NANOCOLOR Sulfate LR 200 Sulfur Analyzer Instructions
- September 27, 2024
- NANOCOLOR
Table of Contents
Sulfate LR 200 Sulfur Analyzer
“`html
Product Information
Specifications:
- Product Code: 985062
- Test Type: 0-62
- Reagent: Sulfate LR 200
- Method: Photometric determination as barium sulfate
- Sample pH Range: 2 to 11
- Not suitable for sea water analysis
Product Usage Instructions
Procedure:
-
Prepare requisite accessories: piston pipette with tips.
-
Open a test tube and add 1.0 mL of Sulfate LR 200 R2, then
close and mix. -
Add 4.0 mL of the sample solution (ensure the pH is between 2
and 11), close and mix. -
Clean the outside of the test tube and measure the solution
after 15 minutes.
Method:
The method involves photometric determination using barium
sulfate.
Contents of Reagent Set:
- 20 test tubes of Sulfate LR 200 R1
- 2 test tubes of Sulfate LR 200 R2
Note:
This method is not suitable for analyzing sea water samples.
FAQ (Frequently Asked Questions)
Q: Can this product be used for sea water analysis?
A: No, this product’s method is not suitable for sea water
analysis.
Q: What is the pH range suitable for the sample solution?
A: The sample’s pH must be between 2 and 11 for accurate
results.
“`
REF 985062
Test 0-62
03.23 Sulfate LR 200
en
Procedure: Requisite accessories: piston pipette with tips
Open test tube, add 1.0 mL Sulfate LR 200 R2, close and mix. Add 4.0 mL sample
solution (the pH value of the sample must be between pH 2 and 11) close and
mix.
Clean outside of test tube and measure after 15 min.
Measurement: For MACHEREY-NAGEL photometers see manual, test 0-62. Measurement
when samples are colored or turbid: For all NANOCOLOR ® photometers see
manual, use key for correction value. Photometers of other manufacturers: For
other photometers check whether measurement of round glass tubes is possible.
Verify factor for each type of instrument by measuring standard solutions.
Analytical quality control: NANOCONTROL Standard Sulfate LR 200 (REF 92562),
NANOCONTROL Multistandard Metals 1 (REF 925015) or NANOCONTROL Multistandard
Drinking water (REF 925018).
Method: Photometric determination as barium sulfate.
Range: Wavelength (HW = 5 12 nm): Reaction time: Reaction temperature:
20 200 mg/L SO42 585 nm 15 min (900 s) 20 25 °C
Contents of reagent set: 20 test tubes Sulfate LR 200 R1
2 test tubes Sulfate LR 200 R2
Hazard warning: Reagent R1 contains barium chloride 3 25 %. H302, H412
Harmful if swallowed. Harmful to aquatic life with long lasting effects. P273,
P301+312, P330 Avoid release to the environment. IF SWALLOWED: Call a POISON
CENTER / doctor / … / if you feel unwell. Rinse mouth.
Preliminary tests: If the order of magnitude of the concentration in a sample
is not known, a preliminary test with QUANTOFIX® Sulfate (REF 91329) or with
VISOCOLOR® ECO Sulfate (REF 931092) rapidly gives this information. From the
order of magnitude the required dilution can be calculated and prepared
directly.
Interferences: Turbidities of sample interfere and test sample must first be
filtered before the determination. In drinking, surface and ground water the
test results are accurate. The following quantities of ions will not
interfere: 100 mg/L Fe, CO32, 500 mg/L Zn2+ , 1000 mg/L Ca2+ , Cl.
The method can not be applied for the analysis of sea water.
MACHEREY-NAGEL GmbH & Co. KG · Valencienner Str. 11 · 52355 Düren · Germany
Tel.: + 49 24 21 969-0 · info@mn-net.com · www.mn-net.com
PD 14122 / A054780 / 985062 / xxxx