Di-LOG DL9206 Autoranging Multimeter User Manual
- June 3, 2024
- Di-LOG
Table of Contents
Di-LOG DL9206 Autoranging Multimeter
Safety Information
This manual contains information that must be followed for operating the meter safely and maintaining the meter in a safe operating condition. If this meter is not used in the manner specifi ed, the protection provided may be impaired.
Warning! Warns of potential danger, refer to the instruction manual to avoid personal injury or damage to the meter.
Caution! Dangerous voltage. Danger of electrical shock.
Continuous double or reinforced insulation complies with IEC536, class II.
Symbol of conformity, confi rms conformity with relevant EU directives. The
meter complies with EMC directives (89/336/EEC). Specifi cally stand-ards EN
50081-1 and EN 50082-1 as well as the Low Voltage Directive (73/23/EEC)
described in the standard EN 61010-1.
The meter has been designed in accordance with the safety regulations for
electronic measuring instruments, EN 61010-1, IEC 61010.
Voltages above 75V DC or 50V AC may constitute a serious shock hazard.
Before using the meter check for physical damage to the casing in particular
around the connectors. If the case is damaged do not use the meter.
Check the test leads for damaged insulation or exposed metal. Check the leads
for continuity. Replace damaged leads with identical model or specifi cation
before using the meter.
Where applicable use GS38 approved leads (not supplied) these are available
from Di-Log. When using test leads keep fi ngers behind the fi nger guards.
Do not apply more than the rated voltage, as marked on the meter between the
terminals or between any terminal and ground.
Before making a measurement ensure that the rotary switch is set to the
appropriate range. Do not turn the rotary switch whilst making a measurement.
Use the appropriate terminals, function and range for your measurements. If
the value to be measured is not known use the maximum measurement position and
reduce the range step by step until a satisfactory reading is obtained.
Do not use or store the meter in an environment of high temperature, humidity,
fumes, vapour, gaseous, infl ammable and strong magnetic fi eld. The perform-
ance and safety of the use may be compromised in such circumstances.
Disconnect circuit power and discharge all high voltage capacitors before
testing resistance, continuity, diodes, capacitance or current.
Before measuring current check the meters fuses and turn off power to the
circuit before connecting the meter to the circuit.
Replace the battery as soon as the low battery indicator appears. If the
battery is low the meter may give false readings.
Turn the meter power off when not in use,. Remove the battery if the meter is
in use for a long period. Constantly check the battery as it may have leaked.
A leaking battery will damage the meter.
The meter may only be opened by a qualifi ed service technician for calibration
and repair.
Input Limits
Never apply voltage or current to the meter that exceeds the specifi ed maximum:
Function | Maximum |
---|---|
V DC or V AC | 1000V DC, 750V AC |
A DC/AC | 10A DC/AC (30 seconds max every 15 minutes) |
Frequency, Resistance, Capacitance, Duty Cycle, Diode test, Continuity | 250V |
DC/AC
Temperature| 250V DC/AC
Symbols and Annunciators
- Continuity
- BAT Low Battery
- Diode
- HOLD Data Hold
- AUTO AutoRanging
- AC Alternating Current or Voltage
- DC Direct Current or Voltage
Controls and Inputs
- 6000 count Liquid Crystal Display with symbolic signs.
- Function rotary switch.
- 10A (positive) input terminal for 10A DC or AC measurements.
- COM (negative) input terminal.
- Positive input terminal.
- HOLD pushbutton.
- Max/Min pushbutton.
- Range/ Diode pushbutton.
- Hz/Duty pushbutton.
Specifications
- The instrument complies with: EN61010-1. Insulation: Class2, Double insulation.
- Overvoltage category: CATIII 600V.
- Display: 6000 counts LCD display with function indication.
- Polarity: Automatic, (-) negative polarity indication.
- Overrange: “OL” mark indication.
- Low battery indication: The “BAT” is displayed when the battery voltage drops below the operating level.
- Measurement rate: 2 times per second, nominal.
- Auto power off: Meter automatically shuts down after approx. 15 minutes of inactivity.
- Operating environment: 0˚C to 50˚C (32˚F to 122 ˚F) at < 70 % relative humidity.
