Operating Instructions Liquiline Mobile CML18
Multiparameter mobile device
Liquiline Mobile CML18
Table of contents
Table of contents
1 About this document . . . . . . . . . . . 4
1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.2
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.3 Symbols
on the device . . . . . . . . . . . . . . . . 5
2 Basic safety instructions . . . . . . . 6
2.1 Requirements for the personnel . . . . . . . . 6 2.2 Intended use . . . .
. . . . . . . . . . . . . . . . . . . . 6 2.3 Workplace safety . . . . . . . .
. . . . . . . . . . . . 6 2.4 Operational safety . . . . . . . . . . . . . . .
. . . . 7 2.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . 7
3 Product description . . . . . . . . . . . . . 8
3.1 Product design . . . . . . . . . . . . . . . . . . . . . . . 8
4 Incoming acceptance and product identification . . . . . . . . . 10
4.1 Incoming acceptance . . . . . . . . . . . . . . . . 10 4.2 Product
identification . . . . . . . . . . . . . . . . 10 4.3 Scope of delivery . . .
. . . . . . . . . . . . . . . . . 11
5 Electrical connection . . . . . . . . . . 12
5.1 Connecting the sensor . . . . . . . . . . . . . . . 12 5.2 Charging the
device . . . . . . . . . . . . . . . . . 13 5.3 Ensuring the degree of
protection . . . . . 15
10 Diagnostics and troubleshooting . . . . . . . . . . . . . . . 42
10.1 Diagnostics information via the local display . . . . . . . . . . . . . .
. . . . . . . . . . . . . . 42
11 Maintenance . . . . . . . . . . . . . . . . . . 43
11.1 Maintenance tasks . . . . . . . . . . . . . . . . . . 43 11.2 Measuring
and test equipment . . . . . . . . 43
12 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . 44
12.1 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 12.2
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
13 Accessories . . . . . . . . . . . . . . . . . . . . 44
13.1 M12 USB data + charging cable . . . . . . . 45 13.2 Protective cover . .
. . . . . . . . . . . . . . . . . . . 45
14 Technical data . . . . . . . . . . . . . . . . . 46
14.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 14.2
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 14.3 Power
supply . . . . . . . . . . . . . . . . . . . . . . . 47 14.4 Environment . . .
. . . . . . . . . . . . . . . . . . . . 47 14.5 Mechanical construction . . .
. . . . . . . . . . 48
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6 Operation options . . . . . . . . . . . . . 16
6.1 Overview of operation options . . . . . . . . 16 6.2 Internal operating
menu with keys . . . . . 16 6.3 Operation via SmartBlue app . . . . . . . . .
. 20
7 Commissioning . . . . . . . . . . . . . . . . 25
7.1 Preliminaries . . . . . . . . . . . . . . . . . . . . . . . 25 7.2
Function check . . . . . . . . . . . . . . . . . . . . . 25 7.3 Switching on
the device . . . . . . . . . . . . . . 25 7.4 Setting the display language . .
. . . . . . . . 26 7.5 Configuring the measuring device . . . . . 26 7.6
Advanced settings . . . . . . . . . . . . . . . . . . 26
8 Operation . . . . . . . . . . . . . . . . . . . . . . 32
8.1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 32 8.2
Reading measured values . . . . . . . . . . . . 32
9 Firmware update . . . . . . . . . . . . . . 40
Endress+Hauser
3
About this document
Liquiline Mobile CML18
1 About this document
1.1 Warnings
Structure of information
LDANGER Causes (/consequences) If necessary, Consequences of noncompliance (if
applicable) Corrective action
LWARNING Causes (/consequences) If necessary, Consequences of noncompliance
(if applicable) Corrective action
LCAUTION Causes (/consequences) If necessary, Consequences of noncompliance
(if applicable) Corrective action
NOTICE Cause/situation If necessary, Consequences of noncompliance (if
applicable)
Action/note
Meaning This symbol alerts you to a dangerous situation. Failure to avoid the
dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid the
dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation. Failure to avoid this
situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.
1.2 Symbols
Symbol
Meaning Additional information, tips Permitted or recommended Not permitted or not recommended Reference to device documentation Reference to page Reference to graphic Result of a step
4
Endress+Hauser
Liquiline Mobile CML18
1.3 Symbols on the device
Symbol
Meaning Reference to device documentation
About this document
Endress+Hauser
5
Basic safety instructions
Liquiline Mobile CML18
2 Basic safety instructions
2.1 Requirements for the personnel
· Installation, commissioning, operation and maintenance of the measuring
system may be carried out only by specially trained technical personnel.
· The technical personnel must be authorized by the plant operator to carry
out the specified activities.
· The electrical connection may be performed only by an electrical technician.
· The technical personnel must have read and understood these Operating
Instructions and
must follow the instructions contained therein. · Faults at the measuring
point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided must be carried
out only directly at the manufacturer’s site or by the service organization.
The battery may only be changed directly at the manufacturer’s premises or by
the service organization.
2.2 Intended use
The Liquiline Mobile CML18 is a multiparameter mobile device for the
connection of digital sensors with Memosens technology and optional operation
by smartphone or other mobile devices via Bluetooth.
The device is designed for use in the following industries: · Life science ·
Chemical industry · Water and wastewater · Food and beverages · Power stations
· Other industrial applications
The device contains a lithium ion battery. For this reason, the device may
only be exposed to the operating and storage temperatures indicated.
The device may not be exposed to mechanical shocks of any kind.
The device may not be operated under water.
2.3 Workplace safety
As the user, you are responsible for complying with the following safety
conditions: · Installation guidelines · Local standards and regulations ·
Regulations for explosion protection
6
Endress+Hauser
Liquiline Mobile CML18
Basic safety instructions
2.4 Operational safety
Before commissioning the entire measuring point: 1. Verify that all
connections are correct. 2. Ensure that electrical cables and hose connections
are undamaged. 3. Do not operate damaged products, and protect them against
unintentional operation. 4. Label damaged products as defective.
During operation:
If faults cannot be rectified:
products must be taken out of service and protected against unintentional
operation.
