ZEWA 11110 V2 Digital Infrared Forehead Thermometer Instruction Manual

June 9, 2024
Zewa

11110 V2 Digital Infrared Forehead Thermometer
Instruction Manual
Owner’s Manual

Infrared Forehead Thermometer
Model 11110 (V2)ZEWA 11110 V2 Digital Infrared Forehead
Thermometer

Introduction

Please read all instructions carefully and thoroughly before using this product. Indication for Use: The infrared forehead thermometer is a non- contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead at a distance of 1-5cm by people of all ages. This thermometer is reusable for home use and clinical use and can be reused by many people.
The 11110 infrared forehead thermometer is specifically designed for safe use on the forehead. The Infrared Forehead Thermometer is a device capable of measuring people’s body temperature by detecting the intensity of infrared light emitted from the forehead.
It converts the measured heat into a temperature reading displayed on the LCD. When properly used, it will quickly assess your temperature in an accurate manner.
This appliance conforms to the following standards:
ASTM E 1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature, ISO 80601-2-56 Medical electrical equipment —Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement, IEC 60601-1-11 Medical electrical equipment —Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment and complies with the requirements of IEC 60601-1-2(EMC) , AAMI ANSI ES60601-1(Safety) standards. And the manufacturer is ISO 13485 certified.

Warning

  1.  There is no gender or age limitation for using infrared forehead thermometer.
  2. Do not touch the temperature probe with hands .
  3. Use of this Forehead thermometer is not intended as a substitute for consultation with your physician.
  4. Do not allow children to take their temperatures unsupervised, some parts are small enough to be swallowed.
  5. Never immerse this device in water or other liquids(not waterproof).
  6. Do not modify this equipment without authorization of manufacturer.
  7. Do not expose the thermometer to temperature extremes (below -25°C/-13°F or over 55°C/ 1 31°F) nor excessive humidity (>95%RH).
  8. Keep the battery away from children.
  9. Remove battery from the device when not in operation for a long time.
  10. Do not put the thermometer in direct sunlight or with cotton wool, otherwise the accuracy will be affected.
  11. Portable and mobile RF communications can affect the devise. The device needs special precaution regarding EMC according to the EMC information provided in the accompany documents.
  12. ME equipment should not be cleaned and disinfected while in use.

Product Description

  1. Probe
  2. START Button
  3. Setting Button
  4. Battery Cover

ZEWA 11110 V2 Digital Infrared Forehead Thermometer -
fig

LCD Display Introduction

ZEWA 11110 V2 Digital Infrared Forehead Thermometer -
fig1

Basic Functions

Real Time Clock| The real time clock will be recorded with the memory function and help you to recognize each measurement result.
-. Please see the Real Time Clock Setting.
---|---
Forehead Mode| The thermometer has been designed for practical use. It’s not meant to replace a visit to the doctor. Please also remember to compare the measurement result to your regular body temperature.
-> Please see the Illustration For Use section to learn how to measure the body temperature.
Object Mode| The object mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. It can help you to monitor if the object temperature is suitable for the baby or patient, for example the baby’s milk.
-. Please see the Illustration For Use section to learn how to measure the object temperature.
Memory Mode| The machine can store up to 30 sets of measurements. Each
memory also records the measurement results date/time/mode.
When the data exceeds 30 sets, the coverage starts from the earliest data.
C/ °F Switch| Please see the Real Time Clock Setting to learn how to change between Celsius and Fahrenheit.
Distance Detection| If the object is too far away from the probe, the temperature will not be measured until the probe is moved within the measured range.
Beep Hint| When the testing is done, it will sound one short beep.
Sound Switch| The thermometer can turn on or off sound. -.Please  see the Illustration For Use.

Forehead Thermometer Advantages

Infrared Forehead Thermometer measures core body temperature, which is the temperature of a body’s vital organs. (See Figure 1) This thermometer is designed to measure the temperature of the skin surface over the temporal artery, a major artery of the head. The temporal artery is connected to the heart via the carotid artery, directly leading from the aorta, the main trunk of the arterial system. It offers constant blood flow. Therefore, body temperature changes are reflected sooner in the forehead than they are in other parts of the body such as oral, rectal and underarm.

