3B Scientific 1013526 Analogue Multimeter ESCOLA 30 Instructions

June 9, 2024
3B Scientific

3B Scientific 1013526 Analogue Multimeter ESCOLA 30

3B Scientific
1013526 Analogue Multimeter ESCOLA 30-product

Instruction sheet

12/16 SD/JS

  1. Connection sockets
  2. Display with mirror scale
  3. Slotted screw for zero calibration
  4. Calibration trimmer for setting centre zero point 5 Rotary switch to select the measurement range 6 Three-way switch3B Scientific 1013526 Analogue Multimeter ESCOLA 30-fig-1

Description

The ESCOLA 30 is an electronic meter with a mir-rored dial for measurements of current and volt-age involving analogue techniques for schools and training use. The meter allows measurements of AC and DC current and voltage values and the zero point can be set up in the centre of the scale for measure-ment of DC quantities. AC quantities with frequen-cies of up to 40 kHz can be measured. All the pos-sible measuring ranges can be selected by means of a single rotary switch. The excess current protection of the Escola 30 au-tomatically limits the power when overloaded. Af-ter a brief cooling phase, the multimeter au- tomat-ically switches on again.

The use of a robust moving coil galvanometer and a mechanically resilient casing make it pos-sible to use the equipment even under highly challeng-ing conditions. This means that this equipment is suitable for use in schools by children of all years.

Initial calibration

  • Set up the ESCOLA 30 meter either horizon-tally or vertically.
  • Do not connect measuring leads to begin with.
  • Set the three-way switch to .
  • The needle will point to the zero point of the dial. If it does not, the amount of charge of the battery should be checked.

Safety instructions

The analogue multimeter Escola 30 conforms to safety regulations for electrical measurement, con-trol and laboratory equipment, as specified in DIN EN 61010-1, protection class 2 and to measuring category CAT I for up to 30 V. It is not approved for measurements on low-voltage distribution equip-ment, such as sock-ets, fuses, etc. The nominal voltage must not exceed 30 V. The meter is intended for measurements within its measuring ranges and in a measuring environ-ment as described in detail in the course of this manual. Safe operation of the multimeter is guar-anteed if it is solely used as specified. Safety can-not be guaranteed, however, if the multimeter is used incorrectly or handled without due care and attention.
In order to avoid serious injury due to cur-rent or voltage shocks, the following safety instruc-tions are to be observed at all times:

  • Carefully read the instruction manual before using the multimeter and obey the instruc-tions therein.

The assumption needs to be made that unfore-seen voltages may be present in the vincinity of objects being measured (e.g. faulty equipment).

  • Before using the multimeter, check the hous-ing and measuring leads for damage and if there should be any malfunctions or visible damage, the multimeter is not to be used. Pay specific attention to the insulation for the measuring sockets.
  • The authorised measuring range is not to be exceeded. If measurements are made when the magnitude of the variable is unknown, al-ways select a large measuring range before shifting down to lower ones.
  • Before using the analogue multimeter to check that a voltage source is not exhibiting any actual voltage, check that the meter is working properly by selecting the battery test function.
  • When measuring current, make sure the elec-tricity is turned off before the analogue multi-meter is connected into the circuit.
  • When making measurements, always con-nect the ground lead first. Disconnect the sig-nal measurement lead before unplugging the ground.
  • Turn off the multimeter before opening the casing, disconnect the power to the circuit and the measuring leads from the multimeter.
  • When the multimeter is used by teenagers, trainees etc., a suitable person should super-vise to ensure the equipment is used safely.

Operation

To switch on:
  • Set the three-way switch to or .

Checking battery charge:

  • Set the three-way switch to .
  • Disconnect all measuring leads.
  • Set the rotary switch to .

If the battery is sufficiently charged, the needle will point to the following range indication, . If this is not the case, the battery will need to be re- placed immediately.

Zero point calibration:
  • Set up the analogue multimeter either hori-zontally or vertically
  • Set the three-way switch to .
  • Turn the rotary switch to 30 V=.
  • Connect the sockets together with a short connecting lead.
  • Turn the zero-point trimmer screw to adjust the zero point as needed.
Zero point calibration for centre zero point:

For measurements of DC current and voltage, the zero point of the scale can be moved to the centre of the dial.

