Use the following procedure to update a legacy MSI-4260 Port-A-Weigh Crane Scale to an alternate configuration.
Manuals and additional resources are available from Rice Lake Weighing Systems at www.ricelake.com/manuals Warranty information can be found at www.ricelake.com/warranties
Overview
Figure 1. MSI-4260 Retrofit Kit Parts Diagram
Item No. | Part No. | Description | Qty |
---|---|---|---|
1 | 139424 | Front Casting Assembly 4260 (Kit PN 152123) | 1 |
169202 | Front Casting Assembly with Alarm Option (Kit PN 215531) | ||
2 | 162185 | CONN,M12 4 Pin Male Field Wireable 6-8mm OD Straight | 1 |
3 | 198913 | Conn, IP67, Euro Field Wireable, 4 Cond, Female Straight PG-9 | 1 |
4 | 149724 | O-ring Assy Front Casting Paw/Trans-weigh | 1 |
5 | 145998 | Screw SH 3/8-16 X 1-1/4 SS | 3 |
6 | 133597 | Screw Seal 3/8-16 X 1-1/4 SS DR | 1 |
Remove the Existing Front Casting
WARNING: Ensure that all power has been removed prior to starting the procedure.
- Turn the MSI-4260 off.
- Remove the front casting by loosening the screws securing it. See Figure 2.
- Unplug the load cell and power cables from the front casting or bulkhead. Discard casting and hardware.
- With one hand, support the battery access cover.
- Release the two latches holding the battery access cover.
- Carefully lower the cover while holding the battery in place, it could be loose and fall causing damage.
- Remove the battery by pulling straight back.
- Close the cover and secure with latches.
- Remove the battery casting by loosening the screws securing it. Retain for re-installation
Figure 2. Front Casting Removal
Load Cell Cable
- Locate the cable coming out of the load cell.
- Cut the connector off as close to the connector as possible.
- Strip the cable jacket as needed to allow 1/2 – 3/4 inch of exposed conductors, Red, Green, White and Black.
- Cut the shield/drain wire flush with the end of jacket.
- Strip the Red, Green, White and Black wires approximately 1/4 inch and tin with solder.
- onnect the wires to the connector together as shown in Table Table 2. and Figure 3.
L/C Cable Conductor Color| Function| 162185 Connector Pin Number
---|---|---
Red| +Excitation| 1
Black| -Excitation| 2
Green| +Signal| 3
White| -Signal| 4
Table 2. Load Cell Cable Connections
Figure 3. L/C Cable Connector
Battery Cable
- Locate the power cable coming from the inside of the battery casting.
- Cut the connector off the battery cable as close to the connector as possible.
- Strip the cable jacket as needed to allow 1/2 – 3/4 inch of exposed conductors, Red and Black.
- Cut the shield/drain wire flush with the end of jacket.
- Strip the Red and Back wires approximately 1/4 inch and tin with solder.
- Connect wire to the connector as shown in Table Table 3. and Figure 4.
Battery Cable Conductor Color| Function| 198913 Connector Pin Number
---|---|---
Red| + Battery| 1
Black| – Battery| 2
–| Not Used| 3
–| Not Used| 4
Table 3. Power Cable Connections
Figure 4. Battery Cable Connector
Reassembling the Unit
- Place the o-ring in the groove on the bulkhead of the supplied front casting. Ensure bulkhead is in place on the front casting.
- Plug the updated cables into the cable connectors on the bulkhead. Load cell cable connects to the male connector and power cable connects to the female connector.
- Place the front casting assembly on mid-casting assembly and secure with the supplied 3/8 inch drive head screws, torqued to 20–25 ft-lbs.
- Remove the existing serial number and replace it with the new tag using the rivets included.
- Reinstall the battery casting securing with the retained hardware.
- Reinstall the battery.
- Calibrate the unit, see the operators manual (PN 152159) for the calibration procedure.
The MSI-4260 Port-A-Weigh is now ready to begin weighing. Refer to the operators manual (PN 152159) for further information on setup and use.