Real Time Clock Setting

When installing the battery, please set the parameters of the thermometer. With the thermometer off, press and hold Setting BUTTON to enter into setting mode.

  1.  Set the unit Press and release the START BUTTON to change theuint between t or T. After the unit is set, press Setting BUTTON ,the time format figure will appear.

  2. Set the time format The device can display the time in either an AM/PM (12-hour) or a 24:00(24-hour) format. Press and release START BUTTON to select the format. With the preferred time format on the display, press Setting BUTTON, the Hour figure is flashing automatically.

  3. Set the hour Press and release the START BUTTON to advance one hour until the correct hour appears. After the hour is set, press Setting BUTTON, the Minute figure is flashing automatically.

  4. Set the minute
    Press and release the START BUTTON to advance one minute until the correct minute appears. After the minute is set, press Setting BUTTON, the Year figure is flashing automatically.

  5. Set the year
    Press and release the START BUTTON to advance one year until the correct year appears. After the year is set, press Setting BUTTON, the Month figure will appear.

  6. Set the month
    Press and release the START BUTTON to advance one month until the correct month appears. After the month is set, press Setting BUTTON, the Date figure is flashing automatically.

  7. Set the date
    Press and release the START BUTTON to advance one day until the correct month appears. After the day is set, press Setting BUTTON to exit the setting mode.

Temperature Taking Hints

To ensure that the reading always reflects the body temperature accurately, you need to take account of the following factors which may affect an accurate reading.

  1. It is important to know each individual’s normal temperature when they are well. This is the only way to accurately diagnose a fever. To determine normal temperature, take multiple readings when healthy. Re-measure with a standard digital thermometer for confirmation.

  2. Users must be inside for 30 minutes before taking a measurement.
    Note: Users and the thermometer should be in the same ambient temperature for at least 10 minutes before taking a reading.

  3. Users should not drink, eat, or be physically active such as bathing, showering, shampooing and hair drying before/while taking the measurement. Remove hat and hair and wait 10 minutes before taking a reading.

  4. Oils or cosmetics on the forehead may give a lower temperature reading than the actual one. Remove dirt from the forehead before taking a measurement. Wait at least 10 minutes after washing the forehead area before taking a reading.

  5. Holding a hand on the forehead for any length of time will affect the temperature reading.

  6. Do not take temperature over scar tissue, open sores or abrasions.

  7. Do not use the thermometer on a perspiring or sweating forehead. as this may affect the reading.

  8. Don’t take a measurement while or immediately after nursing a baby.

  9. Do not take temperatures with this thermometer near places that are very hot, such as fireplaces and stoves.

  10. The probe window of the thermometer is the most delicate part of the device. Do not touch the probe window. The accuracy of the reading may be affected if the probe window is damaged or dirty.

  11. If the thermometer is stored in a significantly different environment than testing location, place it in the testing location for approximately 30 minutes prior to use.

  12. It is not intended for use in the oxygen rich environment and presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.

 Illustration For Use

▶ To measure forehead temperature:

  1. Rotate the probe to the appropriate position, The display is activated to show all segments. After self-checking Figure 2 appears on the display screen with beeps, so you can start a new measurement.
  2. Aim the thermometer at the center of the forehead with a distance less than 5cm (See figure 3) and then press the START BUTTON. Note 1: Do not remove the thermometer from the forehead before hearing the final beep. A waiting period of 1 seconds between testing is recommended to ensure complete accuracy. Note2:If the object to be tested is too far from the probe, a prompt will appear(See figure 4).
  3. Read the temperature on the display.
  4. Rotate the probe to its original position and it will turn off.ZEWA 11110 V2 Digital Infrared Forehead Thermometer - fig10

Rotate the temperature sensor to the appropriate position (facing forward or backward) which turns the thermometer ON.
▶ How to turn on or off sound:
You can press Setting BUTTON to turn on or off sound.
▶ How to change the forehead mode and object mode:
You can press and hold Setting BUTTON to switch the mode between Forehead mode and Object mode.
▶ To measure object temperature:

  1. Rotate the probe to the appropriate position, you can take the object temperature after hearing beep.(see figure 5)
  2. Aim the thermometer at the center of the object you want to measure with The distance less than 5cm.
  3. Press the START BUTTON and then read the temperature on the display.
  4. Rotate the probe to its original position and it will turn off.