  • Disconnect all measuring leads.
  • Set the three-way switch to and turn the rotary switch to a DC current or voltage range,
  • Use the zero-point trimmer to line up the nee-dle precisely in the centre of the dial.
To switch off:
  • Set the three-way switch to .
    When the meter is turned off, the needle points to .
If a measurement is interrupted by battery cut-out:

After 45 minutes of use, the multimeter is auto-matically shut off and the needle will then point to .
To switch back on:

  • Set the three-way switch of the analogue mul-timeter to off and then use it to turn the meter back on.

Current measurements

**| The Escola 30 is not permitted to be used for measurements on low- volt- age mains distribution equipment, such as mains sockets, circuit breakers and fuses etc.
---|---
**
| The nominal voltage of the electricity source must not exceed 30 V and its nominal current must be no higher than 3 A.

·         The multimeter is to be connected in series with the load on the side where the voltage with respect to ground is lowest!

·         Turn off the circuit before the Es- cola 30 meter is connected into it.

3B Scientific 1013526 Analogue Multimeter ESCOLA
30-fig-8

  • Connect the side with the lower potential to the left-hand socket.
  • Connect the common/ground lead first and only then the signal lead.
DC currents (mA=):
  • Set the rotary switch to the desired DC current range. If measurements are made when the magnitude of the variable is unknown, always select a large measuring range before shifting down to lower ones.
  • Set the three-way switch to or, if the zero point is to be set in the centre, to .
5.2 AC currents (mA~):
  • Set the rotary switch to the desired AC current range. If measurements are made when the magnitude of the variable is unknown, always select a large measuring range before shifting down to lower ones.
  • Set the three-way switch to

Voltage measurements

**| The Escola 30 is not permitted to be used for measurements on low- volt- age mains distribution equipment, such as mains sockets, circuit breakers and fuses etc.
---|---
**
| The nominal voltage of the voltage source must not exceed 30 V.

3B Scientific 1013526 Analogue Multimeter ESCOLA
30-fig-9

  • Connect the side with the lower potential to the left-hand socket.
  • Connect the common/ground lead first and only then the signal lead.
DC voltages (V=):
  • Set the rotary switch to the desired DC voltage range.If measurements are made when the magnitude of the variable is unknown, always select a large measuring range before shifting down to lower ones.
  • Set the three-way switch to or, if the zero point is to be set in the centre, to .
AC voltages (V~):
  • Set the rotary switch to the desired AC voltage range.If measurements are made when the magnitude of the variable is unknown, always select a large measuring range before shifting down to lower ones.
  • Set the three-way switch to .

Symbols for measuring ranges

mA= DC (unit mA)
mA~ AC (unit mA)
V= DC voltage (unit V)
V~ DC voltage (unit V)

Technical data

Display:
  • Scales:
    • 0 … 10, linear
    • 0 … 3, linear
  • Type:
    • Mirror scale
  • Scale length:
    •   80 mm
  • Pointer deflection:
    • 0…90°

Electrical

  • zero-point offset:
    •   in all DC ranges

Measurements:

  • Voltage ranges:
    • 0.3, 1, 3, 10, 30 V AC/DC
  • Internal resistance:
    • 10 k/V
  • Current ranges:
    • 1, 10, 100, 1000,
    • 3000 mA AC/DC
  • Voltage drop when measuring current:
    • 100 mV approx. AC/DC

Reference conditions:

  • Ambient temperature:
    • 23 °C
  • Operating alignment:
    • Vertical/horizontal
  • Signal form:
    • Sine (1% max. discrepancy)
  • Peak factor:
    • 2
  • Frequency range:
    • 40 Hz … 50 Hz … 5 kHz b

Accuracy (at reference conditions):

  • DC quantities:
    • Class 2
  • DC with
  • zero-point offset:
    • Class 5
  • AC quantities:
    • Class 3

Extended frequency range (class 10):

  • 0.3 – 30 V:
    • 40 Hz … 50 Hz … 40 kHz
  • 1 – 3000 mA:
    • 40 Hz … 50 Hz … 40 kHz

Overload protection:

  • Current and voltage ranges:
    • reversible for up to ±50 V AC/DC peak-to-peak value and max. 40 A.

