Jamar® Smart Hand Dynamometer
USER’S GUIDE
Download the Free App
OVERVIEW
The Jamar Hand Dynamometer has been used for over 40 years to measure grip strength. Now Patterson Medical has redesigned it to make it easier to use with the introduction of the Jamar Smart Hand Dynamometer.
FEATURES
Precision Electronic Load Cells eliminate the possibility of oil leakage
that sometimes occurred with hydraulic units. The unit allows measurement from
0 to 200 lbs. (90Kg).
Easy-to-Read Display helps configure a specific test protocol, and
provides step-by-step prompts throughout the testing process. The display will
always show the next step in the test, even for rapid exchange protocols. In
addition, the Jamar Smart Hand Dynamometer saves the previous test, saving
time if the same test protocol is used each time. Up to 5 tests/trials can be
conducted with each hand without reset.
AVERAGE, STANDARD DEVIATION, and COEFFICIENT OF VARIATION are
automatically calcu-lated. Hand to hand rapid exchange tests are separately
calculated for each hand.
Individual test readings and all calculations can be reviewed multiple times.
The display shows which reading is being displayed to reduce error.
These features, along with the rugged design of the new Jamar Smart Hand
Dynamometer, help complete each test more quickly and efficiently resulting is
less chance for error in performing calcula-tions and transposing data.
DOWNLOAD THE FREE APP
The new Jamar Smart Hand Dynamometer is the industry’s first dynamometer to
offer wireless communication to your mobile device. Go to your favorite App
store and search for “Jamar Smart”. Download the Free App to access many new
capabilities.
CONTROL PANEL
(only active when the App is not being used)
SET-UP/TEST
Power ON
Press the [ON/OFF] key on the face of the Jamar Smart Hand Dynamometer. The
display will appear as shown below, with the display showing the most recent
test protocol. The default test protocol is “R” (Right) “3”, which means the
test protocol is set for the Right Hand with three (3) repeats.
CHOOSING THE HAND TO BE TESTED
Press the [RESET] key to clear any test which might be in progress. Then press the [SELECT TEST] key to step through the available choices. Press [SELECT TEST] for these choices:
Display | Test |
---|---|
1 | Standard Test Single Hand |
L1 | Standard Test Left Hand |
R1 | Standard Test Right Hand |
LR1 | Standard Test Left and Right Hand |
LR 1.5R x 10 | Rapid Exchange Grip Test 1.5 second intervals |
LR 3.0 R x 10 | Rapid Exchange Grip Test 3.0 second intervals |
CHANGING THE NUMBER OF TRIALS
Press the [RESET] key to clear any test which might be in progress. Then press the [# OF TRIALS] key to step through the available choices. These choices are:
Test | Choices |
---|---|
Standard Hand Test Either Hand, Left, Right, or Both | 1 to 5 trials |
Rapid Exchange Grip Test 1.5 or 3.0 Intervals | 10 or 20 total trials |
Practice changing the hand selection and number of grip repeats to see how
these functions work.
Once test protocol is changed, the Jamar Smart Hand Dynamometer retains the
current protocol until changed.
NOTE: If the batteries are allowed to run down after the Low Battery
indicator comes on, or if they are removed from the instrument for a period of
more than approximately one (1) minute, the instrument may return to the
default factory test protocol, “R 3”. If this happens, simply replace the
batteries and reset the desired test protocol.
SELECTING POUNDS OR KILOGRAMS
ADJUSTING GRIP POSITION
The grip handle may be placed in any of five (5) positions to accommodate the size and comfort of the user’s hand. Push the lower end of the handle so that the slotted portion rotates away from the lower shaft (See diagram below). Making sure not to drop the handle, allow it to separate from the top shaft. Determine the appropriate grip position and replace the top part of the handle on the top shaft. Rotate the lower part of the handle onto the lower shaft until it clicks into place. The special ball detent on the lower part of the handle helps prevent the handle from separating from the shaft during normal use.
INSTALLING AND REPLACING BATTERIES
Turn the instrument so the Key Pad Operator faces front and the back of the Key Pad Housing faces toward you. Open the compartment cover and install two (2)AAA batteries. (Quality, alkaline or re-chargeable batteries are recommended). Replace the cover after installation. Press the [ON/OFF] key on the face of the Jamar Smart Hand Dynamometer to ensure battery installation is correct. As the batteries approach the end of their life cycle, the “BATT” indication will be displayed on the instrument display. Batteries should be replaced as soon as this is displayed to ensure proper unit function.
PERFORMING GRIP STRENGTH MEASUREMENTS
-
[LB/KG]: The Jamar Smart Hand Dynamometer is originally set to display in LBs. To change to KG, open up the battery compartment and move the frequency switch down to KG.
-
Adjust the grip handle to best suit patient comfort and test requirement (See Adjusting Grip Posi-tion, page 7).
-
Press the [ON/OFF] key to start the instrument.
-
Determine which hand will be used. Using the [SELECT TEST] key, scroll to the “L” (Left) or “R” (Right) icon to select the hand to be tested.
-
To select the number of test repetitions (up to 5) to be recorded for that hand, press the [# OF TRIALS] key. Each of the five numbers (1 through 5) will be displayed on the upper line of the display. Stop scrolling on the number which represents the number of tests to be taken.
-
Have the patient grasp the Jamar Smart Hand Dynamometer gently so that the palm fits comfort-ably to the rear of the instrument and the fingers curl around the adjustable grip at the front of the instrument.
-
Press the [TEST] key to begin the test. Under the present test protocol, start with the first test trial and the number “1” will appear and flash at the top of the display, indicating that the first test for the chosen hand is ready. Have the patient squeeze the grip and be ready to enter the second trial (if test protocol setting has more than 1 trial).
-
When the first test is completed, press the [TEST] key one time to scroll to the next test. Repeat the number of trials to complete the test cycle for the chosen hand.
Example: The right hand is chosen to perform four tests. Press the [SELECT TEST] key and scroll to the “R” (Right) icon. Press the [# OF TRIALS] key until the number “4” is displayed. Press the [TEST] key to begin the test. “R” and “1” will be displayed on the upper line. Have the patient perform the first test. Press the [TEST] key to move the display to “R2”. Have the patient perform the second test. Press the [TEST] key after each test until all four tests have been performed. -
When testing is completed for the chosen hand and the required data is recorded, press the [RESET] key to clear the previous test data. Repeat steps 4 through 9 for the remaining hand if desired.
-
To alternate left and right hand tests (i.e., L1 – R1, L2 – R2, L3 – R3, etc) press [SELECT TEST] until both the “L” and “R” are displayed. The instrument will automatically alternate “L” and “R” hand test cycles.
-
Press the [ON/OFF] key to turn the instrument OFF when testing is complete and return it to a safe storage location.
RECORDING GRIP STRENGTH MEASUREMENTS
Upon completion of the desired tests for a patient’s left and/or right hand,
the therapist can select one of two methods for recording the data obtained by
those tests.
- After each individual test, the therapist may choose to record the data displayed on the instrument and perform subsequent calculations at a later date.
- After all the test trials are completed, the therapist may choose to review the test results stored in the Jamar Smart Hand Dynamometer. All the test and calculation results can be viewed on the display by repeatedly pressing the [TEST] key. All results will be displayed in the sequence of: Test Trial 1, Test Trial 2…Test Trial N, AVG, STDEV, CV. There will be an indicator icon shown in the display for each respective result.
CALCULATIONS
Three important calculations are performed automatically by the Jamar Smart
Hand Dynamometer: Test Average, Standard Deviation, and Coefficient of
Variation.
Test Average• Each reading taken for the left hand or right hand is
recorded in memory for that hand. When the test is completed, the Therapist
uses the [TEST] key, located in the lower left of the keypad, to review the
average measure of the multiple test trials.
Standard Deviation: The Jamar Smart Hand Dynamometer calculates the
standard deviation of reading taken. Use the [TEST] key to review the standard
deviation of the multiple test trials.
Coefficient of Variation: To obtain disbursement data, the instrument
divides the standard deviation by the average as shown below. Use the [TEST]
key to scroll to the “CV” icon to display the result.
NORMATIVE GRIP STRENGTH DATA:
| | Males| | Females| | Males| | Females|
---|---|---|---|---|---|---|---|---|---
Age| Hand| Mean (lbs)| SD| Mean (lbs)| SD| Mean (kg)| SD| Mean (kg)| SD
6-7| R| 32.5| 4.8| 28.6| 4.4| 14.7| 2.2| 13.0| 2.0
| L| 30.7| 5.4| 27.1| 4.4| 13.9| 2.4| 12.3| 2.0
8-9| R| 41.9| 7.4| 35.3| 8.3| 19.0| 3.4| 16.0| 3.8
| L| 39| 9.3| 33| 6.9| 17.7| 4.2| 15.0| 3.1
10-11| R| 53.9| 9.7| 49.7| 8.1| 24.4| 4.4| 22.5| 3.7
| L| 48.4| 10.8| 45.2| 6.8| 22.0| 4.9| 20.5| 3.1
12-13| R| 58.7| 15.5| 56.8| 10.6| 26.6| 7.0| 25.8| 4.8
| L| 55.4| 16.9| 50.9| 11.9| 25.1| 7.7| 23.1| 5.4
14-15| R| 77.3| 15.4| 58.1| 12.3| 35.1| 7.0| 26.4| 5.6
| L| 64.4| 14.9| 49.3| 11.9| 29.2| 6.8| 22.4| 5.4
16-17| R| 94| 19.4| 67.3| 16.5| 42.6| 8.8| 30.5| 7.5
| L| 78.5| 19.1| 56.9| 14| 35.6| 8.7| 25.8| 6.4
18-19| R| 108| 24.6| 71.6| 12.3| 49.0| 11.2| 32.5| 5.6
| L| 93| 27.8| 61.7| 12.5| 42.2| 12.6| 28.0| 5.7
20-24| R| 121| 20.6| 70.4| 14.5| 54.9| 9.3| 31.9| 6.6
| L| 104.5| 21.8| 61| 13.1| 47.4| 9.9| 27.7| 5.9
25-29| R| 120.8| 23| 74.5| 13.9| 54.8| 10.4| 33.8| 6.3
| L| 110.5| 16.2| 63.5| 12.2| 50.1| 7.3| 28.8| 5.5
30-34| R| 121.8| 22.4| 78.7| 19.2| 55.2| 10.2| 35.7| 8.7
| L| 110.4| 21.7| 68| 17.7| 50.1| 9.8| 30.8| 8.0
35-39| R| 119.7| 24| 74.1| 10.8| 54.3| 10.9| 33.6| 4.9
| L| 112.9| 21.7| 66.3| 11.7| 51.2| 9.8| 30.1| 5.3
40-44| R| 116.8| 20.7| 70.4| 13.5| 53.0| 9.4| 31.9| 6.1
| L| 112.8| 18.7| 62.3| 13.8| 51.2| 8.5| 28.3| 6.3
45-49| R| 109.9| 23| 62.2| 15.1| 49.8| 10.4| 28.2| 6.8
| L| 100.8| 22.8| 56| 12.7| 45.7| 10.3| 25.4| 5.8
50-54| R| 113.6| 18.1| 65.8| 11.6| 51.5| 8.2| 29.8| 5.3
| L| 101.9| 17| 57.3| 10.7| 46.2| 7.7| 26.0| 4.9
55-59| R| 101.1| 26.7| 57.3| 12.5| 45.9| 12.1| 26.0| 5.7
| L| 83.2| 23.4| 47.3| 11.9| 37.7| 10.6| 21.5| 5.4
60-64| R| 89.7| 20.4| 55.1| 10.1| 40.7| 9.3| 25.0| 4.6
| L| 76.8| 20.3| 45.7| 10.1| 34.8| 9.2| 20.7| 4.6
65-69| R| 91.1| 20.6| 49.6| 9.7| 41.3| 9.3| 22.5| 4.4
| L| 76.8| 19.8| 41| 8.2| 34.8| 9.0| 18.6| 3.7
70-74| R| 75.3| 21.5| 49.6| 11.7| 34.2| 9.8| 22.5| 5.3
| L| 64.8| 18.1| 41.5| 10.2| 29.4| 8.2| 18.8| 4.6
75+| R| 65.7| 21.| 42.6| 11| 29.8| 9.5| 19.3| 5.0
| L| 55| 17| 37.6| 8.9| 24.9| 7.7| 17.1| 4.0
CALIBRATION
The unit should be returned to Patterson Medical after 12 months of use for recalibration. See below for contact details.
CARE AND STORAGE
The Jamar Smart Hand Dynamometer is an electronic instrument and should be
cared for in a manner suitable for its function and components. Store the
instrument when not in use in a safe location to prevent inadvertent damage.
Use a clean, damp cloth and mild detergent for cleaning.
When not in use, the instrument should be stored in the black case provided.
Care should be taken, however, to prevent undo stress to the grip posts, as
this may damage the post mountings or electronic components.
SERVICE AND WARRANTY
The Jamar Smart Hand Dynamometer is a fully self-contained unit which should provide many years of accurate, dependable service to the user, and contains a (1) year warranty from its date of purchase. Should you experience any questions or problems regarding this instrument please contact Performance Health at 800-323-5547.
Quick Start Guide
- Press the [ON/OFF] key to start the instrument.
- Determine which hand will be used. Using the [SELECT TEST] key, scroll to the “L” (Left) or “R” (Right) icon to select the hand to be tested.
- To alternate left and right hand tests (i.e., L1 – R1, L2 – R2, L3 – R3, etc.) press [SELECT TEST] until both the “L” and “R” are displayed. The instrument will automatically alternate “L” and “R” hand test cycles.
- To select the number of test repetitions (up to 5) to be recorded for that hand, press the [# OF TRIALS] key. Each of the five numbers (1 through 5) will be displayed on the upper line of the display. Stop scrolling on the number which represents the number of tests to be taken.
- Have the patient grasp the Jamar Smart Hand Dynamometer gently so that the palm fits comfortably to the rear of the instrument and the fingers curls around the adjustable grip at the front of the instrument.
- Press the [TEST] key to begin the test. Under the present test protocol, start with the first test trial and the number “1” will appear and flash at the top of the display, indicating that the first test for the chosen hand is ready. Have the patient squeeze the grip and be ready to enter the second trial (if test protocol setting has more than 1 trial).
- When the first test is completed, press the [TEST] key one time to scroll to the next test. Repeat the number of trials to complete the test cycle for the chosen hand.
- When testing is completed for the chosen hand and the required data is recorded, press the [RESET] key to clear the previous test data. Repeat steps 2 through 9 for the remaining hand if desired.
- After all the test trials are completed, the test and calculation results can be viewed on the display by repeatedly pressing the [TEST] key. All results will be displayed in the sequence of: Test Trial 1, Test Trial 2…Test Trial N, AVG, STDEV, CV. There will be an indicator icon shown in the display for each respective result.
- Hold the [ON/OFF] button for two seconds to turn the instrument OFF when testing is com-plete and return it to a safe storage location.