TM2450 Remler 24h Blood Pressure Monitor
Instruction Manual Instructions for use
boso TM series
24-hour blood pressure monitor
0124
Device overview
**Scope of supply with complete accessories***
-
24-hr blood pressure monitor
-
Transport case
-
battery charger
-
2x battery sets with two batteries each(one set already inserted in the unit)
-
Cuffs for adults
– Size M CA91 washable
– Size L CA92 washable -
Waist bag with removeable carrying belt and strap
-
Instruction manual for:
– boso TM series
– XD profile manager -
Important notes
-
Medical Devices Book
-
CD-ROM
– boso XD profile manager -
USB connection cable
Icon explanation
| Observe the electronic instructions for use
---|---
| Important notes/warnings
| Order number
| CE marking
| Unique Device Identifier
| Switzerland – Authorisation
| Medical device
| Store in a dry place
| Fragile, handle with care
| Temperature limits
| Humidity limit
| Action note for the user
Symbols on the measurement device
| START/STOP button
---|---
| AUTO button (DAY/NIGHT BUTTON)
| Automatic mode active
| Sleep mode active
| Battery charged
| Battery partially charged
| Battery empty no further measurement or data transfer possible
| Storage full, 600 measurements, no further measurements possible
**| Serial number
IP22| Protection against foreign bodies and water:
The IP classification is the degree of protection provided by enclosures
according to IEC 60529. This unit is protected against solid foreign objects
with a diameter of 12 mm and larger, such as fingers. This unit is protected
against falling dripping water when the housing is tilted up to 15°.
| Do not dispose of the appliance in the household waste .
| Defibrillation protected device type BF
| Manufacturer
**| Date of manufacture
| Read instruction manual
OLED display
| Clock display,
if no measurement is taking place
---|---
| Display of the measurement values:
SYS – Systolic blood pressure
DIA – Diastolic blood pressure
PUL – Pulse
mmHg – Unit for blood pressure
/min – Unit for pulse
| Fault display
Introduction
Dear customer, we are very pleased that you have decided to purchase a boso
long-term blood pressure monitor. The boso brand stands for the highest
quality and precision. Currently, 96 % of all German general practitioners,
practitioners and internists work with boso blood pressure monitors in their
practice (API study by GfK 01/2016). This device has passed our strict quality
control and is your safe partner for checking your patients’ blood pressure
values.
Please read these instructions for use carefully before using the device for
the first time, as correct blood pressure measurement is only possible if the
device is handled correctly.
These instructions for use for the devices of the boso TM series familiarise
you with the use of the ambulatory blood pressure monitor and the associated
accessories. To be able to use all device settings, measurement protocols as
well as all evaluation options of the recorded blood pressure measurements,
you also need the medical software
boso XD profile manager. For instructions on how to use the software, please
refer to the separate software instructions for use.
Please familiarise yourself with both instructions for use before first use.
The manufacturer reserves the right to change the information in these
instructions for use without notice. The current version can be downloaded
from the website: https://www.boso.de/downloads
The instruction manual must be kept with the product to have it available at
all times.
In these instructions for use is used for an action of the user.
For help with commissioning, use or maintenance, please contact your
specialist dealer or the manufacturer (contact details on the back cover of
this instruction manual).
These instructions for use must be enclosed with the unit when it is sold.
This blood pressure monitor complies with the current European regulations and
the international standard IEC 80601-2-30:
“Particular requirements for the safety, including essential performance, of
automated non-invasive blood pressure measurement devices”.
The use of the device in pregnant women or in pre-eclampsia is not intended.
If the device is used for medical purposes (in accordance with the Medical
Devices Operator Ordinance), metrological checks must be carried out at
regular intervals (see section Test Instructions).
Intended purpose
Non-invasive recording of systolic and diastolic blood pressure values and pulse rate of individuals over a period of usually 24 hours.
Side-effects of the blood pressure measurement over 24 hours
Petechiae, bleeding or subcutaneous haematomas on the measuring arm can occur
with any blood pressure measurement, even if the cuff fits correctly.
Patient-dependent risk as a result of treatment with anticoagulants or
patients with coagulation disorders occurs regardless of the type of meter.
Always check if the patient has coagulation disorders or is being treated with
anticoagulants.
Scope of application
The blood pressure monitors of the boso TM series work according to the
oscillometric measuring principle. The device is intended for 24h measurement
in the patient’s usual environment and is to be used only under medical
supervision and after precise instruction by doctors or medical professionals.
The device is not suitable for infants,
newborns, or for unsupervised use with unconscious patients or patients with
impaired cognitive abilities.
Notes/safety instructions
Compression or a reduction in the cross-section of the air hose must be
avoided.
Excessively frequent measurements can lead to injuries by impairing the blood
flow.
The cuff must not be applied over wounds as this may cause further injury.
Make sure that the cuff is not applied to an arm whose arteries or veins are
or have been under medical treatment (e.g. shunt).
For women with a mastectomy, do not apply the cuff to the arm on the amputated
side of the body.
During the measurement, malfunctions may occur in medical devices that are
used simultaneously on the same arm.
The unit has no protection against possible influences from high frequency
(HF) surgical equipment.
Safety information
If liquid has been spilled on the unit, remove the batteries immediately and
send the unit to the customer service address (section Warranty
Conditions/Customer Service) for inspection.
Watch out for damage to the rechargeable batteries or batteries. Never use
damaged batteries.
There is a risk of strangulation with the shoulder strap and cuff tube.
A patient with impaired cognitive abilities may only use the device under
supervision.
Do not place the shoulder strap and cuff tubing around the patient’s neck.
The cuff hose is always laid under the clothing (also at night).
If the appliance is used with children, this must be done with special care
and under constant supervision.
Instruct the patient to turn off the device, remove the cuff and notify the
doctor if they experience pain, swelling, redness or numbness in the arm
around which the cuff is placed. (It is likely that the patient may experience
mild to moderate discomfort when having their blood pressure measured.)
The measuring process can be interrupted at any time by pressing one of the
buttons. This deflates the cuff and the device can be removed.
Instruct the patient to protect the unit from liquid penetration.
In particular, the patient must be advised not to wear the device while
showering.
If the unit has been exposed to moisture or if liquid has entered during
cleaning/use, it must no longer be fitted to patients.
Medical electrical equipment is subject to special precautions regarding
electromagnetic compatibility and must be installed and commissioned in
accordance with the EMC instructions section.
Maintenance work on this unit must be carried out by trained and authorised
personnel.
Due to the risk of strangulation by the tube and cuff, the device must not be
within the reach of unsupervised children or used on unsupervised patients
with impaired cognitive abilities or on patients under anaesthesia.
The device must not be used by children without supervision.
Do not use the device near babies. This can lead to accidents or damage.
The manufacturer is only responsible for the impact on the safety, reliability
and performance of the device if:
Assembly, extensions, new settings, changes or repairs have been carried out
by persons authorised by him.
The appliance is used in accordance with the instructions for use.
Do not start the machine without putting the cuff on.
The appliance contains small parts; these can cause a choking hazard if
accidentally swallowed by babies.
The performance of the unit may be affected by excessive temperature, humidity
or altitude.
Commissioning
Before you start working with boso TM series devices, you should charge the
batteries supplied. To do this, proceed as described in section ”Changing and
charging the batteries”. Then install the boso XD profile manager. This
software enables the programming of the blood pressure monitor and the
evaluation of the stored data.
Selection and connection of the cuff on boso TM series devices
Cuff selection
Only original cuffs CA91, CA91R, CA92, CA93 and CA94 must be used.
The cuff must be selected according to the printed arm circumference.
Connecting the cuff
The air connection plug of the cuff tubing is screwed directly into the air
connection socket of the blood pressure monitor (see Fig. 4).
Attachment of the protective covers (optional)
If necessary, you can also use protective covers (see section Accessories) to
protect against soiling.
