APERA EC700 Benchtop Conductivity Meter Instruction Manual
- June 3, 2024
- APERA
Table of Contents
EC700 Benchtop Conductivity Meter
Instruction Manual
Brief Introduction
Thank you for purchasing Apera Instruments EC700 Benchtop Conductivity Meter.
This instrument is an outstanding combination of advanced electronics desin
and sensor technology. It’s the most economical choice of a reliable lab
benchtop conductivity meter.
Please read this maual carefully to properly use and maintain the meter. Apera
Instruments reserves the right to update the contents of this manual without
giving prior notices.
1.1 Main Features
- The meter has a built-in microprocessor chip, which enables intelligent functions such as automatic calibration, automatic temperature compensation, data storage, parameter setup, max/min reading display, etc.
- The advanced automatic calibration mode has calibration guide and self-diganosis functions, automatically recognizing standard calibration solutions.
- The advanced digital processing technology intelligently improves the meter’s response time and measuring accuracy with stable reading display mode available.
- The kit comes with a combination conductivity electrode with temperature sensor, measuring conductivity and temperature simultaneously, and allows for automatic temperature compensation.
- IP54 dustproof and spillproof. Connectors are protected by silicone seal caps, effectively ensuring the reliability and service life of the meter under harsh environment.
- Detachable electrode holder and 3 bottles of conductivity calibration solutions come with the kit for convenient use.
Technical Specifications
Conductivity| Range| Conductivity: 0-200.0 mS/cm, divided into 4 ranges:
(0-199.9) μS/cm, (200-1999) μS/cm, (2.00-19.99) mS/cm,(20.0-199.9) mS/cm
TDS: (0-100) g/lL devided into 4 ranges: (0-99.9) mg/L, (100-999) mg/L,
(1.00-9.99) g/L, (10.0-100.0) g/L
---|---|---
Resolution| Conductivity: 0.1/1 μS/cm; 0.01/0.1 mS/cm TDS: 0.1/1 mg/L,
0.01/0.1 g/L
Accuracy| ±1.0% F.S ±1 digit
Temperature Compensation| 0 to 50 ˚C (Auto or Manual)
Electrode Constant| 0.1/1/10 cm-1
Reference Temperature| 25 ˚C
Temperature Compensation Coefficient| 0.00-9.99%/ ˚C, default value:
2.00%/ ˚C
Calibration| 1-4 point Auto Calibration
Temp.| Range| 0 to 100˚C; 32 to 212°F
---|---|---
Resolution| 0.1˚C; 0.1/1°F
Accuracy| ±0.5˚C ±1 digit
Others| Data Storage| 50 groups
Storage Content| Numberings, Measurement, Unit, Temperature, Temperature
Compensation Status
Power| DC9V/300mA
IP Ratings| IP54 dustproof and spillproof
Size and Dimension| (240235103) mm/1kg
Instrument Description
3.1 LCD Display
(1) — Measurement mode
(2) — Measurement value
(3) — Calibration icons and numberings
(4) — Measurement unit
(5) — Temperature unit
(6) — The unit displayed during calibration
(7) — The value displayed during calibration, numberings for storage and
recall, and reminder icons
(8) — Icons for data storage and recall — M+: store measuring data; RM: Recall
measuring data
(9) — Temperature and reminder icons
(10) — Self-Diagnosis icons and numberings
(11) — Temperature compensation mode icons — ATC: Auto Temperature
Compensation; MTC: Manual Temperature Compensation
(12) — Completed calibration icons
(13) — Stable reading icon
(14) — Icons for maximum and minimum values
3.2 Keypad
3.2.1. Keypad operations
Short press —– <1.5 seconds;
Long press —– >1.5 seconds.
3.2.2 For keypad operation, please see Table-1.
Table – 1 Keypad operations and descriptions
Keypad | Operation | Functions |
---|---|---|
Short Press | • Power on/off | |
Short Press | • Select measuring mode: Cond→TDS. | |
Long Press | • Enter parameter setting:P1→P2→…P6. | |
Short Press | • Enter Calibration mode |
• Cancel any operation and go back to measurement mode
| Short Press| • In calibration mode, press to conduct calibration
• In parameter setting, press to confirm selection
| Short Press| • Store measuring data
Long Press| • Recall stored measuring data
| Short Press or Long Press| 1. In manual temperature compensation mode,
short press to change temperature, long press to change quickly
2. In parameter setting mode, short press to change parameter and settings.
3. In recall mode, short press to change numberings, long press to change
quickly.
| Short Press| 4. Buttons for max/min values (refere to section 3.6)
3.3 Connectors
Symbol| Connector Name| Connector Type
COND| Socket for conductivity electrode| BNC
TEMP| Socket for temperature sensor| RCA
DC9V| DC9V power socket| Φ2.5 direct type
3.4 Stable reading display mode
When the measuring value is stable, smiley face icon appears on LCD, see
Figure-3. Please do not get the reading value or make calibration until the
icon appears and stays.
3.5 Data storage, recall, and deletion
3.5.1 Storage
When the measurement is stable, press key, the meter displays M+ icon and
storage numberings on LCD and stores the measuring data. In Figure-4, the
meter stores the first group of the measuring data. Press again to keep
storing.
3.5.2 Recall stored data
In the measurement mode, long press **key to recall the last stored
Measuring data. The meter displays RM** icon and storage numberings. In
Figure-5, the meter is recalling the 28th stored measuring data. Keep
pressing to recall successively the stored measuring data. Press and hold to
quickly recall the stored measuring data. Short press to go back to
measurement mode.
3.5.3 Clear stored data
The meter can store 50 groups of data. Once the storage is full, if user
presses again, the FUL icon will be displayed on LCD (see Figure-6).
The storage needs to be cleared in parameter setting P4 so as to keep storing
data. The numberings in M+ indicates the total number of stored data; The
numberings in RM indicates the number of stored data in current mode
(cond/TDS).
For example, if 20 groups of data is stored in conductivity mode, and 10 groups in TDS mode, then the numberings for M+ will be “M+30” in both conductivity and TDS mode; the numberings for RM in conductivity mode would be “RM20”, in TDS mode “RM10”.
3.6 Max/Min Function
Use this function to display the maximum or minimum records out of groups of
measuring data. In measurement mode, press **, LCD will display flashing
icons of “MAX/MIN”,** indicating the meter has entered the mode of recording
maximum and minimum data. After testing, press again, LCD will alternately
display the maximum and minimum value in the past measurements. Press to go
back to regular measurement mode.
Conductivity Measurement
4.1 Preparation
4.1.1 Install the electrode holder
The electrode holder is comosed of a metal stand base, a supporting pole, and
a electrode clip. First, install the metal stand base to the meter (screw it
on). Please note that the metal base can be installed on either side of the
meter. Then install the supporting pole on the metal stand base, and then
install the electrode clip on the supporting pole.
4.1.2 Connect the power adapter
Plug the power adapter tightly into the DC9V socket. The power’s voltage
should meet the requirements listed on the power adapter.
4.2 Conductivity Electrode Information
4.2.1 Conductivity Electrode
The meter comes with the 2301T-F combination conductivity electrode (K=1.0)
with built-in temperature sensor, which enables the automatic temperature
compensation. The electrode’s housing adopts polycarbonate materials, which
is resistant to shocks and corrosion.
4.2.2 Electrode Connectors
The conductivity electrode has two connectors: the BNC connector connects the
conductivity sensor, the RCA connector connects the temperature sensor. Plug
these two connectors into ‘COND” and “TEMP” sockets. Please note do not pull
the cables in case of poor contact. Please keep the connectors clean and dry.
Refer to section 4.8 regarding how to properly maintain the Conductivity
electrode.
4.2.3 Use the Electrode
Stir the electrode for a few seconds after it’s dipped into the solution and
then let it stand still such that the bubbles inside the probe can be removed
and a stable measurement can be quickly obtained.
4.3 Information regarding Calibration
4.3.1 Conductivity standard solutions
The meter automatically recognizes calibration solutions. 1 point or multi-
points calibration (up to 4 points) can be adopted. The calibration icons
stand for 4 different calibration points ( stands for two standard
solutions), see Table-2:
Table-2
Calibration icon | pH standard buffer series |
---|---|
4-Point calibration | |
1413 μS/cm | |
12.88 mS/cm |
111.8 mS/cm
4.3.2 How often to calibrate
- We recommend to calibrate the conductivity meter once per month under normal conditions.
- If the requirement for accuracy is high or the temperature of the test sample has a big difference to the reference temperature (25 ˚C), we recommend to do the calibration once a week.
- Users can tell if the meter needs a calibration by testing the standard calibration solutions. If the measurements you get have a big difference to the standard value of the calibration solutions, then a calibration is necessary.
- For new electrodes or the meter that has been restored to factory default setting, we recommend to conduct 3-4 points calibration. For general uses, users can choose the calibration solutions that are close to the test samples to conduct 1 or 2 points calibration.
4.4 Conductivity Calibration (take 1413 μS as an example)
4.4.1. Rinse the electrode in distilled or pure water, allow it to dry, wash
with a little of standard solution and submerge it in standard solution. Stir
the solution briefly and allow it to stay in the solution until a stable
reading is reached.
4.4.2. Press key to enter the calibration mode. The “CAL” icon will flash
at the top right, and the scanning and locking process of calibration solution
will appear at the bottom right.
4.4.3. When the meter locks 1413 μS, stable icon will display on LCD. Press
key to complete calibration. “End” icon will show up and the meter will
return to measuring mode and will show up on the bottom left (see
Figure-7).
Note: pressing before the reading being stable ( shows up and stays), the
Err4 self-diagnosis icon will appear (see section 6.5.2).
4.4.4. For multi-point calibration, please repeat clause 4.3.1-4.3.2 until all
the calibration is done. The meter can repeat calibration in the same
calibration solution until the stable value is reached.
4.5 TDS and Conductivity
4.5.1. TDS and conductivity is linear related. The conversion factor is
0.40-1.00. Adjust the factor from parameter P2.6. The factory default setting
is 0.71 and please refer to Section 7.4. The meter only needs to be
calibrated in Conductivity mode, then after calibration of conductivity, the
meter can switch from conductivity to TDS.
4.5.2. Adjust TDS conversion factor in parameter setting P4 according to the
data collected during testing and experience. Table – 10 lists some commonly
used Conductivity and TDS conversion factors. This is for your reference
only.
Table – 10 Conductivity and TDS conversion factors
Conductivity of solution | TDS conversion factor |
---|---|
0-100 μS/cm | 0.60 |
100-1,000 μS/cm | 0.71 |
1-10 mS/cm | 0.81 |
10-100 mS/cm | 0.94 |
4.6 Sample Test
4.6.1. Rinse conductivity electrode in distilled or pure water, dry it, and
submerge it in the sample solution. Stir the solution briefly and allow it to
stay in the sample solution until a stable reading is reached and icon
appears on LCD, then get the reading value, which is the conductivity value of
the solution.
4.6.2. During the process of calibration and measurement, the meter has self-
diagnosis functions, indicating the relative information as below: Table – 8.
Table – 8 Self-diagnosis information of conductivity measurement mode
Display Icons | Contents | Checking |
---|---|---|
Wrong conductivity calibration solution or the meter recognition of | ||
calibration solution is out of range. | 1. Check whether conductivity |
calibration solution is correct.
2. Check whether the meter connects the electrode well.
3. Check whether the electrode is damaged.
| Press key when measuring value is not stable during calibration.| Press
key after icon appears.
| During calibration, the measuring value being unstable for over 3 minutes.|
1. Shake the electrode to eliminate bubbles in electrode head.
2. Replace with new conductivity electrode.
4.6.3. Factory default setting
For factory default setting, please refer to parameter setting P7 (Section
4.7.2). With this function, all calibration data is deleted and the meter
restores to the theory value. Some functions restore to the original value.
When calibration or measurement fails, please restore the meter to factory
default setting and then perform recalibration or measurement. Please note
once set the factory default, all the data deleted will be irretrievable.
4.7 Parameter Setting
Table-9
Symbol | Parameter setting content | Parameter | Factory default setting |
---|---|---|---|
P1 | Select electrode constant | 0.1, 1.0, 10 | 1.0 |
P2 | Select reference temperature | 15 – 30 ℃ | 25℃ |
P3 | Adjust temperature compensation coefficient | 0.00 – 9.99% | 2.00% |
P4 | Adjust TDS Factor | 0.40 – 1.00 | 0.71 |
P5 | Select temperature unit | ℃ – ℉ | ℃ |
P6 | Clear stored data | No – Yes | No |
P7 | Restore to factory default | No – Yes | No |
4.7.1 How to do the parameter setting:
In measurement mode, long press to enter parameter setting P1. Then press or
to change menu from P1 to P2…P7. For details, please see Table-10.
Table-10 Parameter Setting
| P1 — Select electrode’s constant (1.0—10.0—0.1)
1. Long press to enter P1
2. Press , “1.0” flashes, press to choose 1.0→10.0→0.1; press to confirm
3. Press to enter P2, or press to return to measurement mode.
---|---
| P2 —Select reference temperature (25.0℃—18.0℃ — 20.0℃)
1. Press , “25.0℃” flashes, press or to adjust temp. from 15℃ to 30℃;
press to confirm.
2. Press to enter P3, or press to return to measurement mode.
| P3 —Adjust temp. compensation coefficient (0.00-9.99%)
1. Press , “2.00” flashes, press or to adjust from 0.00-9.99;
press to confirm.
2. Press to enter P4, or press to return to measurement mode.
| P4 —Adjust TDS Factor (0.40-1.00)
1. Press , “0.71” flashes, press or to adjust from 0.40-1.00; press to
confirm.
2. Press to enter P5, or press to return to measurement mode.
| P5 — (°C—°F)
1. Press , “°C” flashes, press , “°F” flashes; when parameter flashes,
press to confirm.
2. Press to enter P6, or press to return to measurement mode.
| P6 — Clear storage (No—Yes)
1. Press , “No” flashes, press to choose No→Yes; press to confirm
No — not to clear data storage; Yes—clear all data storage
2. Press to enter P7, or press to return to measurement mode.
| P7 — Return to factory default mode (No—Yes)
Press , “No” flashes, press to choose No→Yes; Press to confirm, the
meter returns to measurement mode.
No — not return to factory default mode; Yes—return to factory default
mode
4.8 Conductivity Electrode’s Maintenance
4.8.1. Always keep the conductivity electrode clean. Before taking a
measurement, rinse the electrode in pure water. It is recommended to rinse it
again in the sample solution. When submerge the electrode in solution, stir
the solution briefly to eliminate air bubbles and allow it to stay until a
stable reading is reached. Conductivity electrodes are usually stored dry. For
conductivity electrodes that haven’t been used for a long time, users should
soak the electrode in 12.88 mS calibration solution for 5-10 minutes, or to
soak it in tap water for 1 to 2 hours. Rinse the electrode in pure water after
measurement.
4.8.2. The sensing rod of Model 2301T-F conductivity electrode is coated
with platinum black to minimize electrode polarization and expand measuring
range. The platinum black coating adopted our special processing technology,
which improves the electrode performance and the firmness of the coating. If
the platinum black electrode is stained, gently clean the electrode with soft
brush in warm water containing detergent or alcohol.
Complete Kit
| Content| Quantity
---|---|---
5.1| EC700 Benchtop Conductivity Meter| 1
5.2| 2301T-F Plastic Combination Conductivity Electrode| 1
5.3| Conductivity standard calibration solutions: (84μS/1413μS /12.88mS, 50ml
for each)| 1 for each
5.4| 9V Power Adapter| 1
5.5| Electrode Holder| 1 set
5.6| Instruction Manual| 1
Warranty
We warrant this instrument to be free from defects in material and workmanship
and agree to repair or replace free of charge, at option of APERA INSTRUMENTS
(Europe) GmbH, any malfunctioned or damaged product attributable to
responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of TWO YEARS
(SIX MONTHS for the probe) from the delivery.
This limited warranty does not cover any damages due to:
Transportation, storage, improper use, failure to follow the product
instructions or to perform any preventive maintenance, modifications,
combination or use with any products, materials, processes, systems or other
matter not provided or authorized in writing by us, unauthorized repair,
normal wear and tear, or external causes such as accidents, abuse, or other
actions or events beyond our reasonable control.