- Storage temperature: -20˚C to 60˚C (-4˚F to 140˚F) at < 80 % relative humidity.
- For inside use, max height: 2000m
- Pollution degree: 2
- Power: One 9V battery , NEDA 1604, IEC 6F22.
- Dimensions: 150 (H) x 70 (W) x 48 (D) mm
- Weight: Approx.: 255g.
- Accuracy is given at 18˚C to 28˚C (65˚F to 83˚F), less than 70 % RH
DC Voltage
Range | Resolution | Accuracy |
---|---|---|
600.0mV | 0.1mV | ±0.5% of rdg + 2 dgts |
6.000V | 1mV | **** |
±1.2% of rdg + 2 dgts
60.00V| 10mV
600.0V| 100mV
1000V| 1V| ±1.5% of rdg + 2 dgts
- Input Impedance: 7.8MΩ.
- Maximum Input: 1000V dc or 750V ac rms.
AC Voltage
Range | Resolution | Accuracy |
---|---|---|
6.000V | 1mV | +1.2% of rdg + 3 dgts |
60.00V | 10mV | +1.5% of rdg + 3 dgts |
600.0V | 100mV | |
750V | 1V | +2.0% of rdg + 4 dgts |
- Input Impedance: 7.8MΩ.
- Frequency Range: 50 to 60Hz .
- Maximum Input: 1000V dc or 750V ac rms..
DC Current (Auto-ranging for uA and mA)
Range | Resolution | Accuracy |
---|---|---|
10A | 1mA | ±2.5% of rdg + 5 dgts |
- Overload Protection: 10A / 250V Fuse.
- Maximum Input: 10A dc or ac rms on 10A DC range..
AC Current (Auto-ranging for uA and mA)
Range | Resolution | Accuracy |
---|---|---|
10A | 1mA | ±3.0% of rdg + 7 dgts |
- Overload Protection: 10A / 250V Fuse.
- Frequency Range: 50 to 60 Hz.
- Maximum Input: 10A dc or ac rms on 10A AC range
Resistance (Auto-ranging)
Range | Resolution | Accuracy |
---|---|---|
600.0Ω | 0.1Ω | ±1.2% of rdg + 4 dgts |
6.000kΩ | 1Ω | ±1.0% of rdg + 2 dgts |
60.00kΩ | 10Ω | ±1.2% of rdg + 2 dgts |
600.0kΩ | 100Ω | |
6.000MΩ | 1kΩ | |
60.00MΩ | 10kΩ | ±2.0% of rdg + 3 dgts |
- Input Protection: 250V dc or 250V ac rms.
Capacitance (Auto-ranging)
Range | Resolution | Accuracy |
---|---|---|
40.00nF | 10pF | ±5.0% of rdg + 7 dgts |
400.0nF | 0.1nF | **** |
±3.0% of rdg + 5 dgts
4.000uF| 1nF
40.00uF| 10nF
400.0uF| 0.1uF| ±5.0% of rdg + 5 dgts
4000uF| 1uF| ±5.0% of rdg + 5 dgts
- Input Protection: 250V dc or 250V ac rms.
Frequency (Auto-ranging)
Range | Resolution | Accuracy |
---|---|---|
6.999Hz | 0.001Hz | |
69.99Hz | 0.01Hz | ±1.5% of rdg + 5 dgts |
699.9Hz | 0.1Hz | ±1.2% of rdg + 3 dgts |
6.999kHz | 1Hz | |
69.99kHz | 10Hz | |
699.9kHz | 100Hz | |
2.999MHz | 1kHz | ±1.5% of rdg + 4 dgts |
- Sensitivity: >0.5V RMS while ≤ 1MHz ;
- Sensitivity: >3V RMS while >1MHz ;
- Overload protection: 250V dc or ac rms
Duty Cycle
Range | Resolution | Accuracy |
---|---|---|
0.1%~99.9% | 0.1% | ±1.2% of rdg + 2 dgts |
- Pulse width: >100us, <100ms;
- Frequency width: 5Hz – 150kHz
- Sensitivity: >0.5V RMS
- Overload protection: 250V dc or ac rms.
