U S SOLID USS-DBS62 Analytical Balance Instruction Manual
- June 8, 2024
- U S SOLID
Table of Contents
U S SOLID USS-DBS62 Analytical Balance
INTRODUCTION
This analytical balance works on electromagnetic force compensation technology and micro processor which implements high speed stabilization and high reliability. It can be widely used in industry, agriculture, commerce, schools, scientific research and other institutions to quickly weighing the quality and quantity of objects.
Safety Precautions
Verify that the AC Adapter input voltage matches the local AC power supply.
Use the scale only in dry locations.
For best results, use the scale in the specified optimum operating
temperature. Use your scale with care, gently loading items to be weighed onto
the center of the pan. Avoiding rough treatment will aid in the life of your
scale.
Do not suggest to use in below environment.
DISPLAY INFORMATION
Data sheet
Electronic Analytical Balance(0.1mg external calibration)
Model| 1004| 1204| 2004| 2204| 3004| 3204|
4204| 5204
---|---|---|---|---|---|---|---|---
Capacity| 0–100g| 0–120g| 0–200g| 0–220g| 0-300g| 0–320g| 0-420g| 0-520g
Readability| 0.1mg
Stable time| 2s
Preheating| 30-60 mins
Weighing units| g/oz/ct/mg
Min weighting| 0.1mg
Repeatability| ±0.1mg| ±0.2mg
Linearity| ±0.2mg| ±0.3mg
Interface| RS232/USB(optional)
Pan size| φ90mm
Product size| 31cm21.2cm35cm
N.W.| 5500g
Power| AC110V-240V
Electronic Analytical Balance(0.1mg internal calibration)
Model| 1004N| 1204N| 2004N| 2204N| 3004N|
3204N| 4004N| 5004N
---|---|---|---|---|---|---|---|---
Capacity| 0 – – 1 0 0 g| 0 – – 1 2 0 g| 0 – – 2 0 0 g| 0 – – 2 2 0 g| 0-300g|
0–320g| 0-400g| 0-500g
Readability| 0.1mg
Stable time| 2s
Preheating| 30-60 mins
Weighing units| g/oz/ct/mg
Min weighting| 0.1mg
Repeatability| ±0.1mg| ±0.2mg
Linearity| ±0.2mg| ±0.3mg
Interface| RS232/USB(optional)
Pan size| φ90mm
Product size| 31cm21.2cm35cm
N.W.| 5500g
Power| AC110V-240V
Electronic Analytical Balance(0.01mg external calibration)
Model| 10035| 10055| 10085| 20035X| 20055|
20085
---|---|---|---|---|---|---
Capacity| 100g/30g| 100g/50g| 100g/80g| 200g/30g| 200g/50g| 200g/80g
Readability| 0.1mg/0.01mg
Stable time| 3s
Preheating| 30-60 mins
Weighing units| g/oz/ct/mg
Min weighting| 0.1mg/0.05mg
Repeatability| ±0.1mg/±0.05mg
Linearity| ±0.2mg/±0.05mg
Interface| RS232/USB ( optional)
Pan size| φ90mm
Product size| 31cm21.2cm35cm
N.W.| 5500g
Power| AC110V-240V
Electronic Analytical Balance(0.01mg internal calibration)
Model| 10035N| 10055N| 10085N| 20035N| 20055N|
20085N
---|---|---|---|---|---|---
Capacity| 100g/30g| 100g/50g| 100g/80g| 200g/30g| 200g/50g| 200g/80g
Readability| 0.1mg/0.01mg
Stable time| 3s
Preheating| 30-60 mins
Weighing units| g/oz/ct/mg
Min weighting| 0.1mg/0.05mg
Repeatability| ±0.1mg/±0.05mg
Linearity| ±0.2mg/±0.05mg
Interface| RS232/USB ( optional)
Pan size| φ90mm
Product size| 31cm21.2cm35cm
N.W.| 5500g
Power| AC110V-240V
Electronic Analytical Balance(1mg external calibration)
Model| 1003| 2003| 3003| 5003| 6003| 10003|
20003
---|---|---|---|---|---|---|---
Capacity| 0–100g| 0–200g| 0–300g| 0–500g| 0-600g| 0–1000g| 0–2000g
Readability| 1mg| 1mg
Stable time| 1s| 3s
Preheating| 30-60 mins
Weighing units| g/oz/ct/mg
Min weighting| 1mg
Repeatability| ±1mg| ±3mg
Linearity| ±2mg| ±4mg
Interface| RS232/USB ( optional)
Pan size| φ90mm
Product size| 31cm21.2cm35cm
N.W.| 5500g
Power| AC110V-240V
Electronic Analytical Balance(1mg internal calibration)
Model| 1003N| 2003N| 3003N| 5003N| 6003N|
10003N
---|---|---|---|---|---|---
Capacity| 0–100g| 0–200g| 0–300g| 0–500g| 0-600g| 0–1000g
Readability| 1mg
Stable time| 2s
Preheating| 30-60 mins
Weighing units| g/oz/ct/mg
Min weighting| 1mg
Repeatability| ±1mg
Linearity| ±2mg
Interface| RS232/USB ( optional)
Pan size| φ90mm
Product size| 31cm21.2cm35cm
N.W.| 5500g
Power| AC110V-240V
PREPARATION
Put the balance on the stable desk, the desk can not be moved. Adjust the 2 Leveling Feet on the bottom, until the bubble is centered in the circle, then install the scale pan.
