RightSign ICOVN-C81 COVID-19 Antigen Rapid Test Cassette User Manual
- June 3, 2024
- RightSign
Table of Contents
RightSign ICOVN-C81 COVID-19 Antigen Rapid Test Cassette
COVID-19 Antigen Rapid Test Cassette
(Nasal Swab) Package Insert
A rapid test for the qualitative detection of COVID-19 antigen in a Nasal
Swab.
For use only by individuals who have been given appropriate training for in
vitro diagnostic use.
INTENDED USE
The COVID-19 Antigen Rapid Test Cassette is a rapid chromatographic immunoassay for the qualitative detection of SARS-CoV-2 antigen in Nasal swab. The identification is based on monoclonal antibodies specific to the Nucleocapsid
N protein of SARS-CoV-2. It is intended to aid in the rapid differential diagnosis of COVID-19 infection.
SUMMARY
The novel coronaviruses belong to the β genus. COVID-19 is an acute respiratory infectious disease. Currently, patients infected by the novel coronavirus are the main source of other people getting infected; asymptomatic and symptomatic infected people can both be an infectious source. Based on the current epidemiological investigations, the incubation period is 1 to 14 days, mostly 3 to 7 days. The main manifestations include fever, fatigue and dry cough. Nasal congestion, runny nose, sore throat, myalgia and diarrhea are found in a few cases.
PRINCIPLE
The COVID-19 Antigen Rapid Test Cassette (Nasal Swab) is a qualitative, lateral flow immunoassay for the detection of the N protein of SARS-CoV-2 in Nasal Swab. In this test, antibody specific to the N protein of SARS-CoV-2 is separately coated on the test line regions of the test cassette. During testing, the extracted specimen reacts with the antibody to N protein of SARS- CoV-2 that are coated onto particles. The mixture migrates up the membrane to react with the antibody to N protein of SARS-CoV-2 on the membrane and generate one colored line in the test regions. The presence of this colored line of the test regions indicates a positive result. To serve as a procedural control, a colored line will always appear in the control region if the test has performed properly.
REAGENTS
The test cassette contains anti-SARS-CoV-2 Nucleocapsid protein conjugated gold colloid particles and anti-SARS-CoV-2 Nucleocapsid protein antibody coated onto the membrane.
PRECAUTIONS
Please read all the information in this package insert before performing the test.
- For use only by individuals given appropriate training for in vitro diagnostic use. Do not use after the expiration date.
- The test should remain in the sealed pouch until ready to use.
- All specimens should be considered potentially hazardous and handled in the same manner as an infectious agent.
- Used tests should be discarded according to the local regulations.
- Avoid using bloody samples.
- Wear gloves if possible. Alternatively, sterilize hands with alcohol gel before and after handling the samples, avoid touching the reagent membrane and sample well.
STORAGE AND STABILITY
Store as packaged at room temperature, or refrigerated (2-30℃). The test is stable through the expiration date printed on the sealed pouch. The test must remain in the sealed pouch until use. DO NOT FREEZE. Do not use beyond the expiration date.
MATERIALS
Materials provided
- Test cassettes
- Exaction Reagents
- Extraction Tubes
- Dropper tips
- Sterile Nasal Swabs
- Package Insert
- Workstation
Materials required but not provided
- Timer and Gloves/PPE
SPECIMEN COLLECTION AND PREPARATION
- l Nasal Swab
Use the nasal swab supplied in the kit. Prior to collecting the nasal swab, the patient should be instructed to blow their nose. To collect a nasal swab sample, insert the entire absorbent tip of the nasal swab (usually ⅗ to 1 of an inch (1.5 to 2.5cm) )inside the nostril and firmly sample the nasal wall by rotating the swab in a circular path against the nasal wall at least 5 times. Take approximately 15 seconds to collect the sample per nostril. Be sure to collect any nasal drainage that may be present on the swab. Sample both nostrils with the same swab before testing.
DIRECTIONS FOR USE
Allow the test, specimen, and extraction buffer to equilibrate to room temperature (15-30℃) prior to testing.
- Remove the test cassette from the sealed foil pouch and use it within one hour. Best results will be obtained if the assay is performed immediately after opening the foil pouch.
- Place the extraction tube in the workstation. Turn the hand over and add all extraction buffer to the extraction tube. See illustration 1&2.
- Place the swab specimen in the extraction tube. Rotate the swab for approximately 10 seconds while pressing the head against the inside of the tube to release the antigen in the swab. See illustration 3.
- Remove the swab while squeezing the swab head against the inside of the Extraction Tube as you remove it to expel as much liquid as possible from the swab. Discard the swab in accordance with your biohazard waste disposal protocol. See illustration 4.
- Fit the dropper tip on top of the extraction tube. Place the test cassette on a clean and level surface. See illustration 5
- Add 3 drops of the solution (approx.80μL) to the sample well and then start the timer. Read the result at 10 minutes. Do not interpret the result after 20 minutes
INTERPRETATION OF RESULTS
(Please refer to the illustration above)
**POSITIVE:* Two lines appear. One coloured line should be in the control
line region (C) and another apparent coloured line should be in the test line
region (T). A positive result indicates that SARS-CoV-2 was detected in the
specimen.
NOTE: The intensity of the colour in the test line region (T) will vary
depending on the concentration of SARS-CoV-2 Antigen present in the specimen.
Therefore, any established using limiting dilutions of an inactivated viral
sample. The control (ZeptoMetrix, 0810587CFHI) was supplied at a concentration
of 1.15 x shade of colour in the test line region (T) should be considered
positive.
NEGATIVE: One coloured line appears in the control line region (C). No
line appears in the test line region (T). A negative result indicates that
SARS-CoV-2 antigen is not present in the specimen, or is present below the
detectable level of the test.
INVALID:** Control line fails to appear. Insufficient specimen volume or
incorrect procedural techniques are the most likely reasons for control line
failure. Review the procedure and repeat the test with a new test. If the
problem persists, discontinue using the test kit immediately and contact your
local distributor.
QUALITY CONTROL
A procedural control is included in the test. A coloured line appearing in the control line region (C) is considered an internal procedural control. It confirms adequate membrane wicking.
Control solutions are not provided in this kit. It is however recommended where possible, for positive and negative controls to be tested as good laboratory practice to confirm the test procedure and to verify proper test performance.
LIMITATIONS
- The COVID-19 Antigen Rapid Test Cassette is for use only by individuals who have been given appropriate training for in vitro diagnostic use. Neither the quantitative value nor the rate of increase in SARS-CoV-2 virus concentration can be determined by this qualitative test.
- The accuracy of the test depends on the quality of the swab sample. False negatives may result from improper sample collection or storage.
- The COVID-19 Antigen Rapid Test Cassette (Nasal Swab) will only indicate the presence of SARS-CoV-2 in the specimen from both viable and non-viable SARS-CoV-2 coronavirus strains.
- As with all diagnostic tests, all results must be interpreted together with other clinical information available to the physician.
- A negative result obtained from this kit should be confirmed by PCR, and/or should be interpreted and followed up in line with national/regional guidance. A negative result may be obtained if the concentration of the SARS-CoV-2 virus present in the swab is not adequate or is below the detectable level of the test.
- Excess blood or mucus on the swab specimen may interfere with test performance and may yield a false positive result.
- A positive result for SARS-CoV-2 does not preclude an underlying co-infection with another pathogen, therefore the possibility of an underlying bacterial infection should be considered.
- Negative results do not rule out SARS-CoV-2 infection, particularly in those who have been in contact with the virus. Follow-up testing with a molecular diagnostic should be considered to rule out infection in these individuals.
- Positive results may be due to current infection with acute non-SARS-CoV-2 coronavirus strains, such as coronavirus HKU1, NL63, OC43 or 229E.
- Results from antigen testing should not be used as the sole basis to diagnose or exclude SARS-CoV-2 infection or to inform infection status.
- The extraction reagent has the ability to kill the virus, but it cannot inactivate 100% of the virus. The method of inactivating the virus should be referred to as recommended by WHO/CDC, or according to local regulations.
PERFORMANCE CHARACTERISTICS
Sensitivity, Specificity and Accuracy
Nasal Swab:
The COVID-19 Antigen Rapid Test Cassette (Nasal Swab) has been evaluated with specimens obtained from different clinical sites where the specimens were collected with Nasal Swabs. The Nasal Swabs were randomised and blind tested by operators following the instructions for use. RT-PCR was used as the reference method for the COVID-19 Antigen Rapid Test Cassette. Specimens were considered positive if PCR indicated a positive result.
Method | RT-PCR | Total Results |
---|---|---|
COVID-19 Antigen Rapid Test Cassette | Results | Positive |
Positive | 96 | 2 |
Negative | 7 | 248 |
Total Results | 103 | 250 |
Relative Sensitivity: 93.2% (95%CI:86.5%-97.2%) Relative Specificity: 99.2%
(95%CI:97.1%-99.9%) Relative accuracy: 97.5% (95%CI:95.2%-98.8%)
Confidence Intervals
Detection Limit
The LOD for the COVID-19 Antigen Rapid Test Cassette (Nasal Swab) was 107TCID50/mL. The Estimated LOD is 1000 TCID50/mL.