▶ After measurement:

  1. Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life.
  2. Clean the probe after each use to ensure an accurate reading and avoid cross contamination. (See the section of Care and Cleaning for details.)

▶ BackLight:
In Forehead mode:

  1. The display will be lighted GREEN for 3 seconds when the unit is ready for measurement and a measurement is completed with a reading less than 37.3°C(99.1°F).
  2. The display will be lighted YELLOW for 3 seconds when a measurement is completed with a reading less than 37.8°C(100.07).
  3. The display will be lighted RED for 3 seconds when a measurement is completed with a reading equal to or higher than 37.8°C(100.0°F).

In Object mode:
The display will only be lighted GREEN for 3 seconds when the unit is ready for measurement and a measurement is completed.

Memory Mode

  1. The Memory Mode can be accessed either in forehead mode or object mode: Rotate the temperature sensor to the appropriate position (facing forward or backward) which turns the thermometer ON. Press the START BUTTON and the SETTING BUTTON as the same time. The letter M will appear in the lower right corner of the display (see figure 6). Press the SETTING BUTTON to scroll through the memory.
  2. The thermometer will automatically memorize the last 30 temperature readings. Each memory also records the measurement date/time/mode icon. Each time the Setting BUTTON is pressed, the screen displays past readings that correspond with a number 1-30. The number 1 reflects the most recent reading, while the number 30 reveals the oldest reading stored in memory.(See Figure 7)
  3. In the memory mode, mark or lif mark will not change. The user can press the START BUTTON to take new measurements.

Care And Cleaning

  1. The probe window must be kept clean, dry, and undamaged at all times to ensure accurate readings. The accuracy of temperature readings can be affected by damage to the probe window, or the presence of dirt, fingerprints, earwax, dust and other soiling compounds on the probe window. Degraded sensors can degrade performance or cause other problems.
  2. For cleaning:
  1. Soak a clean soft cloth in drinking water,wring it out,and then wipe the thermometer(including probe) no less than 3 times;
  2. Visual or use magnifying glass to observe the thermometer have no visible dirt and stains ,then use another clean soft cloth to wipe the thermometer residue water;
  3. Put the thermometer in the original packaging.
  1. For disinfection:
  1. Soak a clean soft cloth in drinking water,wring it out,and then wipe the thermometer(including probe) no less than 3 times ;
  2. Visual or use magnifying glass to observe the thermometer have no visible dirt and stains ,then use another clean soft cloth to wipe the thermometer residue water;
  3. Using a clean soft cloth dipped in 70% medical alcohol,wipe the probe for 3 Times,each time 1 minute.
  4. Using a clean cotton swab dipped in 70% medical alcohol,wipe the sensor window 3 times;
  5. Wait at least 10 minutes to let the alcohol volatilization and put it in the original packaging.
  1. Do not put the thermometer into water directly.
  2. Store the thermometer in a dry location, free from dust and contamination and away from direct sunlight.
  3. Put the thermometer back to the original packaging after using.

Battery Replacement

  1. Replace battery when ” ” appears in the upper right corner of LCD display. (See Figure 8)
  2. Slide battery cover down as shown in Figure 9.
  3. Remove battery and install 2 new AAA alkaline batteries as shown in Figure 10 .
  4. Slide battery cover back on.