Plug in
the AC adapter
Press “ON” button and the display shows“0.0000”, the scale enters into the
weighing mode.
When the operating temperature changes, put the balance in the new place for 2 hours under the “power on” state to adapt the balance to the new temperature.
Attention: If the number on display is not stable at the first usage, it is caused by the temperature difference of the environment, you can press ”TARE”button repeatedly and let it stand for 30 minutes.
CALIBRATION
- In order to get accurate weighing result, balance should be calibrated before the below usage scenarios.
- Before first use
- Balance power off for long time or power error.
- After changing the operating environment.
- Regularly in weighing procedure.
- Choose the calibration (follow the sequence) Press and hold“ON”
- when display show “SET-C” release the button.
- Press “TARE”to save and exit the setup.
Single-point calibration follow the sequence
- Press the button “TARE”, the display shows “0.0000”.
- Press “CAL” button, balance enters into calibration procedure, the display shows “CAL–0”, internal calibration balance will make the calibration automatically
- The display shows from“CAL-0”to “CAL-200”and flashes, put on the 200g weight for FA2204X model
- Display shows“200.0000”,remove the weight and finish the calibration.
Internal calibration weight value setting (internal calibration balance)
- Choose the calibration“CAL-2”(follow 3.2.2)
- Press “CAL” button, the display shows “CAL-DN”
- Put on the standard calibration weight accordingly the display shows “CAL-UP”, remove the weight.
- Wait for the calibration finished, display shows“0.0000”
Linearity calibration
Balance without loading, and display shows “0.0000”, it can be operated for linear calibration.
- Press and hold “ON” button, balance enters into main menu mode, start to change function
- Wait to the display shows “SET-C”, release the “ON” button
- Press and hold the “CAL” button for 3 seconds and release the button, the display show“LIN-0”.
- Put on the calibration weight accordingly, finish the calibration procedure.
OPERATION
Unit conversion
Under the weighing interface press and hold “PRINT” to change the unit the default unit:g,oz,ct,mg
Other unit is optional
piece counting setting
- Press and hold “ON” button under weighing interface, balance enter into main menu. wait until the display to show “SEt – 0”, release the “ON” button.
- Press and hold “PRINT”, enter into piece counting sample quantity setting, the display shows “-COU-*-”
- Press and hold “PRINT” to set up piece counting sample quantity, the display will show
Piece counting weighing
- Press and hold “PRINT” button under weighing interface, the display shows “0 PCS” and , release the “PRINT” button
- Put the sample quantity which has been set in SET-0.
- Press “ON” button to confirm the sample weight, and start weighing.
If the same sample is used the second time, no re-sampling is required.
Long press “PRINT” button to exit the piece counting and return to the
weighing mode.
Percent weighing
- Press and hold “PRINT” button under weighing interface, the display show“0.00” and , release the “PRINT” button
- Put the contrast target sample on the pan (e.g. 100g weight), press “ON”button to save the sample weight, the display shows “100.00”.
- Put the contrast on the pan (e.g.50gweight), the figure on the display is the percentage of contrast
If the
same contrast is used the second time, no re-sampling is required.
Long press “PRINT” button to exit the percent weighing and return to the
weighing mode.
Density determination (need gravity kit, optional) density weighing setting
- Press and hold “ON” button, the balance enters into the main menu mode.
- When the display shows “SET-d”, release the button.
- Press and hold “PRINT”, enter into density weighing setting, the display shows “-dEn-*-” press and hold “PRINT“ to set up density determination, display shows:
- Press “TARE” to save and exit the setup, display shows W1 in the right corner
Solid density determination:
- Set up auxiliary liquid density value
- Press and hold “ON” button under , balance will enter into the main menu and change
- Release the “ON” button until the display shows “SEt-d”,
- Press and hold “CAL” button, enter into auxiliary liquid density value setting
- Press and hold “CAL” button to increase, press and hold “PRINT” button to decrease press “TARE” to save the density value and exit
- Display shows W1, put the sample on the pan in air.( e.g.screw sample)
-
- Weighing the sample in liquid – 2 press and hold “PRINT” button to collect the data
- Display will show the result of density (W1\W2 no display
- Press and hold “RPINT” to exit density result
Follow 4.4 setting,set the density weighing to“-dEn-0-“, press “TARE“ button to exit density determination.
