DS INSTRUMENT SG6000 RF Signal Generator User Manual

June 3, 2024
DS INSTRUMENT

DS INSTRUMENT SG6000 RF Signal Generator

DS-INSTRUMENT-SG6000-RF-Signal-Generator-User-Manual-
PRODUCT

DS Instruments’ various signal generator models are all designed to be as intuitive as possible, either being controlled by the front panel, or from the USB/Ethernet interface. Buttons and front panel modes are similar across all models.

Stand-Alone Use Quick Setup

  •  Connect your device to a standard USB port or 5 Volt power source using the included USB cable. A desktop, laptop computer, or smartphone charger should work. Avoid using unpowered USB hubs or long USB cables. Blue USB 3.0 ports are preferred due to their high current capabilities.
  •  Connect an external 10MHz reference source to the rear reference MCX connector if desired. If no external source is detected, the unit will use its own internal source. Searching for an external frequency only occurs at boot-up, do not connect an external reference after the unit is already powered up.
  •  Press the red power button if available on your device. The screen will display the model number and firmware version. The unit will beep indicating boot-up and the green power LED should light up.
  •  The unit is now ready to accept front panel changes or a data connection from the control software.
  •  To cycle between front control modes, hold the MODE button down until there is a beep (~0.5 seconds). The default mode is integer frequency, followed by fractional frequency, power output level, power vernier, RF output on-off, and save system state.
  •  The RFO led indicates that the device is successfully locked to the reference and provides the desired frequency on the rear SMA output port.

SCPI/COM/Computer Control Use Quick Setup

  • Connect your device to a standard computer USB port using a high-quality USB cable.
  • Connect an external 10MHz reference source to the rear MCX connector if desired. If no external source is detected, the unit will use its own internal source. Reference auto-detection can be overridden in the GUI if desired.
  • Press the red power button if available on your device. The screen will display the model number and firmware version. The unit will beep indicating boot-up and the green power LED should light up.
  •  Windows should now install the standard virtual COM port driver. Wait for this to complete.
  • Download and open the DS Instruments Signal Generator Control GUI from the website. Install any .NET framework updates from Microsoft if needed.
  •  Select a COM port from the top select list and press CONNECT to try to control the unit. If the select box turns green then a connection has been established. The model number and firmware versions will appear in the green info box. If this box turns red, the wrong COM port was selected or an error has occurred.
  •  At this point, a user can enter a frequency and click APPLY or set up a frequency sweep.
  •  The blue RFO led on the front of the SG unit indicates that it is successfully locked and outputting the desired frequency on the output port
  •  The physical control buttons will be disabled when a COM port connection is established to prevent any accidental bumping of the buttons when under computer control.DS-INSTRUMENT-SG6000-RF-Signa

Control Software Feature Descriptions

  1.  COM / Ethernet search and connect – Lists any COM ports or DSI IP addresses found
  2.  Info Box – Displays information about the connected device
  3.  Custom Name – Lets the user save a TAG ID, Name, or inventory number to non-volatile memory
  4.  Reference frequency source – Sets internal, external or auto-detect for the 10MHz reference input
  5.  Custom Command – Type or select a custom SCPI serial command to send to the connected device
  6.  Frequency – Step or enter a frequency to generate
  7.  RF Output Power – Adjust the output power in calibrated 0.5dB steps
  8.  RF Enabled – Set the RF output port to ON or OFF
  9.  Sweep Settings – Set the bounds, number of points, direction, and type of frequency sweep
  10.  Sweep Control – Save the current sweep settings to the device, start or abort the sweep
  11.  Fine Power Tune – Adjust output power in fine (~0.05dB) increments between the 0.5dB steps.
  12.  Message Bar – Displays feedback information sent from the device

Control Software Notes

  •  If reference auto-detection is misbehaving with your input signal, it can be forced to external or internal using the select box and “Change Ref” button. An SG reset is recommended after this setting is changed.
  •  “Send Command” allows a user to send a raw SCPI command to the Signal Generator, in USB mode the GUI will also display any response from the device in the message bar at the bottom.
  •  The Vernier setting allows fine power output level setting between the 0.5dB steps. Highest power output results when the slider is farthest to the right.
  •  If the USB voltage box goes into the red, performance will be unpredictable or an error will be indicated by the device. Try finding a better USB power source or cable.
  •  If the device encounters any errors it will start blinking red, to query what the error is, send the SCPI command “SYST: ERR?”.

Hardware

  •  The MCX 10MHz reference input will accept sine or square wave input between -10 and +10dBm.
  •  The sweep trigger MCX input is pulled high to +3.3V internally, when pulled low the saved sweep will be initiated. Sweep parameters must be loaded from the GUI for this trigger to be activated. This input is 5-volt tolerant. Bringing the input back high will not stop the sweep or have any other effect.

Sweeping Details

  •  The Sweep mode control lets the user configure a continuous or single sweep.
  •  Direction UP increases frequency from Start to Stop, and DOWN starts the sweep at the stop frequency.
  •  The Trigger option changes the functionality of the physical trigger input pin between activating the complete sweep, or just the next STEP. This is only available in certain firmware versions.
  •  The Points field determines how many independent frequency points are calculated in the total. This can be from 2 to 10000.
  •  The Dwell time field controls how long the synthesizer will pause at each new frequency point. The minimum value is ~3ms.
  •  The Start and Stop frequency settings control the beginning and end of the sweep.
  •  The Step Size and Run Time are system-calculated values based on the above values. Run time is only a close estimate of the ideal value, not including actual minor system overhead and phase lock loop delays.
  •  The Save Sweep button transfers these values from the GUI to the signal generator CPU. This button is critical to the press after values are updated in the control software.
  •  The Abort button immediately stops the device mid-sweep.
  •  The Trigger button emulates the physical trigger input pin by sending a command to start the sweep now.

Notes

Version 4.0 – DS Instruments 2022 – www.dsinstruments.com

References