ZEWA 11110 V2 Digital Infrared Forehead Thermometer -
fig13Specifications

Measuring range| Forehead mode: 34.0°C-43.0°C(93.2°F-109.4°F) Object mode: 0t-100°C(32T-212°F)
---|---
Measuring site| Forehead(Forehead Mode)
Reference body site| Oral (This thermometer converts the forehead temperature to display its “oral equivalent.”)
Operation mode| Forehead mode(Adjust mode)
Laboratory accuracy| Forehead mode: ±0.2t (0.4°F) during 35.5°C-42.0t (95.9°F-107.6°F) at 1st-35r (59.07-95.0°F) operating temperature range ±0.3°C (0.5°F) for other measuring and operating temperature range
Object mode: ±4% or ±2C(4°F) whichever is greater
Display resolution| 0.1°C or 0.1°F
Measure time| 1 second
Operating temperature range:| I 0°C-40t(507-104°F), I 5%-85%RH, non-condensing Atmospheric Pressure : 700hPa — 1060hPa
Storage and transport temperature range| -25°C— 55°C (-13°F-131°F), 15%-95%RI-1, non-condensing Atmospheric Pressure : 700hPa — 1060hPa
Clinical accuracy| Clinical bias: -0.09r (-0.16°F)
Clinical repeatability: 0.13°C (0.23°F) Limits of agreement: 0.87C (1.57°F)
Shock| withstands drop of 3 feet
Dimension| 153.941.425.9mm
Weight| Approx. I 02grams(with batteries)
Battery| DC3V(2 X AAA battery)
Battery life| Approx. 1 year/6000 readings
Expected service life| Three years
Ingress protecting rating| IP22
Contraindication| No contraindication

Troubleshooting

Error message Problem Solution
The thermometer is not functioning properly. Unload the battery, wait for

1 minute and repower it. If the message reappears, contact the retailer for service.
| The ambient temperature is not within the range between 10°C and 40°C (50°F-104°F).| Place the thermometer in a room for at least 30 minutes at room temperature between 10°C and 40°C  (50T-104T)
Error message| Problem| Solution
---|---|---
| In Forehead mode: Temperature taken is higher than 43.0 °C (109.47).
In Object mode: Temperature taken is higher than 100 °C (212°F).| Read Temperature Taking Hints Thoroughly, then take a new temperature measurement.
| In Forehead mode: Temperature taken is lower than 34.0 °C (93.2°F).
In Object mode: Temperature taken is lower than 0°C
(327).| Read Temperature Taking Hints thoroughly , then make sure the lens filter are clean, then take a new temperature measurement.
| The thermometer works properly.| Use the thermometer normally
| When battery outline flashes, it indicates that the power is low, but you can continue to measure.| The thermometer will take a proper measurement but batteries must be replaced soon.
| The thermometer could not work due to low battery.| Replace two new alkaline batteries size AAA.

Calibration

The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to the use instruction, periodic readjustment is not required. However, We recommends checking calibration every two years or whenever clinical accuracy of the thermometer is in question. Please send the complete device to the dealers or manufacturer.
The above recommendations do not supersede the legal requirements. The user must always comply with legal requirements for the control of the measurement, functionality, and accuracy of the device which are required by the scope of relevant laws, directives or ordinances where the device is used.
A clinical summary and procedures for checking calibration are available upon request.(Turn on the thermometer and press the START button long time until entering into calibrate mode, software version will be displayed.)

Symbol Explanation

| Caution
---|---
| Direct Current
| Batch Code
| Storage and Transportation Temperature Limit: -13°F —131°F (-25°C-55°C )
| TYPE BF APPLIED PART
| Refer to instruction manual/booklet
| Generalsymbol for recovery/ recyclable
| Disposal of this product and used batteries should be carried out in accordance with the national regulations for the disposal of electronic products.
| Atmospheric pressure limitation
| Storage and Transportation Humidity limitation: 15%-95%R1-1
| Manufacturing Date
| Manufacturer
| The first num.2:Protected against solid foreign objects of 12,5 mm∅and greater.
The second num.2:Protection against vertically falling water drops when ENCLOSUREtilted up to 15° .