Cross Reactivity
Cross reactivity of COVID-19 Antigen Rapid Test Cassette was evaluated by testing commensal and pathogenic microorganisms that may be present in the nasal cavity. Each of the organism, viruses, or yeast were tested in the absence or presence of inactivated SARS-CoV-2 virus. No cross-reactivity result was found with the following Cross-reactive substances when tested at the concentration or viral load presented in the table below
Cross-reactive substances | Test Concentration or Viral Load |
---|---|
Parainfluenza Virus Type4a | 1.6 x 104TCID50/mL |
Human Coxsackievirus | 2.8 x 106 TCID50/mL |
Mumps virus | 1.6 x 105TCID50/mL |
Rhinovirus | 2.8 x 107TCID50/mL |
Haemophilusparainfluenzae | 6 x 107 bacterial/ml |
Staphylococcus aureus | 6 x 107 bacterial/ml |
Neisseria meningitides | 6 x 107 bacterial/ml |
Streptococcus sp. group A | 1 x 109org/ml |
Streptococcus sp. group B | 6 x 107 bacterial/ml |
Streptococcus sp. group C | 6 x 107 bacterial/ml |
Influenza A virus (H3N2)A/Aichi/2/68 | 2.3 x 108CEID50/ mL |
Adenovirus (e.g. C1 Ad. 71)-Type 7A | 1.41 x 105 U/ml |
Enterovirus (e.g. EV68) | 5.01 x 105 TCID50/ml |
Human Metapneumovirus (hMPV) | 3.80 x 106 TCID50/ml |
Influenza A H1N1 (New Cal/20/99) | 1.15 x 107 U/ml |
Influenza B (Florida/02/06) | 1.41 x 105 U/ml |
Parainfluenza virus 1 | 9.12 x 108 TCID50/ml |
Parainfluenza virus 2 | 1.15 x 107 U/ml |
Parainfluenza virus 3 | 6.61 x 106 U/ml |
Parainfluenza virus 4 | 2.82 x 107 U/ml |
Respiratory syncytial virus-Type A | 3.80 x 106 U/ml |
Rhinovirus (Type 1A) | 3.55 x 105 U/ml |
Bordetella pertussis | 1.13 x 1010 CFU/ml |
Candida albicans | 6.27 x 108 CFU/ml |
Haemophilus influenzae | 5.43 x 108 CFU/ml |
Legionella pneumophila | 1.88 x 1010 CFU/ml |
Mycobacterium tuberculosis | 6.86 x 107 CFU/ml |
Mycoplasma pneumoniae | 3.16 x 108 CCU/ml |
Pneumocystis jirovecii (PJP)-S. | 3.45 x 108 CFU/ml |
Pseudomonas aeruginosa | 8.44 x 109 CFU/ml |
Staphylococcus epidermis | 1.21 x 1010 CFU/ml |
Streptococcus pneumoniae | 3.23 x 108 CFU/ml |
Streptococcus pyogenes | 1.64 x 109 CFU/ml |
Streptococcus salivarius | 8.17 x 108 CFU/ml |
Human coronavirus 229E | 4.17 x 105 TCID50/ml |
Human coronavirus OC43 | 1.05 x 106 TCID50/ml |
Human coronavirus NL63 | 1.70 x 105 TCID50/ml |
MERS-coronavirus | 3.16 x 106 TCID50/ml |
Staphylococcus aureus, Titered | 1.84 x 1010CFU/ml |
Chlamydophila pneumoniae, Titered | 1.75 x 108IFU/ml |
BIBLIOGRAPHY
- Weiss SR, Leibowitz JL. Coronavirus pathogenesis. Adv Virus Res 2011;81:85-164.
- Cui J, Li F, Shi ZL. Origin and evolution of pathogenic coronaviruses. Nat Rev Microbiol 2019;17:181-192.
- Su S, Wong G, Shi W, et al. Epidemiology, genetic recombination, and pathogenesis of coronaviruses. Trends Microbiol 2016;24:490-502.
Index of Symbols
Hangzhou Biotest Biotech Co., Ltd. 17#, Futai Road, Zhongtai Street, Yuhang
District, Hangzhou, P. R. China
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537
Hamburg, Germany