2.5 Product safety
2.5.1 State-of-the-art technology The product is designed to meet state-of-
the-art safety requirements, has been tested, and left the factory in a
condition in which it is safe to operate. The relevant regulations and
international standards have been observed.
Endress+Hauser
7
Product description
3 Product description
3.1 Product design
1
8 7
6
5
1 CML18 1 Protection cap 2 Display screen with automatic screen rotation 3
“Select” button 4 “Next” button 5 Memosens connection 6 Area for wireless
charging 7 Status LED 8 M12 connection
8
Liquiline Mobile CML18
2 3 4
A0040968
Endress+Hauser
Liquiline Mobile CML18
Product description
3.1.1 Measuring parameters
The mobile device is designed for digital Memosens sensors with an inductive
plug-in head and fixed cable sensors with the Memosens protocol and no
external power supply: · pH · ORP · pH/ORP combination sensors · Conductive
conductivity · Inductive conductivity · Dissolved oxygen
(optical/amperometric)
In addition to measuring the main parameters, Memosens sensors can be used to
measure temperature.
The measuring range is adapted to the individual sensor type.
Endress+Hauser
9
Incoming acceptance and product identification
Liquiline Mobile CML18
4 Incoming acceptance and product identification
4.1 Incoming acceptance
1. Verify that the packaging is undamaged. Notify the supplier of any damage
to the packaging. Keep the damaged packaging until the issue has been
resolved.
2. Verify that the contents are undamaged. Notify the supplier of any damage
to the delivery contents. Keep the damaged goods until the issue has been
resolved.
3. Check that the delivery is complete and nothing is missing. Compare the
shipping documents with your order.
4. Pack the product for storage and transportation in such a way that it is
protected against impact and moisture. The original packaging offers the best
protection. Make sure to comply with the permitted ambient conditions.
If you have any questions, please contact your supplier or your local Sales
Center.
4.2 Product identification
4.2.1 Nameplate The nameplate contains the following information: ·
Manufacturer identification · Device designation · Order code · Serial number
· Protection class · Ambient and process conditions · Input and output values
Compare the information on the nameplate with the order.
4.2.2 Product identification
Product page www.endress.com/CML18
Interpreting the order code The order code and serial number of your product
can be found in the following locations: · On the nameplate · In the delivery
papers
Obtaining information on the product 1. Go to www.endress.com.
10
Endress+Hauser
Liquiline Mobile CML18
Incoming acceptance and product identification
2. Call up the site search (magnifying glass). 3. Enter a valid serial
number. 4. Search.
The product structure is displayed in a popup window. 5. Click on the product
image in the popup window.
A new window (Device Viewer) opens. All of the information relating to your
device is displayed in this window as well as the product documentation.
Manufacturer’s address Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24
D-70839 Gerlingen
4.3 Scope of delivery
The scope of delivery comprises: · 1 Liquiline Mobile CML18 · 1 set of
Operating Instructions in German · 1 set of Operating Instructions in English
Inductive charger and power unit are available separately.
If you have any queries:
Please contact your supplier or local sales center.
Endress+Hauser
11
Electrical connection
5 Electrical connection
5.1
5.1.1
Connecting the sensor
Connecting the Memosens sensor directly
Liquiline Mobile CML18
1. 2.
2 Sensor connection 1. Insert the sensor into the Memosens connection. 2.
Click the Memosens connection into place.
5.1.2 Connecting the Memosens sensor with M12 fixed cable connection
A0040973
3. 1.
2.
1. Remove the protection cap. 2. Insert the M12 fixed cable. 3. Screw on the M12 fixed cable.
A0040974
12
A0040972
Endress+Hauser
Liquiline Mobile CML18
5.1.3 Connecting the sensor via the Memosens M12 cable The M12 cable has two
different connectors: · M12 connector for connecting to the device · Memosens
connection for connecting a Memosens sensor
3.
1. 2.
Electrical connection
5. 4.
3
A0040974
Cable connection
A0040972
1. Remove the protection cap. 2. Insert the M12 connector. 3. Screw on the
M12 connector. 4. Insert the sensor into the Memosens connection. 5. Click the
Memosens connection into place.
5.2 Charging the device
Charge the device fully before initial commissioning.
There are two possible ways to load the device: · wirelessly via Qi-certified
charger · via cable with M12 USB data + charging cable
A0041005
Endress+Hauser
13
Electrical connection
Liquiline Mobile CML18
The following applies to both options: · When device is switched on:
· When charging starts, a flash symbol appears on the display and a
confirmation tone sounds.
· If charging stops before the battery is fully charged, another confirmation
tone sounds. · When charging is complete, the “charging complete” melody
sounds. · When device is switched off: · The green LED flashes during
charging. · When charging is complete, the “charging complete” melody sounds
and the LED is
continuously lit green for 10 minutes. · The device then switches off.
5.2.1 Charging via Qi charger
Only use Qi-certified chargers (Qi version 1.2)!
Further information: www.wirelesspowerconsortium.com
-
4 Inductive charging 1. Connect the charger to the power source. 2. Place the device with the charging side on the charger.
Charging begins and the charging state is indicated on the display.
A0044052
14
Endress+Hauser
Liquiline Mobile CML18
Electrical connection
An acoustic signal indicates that charging is complete.
During inductive charging, measurement via the integrated Memosens connection
on the device is not possible. A message to this effect is shown on the
display. Measurement via M12 cable is still possible.
5.2.2 Charging via M12 USB data + charging cable The M12 USB data + charging
cable has two different connectors: · M12 connector for connecting to the
device · USB connector for connecting to a computer or USB charger
3. 4.
A0040974
A0040972
A0044054
1. Remove the protective cap. 2. Attach the M12 connector of the cable to the
device connection. 3. Screw on the M12 connector of the cable. 4. Connect the
USB connector to a USB charger or USB port on a computer.
5.3 Ensuring the degree of protection
Only the mechanical and electrical connections which are described in these
instructions and which are necessary for the required, designated use, may be
carried out on the device delivered.
Exercise care when carrying out the work.
Otherwise, the individual types of protection (Ingress Protection (IP),
electrical safety, EMC interference immunity) agreed for this product can no
longer be guaranteed due, forexample to covers being left off or cable (ends)
that are loose or insufficiently secured.
Endress+Hauser
15
Operation options
6 Operation options
6.1 Overview of operation options
There are two options for operating and configuring the device: · Internal
operating menu with keys · SmartBlue app via Bluetooth® LE wireless technology
20
6.2
6.2.1
Internal operating menu with keys
Display and operating elements
Liquiline Mobile CML18
1
3
2
5 Overview of display and operating elements
1 Display 2 “Select” button 3 “Next” button
A0040996
Button functions
Button
Device switched off Switching on
On measuring screen
In the menu
Scroll through measuring screens Scroll down
Switching on
Save current measured values (Grab Sample)
Confirm/select
(long hold)
Open the menu
Switch to previous menu level/measuring screen
Forced hardware reset Forced hardware reset
(pressed for longer than 7
seconds)
Forced hardware reset
16
Endress+Hauser
Liquiline Mobile CML18
6.2.2 Structure and function of the operating menu
Power-off Power-off
Application Data logger
Data logger Log interval Cond. unit Res. unit Erase data
Erase grab values
Erase continuous logs
Data logger plot Units
Diagnostics Sensor info Calibration info Diagnostics list Data logger entries Display test Device info
Manufacturer Software version Serial number Description Extended order code
System/Language Display language Bluetooth Display brightness Signal sounds
M12 CSV
Endress+Hauser
Operation options
Abort Erase Abort Erase
17
Operation options
System/Language Power management
Regulatory information
Support links Support links
Guidance 1 point calib. (ORP) 2 point calibration (pH and ISFET) Cell constant
(inductive/ conductive conductivity) Installation factor (conductive
conductivity) Air 100% rh (oxygen) Air variable (oxygen) 1 point calib.
(oxygen)
Power save w. charger Power save w/o charger Power-off w. charger Power-off w/o charger
Liquiline Mobile CML18
18
Endress+Hauser
Liquiline Mobile CML18
Display structure
1
Operation options
2 3 4
6 5
6 Schematic representation of the display structure
1 Menu path/title of measuring screen 2 Bluetooth status 3 Battery level,
charging information 4 NAMUR indicator 5 Measuring screen 6 Date and time
(displayed in main menu and if no sensor is connected)
A0044047
Status according to NAMUR NE107 categories:
NAMUR indicator OK F
M
C
S
Status
The device and sensor are working reliably.
Failure of device or sensor. F status signal as per NAMUR NE107
Device or sensor requires maintenance. M status signal as per NAMUR NE107
Device or sensor undergoing function check. C status signal as per NAMUR NE107
Device or sensor being operated out of specification. S status as per NAMUR
NE107
Structure of the measurement window The measurement window has 3 measuring screens, which the user can scroll through:
Measuring screen (1 of 3) Main value
Endress+Hauser
Measuring screen (2 of 3) Main and secondary measured value
Measuring screen (3 of 3)
All measured values of the sensor input
19
Operation options
Liquiline Mobile CML18
6.2.3 LED status indicator The status LED is used for the quick visualization of the sensor status.
LED behavior Solid green Solid red Flashes red
Status Sensor working correctly No sensor connected Sensor error
6.3 Operation via SmartBlue app
The SmartBlue App is available for download from the Google Play Store for
Android devices and from the Apple App Store for iOS devices.
Download the SmartBlue App.
Use the QR codes to download the app.
7 Download links
A0033202
System requirements · iOS devices: iPhone 4S or higher from iOS9.0; iPad2 or
higher from iOS9.0; iPod Touch 5th
generation or higher from iOS9.0 · Devices with Android: from Android 4.4
KitKat and Bluetooth® 4.0 · Internet access
Open the SmartBlue App.
20
Endress+Hauser
Liquiline Mobile CML18
8 SmartBlue App icon Bluetooth must be enabled on both devices. Enable
Bluetooth 26
Operation options
A0029747
9 SmartBlue App Livelist
A0044142
The Livelist displays all of the devices that are within range.
Tap the device to select it.
To be able to use the device with the SmartBlue App, the Bluetooth connection
must be confirmed by entering a user name and password.
Endress+Hauser
21
Operation options
Liquiline Mobile CML18
1. User name >> admin 2. Initial password >> device serial number Change the
user name and password after logging in for the first time.
The current measured values are displayed in the Home view. The device
information (device tag, serial number, firmware version, order code) is also
displayed.
22
Endress+Hauser
Liquiline Mobile CML18
4 3
Operation options
1 2
10 Home view of SmartBlue App with current measured values 1 CML18 system and
device information 2 Shortcut to diagnostic list 3 Overview of measured values
of connected sensor 4 General information and sampling option
Operation is via 4 main menus:
Endress+Hauser
A0048102
23
Operation options
Liquiline Mobile CML18
1
CML18_RA010905MHO
2
3
4
11 Main menus of the SmartBlue App
1 Guidance 2 Diagnostics 3 Application 4 System
A0048103
Menu Guidance Diagnostics Application System
Function
Contains functions involving a self-contained sequence of activities, e.g. for
calibration (= “Wizard”, guided operation).
Contains information on operation, diagnostics and troubleshooting, as well as
configuration of the diagnostic behavior.
Sensor data for specific optimization and for detailed process adjustment.
Adapts the measuring point to the application.
These menus contain parameters for configuring the overall system, e.g. Time
and date options.
24
Endress+Hauser
Liquiline Mobile CML18
Commissioning
7 Commissioning
7.1 Preliminaries
Charge the device fully before initial commissioning. 12 Connect the sensor.
12
7.2 Function check LWARNING
Connection errors The safety of people and of the measuring point is at risk!
Put the device into operation only if you can answer yes to all the following
questions.
Device condition and specifications
Are the device and all the cables free from damage on the outside? Are the
mounted cables strain relieved? Are the cables routed without loops and cross-
overs?
7.3 Switching on the device
12 Switching on the device
A0040976
Press or .
The device starts up.
A connected sensor is recognized automatically.
The time required before a measured value is displayed depends on the sensor
type and measuring principle and can vary.
Endress+Hauser
25
Commissioning
Liquiline Mobile CML18
7.4 Setting the display language
1. Navigate to: Display language Main menu >> System/Language >> Display
language
2. Press to scroll through the predefined values.
Description of setting Change the language of the operating menu.
Configuration options
· Deutsch · English
7.5 Configuring the measuring device
7.5.1 Configuring the Bluetooth connection 1. Navigate to: Bluetooth Main menu
System/Language >> Bluetooth 2. Press to scroll through the predefined values.
Description of setting Switch Bluetooth connection on/off
Configuration options
· Enabled · Disabled
If the Bluetooth connection is disabled, operation via the SmartBlue App is not possible.
7.5.2 Setting date and time Preparatory steps
1. Enable Bluetooth. 26 2. Link the device to a mobile terminal via the
SmartBlue app. 20
1. Select the device in the SmartBlue App. 2. Select System. 3. Select
date/time. 4. Select Take over from mobile device.
or: 5. Manually configure the date and time.
7.6 Advanced settings
7.6.1 Displaying device information 1. Navigate to: Device info Main menu >>
Diagnostics >> Device info
26
Endress+Hauser
Liquiline Mobile CML18
Commissioning
2. Press to scroll through the Device info.
The following information about the device is shown on the display: · Manufacturer identification · Software version · Serial number · Designation · Extended order code
7.6.2 Adjusting the energy settings
A maximum battery life of 48 h can be achieved via the energy settings.
For measurements with oxygen sensors, the device remains permanently switched
on, regardless of the energy settings selected.
1. Navigate to: Power management Main menu >> System/Language >> Power
management
2. Press to scroll through the predefined values.
The following energy settings are available: · Power save w. charger · Power
save w/o charger · Power-off w. charger · Power-off w/o charger
The power save mode is activated after the set time if there is no user
interaction.
In the power save mode, the display is switched off and the device remains on
standby.
There are 2 power save settings:
Power save w. charger
Description of setting
Set the time until the power save mode is activated if the device is connected
to the mains.
Configuration options
· 1 min · 5 min · 15 min · 30 min · 1h · 2h · Never
Endress+Hauser
27
Commissioning
Liquiline Mobile CML18
Power save w/o charger
Description of setting
Configuration options
Set the time until the power save mode is activated if the device is running on the battery.
· 1 min · 5 min · 15 min · 30 min · 1h
The device is automatically switched off after the selected time. The device is not switched off automatically if the Bluetooth connection is enabled. There are 2 power-off settings:
Power-off w. charger
Functional description
Configuration options
Set the time until the device switches off automatically if it is connected to the mains.
· 1 min · 5 min · 15 min · 30 min · 1h · 2h · Never
Power-off w/o charger
Functional description
Set the time until the device switches off automatically if it is running on
the battery.
Configuration options
· 1 min · 5 min · 15 min · 30 min · 1h · 2h · Never
7.6.3 Signal sounds 1. Navigate to: Signal sounds Main menu >> System/Language
Signal sounds 2. Press to scroll through the predefined values. Other settings are possible via the SmartBlue App.
28
Endress+Hauser
Liquiline Mobile CML18
Commissioning
Description of setting Switch signal sounds on/off
Configuration options
· Enabled · Disabled
Additional changes to the signal sounds can be made via the SmartBlue App.
7.6.4 Configuring M12 CSV Measured values can be output to other devices via
the device’s M12 connection. The M12 USB data + charging cable 45 is used for
this purpose. The transmitted data can, for example, be further processed in
real-time in an external computer program. A data rate of 9600 bit/s in 8N1
configuration must be used as the connection parameter at the receiving
system.
1. Navigate to: M12 CSV Main menu >> System/Language >> M12 CSV
2. Press to scroll through the predefined values.
Description of setting Switch on/off M12 CSV
Configuration options
· On · Off
When the M12 CSV option is enabled, no sensor can be operated via cable.
Operation via the Memosens connection on the device is still possible. A
message to this effect is shown on the display.
7.6.5 Adjusting the display brightness 1. Navigate to: Display brightness Main
menu >> System/Language >> Display brightness 2. Press to adjust the display
brightness.
Description of setting Set the display brightness
Configuration options
· Low · Medium · High · Maximum
7.6.6 Hardware reset in an emergency
This type of restart should only be performed in an emergency if the device
does not respond to any other input.
Press and hold and simultaneously for at least 7 seconds.
The device restarts.
Endress+Hauser
29
Commissioning
Liquiline Mobile CML18
7.6.7 Displaying regulatory information and approvals 1. Navigate to:
Regulatory information Main menu >> System/Language >> Regulatory information
2. Press to display regulatory information and approvals.
7.6.8 Data logger
Defining the log interval The log interval can only be changed if the data
logger is deactivated.
1. Navigate to: Log interval Main menu >> Application >> Data logger >> Log
interval
2. Press to scroll through the predefined values.
Description of setting
Set the time until the next measured value is saved automatically.
Configuration options
· 1s · 2s · 10 s · 20 s · 30 s · 1 min · 5 min · 30 min · 1h
If the device is woken up to record a log value, any existing switch-on /
settling times of the connected sensor are not taken into account.
When measuring with oxygen sensors, such as the Oxymax COS51D or COS22D, the
device with activated data logger remains permanently switched on, regardless
of the energy settings selected.
Adjust the energy settings: 27
Enabling/disabling the data logger
The data logger must be deactivated in the following cases: · if changes are
made to the measurement settings · if the measured values are exported · if
the sensor is replaced
1. Navigate to: Data logger Main menu >> Application >> Data logger >> Data
logger
2. Press to scroll through the predefined values.
30
Endress+Hauser
Liquiline Mobile CML18
Commissioning
Description of setting Enable/disable automatic data logger
Configuration options
· On · Off
3. Exit the menu. 4. Once activated, the data logger automatically starts
recording the measured values.
If the data logger is activated, the display flashes alternately between the
“Logging…” message and the current menu path/measuring screen title.
4. Press to change the active measuring window.
Configuring the data logger for ultrapure water Prior to activating the data
logger, the measured value units can be adjusted for conductivity measurement
with the data logger in ultrapure water. An adjustment is necessary to
eliminate rounding errors in the smallest measured values. The units for
conductivity and resistance can be permanently configured.
1. Navigate to: Cond. unit Main menu >> Application >> Data logger >> Data
logger >> Cond. unit
2. Press to scroll through the predefined values.
1. Navigate to: Res. unit Main menu >> Application >> Data logger >> Data
logger >> Res. unit
2. Press to scroll through the predefined values.
7.6.9 Switching units Only the units used by the sensor are displayed.
1. Navigate to: Units Main menu >> Application >> Units
2. Press to scroll through the predefined values.
Endress+Hauser
31
Operation
Liquiline Mobile CML18
8 Operation
8.1 Calibration
To calibrate sensors, the calibration settings are first configured via the
SmartBlue app. The calibration can then be started from the device.
Configure the calibration settings in the SmartBlue app: 1. Enable Bluetooth. 26 2. Link the device to a mobile terminal via the SmartBlue app. 20 3. Select the device in the SmartBlue App. 4. Navigate to: Calibration settings Application >> Sensor >> Advanced settings >> Calibration settings 5. Configure the calibration settings. e.g. Manufacturer and calibration buffer.
Perform calibration on the device: 1. Navigate to: Guidance Select the desired calibration. 2. Press to navigate through the calibration.
The following calibrations can be performed:
Type of calibration 1-point calibration 2-point calibration
Measuring parameters ORP pH or ISFET
Cell constant calibration Installation factor calibration
Inductive/conductive conductivity Conductive conductivity
Air 100%rH calibration Air variable calibration
Oxygen Oxygen
1-point calibration
Oxygen
Navigate to: >> 1 point calib. >> 2 point calibration >> Cell constant >> Installation factor >> Air 100% rh >> Air variable >> 1 point calib.
8.2 Reading measured values
Measuring screens are shown on the display when a sensor is connected. For
each sensor, there are 3 measuring screens with different measured variables
19.
To scroll through the measuring screens:
Press .
After the last measuring screen, the display returns to the first measuring
screen.
32
Endress+Hauser
Liquiline Mobile CML18
Operation
8.2.1 Saving the sample (Grab Sample) Samples can be assigned IDs and a user-
definable text. By assigning an ID, samples can be assigned more easily to a
measuring point, for example.
IDs and the associated texts can be changed via the SmartBlue App. 33
1. In the measurement window, press . A new window appears.
2. Give the sample an ID. Press to scroll through the available IDs.
3. Press to save the sample with the selected ID. Or: Press and hold to
discard the sample.
8.2.2 Changing the sample IDs The 10 pre-set IDs for samples can be changed
via the SmartBlue App.
Preparatory steps 1. Enable Bluetooth. 26 2. Link the device to a mobile
terminal via the SmartBlue app. 20
Transferring the data 1. Select the device in the SmartBlue App. 2. Select
Grab sample. 3. Select ID text. Click in the line of text to assign an
individual text for the selected ID.
Depending on the input language selected, there are up to 32 characters
available for the assignment of the individual ID.
8.2.3 Saving measured values automatically (Data logger) Configure the data
logger 30.
8.2.4 Displaying saved measured values
Navigate to: Log entries
Main menu >> Diagnostics >> Log entries
This menu displays the number of saved entries for the different log
procedures.
8.2.5 Exporting measured values
Export to mobile terminal device Saved data can be transferred from the
internal device memory to mobile terminals.
Endress+Hauser
33
Operation
Liquiline Mobile CML18
Preparatory steps 1. Install the SmartBlue app on a mobile terminal. 20 2.
Enable Bluetooth. 26 3. Link the device to a mobile terminal via the SmartBlue
app. 20
Transferring the data 1. Select the device in the SmartBlue app. 2. Select in
the SmartBlue app.
3. Select Guidance.
A0042257
34
Endress+Hauser
Liquiline Mobile CML18
4. 4. Select Data transfer.
5. 5. Continue with Next.
Endress+Hauser
Operation
A0042258
A0042261
35
Operation
Liquiline Mobile CML18
-
A0042260
6. Select Data source. Select Grab sample logger for saved samples. Select Cont. data logger for data records of data logger.
7. Press Ok to confirm. Press to discard changes and close the drop-down menu.
8. Select File name. Click the text line to enter an individual name for the generated data package.
9. Press Ok to confirm. Press to discard changes and close the drop-down menu.
10. Press Next to continue. Data transfer starts. A progress bar indicates the progress percentage.
36
Endress+Hauser
Liquiline Mobile CML18
Operation
- 11. When the transfer is complete, press Next to continue. The result of the data transfer is displayed.
A0042263
12. 13.
Endress+Hauser
A0042265
37
Operation
Liquiline Mobile CML18
12. Use Share ZIP file… to send the exported data records or save them
locally. 13. Complete export by pressing Finish.
Exporting to a computer Preparatory steps:
1. Download the CML18 readout tool to the target computer and save it. The
current readout tool can be found in the Download area on the product page
under www.endress.com/CML18.
2. Deactivate the data logger. 30
1. Remove all the sensors from the device. 2. Connect the device to a
computer via the M12 USB data + charging cable. 15 3. Run the CML18 readout
tool on the computer. 4. Follow the instructions given by the tool.
The measured values are exported to an .xlsx file for table programs such as
Microsoft Excel.
The export files of the grab sample and data logger measured values have a
different display format.
Elements of the export file
Data logger export file
Grab sample export file
Elements in the general information section of the export file: · Filename ·
File content · Format version · Device type · Device tag · Device serial
number · Device firmware version · Sensor serial number · PV name · PV unit ·
SV name · SV unit · TV name · TV unit
Elements of the individual measured value entries: · Sample number · Status ·
PV value · SV value · TV value · Timestamp
Elements in the general information section of the export file: · Filename ·
File content · Format version · Device type · Device tag · Device serial
number · Device firmware version
Elements of the individual measured value entries: · Sample number · Status ·
PV name · PV value · PV unit · SV name · SV value · SV unit · TV name · TV
value · TV unit · Timestamp · Sensor serial number · Sample ID
38
Endress+Hauser
Liquiline Mobile CML18
Operation
Description of the individual elements of the export files
Filename
Name of the export file, based on the date/time of the first logged entry. If the sensor, sensor type or unit settings are changed, a new export file is created.
File content
Contents of the export file: · Data logger always “Continous log” · Sample always “Grab sample logs”
Format version
Version of the format structure of the generated export file. The number increases if the structure changes with a new firmware.
Device type
Type of device used for logging. “Liquiline Mobile” in the case of CML18.
Device tag
Tag of the device used for logging.
Device serial number
Serial number of the device used for logging.
Device firmware version
Firmware version of the device used for logging.
Sample number
Unique entry number. This value is increased for each logged entry. It is reset if the entries are deleted.
Status
NAMUR device status when the entry is logged.
PV name
Name of the primary value.
PV value
Numerical display of the primary value of the logged entry.
PV unit
Unit of the primary value.
SV name
Name of the secondary value.
SV value
Numerical display of the secondary value of the logged entry.
SV unit
Unit of the secondary value.
TV name
Name of the tertiary value.
TV value
Numerical display of the tertiary value of the logged entry.
TV unit
Unit of the tertiary value.
Timestamp
Time and date stamp of the logged unit.
Sensor serial number
Serial number of the sensor used for logging.
Sample ID
User-defined text to identify the entry.
8.2.6 Deleting measured values
Navigate to: Erase data
Main menu >> Application >> Data logger >> Erase data
The data are divided into 2 categories: · Erase continuous logs
Selects all data logger entries for deletion. · Erase grab values
Selects all grab values (samples) for deletion.
Endress+Hauser
39
Firmware update
Liquiline Mobile CML18
NOTICE
Deletion of data! Once data are deleted, they cannot be restored. The deletion
of data must be confirmed.
Save data before deletion.
1. Press 2. Press 3. Press 4. Press
to navigate to the desired category. to select the category to be deleted. to select Erase or Abort. to select Erase or Abort.
8.2.7 Switching off the device 1. Navigate to: Power-off Main menu >> Power- off 2. Press to switch off the device.
9 Firmware update
The firmware of the device can be updated to the latest version via the
Smartblue App.
All saved data logger entries must be exported before each firmware update.
A firmware update can take up to one hour.
There must be sufficient battery charge; if necessary, connect the device to
the mains. 13
The device is prevented from switching off automatically if is connected to
the SmartBlue app.
NOTICE Damage to firmware! Risk of incomplete update and limited functionality
of the device.
During a firmware update, do not switch off the device manually or disconnect
it from the
mobile terminal.
A video tutorial on how to update the firmware is available on
Endress+Hauser’s YouTube channel via the following link or QR code:
Firmwareupdate CML18
40
Endress+Hauser
Liquiline Mobile CML18
Firmware update
13 Scan the QR code to access the instruction video
A0045926
Preparatory steps
1. Download the firmware update package and save to the terminal. The current
firmware update package can be found in the Downloads area on the product page
at www.endress.com/CML18.
2. Enable Bluetooth. 26
3. Link the device to a mobile terminal via the SmartBlue app. 20
Starting a firmware update 1. Select the device in the SmartBlue App. 2. Select in the SmartBlue app. 3. Select System. 4. Select Firmware update. 5. Search for the available firmware update package on the terminal device and select it. If the update is not displayed, the firmware update package must be opened once using the SmartBlue App. 6. Start the update. 7. After updating the firmware successfully, update the time and date. 26
Following a firmware update, Bluetooth functionalities are restarted in the background. This process may take some time. All other functions of the device can be used immediately.
Endress+Hauser
41
Diagnostics and troubleshooting
Liquiline Mobile CML18
10 Diagnostics and troubleshooting
10.1 Diagnostics information via the local display
10.1.1 Accessing sensor information 1. Navigate to: Sensor info Main menu >>
Diagnostics >> Sensor info 2. Press to access the sensor information.
10.1.2 Accessing calibration information 1. Navigate to: Calibration info Main
menu >> Diagnostics >> Calibration info 2. Press to access the calibration
information.
10.1.3 Opening the diagnostics list 1. Navigate to: Diagnostics list Main menu
Diagnostics >> Diagnostics list 2. Press to open the diagnostics list.
10.1.4 Display testing 1. Navigate to: Display test Main menu >> Diagnostics Display test 2. Press to invoke the screen test. 3. Press to scroll through the test windows and check the display for damage.
42
Endress+Hauser
Liquiline Mobile CML18
Maintenance
11 Maintenance
11.1 Maintenance tasks
11.1.1 Cleaning
Only clean with a damp cloth and commercially available cleaning agents.
The device is resistant to: · Ethanol (for a short time) · Soap-based
household cleaning agents · Dishwashing detergent
NOTICE Cleaning agents not permitted Damage to the housing surface or housing
seal
Do not use concentrated mineral acids or alkaline solutions for cleaning. Do
not use organic cleaners such as acetone, benzyl alcohol, methanol, methylene
chloride,
xylene or concentrated glycerol cleaner for cleaning.
Do not use high-pressure steam for cleaning.
11.2 Measuring and test equipment
Calibrated and adjusted sensors with Memosens technology save their
calibration data directly in the sensor. The sensors can be used as testing
equipment thanks to this functionality. The device can be used to display the
measured values of such test equipment. Each connected sensor uses its own
calibration data. Using the SmartBlue App, a sensor can be calibrated,
recalibrated and adjusted in a suitable testing medium directly at the device.
Endress+Hauser
43
Repair
Liquiline Mobile CML18
12 Repair
12.1 Return
The product must be returned if repairs or a factory calibration are required,
or if the wrong product was ordered or delivered. As an ISO-certified company
and also due to legal regulations, Endress+Hauser is obliged to follow certain
procedures when handling any returned products that have been in contact with
medium.
To ensure the swift, safe and professional return of the device:
Refer to the website www.endress.com/support/return-material for information
on the
procedure and conditions for returning devices.
12.2 Disposal
The device contains electronic components. The product must be disposed of as
electronic waste.
Observe the local regulations.
If required by the Directive 2012/19/EU on waste electrical and electronic
equipment (WEEE), the product is marked with the depicted symbol in order to
minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of
products bearing this marking as unsorted municipal waste. Instead, return
them to Endress+Hauser for disposal under the applicable conditions.
The battery cannot be replaced or removed by the end customer! It may only be
disposed of by trained staff.
13 Accessories
The latest list of accessories and all compatible Memosens sensors is provided
on the product page: www.endress.com/CML18
44
Endress+Hauser
Liquiline Mobile CML18
13.1 M12 USB data + charging cable
Order code: 71496600 · Charging via cable · Data backup · Live data transfer
13.2 Protective cover
Order code: 71530939 · Comprehensive protection · Extremely robust · Tabs and
eyelets provide a range of securing options
Accessories
A0047709 A0047710
Endress+Hauser
45
Technical data
Liquiline Mobile CML18
14 Technical data
14.1 Input
14.1.1 Input power
Wireless charging M12 connection
5 W 5 V; 0.6 A
14.1.2 Measured variables
· pH · ORP · pH/ORP · Oxygen · Conductivity · Temperature
14.1.3 Measuring range
Documentation of the connected sensor
14.1.4 Type of input
Memosens connection for sensors with Memosens technology
M12 connection for digital measuring cable CYK10, CYK20 for sensors with
Memosens technology
A complete list of supported sensors is provided on the device product page:
www.endress.com/CML18 -> Documents/Manuals/Software -> Certificates …
Supported sensors from the lab portfolio include: · CPL51E, CPL53E, CPL57E,
CPL59E · CLL47E · COL37E
Supported sensors from the process portfolio include: · CPS11D, CPS12D,
CPS16D, CPS31D, CPS41D, CPS42D, CPS47D, CPS71D, CPS72D, CPS76D,
CPS77D, CPS91D, CPS92D, CPS96D, CPS97D · CPS171D, CPS341D, CPS441D, CPS471D,
CPS491D · CPF81D, CPF82D · CLS15D, CLS16D, CLS21D, CLS82D · CLS50D, CLS54D ·
COS21D, COS22D, COS51D, COS81D
14.2 Output
14.2.1 Output signal Memosens M12 (maximum 80 mA)
46
Endress+Hauser
Liquiline Mobile CML18
Technical data
14.3 Power supply
14.3.1 Supply voltage Inductive charging: use Qi-certified devices (min. 5 W
output power) The power supply unit must supply an output current of at least
1500 mA.
14.3.2 Battery rated capacity 1 000 mAh (min. 950 mAh)
14.3.3 Battery life Max. 48 h (with adapted energy settings)
14.3.4 Overvoltage protection IEC 61 000-4-4 with 0.6 kV IEC 61 000-4-5 with
2.0 kV
14.3.5 Sensor connection Sensors with Memosens technology
14.3.6 Cable specification Digital measuring cable CYK10-Axx2+x Digital
measuring cable CYK20-AAxxC1 M12 USB data + charging cable
14.4 Environment
14.4.1 Ambient temperature range Charging: 0 to +45 °C (32 to 113 °F)
Operation: 10 to +60 °C (14 to 140 °F)
The maximum ambient temperature depends on the process temperature and the
installation position.
14.4.2 Storage temperature 20 to +45 °C (4 to 113 °F)
Elevated storage temperatures reduce the battery capacity.
14.4.3 Relative humidity 0 to 95 %
14.4.4 Degree of protection IP66
Endress+Hauser
47
Technical data
14.4.5 Electrical safety EN 61010-1 14.4.6 Pollution degree
Complete device: Internal:
14.5 Mechanical construction
14.5.1 Dimensions
Pollution level 4 Pollution level 2
Liquiline Mobile CML18
86 (3.38)
48 (1.89)
14 Dimensions: mm (in) 48
54 (2.13)
A0044044
Endress+Hauser
Liquiline Mobile CML18
Technical data
14.5.2 Materials
Components Housing Display window, light guide Buttons, cap M12 connection
Material PBT PMMA TPE CuZn, nickel-plated
14.5.3 Materials not in contact with the medium Information according to REACH
Regulation (EC) 1907/2006 Art. 33/1: The device battery contains the SVHC 1.3
propane sulton ; ethylene glycol dimethyl ether (CAS number 1) 110-71-4) with
more than 0.1% (w/w). The product does not present a hazard if it is used as
designated.
14.5.4 Impact loads The product is designed for mechanical impact loads of 1 J
(IK06) as per the requirements of EN 61010-1.
14.5.5 Weight
Liquiline Mobile CML18
155 g (5.5 oz)
- CAS = Chemical Abstracts Service, international identification standard for chemical substances
Endress+Hauser
49
Index
Liquiline Mobile CML18
Index
A
Accessories . . . . . . . . . . . . . . . . . . . . . 44 Ambient temperature .
. . . . . . . . . . . . . . 47
B
Battery life . . . . . . . . . . . . . . . . . . . . . 47 Bluetooth connection
. . . . . . . . . . . . . . . 26
C
Cable specification . . . . . . . . . . . . . . . . . 47 Charging the device .
. . . . . . . . . . . . . . . 13 Cleaning . . . . . . . . . . . . . . . . . .
. . . . . 43 Commissioning . . . . . . . . . . . . . . . . . . . 25 Connection
Fixed cable sensor . . . . . . . . . . . . . . 12 Measuring cable . . . . . .
. . . . . . . . . 13 Sensor . . . . . . . . . . . . . . . . . . . . . 12
Sensors . . . . . . . . . . . . . . . . . . . . 47 Supply voltage . . . . . .
. . . . . . . . . . 47
D
Data logger . . . . . . . . . . . . . . . . . . . . . 30 Enable/disable . . .
. . . . . . . . . . . . . 30 Log interval . . . . . . . . . . . . . . . . . .