Temperature
Range | Resolution | Accuracy |
---|---|---|
-20˚C~+760˚C | 1˚C | ±3% of rdg |
-4˚F~+1400˚F | 1˚F | ±5˚C/9˚F |
- Sensor: Type K Thermocouple.
- Overload protection: 250V dc or ac rms.
Diode Test
Range | Resolution | Accuracy |
---|---|---|
0.3mA typical | 1 mV | ±10% of rdg + 5 dgts |
- Open circuit voltage: 1.5V dc typical.
- Overload protection: 250V dc or ac rms
Audible continuity
- Audible threshold: Less than 30Ω;
- Test current: <0.3mA
- Overload protection: 250V dc or ac rms.
Operation
Warning: RISK OF ELECTROCUTION. HIGH-VOLTAGE CIRCUITS, BOTH AC AND DC, ARE VERY DANGEROUS AND SHOULD BE MEASURED WITH GREAT CARE.
- ALWAYS turn the function switch to the OFF position when the meter is not in use. This meter has Auto OFF that automatically shuts the meter OFF if 15 minutes elapse between uses.
- If “OL” appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range.
Note: On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random, changing reading. This is normal and is caused by the high-input sensitivity. The reading will stabilize and give a proper measurement when connected to a circuit.
Range Button
When the meter is first turned on, it automatically goes into AutoRanging. This
automatically selects the best range for the measurements being made and is
generally the best mode for most measurements. For measurement situations
requiring that a range be manually selected, perform the following:
- Press the RANGE button. The “AUTO” display indicator will turn off.
- Press the RANGE button to step through the available ranges until you select the range you want.
- Press and hold the RANGE button for 2 seconds to exit the Manual Ranging mode and return to AutoRanging.(If backlight is turn on, please press BACKLIGHT button it will turn off).
- To select Diode/continuity.
Data Hold Button
The Hold function allows the meter to “freeze” a measurement for later
reference.
- Press the HOLD button to “freeze” the reading on the indicator. The indicator “HOLD” will be appear in the display.
- Press the HOLD button to return to normal operation.
Hz/duty
- Switch to Hz/Duty range.
- Press the Hz/Duty button to show the reading in the display and the “Hz/Duty” indicator will appear on the display.
DC Voltage Measurements
Caution: DO NOT MEASURE DC VOLTAGES IF A MOTOR ON THE CIRCUIT IS BEING SWITCHED ON OR OFF. LARGE VOLTAGE SURGES MAY OCCUR THAT CAN DAMAGE THE METER.
- Set the function switch to the V DC position .
- Insert the black test lead banana plug into the negative (COM) terminal and the red test lead banana plug into the positive (V) terminal.
- Connect the test probe tips to the circuit under test. Be sure to observe the correct polarity (red lead to positive, black lead to negative).
- Read the voltage in the display. The display will indicate the proper decimal point and value. If the polarity is reversed, the display will show (-) minus before the value.
AC Voltage Measurements
Warning:
RISK OF ELECTROCUTION. THE PROBE TIPS MAY NOT BE LONG ENOUGH TO CONTACT THE
LIVE PARTS IN-SIDE SOME 240V OUTLETS FOR APPLIANCES BECAUSE THE CONTACTS ARE
RECESSED DEEP IN THE OUTLETS. AS A RESULT, THE READING MAY SHOW 0 VOLTS WHEN
THE OUTLET ACTUALLY HAS VOLTAGE ON IT. MAKE SURE THE PROBE TIPS ARE TOUCHING
THE METAL CONTACTS INSIDE THE OUTLET BEFORE ASSUMING THAT NO VOLTAGE IS
PRESENT.
Caution: DO NOT MEASURE AC VOLTAGES IF A MOTOR ON THE CIRCUIT IS BEING
SWITCHED ON OR OFF.
LARGE VOLTAGE SURGES MAY OCCUR THAT CAN DAMAGE THE METER.
- Set the function switch to the V AC position.
- Insert the black test lead banana plug into the negative (COM) terminal and the red test lead banana plug into the positive (V) terminal.
- Connect the test probe tips to the circuit under test.
- Read the voltage in the display. The display will indicate the proper decimal point, value and symbol (AC, V, etc.).
DC Current Measurements
Caution: DO NOT MAKE CURRENT MEASUREMENTS ON THE 10A SCALE FOR LONGER THAN 30 SECONDS. EXCEEDING 30 SECONDS MAY CAUSE DAMAGE TO THE METER AND/OR THE TEST LEADS.
- Insert the black test lead banana plug into the negative (COM) terminal.
- For DC current measurements , set the function switch to the DC 10A position and insert the red test lead banana plug into the 10A terminal.
- Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
- Connect the black test probe tip to the negative side of the circuit. Connect the red test probe tip to the positive side of the circuit.
- Apply power to the circuit.
- Read the current in the display. The display will indicate the proper decimal point, value and symbol.
AC Current Measurements
Warning: TO AVOID ELECTRIC SHOCK, DO NOT MEASURE AC CURRENT ON ANY CIRCUIT WHOSE VOLTAGE EXCEEDS 250V AC.
Caution: DO NOT MAKE CURRENT MEASUREMENTS ON THE 10A SCALE FOR LONGER THAN 30 SECONDS. EXCEEDING 30 SECONDS MAY CAUSE DAMAGE TO THE METER AND/OR THE TEST LEADS.
- Insert the black test lead banana plug into the negative (COM) terminal.
- For AC current measurements , set the function switch to the AC 10A position and insert the red test lead banana plug into the 10A terminal.
- Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
- Connect the black test probe tip to the negative side of the circuit. And connect the red test probe tip to the positive side of the circuit.
- Apply power to the circuit.
- Read the current in the display. The display will indicate the proper decimal point, value and symbol.
Resistance Measurements
Warning: TO AVOID ELECTRIC SHOCK, DISCONNECT POWER TO THE UNIT UNDER TEST AND DISCHARGE ALL CAPACITORS BEFORE TAKING ANY RESISTANCE MEASUREMENTS. REMOVE THE BATTERIES AND UNPLUG THE LINE CORDS.
- Set the function switch to the Ω position.
- Insert the black test lead banana plug into the negative (COM) terminal and the red test lead banana plug into the positive Ω terminal.
- Connect the test probe tips across the circuit or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading.
- Read the resistance in the display. The display will indicate the proper decimal point, value and symbol.
Continuity Check
Warning : TO AVOID ELECTRIC SHOCK, NEVER MEASURE CONTINUITY ON CIRCUITS OR WIRES THAT HAVE VOLTAGE ON THEM.
-
Set the function switch to the position.
Insert the black lead banana plug into the negative (-) terminal (COM) and the red test lead banana plug into the positive (+) terminal ( ). -
Pressthe button until the symbol appears in the display.
-
Connect the test probe tips to the circuit or wire you wish to check.
-
If the resistance is less than approximately 30 , the audible signal will sound. The display will also show the actual resistance.
Diode Test
Warning: TO AVOID ELECTRIC SHOCK, DO NOT TEST ANY DIODE THAT HAS VOLTAGE ON IT.
- Set the function switch to position.
- Press the button until the symbol appears in the display.
- Insert the black test lead banana plug into the negative (-) terminal (COM) and the red test lead banana plug into the positive (+) terminal(Ω).
- Connect the test probe tips to the diode or semiconductor junction you wish to test. Note the meter reading.
- Reverse the probe polarity by switching probe position. Note this reading.
- The diode or junction can be evaluated as follows:
- If one reading shows a value and the other reading shows OL, the diode is good.
- If both readings show OL, the device is open.
- If both readings are very small or 0, the device is shorted.
Note: The value indicated in the display during the diode check is the forward voltage.
Frequency Measurements
- Set the function switch to the FREQ position.
- Insert the black test lead banana plug into the negative (-) terminal (COM) and the red test lead banana plug into the positive (+) terminal (F).
- Connect the test probe tips to the circuit under test.
- Read the frequency in the display. The digital reading will indicate the proper decimal point, symbols (Hz, kHz) and value.
Capacitance Measurements
Warning: TO AVOID ELECTRIC SHOCK, DISCONNECT POWER TO THE UNIT UNDER TEST AND DISCHARGE ALL CAPACITORS BEFORE TAKING ANY CAPACITANCE MEASUREMENTS. REMOVE THE BATTERIES AND UNPLUG THE LINE CORDS.
- Set the function switch to the CAP position. (“nF” and a small value will appear in the display).
- Insert the black test lead banana plug into the negative (-) terminal (COM) and the red test lead banana plug into the positive (+) terminal(CAP).
- Connect the test leads to the capacitor to be tested. The display will indicate the proper decimal point, value and symbol.
Temperature Measurement
Warning: TO AVOID ELECTRIC SHOCK, DISCONNECT BOTH TEST PROBES FROM ANY SOURCE OF VOLTAGE BEFORE MAKING A TEMPERATURE MEASUREMENT.
- If you wish to measure temperature in ˚F, set the function switch to the ˚F range. If you wish to measure temperature in ˚C, set the function switch to the ˚C range.
- Insert the type K thermocouple probe black test lead banana plug into the negative COM terminal and the red test lead banana plug into the positive Temp terminal.
- Connect the Temperature Probe head to the part whose temperature you wish to measure. Keep the probe connected to the part under test until the reading stabilizes (about 30 seconds).
- Read the temperature in the display. The digital reading will indicate the proper decimal point and value.
Warning: TO AVOID ELECTRIC SHOCK, BE SURE THE THERMOCOUPLE HAS BEEN REMOVED BEFORE CHANGING TO ANOTHER MEASUREMENT FUNCTION.
Replacing the Battery
Warning: TO AVOID ELECTRIC SHOCK, DISCONNECT THE TEST LEADS FROM ANY SOURCE OF VOLTAGE BEFORE REMOVING THE BATTERY COVER.
- When the batteries become exhausted or drop below the operating voltage, “BAT” will appear in the right-hand side of the LCD display. The battery should be replaced.
- Disconnect the test leads from the meter.
- Open the battery cover by loosening the screw using a Phillips head screwdriver.
- Insert the battery into battery holder, observing the correct polarity.
- Put the battery cover back in place. Secure with the two screws.
- Follow instructions for installing battery. See the Battery Installation section of this manual.
Warning: TO AVOID ELECTRIC SHOCK, DO NOT OPERATE THE METER UNTIL THE BATTERY COVER IS IN PLACE AND FASTENED SECURELY.
Note: If your meter does not work properly, check the fuses and battery to make sure that they are still good and that they are properly inserted.
Replacing the Fuses
Warning: TO AVOID ELECTRIC SHOCK, DISCONNECT THE TEST LEADS FROM ANY SOURCE OF VOLTAGE BEFORE REMOVING THE FUSE COVER.
- Disconnect the test leads from the meter and any item under test.
- Open the fuse cover by loosening the screw on the door using a Phillips head screwdriver.
- Remove the old fuse from its holder by gently pulling it out.
- Install the new fuse into the holder.
- Always use a fuse of the proper size and value (0.5A/250V fast blow for the 400mA range, 10A/250V fast blow for the 10A range).
- Put the fuse cover back in place. Insert the screw and tighten it securely.
Warning: TO AVOID ELECTRIC SHOCK, DO NOT OPERATE YOUR METER UNTIL THE FUSE COVER IS IN PLACE AND FASTENED SECURELY.
Warranty & Maintenance
24 Month Warranty
Di-Log instruments are subject to stringent quality controls. If in the course
of normal daily use a fault occurs we will provide a 24 month warranty (only
valid with invoice).
Faults in manufacture and materials defect will be rectifi ed by us free of
charge, provided the instrument has not been tampered with and returned to us
unopened.
Damage due to dropping abuse or misuse is not covered by the warranty.
Outside the warranty period we offer a full repair and re-calibration service.
Maintenance
WARNING Do not attempt to repair or service you meter unless you are
qualified to do so and have the relevant calibration, performance test and
service information.
To avoid electrical shock or damage to the meter do not get water inside the
case.
Periodically wipe the case with a damp cloth and mild detergent. Do not use
chemical solvent.
Clean the input terminal with cotton bud, as dirt or moisture in the terminal
can affect readings.