Put on the protective covers as shown below:
Pull the cuff through the tab of the protective cover.
Attach the protective cover to the cuff with the Velcro fasteners on the
inside.
Care instructions for protective covers: machine wash at max. 60 °C
Attaching the cuff
Place the cuff on the unclothed upper arm so that the white marking lies on
the arteria brachialis. For most people, blood pressure is higher on the left
arm, so blood pressure is measured on the left arm. If the blood pressure is
higher on the right arm, it should be measured on the right arm.
The cuff should be approx. 2-3 cm above the crook of the elbow. The cuff
should not be too tight, about two fingers should still fit between the arm
and the cuff.
After the end of the measurement, the blood circulation must not be impaired
by the cuff. Place the cuff tube over the shoulder (see fig. 4). Fix the cuff
tube to the shoulder with tape. The blood pressure monitor is carried in the
pouch either on an existing belt or with the included carrying strap.
Carrying out measurements with boso TM series devices
After the cuff has been properly put on, a test measurement can be triggered
on the boso TM series device by means of the START/STOP button (measurement is
only displayed if the device has been programmed accordingly). If the display
is off, activate it by pressing any button. If this measurement is successful,
the automatic interval function (see below) can be started. The sample
measurement is included in the evaluation.
Please note that the oscillometric measurement method can lead to measurement
inaccuracies in some types of patients. Persons with cardiac arrhythmia,
arteriosclerosis, circulatory disorders, diabetes or pacemaker wearers should
have a comparative measurement taken with an auscultatory device before
starting the measurement. This also applies to women during pregnancy.
External disturbing influences, such as movements of the measuring arm,
disturbing vibrations, e.g. due to driving or the use of public transport
during the measurement, can lead to incorrect measurements. For this reason,
the record kept by the patient must be viewed and included in the evaluation
of the measurement results.
Starting the automatic interval
To start the automatic interval, press and hold the black AUTO button until “
” appears in the blood pressure monitor display and is acknowledged by a short
signal tone (after approx. 5 seconds).
If the device is operated in “Sleep button” mode, the patient must press the
black AUTO button before going to sleep. The display shows ” ” for automatic
and “ ” for sleep mode. After getting up, the black AUTO button must be
pressed again. The “ ” in the display disappears.
Automatic adjustment of the pump-up level (only in automatic interval mode)
The boso TM series device automatically pumps up to the required pressure
level. If this inflation height is not sufficient, the unit automatically
inflates again approx. 60 mmHg above the original inflation level.
Limiting the maximum pumping level
The device of the boso TM series has the option of limiting the inflation
height. For the corresponding procedure, please refer to the boso XD profile
manager instructions for use. Performing a manual measurement A manual
measurement can be started by the patient at any time in addition to the
automatic measurements. This can be sensible following physical or emotional
stress. To do this, activate the display, then press the white START/STOP
button.
Aborting measurements
To cancel the measurement, the white START/STOP button on the boso TM series
must be pressed.
If the measurement is to be taken at a later time, a manual measurement can be
started at any time using the white START/ STOP button.
Ending the measurement and transferring the measurement data
As soon as the device is removed from the patient after the 24-hour
measurement has been completed, the automatic function must be switched off.
To do this, press and hold the black AUTO button until the ” ” disappears from
the blood pressure monitor display (approx. 5 seconds).
Then connect the boso TM series device to the computer using the USB
connection cable. Transfer the data according to the instructions for use boso
XD profile manager.
After transferring the measured data, it his highly suggested to delete the
data memory.
Changing the batteries
We recommend changing the used battery pack after each 24-hour measurement and
replacing it with the freshly charged battery pack. To prevent data loss, the
data stored in the boso TM series device is buffered by an internal battery.
This battery is automatically charged via the battery pack. With a fully
charged battery pack, the data remain
stored for approx. 10 days. To fully charge the battery when using the unit
for the first time, switch on the unit with fully charged batteries for
approx. 24 hours.
To change the batteries, proceed as follows (see fig. 5):
Open the battery compartment cover (step 1)
Remove the empty batteries (step 2) and reinsert the charged batteries (step
3) (note polarity!) (Step 4))
Close the battery compartment cover (Step 5)
Insert the rechargeable batteries into the charger and plug it into
a power socket. If the blue LED lights up, the battery packs are being
charged. The charging process takes roughly 11 hours for a fully charged
battery pack. After 13 hours, the charger automatically switches off the
charging process.
Important note regarding battery charging
To ensure proper functioning of the boso TM series device for more than 24
hours, use only rechargeable batteries with the following ratings: min. 1900
mAh; 1.2 V; NiMH or batteries (Type AA 1.5 V).
In addition to the two rechargeable batteries required for the power supply,
the device of the boso TM series also contains an internal battery to save the
programme setting in the device.
To avoid loss of programming and stored readings when the internal battery is
discharged, please observe the following procedure:
Insert charged batteries into the unit even when it is not in use.
The state of charge of the internal battery is thus constantly kept at a high
level. If the power supply to the internal battery is interrupted, the
settings of the boso TM series device will be lost after approx. 10 days.
Before placing the unit on a patient, please replace the batteries in the unit
with a set of freshly charged batteries.
If the batteries are short-circuited, they may become hot and cause burns and
scorching damage to the unit.
Do not touch the batteries and the patient at the same time.
Longer storage of the device
If the unit will not be used for a long time (4 weeks or more), remove the
batteries to prevent possible damage due to leakage.
Before the device is then attached to a patient again, the internal battery
must be charged and the device reprogrammed.
Insert freshly charged batteries.
Please leave the battery packs in the device for two hours. The internal
battery is recharged during this time.
Reprogram the device.
Before attaching the device to a patient, replace the batteries with a set of
freshly charged ones.
Error messages
Error code | Cause | Remedy |
---|---|---|
0:00 | Time sets to 0:00 when battery is replaced | Device must be reprogrammed. |
E03
E90| Zero point adjustment not possible| Deflate cuff completely.
E04| Empty battery| Charge or replace batteries.
E05| Leakage| Disconnect the cuff from the device and reconnect. If the error
occurs repeatedly, contact your sales partner.
E06| Pressure above 299 mmHg| The arm must be kept still during measurement.
E07| User abort via START/STOP button|
E08
E10| No or non-interpretable oscillations| The arm must be kept still during
measurement.
E09| Error of the activity sensor| Remove and re-insert the battery packs.
E20| Pulse < 30 or > 200| Check the position and fit of the cuff.
E21
E22| No interpretable oscillations in the area of diastole (E21) or systole
(E22)
E23| Systole-diastole < 10 or > 150 mmHg
E30| Measuring time longer than 180 seconds| Contact your sales partner.
E31| Deflation longer than 90 seconds| Contact your sales partner.
E48| Pulse cannot be measured| The arm must be kept still during measurement.
E52| Storage error| Contact your sales partner.
E91| Pressure inside the cuff too high / maximum pressure set too low| Set a
higher maximum pressure. The arm must be kept still during measurement.
After use
Cleaning and disinfection
To clean the boso TM series device and the cuff, please use a soft cloth that
may be moistened with soapy water.
For the protective covers: machine wash at max. 60 °C.
Never use solvents, benzine, spirits or abrasive cleaners for cleaning!
Disinfection:
For disinfection by wiping (exposure time at least 5 minutes) the device, we
recommend the disinfectant Antifect Liquid (Schülke & Mayr). We recommend
using spray disinfection for disinfecting the cuff. Especially if the device
is used by several patients, make sure that the cuff is cleaned and
disinfected regularly.
Disposal information
Used batteries and rechargeable batteries must not be disposed of in household
waste. You can take them to a collection point for used batteries or to
hazardous waste. Please contact your municipality for more information.
Customer information on the recycling of commercial waste electrical
equipment
1) Purpose
Based on the EU Directive 2012/19/EU, the German implementation of the
ElektroG was revised in 2021. The amended ElektroG3 came into force on
01.01.2022. The background to this is to steadily improve the collection rates
of e-waste and to achieve a rate of > 65 %. In this document we inform you
about the return option we have created for your waste electrical equipment
(WEEE) from the commercial sector.
2) Manufacturer’s declaration on the return option
For commercially used appliances, a collection can take place at the end of
the life cycle via our recycling partner (see point 3). For this purpose, a
notification must be sent to our return partner or to “Bosch+ Sohn GmbH u. CO.
KG”, stating the articles and their number. The customer then receives an
offer from our recycling partner for coordinated collection at the point of
generation. The customer is free to opt for this collection or to take the
WEEE to his own disposal system and to comply with the corresponding
obligations.
3) Authorised recycling partner
The authorised recycling company for the company “Bosch+Sohn
GmbH u. CO. KG” is:
WEEE Return GmbH
Lahnstraße 31
12055 Berlin
4) Contact details for the return option
A return can be registered by telephone or by e-mail.
The following options are available to the waste owner:
Phone: +49 (0) 74 77 92 75-0
E-Mail: zentrale@boso.de
Obligation to report incidents
A serious incident shall be reported to the manufacturer and to the
responsible authority of the Member State in which the user and/or patient is
established.
A “serious incident” means an incident that directly or indirectly had, could
have had, or may have had any of the following consequences: the death of a
patient, user, or other person; the temporary or permanent serious
deterioration of the health of a patient, user, or other person; a serious
public health hazard. Please send reports of serious incidents to:
E-Mail: vigilanz@boso.de
Fax: +49 (0) 74 77 92 75-56
Warranty conditions/customer service
This product is covered by a 2-year factory warranty from the date of
purchase.
The date of purchase must be proven by invoice. Within the warranty period,
defects resulting from material or manufacturing faults will be repaired free
of charge. Guarantee services do not extend the guarantee period for the
entire appliance, but only for the replaced components. The warranty does not
cover wear and tear (e.g. cuff), transport damage or any damage caused by
improper handling (e.g. failure to follow the instructions for use) or by
tampering by unauthorised persons. The guarantee does not give rise to any
claims for damages against us. The buyer’s statutory claims for defects
pursuant to Art. 437 of the German Civil Code (BGB) shall not be restricted.
In the event of a claim under the warranty, the unit must be sent together
with the original proof of purchase to: BOSCH + SOHN GmbH u. Co. KG
Bahnhofstr. 64, 72417 Jungingen, Germany
Maintenance work on this unit must be carried out by trained and authorised
personnel.
The unit must not be modified without the permission of the manufacturer.
Accessories
Please use only the accessories recommended by the manufacturer.
Cuffs
Size M| CA91| 20 – 31 cm| 259-4-400
Size M (right)| CA91R| 20 – 31 cm| 259-4-440
Size L| CA92| 28 – 38 cm| 259-4-410
Size XL| CA94| 36 – 50 cm| 259-4-430
Size S| CA93| 15 – 22 cm| 259-4-420
Protective covers (10 pieces)
Size M left and right | 259-7-400 |
---|---|
Size L | 259-7-410 |
Size XL | 259-7-430 |
Size S | 259-7-420 |
5x size M and 5x size L | 259-7-405 |
Other accessories
Charger | 535-7-130 |
---|---|
NiMh batteries (2 pieces, Mignon) | 535-7-131 |
Waist bag with carrying strap | 515-7-116 |
Technical specifications
Product: | Blood pressure measuring device for 24-hour measurement |
---|---|
Type designation: | See equipment labelling |
Rated voltage: | 2x 1.5 V DC or 2x 1.2 V DC |
Power supply: | 2x NiMh batteries (Mignon) |
Measuring range: | Systole: 60 – 280 mmHg |
Diastole: 30 – 160 mmHg
Pulse: 30 – 200 pulse/min
Maximum deviation of the cuff pressure:| ±3 mmHg or 2 % of reading (whichever
is greater)
Maximum deviation of the pulse reading:| ±5 %
Measurement storage:| 600 measurements
Operating conditions:| +10 °C to +40 °C
rel. humidity 30 – 85 % (non-condensing)
Air pressure 700 – 1060 hPa
Storage conditions:| +20 °C to +60 °C
10 – 95 % rel. humidity
Air pressure 700 – 1060 hPa
Weight:| 135 grammes without batteries
Dimensions (W x H x D):| 66 mm x 25 mm x 95 mm
Typical battery life:| 1000 charging cycles (depending on pump-up level +
frequency of use)
Expected service life of the unit:| 10 years
Expected service life of the cuff:| 10,000 measuring cycles
Clinical test:| The measurement accuracy meets the requirements of ISO 81060-2
Test instruction for metrological checks
A) Functional check
A functional check of the unit can only be carried out on humans or with a
suitable simulator.
B) Checking for tightness of the pressure circuit and deviation of the
pressure indication
Remove the batteries. Then create a test setup as shown in Fig. 6.
Immediately after reinserting the batteries, press and hold the white
START/STOP button. The white START/STOP button must be held down until a “0”
appears in the display of the boso TM series device.
Then carry out the test for deviation of the pressure indication and tightness
of the pressure circuit (observe the setting time of the cuff – at least 30
seconds) in the usual way. To return to the measuring mode after the test has
been completed, the batteries must be removed again and reinserted.
C) Securing
For securing, the housing halves (upper and lower part) are connected with a
securing tag.
EMC information
Medical electrical equipment is subject to special precautions regarding EMC
and must be installed and commissioned in accordance with the guidelines given
below.
Portable and mobile HF devices (e.g. mobile phones) can affect medical
electrical equipment.
The use of third-party accessories (not original boso parts) may result in
increased emissions or reduced immunity of the device.
Guidelines and manufacturer’s declaration – Electromagnetic emissions
The boso blood pressure monitor is intended for operation in the
electromagnetic environment specified below. The customer or the user of the
boso blood pressure monitor should ensure that it is used in such an
environment.
Emission measurements | Compliance | Electromagnetic Environment Guidelines |
---|---|---|
HF emissions according to CISPR 11 | Group 1 | The boso blood pressure monitor |
uses HF energy exclusively for its internal function. Therefore, its HF
emission is very low and it is unlikely to interfere with nearby electronic
equipment.
HF emissions according to CISPR 11| Class B| The boso blood pressure monitor
is intended for use in all establishments, including domestic establishments
and those directly connected to a public supply network which also supplies
buildings used for domestic purposes.
Harmonics according to IEC 61000-3-2| not applicable
Voltage fluctuations/ flicker according to IEC 61000-3-3
Guidelines and manufacturer’s declaration – Electromagnetic interference
immunity
The boso blood pressure monitor is intended for operation in the
electromagnetic environment specified below. The customer or the user of the
boso blood pressure monitor should ensure that it is used in such an
environment.
Interference immunity tests| IEC 60601-test level| Compliance level|
Electromagnetic Environment -Guidelines
---|---|---|---
Static electricity discharge (ESD) according to IEC 61000-4-2| ±6 kV contact
discharge ±8 kV air discharge| ±6 kV contact discharge ±8 kV air discharge|
Floors should be wood, concrete or ceramic tiles.
Fast transient electrical disturbance/bursts according to IEC 61000-4-4| ±2 kV
mains cables
±1 kV for input and output lines| Not applicable
with synthetic material,| If the floor is covered
the relative humidity must be at least 30 %.
Surges according to IEC 61000-4-5| ±1 kV push-pull voltage
±2 kV common-mode voltage| Not applicable|
Voltage dips, short-time interruptions and fluctuations of the supply voltage
according to IEC 61000-4-11| < 5 % U,. for 1/2 period
(> 95 % dip)
40 % UT for 5 periods
(65 % dip)
70 % Li, for 25 periods
(30 % dip)
< 5 % U,. for 5 s
(> 95 % dip)| Not applicable|
Magnetic field at the supply frequency (50/60 Hz) according to IEC 61000-4-8|
3 Alm| 3 Alm|
NOTE: U is the AC mains voltage before the test level is applied.
Guidelines and manufacturer’s declaration – Electromagnetic emissions
The boso blood pressure monitor is intended for operation in the
electromagnetic environment specified below. The customer or the user of the
boso blood pressure monitor should ensure that it is used in such an
environment.
Interference immunity tests| IEC 60601-test level| Compliance level| el-magn.
Environment – Guidelines recommended safe distance
---|---|---|---
| | | Portable and mobile radios are used at no closer distance from the
boso blood pressure monitor, including the leads, than the recommended
protective distance calculated using the equation appropriate to the
transmitting frequency: d = 1.2 P
Conducted HF disturbances according to IEC 61000-4-6| 3 Veff
150 kHz – 80 MHz| 3 Vef
radiated HF disturbances according to IEC 61000-4-3| 3 V/m
80 kHz – 2.5 GHz| 3 V/m| d = 1.2 P for 80 MHz – 800 MHz d = 2.3 P for 800 MHz
– 2.5 GHz
With P as the nominal power of the transmitter in watts (W) according to the
transmitter manufacturer’s specifications and d as the recommended protective
distance in metres (m). The field strength of stationary radio transmitters is
lower than the compliance levelb at all frequencies according to an on-site
investigationa.
Interference is possible in the vicinity of units with this symbol.
NOTE 1: For 80 MHz and 800 MHz, the higher value applies.
NOTE 2: These guidelines may not apply in all situations. The propagation
of electromagnetic waves is influenced by absorption and reflection from
buildings, objects and people.
a The field strength of stationary transmitters, such as base stations of
radio telephones and land mobile services, amateur stations, AM and FM radio
and television transmitters, cannot theoretically be predicted exactly. To
determine the electromagnetic environment due to stationary HF transmitters, a
site survey is recommended. If the determined field strength at the location
of the boso blood pressure monitor exceeds the compliance level specified
above, the boso blood pressure monitor must be observed with regard to its
normal operation at any application location. If unusual performance
characteristics are observed, it may be necessary to take additional measures,
such as reorienting or relocating the boso blood pressure monitor. b Over the
frequency range from 150 kHz to 80 MHz, the field strength is less than 3 V/m.
Recommended safe distances
between portable and mobile HF communication devices and the boso blood
pressure monitor. The boso blood pressure monitor is intended for operation in
an electromagnetic environment in which radiated HF disturbances are
controlled. The customer or the user of the boso blood pressure measuring
device can help prevent electromagnetic interference by maintaining minimum
distances between portable and mobile HF communications equipment
(transmitters) and the boso blood pressure measuring device as recommended
below according to the maximum output power of the communications equipment.
Rated power of the transmitter| Protective distance according to transmission
frequency m
---|---
W| 150 kHz to 80 MHz
d = 1.2 √P| 80 MHz to 800 MHz
d = 1.2 √P| 800 MHz to 2.5 GHz
d = 2.3 √P
0.01| 0.12| 0.12| 0.23
0.01| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters whose rated power is not specified in the above table, the
distance can be determined using the equation associated with the respective
column, where P is the rated power of the transmitter in watts (W) as
specified by the transmitter manufacturer.
NOTE 1: An additional factor of 10/3 was used to calculate the
recommended separation distance of transmitters in the frequency range from 80
MHz to 2.5 GHz to reduce the likelihood that a mobile/portable communication
device inadvertently introduced into the patient area would cause
interference.
NOTE 2: These guidelines may not apply in all situations. The propagation
of electromagnetic waves is influenced by absorption and reflection from
buildings, objects and people.
BOSCH + SOHN GmbH u. Co. KG
Bahnhofstraße 64
72417 Jungingen, Germany
T + 49 (0) 74 77 92 75 – 0
F + 49 (0) 74 77 10 21