UNPACKING AND INSPECTION

Thank you for your purchase of the AE TEGAM GEMINI 5541A RF Power Meter. This Quick Start Guide will provide the essential information necessary to begin using your new instrument right away. For a full explanation of the features and functions, however, please download the Operator’s Manual at www.tegam.com.

Inspection
Each 5541A RF Power Meter is electrically and mechanically inspected before shipment. Upon receiving your new AE TEGAM RF Power Meter, unpack all items from the shipping container and inspect the items for any obvious damage that may have occurred during transit. Notify AE TEGAM immediately if damage, dents, worn connectors, or other irregularities are
found. Use the original packing materials if reshipment is necessary.TEGAM-
5541A-RF-Power-Meter-1

Package Contents

Each GEMINI 5541A RF Power Meter is shipped with:

  • GEMINI 5541A RF Power Meter1,
  • 5541A-901, GEMINI 5541A RF Power Meter Quick Start Guide
  • Calibration Certificate and Documentation

INSTRUMENT DISPLAY

The 5541A RF Power Meter digital direct-reading display, shown in Figure 2, provides forward and reverse power measurements, VSWR, and frequency. These readings are indicated on the display by the designators FWD PWR, REV PWR, VSWR, and FREQ respectively. Units of measurement are displayed on the right side of the display. The displayed measurement data is updated three times per second. TEGAM-5541A-RF-Power-Meter-2

NOTE: The Frequency and VSWR indications are informational and provided only for the user’s convenience. For traceable frequency measurements, use appropriate instruments.

CONNECTING THE RF POWER METER AND TAKING YOUR FIRST MEASUREMENT

  1. To meet published specifications, allow the instrument to acclimate in the laboratory environment for at least 24 hours after shipping, storage, or other exposure to temperatures beyond its operating range of 18 to 28 °C.

  2. Before connecting your 5541A RF Power Meter to any RF power source, identify the INPUT and OUTPUT ports. The INPUT port and power flow direction are designated by the symbol on the instrument housing (see Figure 3). NOTE: The 5541A RF Power Meter may be calibrated to either the input or output port. However, the RF power source must always be connected to the input port regardless of which port is calibrated. Failure to do so may result in incorrect readings. Refer to the Calibration Certificate for port calibration information.

  3. Connect an IEEE 802.3af PoE (Power over Ethernet) compliant device (switch/hub/router/etc.) to the 5541A RF Power Meter The 5541A RF Power Meter display will illuminate.
    NOTE: The 5541A RF Power Meter may be connected to any IEEE 802.3af compliant device for power. However, the meter must be connected to the same network as the host PC for communication.

  4. Connect the 5541A RF Power Meter OUTPUT to a suitable load using torque techniques appropriate for the connector type.
    WARNING: High RF power levels may be present at the output of the 5541A RF Power Meter. Always connect an appropriate load to the instrument output.
    WARNING: The 5541A RF Power Meter should be serviced by qualified personal only. Never attempt to open the instrument casing, disconnect transmission lines, or otherwise service the meter while RF power is applied to the system. Failure to comply with these warnings may result in injury or death to personnel, and damage to equipment.

  5. Connect the 5541A RF Power Meter INPUT to the RF power source using torque techniques appropriate for the connector type.

  6. Allow 10 minutes for the instrument to warm up.

  7. Turn ON the source RF output power. The display will begin updating immediately. Allow sufficient time for the readings to stabilize before recording the measurement results.

GEMINI TOOLS™ RF POWER METER APPLICATION

Description
The GEMINI Tools power meter application is a free software interface available for download to all 5541A RF Power Meter users. The application allows for remote monitoring and data collection from up to five instruments simultaneously. Data is stored in text files and formatted for easy importing into spreadsheet, database, or other data management programs.

Operation
Operation of the GEMINI Tools power meter application is straightforward:

  1. Download and run the installer found at www.tegam.com/geminitools.
  2. Launch the program and set the Configuration Parameters as desired (click Tools -> Configuration).
  3. Connect your new 5541A RF Power Meter to the PC or compatible USB hub on which the GEMINI Tools power meter application is running. The application will automatically detect connected instruments.
  4. Select the desired unit in the Available Meters group (hold the CTRL key to select up to five (5) units).
  5. Click START Data Collection to begin collecting measurement data.