Dynamic weighing function dynamic weighing setting
- Press and hold “ON” button under weighing interface, the balance enters into the main menu mode.
- Release “ON” button until the display shows “SET-1”,
- Press and hold “PRINT” button to enter into filter level setting, the display shows“-Int-*-”.
- Press and hold “PRINT” button to set up filter level, the display shows:
- Press “TARE” to save the setting and exit, the display shows
Dynamic weighing function dynamic weighing
- Put the sample on the pan.
- Press and hold “PRINT” above 3 seconds, will flash, the dynamic weighing starts.
- Stop flashing, weighing end, display 3 shows the final result.
- Press and hold “PRINT” to exit.
Follow 4.5 setting, set the dynamic weighing to “-dEn-0-”, exit dynamic weighing.
Baud rate setting
- Press and hold “ON” button under weighing interface, the balance enters into the main menu mode release “ON” button until the display shows “SET-F”,
- Press and hold “PRINT” to enter into baud rate setting, display shows: “F-****”
- Press and hold “PRINT” to set up baud rate, the display shows:
- Press “TARE” to save the setting and exit.
Print setting
- Press and hold “ON” button under weighing interface, the balance enters into the main menu mode.
- Release“ON”button until the display shows “SET-P”,
- Press and hold“PRINT”to enter into print setting, display shows “-Prt-*-”.
- Press and hold“PRINT”to set up print model, display shows:
- Press “TARE” to save the setting and exit.
sensitivity setting
- Press and hold“ON”button under weighing interface, the balance enters into the main menu mode.
- Release“ON”button until the display shows “SET-A”,
- Press and hold“PRINT”to enter into sensitivity setting, display shows “-ASD-*-”.
- Press and hold“PRINT”to set up sensitivity, the display shows.
- Press “TARE” to save the setting and exit.
REMARK: factory setting sensitivity “-ASD-3-”, suitable for most using
environment. Higher sensitivityÿrequired better environment. Do not adjust the
sensitivity
without consulting the factory.
Filter level setting
- Press and hold “ON” button under weighing interface, the balance enters into the main menu mode.
- Release “ON” button until the display shows “SET-1”,
- Press and hold “PRINT” button to enter into filter level setting, the display shows“-Int-*-”.
- Press and hold “PRINT” button to set up filter level, the display shows:
- Press “TARE” to save the setting and exit.
REMARK: filter level is the internal calculating time of balance, factory setting is “-Int-2-“, do not change it by yourself without professional instruction.
Time setting
- Press and hold “ON” button under weighing interface, the balance enters into the main menu mode.
- Release “ON” button until the display shows “SET-T”,
- Press and hold“PRINT”button to enter into clock setting, the display shows“-CLOC-*-”.
- Press and hold“PRINT”button to set up time, display shows:
- Press “CAL” button to increase, press “ON” button to decrease.
- Press “TARE” to save the setting and exit.
Interface
RS232 interface
- Connection
Balance (9 pins) PC/Printer (9 pins)
RXD (Input) 2 2
TXD (Output) 3 3
GND (Ground) 5___ 5
- The baudrate by default is 9600 bps (see Baudrate setting)
- Data format: 10 bits, 0 as start bit, 1 as stop bit, 8 digits (ASCII code) – No odd and even numbers adjusting
- Data output: by default is continuous mode. The data outputs mode can be changed into press output, timing output and continuous output (see Data output setting)
- Output data format
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 | 15 |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Tyopre Àda ta | data | data | data | data | doarta dot | doarta dot | data | data | ||||||
data | unit | unit | unit | return | Line feed |
TROUBLESHOOTING
Problem | Cause | Solution |
---|---|---|
No display | No power supply; Fuse damaged; |
Power transformer is damaged;
| Plug in adapter; Change the fuse;
Change the power transformer; If problem persists, send the balance to the
Technical Service for repair.
Unstable display|
Bad working conditions; Air flow Something between the scale pan and working table.
The power exceeds its permissible value and is unstable; Static electricity
| Improve the working condition, Close the windshield;
Remove the pan and clean well the balance surface;
Connect the balance to power supply 110-220 V AC Static Elimination
Poor accuracy|
Improper calibration
The weight of the recipient has not been tared Big temperature difference The balance is not horizontal
| Make calibration Make tare
Put into suitable enviroment Adjust level of balance
COMPONENTS LIST
Balance | 1 |
---|---|
Balance pan | 1 |
Adapter | 1 |
Instruction manual | 1 |
Weight(external | 1 |
calibration) glove | 1 |