coVita Micro plus Basic, Baby and Pro Smokerlyzer User Guide

June 3, 2024
coVita

Micro plus Basic, Baby and Pro Smokerlyzer
User Guide

Calibration Tips

coVita Micro plus Basic, Baby and Pro Smokerlyzer - Figure
1

  1. coVita calibration instructional videos can be found under the help section of www.covita.net .
  2. You must calibrate every 6 months to correct for sensor drift using 20 ppm Carbon Monoxide (CO) calibration gas. On occasion your Micro+™ may prompt you for an  intermittent calibration.
  3. Your Micro+™ has built in calibration reminders to prompt you to calibrate every 6 months.
  4. Be sure to identify your flow meter-regulator for correct instructions.
  5. Remove flow meter-regulator from can for proper storage and replace dust cap. (You may hear a pop and this is normal.)
  6. Ensure that the cali-adaptor tubing has no kinks while the gas is flowing.
  7. Be sure you are using the compatible cali-adaptor tubing and D-piece to avoid issues with calibrating.
  8. Be sure you are inserting the D-piece into the monitor correctly. The circle opening must be facing you. (Figure A & B)
  9. Never block the air/vent hole on the Micro+™ with labels etc.

Required Parts

coVita Micro plus Basic, Baby and Pro Smokerlyzer - Figure
2

A. 20 ppm Carbon Monoxide (CO) Calibration Can
B. Fixed Flow 1 lpm meter-regulator
C. Cali-adaptor Tubing
D. D-Piece+™
E. Micro® Smokerlyzer

Calibration Kit Set Up

  1. Remove white or black dust cap from can of gas
  2. Ensure the control valve on the flow meter-regulator is in the off position.
  3. Screw the regulator-flow meter onto the gas can (some slight pressure may be required). This can be done by screwing the gas can into the valve.
  4. Attach the small end of the tubing onto the regulatorflow meter.
  5. Insert the larger end of the cali-adaptor tubing into the D-piece. Ensure a tight fit.
  6. Verify the gauge on the regulator-flow meter is registering gas (arrow is not on zero)

Calibration

IMPORTANT:
Do not connect the cali-adaptor or D-piece to the monitor until after the monitor zeros (Step 5)coVita Micro plus Basic, Baby and Pro Smokerlyzer -
Figure 4

Additional Screens & Tips

coVita Micro plus Basic, Baby and Pro Smokerlyzer - Figure
5

Reset & Calibration

coVita Micro plus Basic, Baby and Pro Smokerlyzer - Figure
6

Additional Screens & Tips

coVita Micro plus Basic, Baby and Pro Smokerlyzer - Figure
7

www.covita.net | 800.707.5751
Copyright 2022 coVita LLC All Rights Reserved v082522

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals