ZEISS Sigma 300 RISE Raman Imaging and Scanning Electron Microscopy Installation Guide

June 4, 2024
Zeiss

ZEISS Sigma 300 RISE Raman Imaging and Scanning Electron Microscopy

ZEISS-Sigma-300-RISE-Raman-Imaging-and-Scanning-Electron-Microscopy-
PRODUCT

© 2022 Without the prior written consent of ZEISS, this document or any part of it must neither be translated nor reproduced or transmitted in any form or by any means – including electronic or mechanical methods, by photocopying, recording or by any information or filing system. The right to make backup copies for archiving purposes shall remain unaffected thereby. Any violations may be prosecuted as copyright infringements.
The use of general descriptive names registered names, trademarks, etc. in this document does not imply that such names are exempt from the relevant intellectual property laws and regulations and therefore free for general use. This shall also apply if this is not specifically referred to. Software programs shall entirely remain the property of ZEISS. No program or subsequent up-grade thereof may be disclosed to any third party, copied or reproduced in any other form without the prior written consent of ZEISS, even if these copies or reproductions are destined for internal use at the customer’s only, the only exception being one single back-up copy for archiving purposes.

Installation Requirements Option

Option-specific installation requirements are provided as separate documents, as the options could be supplied together with a new microscope or as an upgrade on an already installed microscope. This document is to be used together with the main microscope installation requirements document. Also consult ZEISS regarding conflicting information given in the installation requirements for the microscope and the option. It may be required to change an existing installation of a microscope when an option is added, e.g. adequate space is required or an additional mains electricity supply. Furthermore, adding an option on the microscope may introduce interference on the image of the microscope.

Requirements

The following sections describe each requirement in detail.

Environmental Requirements

The RISE option does not require any specific environmental conditions. The environmental specifications for the main microscope apply to the whole system. Please also refer to the installation requirements of the microscope for detailed information about environmental requirements. A site survey can be requested from your local ZEISS service representative prior to the installation.

Location Requirements

If a microscope is to be delivered with additional options, please discuss the additional requirements for these options with a ZEISS service representative. The options may require additional health and safety considerations, space, power points or other resources.
The microscope and options must only be operated in closed rooms. The microscope should not be installed near radiators or windows with direct sunlight.
Compliance with the installation requirements of the microscope and the availability of the requested supplies is the responsibility of the customer and has to be readily available at the time of installation.

  • Installation site Exclusively inside buildings
  • Operating range ≤ 2000 m above sea level

Info
Harmful gases and fumes that may be used on the microscope must be safely extracted from the microscope and site in accordance with local safety regulations. The microscope must not be set up in areas with potential danger by explosives.

Space Requirements

  • Recommended room size Min. 4.0 m x 5.0 m x 2.3 m

  • Service and safety clearance area Min. 0.75 m at each side
    Note: Some detectors require a longer distance between the wall and the left-hand side of the SEM.

  • Entrance Min. 0.8 m wide

The Raman system uses a 2 m USB cable from the control unit to the Raman system PC. The Ra-man power and control unit is placed on the left-hand side of the Sigma SEM and the spectrometer is installed on the left WDX port of the Sigma SEM. Therefore moving the Raman power and control unit to access the left panel of the SEM is limited due to cable and fiber optic lengths. The following layouts are recommended for optimal accessibility, airflow and stability, depending on the options purchased with the system. For possible changes in preferential layout, please discuss with your ZEISS representative. The placing of the user table can vary depending on available space and cable lengths.

OVERVIEW

ZEISS-Sigma-300-RISE-Raman-Imaging-and-Scanning-Electron-Microscopy-
FIG-1

  1. Raman power and control unit
  2. Spectrometer
  3. SEM PC
  4. Raman system PC

Electrical Supplies

It is recommended to install a surge protector on the Raman to prevent intense voltage spikes at startup.

Value

  • Power requirements 2 x L1/N/PE 230 VAC ± 10% 16A
  • Nominal AC voltage L1/N/PE 230 VAC ± 10 %
  • Nominal frequency 50 – 60 Hz
  • System connection Length of cable and fibers required
    • Single mode fibers: 3 m (to connect the lasers with the back of the microscope)
    • Multi-mode fibers: 3 m (to connect the top of the microscope with the spectrometer)
    • USB cable: 2 m (to connect the control unit to the PC)
  • Max. power consumption 2 power points are required
    • Power 1: Control unit, spectrometer and
    • camera: 520 W
    • Per laser: 100 W
    • Power 2: PC and monitor 500 W

Safety and Radiation Information

Laser Safety

The Raman equipment is classified as laser class 3B / IIIB, as laser class 3B radiation is generated in the system. The operator of the laser equipment is responsible for conformance with protective measures and safety requirements. Laser safety measures in accordance with the applicable country-specific legal regulations remain valid. As the responsible body, the operator is responsible for complying with laser safety regulations.

WARNING – CLASS 3B LASER RADIATION WHEN OPEN. AVOID EXPOSURE TO THE BEAM

Transport and Storage

  • Allowable: Allowable temperature during on-site storage and transport:
  • temperature
    • Between 10 °C and 30 °C
    • Relative humidity less than 65 %

Weight and Sizes

Raman equipment stand-alone            140

Weight and Sizes of the Transported Goods

Main Components| Length (mm)| Width (mm)| Height (mm)| Weight (kg)
---|---|---|---|---
Console and accessories**| 1450| 1360| 1180| Appr. 348
Spectrometer| 840| 505| 270| 26
Laser| 785| 305| 325| Appr. 9
Controller| 640| 585| 710| Appr. 50
Optical components, couplers| 490| 400| 325| Appr. 50
Microscope head| 645| 575| 445| Appr. 21
PC| 490| 225| 535| Appr. 10
Keyboard| 495| 210| 45| Appr. 0.5
Screen| 650| 220| 450| Appr. 6.5
Cables| 430| 105| 130| Appr. 1.0
Cables, optical fibre spectrometer| 4350| 300| 75| Appr. 0.5
Tools and power supply, cables, performance sheets| 350| 300| 75| Appr. 2.0
Power supplies, objective| 350| 300| 75| Appr. 2.0
PC accessories| 330| 270| 55| Appr. 0.5
White light source| 265| 210| 125| Appr. 1.0

Console, damper, monitors, cables, pipes &c. are wrapped and packed in separate cartons and shipped in a separate box.

Applicable Standards and Regulations

Observe all general and country-specific safety regulations as well as applicable environmental protection laws and regulations. Please refer to the installation requirements of the microscope for information about compliance with regulations and directives.

Responsibility Checklist for the ZEISS Crossbeam RISE Microscopy

ZEISS-Sigma-300-RISE-Raman-Imaging-and-Scanning-Electron-Microscopy-
FIG-3

The above signature conveys that the Installation Requirements have been read and understood. The customer further understands that some of these requirements are necessary to help assure the safety of personnel, or to prevent damage to the equipment or the facility. The customer further understands that some of the requirements must be fulfilled in order for the equipment to meet its full performance capabilities and for the equipment to be successfully installed.

Installation Requirements ZEISS RISE Microscopy (Raman Imaging and Scanning Electron microscopy) | en-US | Rev. 1 |349500-8193-000