GE Healthcare LOGIQ Series Ultrasound Guided Attenuation Parameter
YASIN INCOMPLET POST
Introduction
The prevalence of metabolic dysfunction-associated steatotic liver disease
(MASLD) is growing worldwide with the increase in obesity.1
Among the many forms of MASLD, metabolic steatohepatitis (MASH) has attracted
attention, as it can progress to liver cirrhosis and hepatocellular carcinoma
due to hepatocyte apoptosis, inflammation and fibrosis.2 Traditionally, liver
biopsy has been the gold standard for the diagnosis and assessment of hepatic
steatosis. However, this method has some problems such as sampling error and
inter-pathologist variability.3 In addition, the invasive nature of the
procedure creates a risk of complications. More recently, MR proton density
fat fraction (PDFF) has been accepted as a non-invasive reference standard,
but limited access and high cost prevent widespread use, especially with
respect to regular follow up exams.
The liver echogenicity on ultrasound B-Mode is widely used for the detection
of hepatic steatosis. However, this technique does not enable a quantitative
assessment since liver texture or brightness may vary depending on the imaging
parameters used or the examiner’s technique. Therefore, an objective
ultrasound quantification method is desired for steatosis grading in current
and potential MASH patients. Recently, a novel non-invasive tool that utilizes
attenuation of the sound wave was developed.
However, it may be susceptible to multi-reflection artifacts from subcutaneous
tissues as well as disruptive structures such as vessels or diaphragm since
the measurement area is not guided by imaging.4
This paper describes Ultrasound-Guided Attenuation Parameter (UGAP), a real- time, image-guided method of measuring the attenuation of the sound wave. The principles of the method as well as the clinical evaluation results are presented.
Ultrasound attenuation
When the ultrasound wave propagates in an organ, such as the liver, it is gradually weakened due to diffusion, scattering and absorption. Known as sound attenuation, this results in less signal returning to the ultrasound transducer, causing the image to get darker with depth. If an image of a healthy liver has uniform image brightness over depth, it is because of the time-gain compensation (TGC) capability of the ultrasound scanner, which applies a different gain for each depth (Figure 1). In the case of fatty liver, the presence of many lipid droplets in the hepatocytes becomes the dominant factor of the attenuation, sometimes causing insufficient echo signals in the deeper area. The amplitude of ultrasound wave u propagating in the x direction is expressed u = u0e-αx , where u0 is the amplitude at x = 0 and α is attenuation rate. Since the sound attenuation increases nearly proportional to the frequency (ranging between 1 MHz to 10 MHz), the attenuation rate can be approximated by α = α0f, where α0 is the attenuation coefficient (dB/cm/MHz) and f is frequency in [MHz]. 5 Attenuation rate and attenuation coefficient are used primarily to evaluate ultrasound attenuation in human tissues.
Figure 1. Ultrasound attenuation compensation by TGC.
Ultrasound attenuation evaluation method by UGAP
Principle of measurement6
As shown in Figure 1, measuring the attenuation slope would provide insight
into liver attenuation. However, the sound profile
in the real signal is not so simple, since it is curved as shown in Figure 2A.
This complexity is caused by a focused sound beam from transmission as well as
reception conditions. To cancel out or compensate for this complexity, several
methods have been reported.7 UGAP performs the compensation based on a
Reference-Phantom Method (RPM)8 (Figure 2). The profile of the echo amplitude
for a tissue-mimicking phantom is measured in the depth direction and stored
in the ultrasound system as a reference.
In this case, a frequency of 3.5 MHz is used. This industry-standard phantom
includes glass bead particles for attenuating materials and the attenuation
coefficient is known. In the UGAP mode, the transmission and reception
conditions are fixed to the same values as were used on the reference phantom,
and the acquired echo profiles of the target (liver) are compensated by the
reference data. As a result, the compensated sound profiles represent only
decay caused by attenuation. If the compensated sound profile is flat, the
attenuation is the same as the reference phantom.
Figure 2. Compensation of ultrasound signal to enable UGAP measurement.
Ultrasound-Guided Attenuation Parameter | LOGIQ E10, LOGIQ E10s, LOGIQ Fortis, LOGIQ Totus and LOGIQ P Series ultrasound systems
Measurement algorithm
Although the system-dependent sound profile is compensated, there are still problems in performing a successful measurement. For example, structures such as large vessels and diaphragm may deform the slope profile. Or, multi- reverberation in the subcutaneous fat may generate artifacts into the liver parenchyma. Or, information needed to determine the slope may be diminished if the attenuation is very large. To avoid these problems, UGAP includes an automated measurement algorithm to find and then analyze the optimum measurement range. The start point of the range is determined by analyzing linearity and discontinuity of the echo profile close to the liver surface, thereby avoiding the multi-reverberation artifacts. In addition, the algorithm automatically detects and avoids depths where the signal-to-noise ratio (SNR) is insufficient. This enables the algorithm to employ the deepest, usable end point. The diaphragm is also automatically excluded. Finally, the angle of the slope is measured across this optimum range to provide a representative attenuation coefficient. Because this measurement takes place on the raw data of a frozen or recalled image, it is not dependent on gain or other post processing settings. The goal of these automations is to make the UGAP measurement less dependent on the ROI position and more robust across various liver sizes and conditions.