30 Ultrapure water . . . . . . . . . . . . . . . 31
Date and time Date . . . . . . . . . . . . . . . . . . . . . . 26 Time . . . .
. . . . . . . . . . . . . . . . . . 26
Degree of protection . . . . . . . . . . . . . 15, 47 Device information
Device name . . . . . . . . . . . . . . . . . 26 Extended order code . . . . .
. . . . . . . . 26 Manufacturer identification . . . . . . . . 26 Serial
number . . . . . . . . . . . . . . . . . 26 Software version . . . . . . . . .
. . . . . . 26 Dimensions . . . . . . . . . . . . . . . . . . . . . 48 Display
language . . . . . . . . . . . . . . . . . . 26
E
Electrical connection . . . . . . . . . . . . . . . . 12 Electrical safety . .
. . . . . . . . . . . . . . . . . 48
F
Firmware update . . . . . . . . . . . . . . . . . . 40
50
H
Hardware reset . . . . . . . . . . . . . . . . . . . 29
I
Impact loads . . . . . . . . . . . . . . . . . . . . 49 Incoming acceptance .
. . . . . . . . . . . . . . 10 Input
Measured variables . . . . . . . . . . . . . 46 Intended use . . . . . . . . .
. . . . . . . . . . . . 6
L
Language . . . . . . . . . . . . . . . . . . . . . . 26
M
Manufacturer’s address . . . . . . . . . . . . . . 11 Materials . . . . . . .
. . . . . . . . . . . . . . . 49 Measured variables . . . . . . . . . . . . .
. . . 46 Measuring parameters . . . . . . . . . . . . . . . 9 Measuring range
. . . . . . . . . . . . . . . . . . 46
N
Nameplate . . . . . . . . . . . . . . . . . . . . . 10
O
Operation . . . . . . . . . . . . . . . . . . . . . . 32 Calibration . . . . .
. . . . . . . . . . . . . 32 Device operation . . . . . . . . . . . . . . . 16
Grab Sample . . . . . . . . . . . . . . . . . 33 LED status indicator . . . .
. . . . . . . . . 20 Operating menu . . . . . . . . . . . . . . . 17 Reading
measured values . . . . . . . . . . 32 Saving the sample . . . . . . . . . . .
. . . 33 SmartBlue app . . . . . . . . . . . . . . . . 20
Operation options . . . . . . . . . . . . . . . . . 16 Order code . . . . . .
. . . . . . . . . . . . . . . . 10 Output signal . . . . . . . . . . . . . . .
. . . . . 46 Overvoltage protection . . . . . . . . . . . . . . 47
P
Pollution degree . . . . . . . . . . . . . . . . . . 48 Power supply . . . . .
. . . . . . . . . . . . . . . 47
Overvoltage protection . . . . . . . . . . . 47 Sensor connection . . . . . .
. . . . . . . . 47 Supply voltage . . . . . . . . . . . . . . . . 47 Product
description . . . . . . . . . . . . . . . . . 8 Product design . . . . . . . .
. . . . . . . . . . . . 8
Endress+Hauser
Liquiline Mobile CML18
Index
Product identification . . . . . . . . . . . . . . . 10 Product page . . . . .
. . . . . . . . . . . . . . . 10 Product safety . . . . . . . . . . . . . . .
. . . . . 7
R
Relative humidity . . . . . . . . . . . . . . . . . . 47 Requirements for the
personnel . . . . . . . . . . 6
S
Safety Operational safety . . . . . . . . . . . . . . . 7 Product . . . . . .
. . . . . . . . . . . . . . . 7 Workplace safety . . . . . . . . . . . . . . .
6
Safety instructions . . . . . . . . . . . . . . . . . . 6 Saving the measured
value
Data logger . . . . . . . . . . . . . . . . . . 33 Scope of delivery . . . . .
. . . . . . . . . . . . . 11 Sensor
Connection . . . . . . . . . . . . . . . . . . 47 Settings . . . . . . . . . .
. . . . . . . . . . . . . 26
Audio . . . . . . . . . . . . . . . . . . . . . 28 Data logger . . . . . . . .
. . . . . . . . . . 30 Display brightness . . . . . . . . . . . . . . 29
Energy settings . . . . . . . . . . . . . . . . 27 Signal sounds . . . . . . .
. . . . . . . . . . 28 Switching units . . . . . . . . . . . . . . . . 31
State-of-the-art technology . . . . . . . . . . . . 7 Storage temperature . .
. . . . . . . . . . . . . . 47 Supply voltage . . . . . . . . . . . . . . . .
. . . 47 Switching off . . . . . . . . . . . . . . . . . . . . 40 Switching on
. . . . . . . . . . . . . . . . . . . . 25 Symbols . . . . . . . . . . . . . .
. . . . . . . . 4, 5
T
Technical data . . . . . . . . . . . . . . . . . . . 46 Environment . . . . .
. . . . . . . . . . . . 47 Input . . . . . . . . . . . . . . . . . . . . . .
46 Mechanical construction . . . . . . . . . . 48 Output . . . . . . . . . . .
. . . . . . . . . . 46
Technical personnel . . . . . . . . . . . . . . . . . 6 Types of input . . . .
. . . . . . . . . . . . . . . . 46
U
Update . . . . . . . . . . . . . . . . . . . . . . . . 40 Use
Intended . . . . . . . . . . . . . . . . . . . . . 6
W
Warnings . . . . . . . . . . . . . . . . . . . . . . . 4 Weight . . . . . . .
. . . . . . . . . . . . . . . . . 49 Workplace safety . . . . . . . . . . . .
. . . . . . 6
Endress+Hauser
51
71558219
71558219
www.addresses.endress.com
References
- Contact | Endress+Hauser
- Handheld for pH/ORP, conductivity, oxygen and temperature | Endress+Hauser
- Equipment returns | Endress+Hauser
- Home | Wireless Power Consortium
- Flow, level, liquid analysis, optical analysis, pressure, temperature measurement, software and system products | Endress+Hauser
- Handheld for pH/ORP, conductivity, oxygen and temperature | Endress+Hauser
- wirelesspowerconsortium.com/qi/
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>