Electrical safety:

  • Safety specifications:
    • EN 61010-1
  • Measuring category:
    • CAT I: 30 V
  • Contamination level:
    • 2
  • Protection type:
    • IP20
  • Connectors:
    • 4-mm safety sockets

Power supply:

  • Battery:
    • 1x 1.5 V, AA IEC LR6
  • Automatic cut-off after:
    • 45 min ± 10 min

Electromagnetic compatibility:

  • Interference emission:
    • EN 55011:2009
  • Interference resistance:
    • EN 61326-1:2013

Operating conditions:

  • Ambient temperature:
    • 5 °C … 23 °C … 40°C
  • Storage temperature:
    • -20 … 70°C
  • Relative humidity:
    • <85% with no condensation

General data:

  • Shock test:
    • max. 147 m/s²
  • Dimensions:
    • 100x150x50 mm3 approx.
  • Weight:
    • 300 g approx..

Measuring categories according to DIN EN 61010-1.

  • CAT I:
    • or unstipulated: Approved for measurements in circuits which are not directly connected to the low voltage mains grid (e.g. batteries).
  • CAT II:
    • Approved for measurements in circuits which are directly connected, by a mains lead and plug for instance, to the low voltage mains grid (e.g. household or office appliance and lab equipment).
  • CAT III:
    • Approved for measurements in circuits which are part of a building’s wiring installation (e.g. stationary consumers, distribution terminals, appliances connected directly to the distribution box).
  • CAT IV:
    • Approved for measurements in circuits which are directly connected to the source of the low voltage mains (e.g. electricity meters, main service feed, primary excess voltage protection).

Battery, care and maintenance

Battery testing:

Batteries which are discharged and have not been used for a while may leak.

 | ·         Check the amount of charge left in the battery from time to time.

·         Flat or leaking batteries should be removed from the meter.

---|---
 | If the Escola 30 goes unused for a long period:

·         Remove the Escola 30’s battery before reusing.

Changing the battery:

****| Before opening the casing:

·         Turn off the Escola 30.

·         Disconnect all measuring leads.

---|---

The polarity is indicated by plus and minus signs inside the fuse holder compartment. A mechanical system ensures the battery makes no contact if it is inserted the wrong way round.

  • Unscrew the back of the casing.
  • Replace flat batteries with 1.5-V alkaline batteries of size AA IEC LR6.
  • Place the negative pole of the battery on the spring.
  • Close casing again.
Cleaning:
  • For cleaning, use a soft cloth, slightly moistened with alcohol, or a brush.

Electrostatic charging of the display window can affect the measurements under certain circumstances:

  • To remove such charge, use a soft cloth slightly soaked in alcohol or a paint brush.

Dirt or moisture in the measurement sockets can affect readings.

  • Shake out any dirt that may be in the measurement sockets.
  • Soak a new swab with isopropyl alcohol and work around the inside of each measurement socket.

Disposal

  • The packaging should be disposed of at local recycling points.
  • Should you need to dispose of the equipment itself, never throw it away in normal domestic waste. If being used in private households it can be disposed of at the local public waste disposal authority.
  • Comply with the applicable regulations for the disposal of electrical equipment.
  • Do not dispose of the batteries in the regular household garbage. Follow the applicable legal regulations (UK: Waste Batteries and Accumulators Regulations, EU: 2006/66/EC).

Symbol legend

3B Scientific 1013526 Analogue Multimeter ESCOLA
30-fig-10

3B Scientific GmbH
Ludwig-Erhard-Str. 20
20459 Hamburg
Germany
www.3bscientific.com
Subject to technical amendments
© Copyright 2022 3B Scientific GmbH

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

3B Scientific User Manuals

Related Manuals