Figure 3. Example of B-Mode/color dual display format.
Color mapping for measurement guidance To find the right scan-plane and ROI
for measurement, B-Mode and two types of new color-mapped images are available
in UGAP mode (Figure 3).
(a) Attenuation map: Displays color-coded local attenuation values for
each pixel. When the measurement area (denoted by a trapezoid with a
centerline in Figure 3) has a uniform color, it is suitable to measure.
The color will become inhomogeneous if the area includes a disruptive
structure such as a large vessel.
(b) Quality map: Displays a color at pixels where signal quality is
sufficiently high to perform a measurement. Even though the B-Mode texture may
look homogeneous, a lack of color could be the result of unseen artifacts.
To aid in the acquisition and measurement of UGAP, various display formats are
selectable: B-Mode only, B-Mode with color map overlay, and a dual display
that shows both images side-by-side. Figure 3 shows an example of the
B-Mode/color map dual display. Examples of the color-coded attenuation for
different degrees of steatosis are shown in Figure 4.
Normal liver
Mild fatty liver (NAFLD)
Moderate fatty liver (NAFLD)
Severe fatty liver (NAFLD)
Reference image B-Mode
Reference image attenuation map
Reference image quality map
Figure 4. Examples of color-mapped attenuation images at different degrees of liver fat. Images courtesy of Prof. Sporea, Victor Babes University of Medicine and Pharmacy of Timisoara.
Evaluation of steatosis grade in chronic liver disease by UGAP
This section summarizes the results of the multicenter prospective study reported by Nayoro City General Hospital, Ogaki Municipal Hospital, Yokohama City University Hospital, Tokyo Medical University Hospital, Iwate Medical University, Musashino Red Cross Hospital and Gifu Kyoritsu University Hospital.9, 10
Materials and methods
A total of 1,010 patients with chronic liver disease who underwent MRI-PDFF
(proton density fat fraction) and UGAP were enrolled. PDFF was measured using
a multi-echo Dixon method (IDEAL-IQ sequence) with a single region of interest
(ROI) (20 × 20 × 20 mm) placed in liver segment VII or VIII. (An example is
shown in Figure 5A). Patients’ sex, age, etiology, body mass index (BMI), skin
capsule distance and steatosis grade are shown in Table 1.
Variable | Value |
---|---|
Subject (n) | 1,010 |
Sex (male/female) | 544/468 |
Age (y) | 61.7 (52.0 – 72.0) |
Etiology (NAFLD/ALD/HBV/HCV/AIH/others) | 515/90/133/124/38/110 |
BMI (kg/m2) | 26.2 (23.1 – 28.9) |
Skin capsule distance (mm) | 19.0 (15.5 – 22.0) |
Steatosis grade (%)
S0 (< 5.2%)| 356 (35.2%)
S1 (5.3% < PDFF < 11.3%)| 281 (27.8%)
S2 (11.4% < PDFF < 17.1%)| 168 (16.6%)
S3 (PDFF > 17.2%)| 205 (20.3%)
Table 1. Patients’ sex, age, etiology, BMI, skin capsule distance and steatosis grade.
UGAP measurements were performed using software equivalent to LOGIQ E10, LOGIQ
E10s and LOGIQ Fortis.
All measurements were performed within three months before or after MRI-PDFF,
in fasting conditions for more than four hours, on patients in a supine
position, with the right arm in maximum abduction, by intercostal approach, in
the right liver lobe. A colored-coded attenuation map was used during
measurement to confirm a homogenous area of the liver, free of large vessels.
(An example is shown in Figure 5B). At least six measurements were performed.
Reliable UGAP measurements were defined as the median value of six
measurements performed in a homogeneous area of liver parenchyma, with an
IQR/M < 0.30. UGAP values are expressed in dB/cm/MHz.
Figure 5. Region of interest (ROI) and attenuation map of (A) MRI-based proton density fat fraction (PDFF) and (B) Ultrasound Guided Attenuation Parameter (UGAP) measurements.
Results
All UGAP measurements were successful for 1010 cases. The mean MRI-PDFF was 10.8 ± 8.05% (range: 0.3 – 44.81%), and the mean UGAP value was 0.69 ± 0.12 dB/cm/MHz (range: 0.32 – 0.99 dB/cm/MHz). The relationship between UGAP and PDFF is shown in Figure 6. Significantly strong correlation was confirmed between UGAP and PDFF, Spearman’s rank correlation coefficient was 0.785 (p < 0.001).
UGAP values in each steatosis grade are shown in Figure 7. The average values of UGAP in S0, S1, S2 and S3 steatosis were 0.59, 0.69, 0.77, and 0.83 dB/cm/MHz, respectively, as UGAP increased with increasing steatosis grade.
Receiver operating characteristics (ROC) curve of UGAP for diagnosis of S1 or higher, S2 or higher and S3 steatosis are shown in Figure 8: (A): ≥ S1, (B): ≥ S2, (C): S3. Area under the ROC curve (AUROC), 95% confidence interval (95% CI), and cutoff value of UGAP are shown in Table 2. The AUROCs of UGAP for the prediction of S1 or higher, S2 or higher and S3 steatosis were 0.901 (95% CI: 0.891 – 0.928), 0.912 (95% CI: 0.894 – 0.929) and 0.894 (95% CI: 0.873 – 0.916), respectively.
Figure 6. Relationship between UGAP and PDFF
Figure 7. Measured UGAP values in each steatosis grade.
Figure 8. ROC of UGAP for diagnosis of steatosis grade (A) ≥S1, (B) ≥S2 and (C) S3.
UGAP
≥ S1 | ≥ S2 | S3 | |
---|---|---|---|
AUROC (95% CI) | 0.901 (0.891 – 0.928) | 0.912 (0.894 – 0.929) | 0.894 (0.873 – |
0.916)
Attenuation coefficient cutoff value (dB/cm/MHz)| 0.65| 0.71| 0.77
Attenuation rate cutoff value (dB/m)| 228| 249| 270
Table 2. AUROCs, 95% CI and cutoff values of UGAP for the prediction of ≥ S1, ≥ S2 and S3 steatosis.
Discussion
In this prospective study based on more than 1,000 examinations – the largest
to date – the results showed that UGAP has high accuracy at diagnosing and
grading steatosis with MRI-PDFF as reference standard. Liver biopsy is still
considered the gold standard for the evaluation of hepatic steatosis, but it
is painful and costly.
In addition, MRI-PDFF has the advantage of accounting for spatial variability
of hepatic steatosis compared to biopsy which has potential for sampling error
and may result in grading inaccuracies.10 Recently, there have been several
reports showing the utility of MRI-based PDFF for diagnosing hepatic
steatosis, including reports showing the superiority of MRI-PDFF to VCTE-based
CAP for diagnosing hepatic steatosis grade in NAFLD patients who underwent
liver biopsy.11 Therefore, MRI-PDFF is a precise and accurate non-invasive
imaging biomarker for the diagnosis of hepatic steatosis.
The result of this study indicated that UGAP has significant linearity and negligible bias with respect to the reference standard of MRI-PDFF measurements over the entire range of observed steatosis severity. Therefore, it can be concluded that UGAP has excellent technical performance characteristics as a quantitative method of steatosis for widespread use in clinical trials and patient care.
LOGIQ Totus
Evaluation of the LOGIQ Totus comparing to LOGIQ Fortis
Alina Popescu, Felix Bende, Ioan Sporea et. al. Department of Gastroenterology
and Hepatology, “Victor Babeș” University of Medicine and Pharmacy Timișoara,
Romania.
Background
The LOGIQ Totus is equipped with the same UGAP technology as the LOGIQ E10, LOGIQ E10s and LOGIQ Fortis.
In addition, the same C1-6-D and C1-6VN-D probes are optimized for UGAP. UGAP workflow features, such as measurement tools, worksheet pages, automatic statistics on the summary pages, and the color mapping for measurement guidance are identical on the LOGIQ E10, LOGIQ E10s, LOGIQ Fortis and LOGIQ Totus.
A clinical study was performed to compare liver measurement values on the LOGIQ Fortis versus LOGIQ Totus and determine whether the LOGIQ E10, LOGIQ E10s, LOGIQ Fortis cutoff chart can be used for LOGIQ Totus.
Materials and methods
Subjects
141 patients were recruited. The average age and BMI of the subjects were 56
years (age range 42–68 years) and 28.7 (BMI range 24.1–33.3), respectively.
The subject demographics and disease etiology are summarized in Table 3.
Variable | Value |
---|---|
Subject (n) | 141 |
Sex (male/female/unknown) | 73/63/5 |
Age (y) | 55.8 (42.7 – 68.9) |
Etiology (NAFLD/ALD/HBV/HCV/ASH/others) | 50/8/27/15/7/17 |
BMI (kg/m2) | 28.7 (24.1 – 33.3) |
Steatosis grade by LOGIQ Fortis cut-off chart (%)
S0| 61 (43.3%)
S1| 27 (19.1%)
S2| 19 (13.5%)
S3| 34 (24.1%)
Table 3. Patients’ sex, age, etiology, BMI and steatosis grade as determined by LOGIQ Fortis cut-off values.
References