep lightning LED-
(3) Zero tracking range
Display | Parameter | Instruction |
---|---|---|
【P5 1 】 | 1 | 0.5e/2s |
2 | NO |
(4) Filtering depth
DISPLAY | Parameter | Instruction |
---|---|---|
EP6 li | 1 | low |
2 | middle | |
3 | high |
(5) Stable time
Display | Parameter | Instruction |
---|---|---|
【P7 1 】 | 1 | fast |
2 | middle | |
3 | slow |
(6) Stable range
Dispaly | Parameter | Instruction |
---|---|---|
【P8 1 】 | 1 | lower |
2 | middle | |
3 | high |
(7) Animal weighing function switch on/off
Display | Parameter | Instruction |
---|---|---|
【P9 1 】 | 1 | Turn off the animal function |
2 | Turn on the animal function |
(8) RS232 output protocol selection
Display | Parameter | Instruction |
---|---|---|
【P10 1 】 | Reference communication data |
Error and correction
ERROR | Reason |
---|---|
Errl | The AD value is too small when calibrated or the load cell capacity is |
too large
Err2| The zero point is out of the range when calibrated
Err3| The zero point is out of range upon starting or there is heavy weights
Err5| There is no load weight in calibration mode
Err7| The load cell cable connection is wrong, please check the connection
method again.
Err8| Load cell signal is not stable
Err9| The maximum range voltage is exceeded, please connect or select the
combined sensor correctly
Err10| Storage error
Using of battery
- as long as the power cord is connected to the AC 220V, that is to charge the battery .
- When electronic scales under voltage indicator blinks to indicate low battery, please plug in the power to charge the device .
- When electronic scale show Lob, said serious low battery. At this time will stop weighing function, please immediately shut down equipment and plugged in to recharge the battery.
- Electronic scales on first use, please charge the battery, it is recommended charging time is more than 24 hours to ensure the battery into safe, stable working conditions
- Recommended to charge immediately when the battery undervoltage in normal use, in order to avoid long time work under the undervoltage state,it may damage battery and shorten the service life of the battery; Suggest that more than 12 hours per charge time . If not use for a long time, we should pay attention to every two months to deal with a single battery charge, recommended charging time 20 to 24 hours a time, in order to make up the deficiency of battery caused by the loss of itself , Protect the battery from damage, prolong the service life of the battery.
Maintenance and precautions
- In order to ensure clarity and service life of electronic scales, electronic scales in this paragraph should not be placed directly under sunshine, or location should be relatively smooth.
- Should not be placed where severe vibration and dust, to avoid the use in wet environments.
- Do not use strong solvents (such as: benzene and Nitro-based oils) to clean the cabinet.
- It should not be injected the liquid or other conductive particles into the electronic scale to prevent damage and electric shock .
- Built-in electronic scale battery is a consumable, does not belong to warranty scope.
- Within one year of the date from the sales of this paragraph electronic scales, under normal conditions of use, non-artificial problems belong to the warranty scope, the user will set the product and warranty card (the correct number) back to our service point or suppliers. Production plant implementation of a maintenance of the equipment. This paragraph does not allow arbitrary electronic scales open, otherwise no warranty.