Warranty

Zewa provides a 2 year limited warranty on this product to be free from defects in material and workmanship only under normal use. This warranty extends only to the original owner.
For warranty claims, please return the product together with satisfactory proof of purchase such as a sales receipt. Be sure to include your return address, phone number and a brief description of the problem and mail it to:
Zewa Inc.
Attn: Service Department 1
2960 Commerce Lakes Drive Suite 29
Ft. Myers, Florida 33913
Include a check for return shipping and insurance. Contact Zewa (US and Canada only) at 1-888-993-3592 for appropriate shipping and insurance costs.
In lieu of return of the product, Zewa offers a free of charge warranty repair service on all defective subassemblies, provided the product fails to perform due to faulty workmanship or faulty materials. Defects on the product due to normal wear and tear, misuse, abuse, improper handling whether negligent or intentional, failure to follow instructions or alterations and repairs of the product other than by an authorized Zewa Service location will void this warranty.
All other repair services not covered under this warranty will be charged to the customer.
Zewa will either repair or replace any parts necessary ( at its option) to correct defects in the materials or workmanship. Any replacement parts or product may be new or refurbished.

The above warranty is complete and exclusive. Except as stated above, Zewa makes no express or implied warranties or representations, including, but not limited to the accuracy or completeness or information delivered to you, or warranties of fitness for a particular purpose, intended use, merchantability, non-infringement, or any implied warranties arising out of a course of performance, dealing, or trade usage and we specifically disclaim such warranties. In no event will Zewa be liable for indirect, exemplary, incidental, special or consequential punitive or similar damages, including without limitation, any damages resulting from loss of use, loss of business, loss of revenue, loss of profits, or arising in connection with this warranty or your purchase or attempt to purchase products, or of any other obligations, even i f we have been advised of the possibility of such damages. Further, the aggregate liability (whether claims arise in contract, tort, personal injury, product liability or otherwise) of Zewa ( and its officers, directors, employees, agents, representatives and affiliates), and your sole and exclusive remedy, arising with respect to or in connection with this warranty, shall not exceed the aggregate of our invoiced purchase price with respect to any products as to which there is a dispute. The foregoing limitation of liability shall apply regardless of the cause of action under which such damages are sought, whether in contract, in tort or otherwise.
This warranty gives you specific legal rights, and you may have other rights, which vary from state to state. For this warranty to be effective the enclosed registration card must be completed and returned to us within 15 days from the date of purchase. This warranty registration is also available online at www.zewa.com. Any failure to return the registration card to Zewa timely will constitute a complete bar to any claim by you with respect to such product. Any failure to make a timely claim with respect to the cuff will also bar your claim. Unauthorized returns will be returned to you at your expense and need not be accepted by Zewa.
For any questions regarding this warranty, please contact us at www.zewa.com or 888-993-3592.

FCC Information

Caution: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user authority to operate the equipment. *Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try and correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the distance between the equipment and the receiver.
  • Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
The device has been evaluated to meet general RF exposure requirement. The device can be used in portable exposure condition without restriction.

Electromagnetic Compatibility Information

The device satisfies the EMC requirements of the international standard IEC 60601-1-2.The requirements are satisfied under the conditions described in the table below.The  device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be published in the instructions for use.Portable and  mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-approved accessories can affect the device negatively and alter the  electromagnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.

Table 1

Guidance and manufacturer’s declaration — electromagnetic emission

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device shou Id
assure that it is used in such an environment.
Emissions test| Compliance| Electromagnetic environment – guidance
RF emissions
CISPR 11| Group I| The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR II| Class B| The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2| Not applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3| Not applicable

Table 2

Guidance and manufacturer’s declaration — electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test| IEC 60601
test level| Compliance level| Electromagnetic environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2| 8 kV contact
2 kV, ±4 kV, .8 kV,±I5 kVair| 8 kV contact
a kV, ±4 kV. ±8 kV, t15 kV air| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrostatic
transient / burst
IEC 61000-4-4| *2 kV for power supply lines
100 kHz repetition frequency

  • I kV for input/output lines| N/A| N/A
    Surge
    11 C 61000-4-5| 0,5 kV, I kV differential
    mode line-line| N/A| N/A
    Voltage dips, short
    interruptions and
    voltage variations
    on power supply
    input lines
    IEC 61000-4-11| 0 % UT
    (100% dip in UT)
    for 0.5 cycle at 0°,45°, 90°, 135°,180°, 225°, 270°, and 315°
    0 % UT
    (100 % dip in UT ) for I cycle at 0°
    70 % UT
    (30% dip in UT )
    for 25/30 cycles at 0°
    0 % UT
    (100% dip in UT)
    for 250/300 cycle at 0°| N/A| N/A
    Power frequency
    (50/60 Hz)
    magnetic field
    IEC 61000.4-8| 30 Alm, 50/6011z| 30 A/m, 50/60Hz
    ro| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
    NOTE: UT is the a. c. mains voltage prior to application of the test level.

Table 3

Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended device should assure that for use in the electromagnetic environment specified below. The customer or the user of the it is used in such an environment.
Immunity test| IEC 60601 test
level| Compliance
level| Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3| 3 Vrms
150 kHz to 80 MHz
6 Vnns 150 kHz to
80 MHz outside
ISM bandsa
10 V/m
80 MHz to 2.7 GHz| N/A
10 V/m| Portable and mobile RF communications
equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to  the frequency of the transmitter.
Recommended separation distance

80MHZto800MHz

where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range

Interference may occur in the vicinity of equipment marked with the following symbol:

Table 3 continued
NOTE
lAt 80 MHz and 800 MHz, the higher frequency range applies.NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a The ISM(industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHzto6,795 MHz;13,553 MHz to 13,567 MHZ;26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz,3,5 MHz to 4,0 MHz,5,3 MHz to 5,4 MHz,7 MHz to 7,3 MHz,10,1 MHz to 10,15 MHz,14 MHz to 14,2 MHz,18,07 MHz to 18,17 MHz,21,0MHz to 21,4MHz,24,89 MHz to 24,99 MHz,28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHZ.
b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2,7 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended separation distance for transmitters in these frequency ranges.
c Field strengths from fixed transmitters, such as base stations for radio(cellular/ cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Table 4

Recommended separation distances between
portable and mobile RF communications equipment and the device

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output of
transmitter| Separation distance according to frequency of transmitter  m

150 kHz to 80 MHz

| 80 MHz to 800 MHz|

800 MHz to 2.7 GHz

0.01| 0.12| 0.04| 0.07
0.1| 0.37| 0.12| 0.23
1| 1.| 0.35| 0.7
10| 4.| 1.| 2.
100| 12.| 4.| 7.0
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres (m) can be estimated using the equation  applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter  manufacturer.
NOTE I At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all  situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Recommended separation distances between RF wireless communications equipment

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between RF wireless communications equipment and the device as recommended below, according to the maximum output power of the communications equipment.
Frequene MHz| Maximum PowWer| Distance| IEC 60601 Test Level| Compliance Level| Electromagnetic Environment – Guidance
385| 2.| 0.3| 27| 27| RF wireless communications equipment  should be used no closer to any part of the  device, including cables, than the recommended separation distance calculated  from the equation applicable to the frequency of the transmitter.
Recommended separation distance

Where P is the maximum output power rating of the ransmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitter, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:

450| 2| 0.3| 28| 28
710| 0.2| 0.3| 9| 9
745
780
810| 2| 0.3| 28| 28
870
930
1720| 2| 0.3| 28| 28
1845
1970
2450| 2| 0.3| 28| 28
5240| 0.2| 0.3| 9| 9
5500
5785
Note I: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

WARNINGS!
– This device should not be used in the vicinity or on the top of other electronic equipment such as cell phone, transceiver or radio control products. If you have to do so, the  device should be observed to verify normal operation.
– The use of accessories and power cord other than those specified, with the exception of cables sold by the manufacturer of the equipment or system as replacement parts  for internal components, may result in increased emissions or decreased immunity of the equipment or system.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals