lbx instruments V10 Oil-Free Vacuum Pump User Manual

June 3, 2024
lbx instruments

lbx instruments V10 Oil-Free Vacuum Pump

Important notice

Please read the User Manual carefully before use, and follow all operating and safety instructions! This instrument is designed for laboratory usage only. Please read this manual carefully before installing or operating this equipment. The instrument shall not be modified in any way. Any modification will void the warranty and may result in potential hazard. We are not responsible for any injury or damage caused by any non-intended purposes and modifying the instrument without authorization.

  1. Check the voltage specified on the name-plate and ensure it matches the line voltage in your location.
  2. The pump is not corrosion-proof. Please do not operate it with corrosive gases directly.
  3. Install the instrument in a clean, dust-less and ventilated area under 40℃.
  4. The ProteTM filter cartridge can prevent a large amount of liquid from overflowing into the pump. Please ensure the liquid level of flask not exceed safety level. Failure to comply can result in serious damage to the pump and void the warranty. If it occurs, please contact the distributor immediately for service.
  5. The filter cartridge is used to absorb moisture and dust. Please replace the cartridge to maintain proper pumping efficiency when it is saturated.
  6. To avoid damaging the pump easily, please vent the pump before starting the pump.
  7. Don’t incline the pump during work to prevent the ProteTM filter cartridge from blocking the air inlet.
  8. This pump has a thermal protection device that automatically shuts-off when it becomes overheated.
  9. The temperature of the surface of pump is very high after use or during work, please don’t touch it to avoid being burnt.
  10. Do not use any lubricant, which may damage the pump.
  11. Never spray flammable or toxic materials.
    If the supply cord is damaged, Please contact the manufacturer or your service agent for replacement to avoid hazard.

Service

In order to guarantee this equipment Works safely and efficiently, it must have a regular maintenance. In case of any faults, do not try to repair it yourself. If help is needed, you can always contact your dealer or Labbox via www.labbox.com
Please provide the customer care representative with the following information:

  • Serial number
  • Description of problem
  • Your contact information

Assembly & Disassembly diagram

Position Designation
1 On/Off switch
2 Muffler (Air outlet)
3 Air Inlet
4 Vacuum Gauge
5 ProteTM filter cartridge
6 Moisture trap
7 Drain valve (Regulator)

Operation

Operating procedure

  1. Press on/off button to activate the pump after making sure all tubing are well connected.
  2. To setup vacuum by adjusting the regulator .

Warning: When pump is used for water filtration, please ensure the liquid level of flask not exceed safety level. Failure to comply could result in serious damage to the pump and void the warranty. If it occurs, please contact the distributor immediately for service.

Maintenance

How to replace cartridge:

  1. Please replace the cartridge when it is saturated with moisture or particle.
  2. Please remove the moisture trap 6 by rotating it counterclockwise.
  3. Replace the cartridge by the same way.

Warranty

This instrument is warranted to be free from defects in materials and workmanship under normal use and service, for a period of 48 months from the date of invoice. The warranty is extended only to the original purchaser. It shall not apply to any product or parts which have been damaged on account of improper installation, improper connections, misuse, accident or abnormal conditions of operation. For claim under the warranty please contact your supplier.
www.labbox.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>