MOOG Infinity Orange Small Volume Enteral Feeding Pump User Guide

September 8, 2024
moog

MOOG Infinity Orange Small Volume Enteral Feeding Pump

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-\(15\)

Specifications:

  • Product Name: Infinity Orange Small Volume Enteral Feeding Pump
  • Feeding System: Enteral-only
  • Optimized for: Small volume feedings and increased safety

Product Usage Instructions

Programming the Pump:
To program the pump, refer to the operator’s manual for detailed instructions. Typically, you will need to set parameters such as rate (mL/hr), dose (mL), and bag volume.

Cleaning:
Follow the cleaning instructions provided in the operator’s manual. Ensure proper sanitation to prevent contamination.

Alarms & Messages:
If the pump triggers an alarm, refer to the manual to understand the message displayed. You can silence the alarm by pressing the RUN/PAUSE button.

Growing with Infinity:
The Infinity Orange pump is designed to adapt to your changing needs. Consult the operator’s manual for guidance on adjusting settings as required.

Troubleshooting:
If you encounter issues with the pump, visit moogmedical.com for troubleshooting assistance or contact Moog’s Clinical Support at 800.970.2337.

FAQ:

  1. Who should I contact for medical concerns?
    For medical concerns, contact your medical team or primary care provider.

  2. How do I obtain supplies?
    Contact your home care provider for supplies. Ensure you have ample supplies on hand and contact your provider ahead of time to avoid running out.

  3. What should I do if the pump alarms?
    If the pump alarms, check the message on the screen, silence the alarm by pressing the RUN/PAUSE button, and refer to Moog’s Clinical Support for assistance.

Infinity Orange is an enteral-only feeding system that is optimized for small volume feedings and increased safety. It is ideal for use in many environments where precise delivery of small volumes of human milk or formula is required.

USING THIS GUIDE

Receiving a medical device for a loved one can be overwhelming. As the manufacturer of the Infinity Orange Small Volume Enteral Feeding Pump, we want to ensure that your experience with our pump is positive. We want to take away as much of the stress and worry about using an enteral feeding pump as we can so that your loved one can focus on being healthy. This guide provides you an overview of your new Infinity Orange pump. Remember, this guide is not intended to be a substitute for the operator’s manual. Be sure to read the operator’s manual thoroughly prior to operating the pump as it includes more in-depth details on using the pump.

Here are some highlights of your Infinity Orange pump: Your Infinity Orange pump is portable, it can operate in any orientation (even upside down), weighs less than one pound, and can be carried in a discrete backpack. It is easy-to- use with a very simple user interface. It is accurate up to ± 5%. Perhaps most importantly, Infinity Orange pumps are designed to be rugged and withstand everyday use, including being washed under running water. If you ever have questions or concerns, call our clinical support hotline. This support line is available
24 hours per day, 7 days per week for anyone who needs assistance with any of our pumps.

800.970.2337
24 HOUR Clinical & Technical Support
For after hours support, dial extension 240
Where training videos are available to supplement the information contained in this guide, you will see a QR code (similar to what is pictured below). You can scan these codes with a smartphone or tablet camera to watch a related video. If you are reading this guide on a computer or phone, clicking the orange buttons will take you to the same video.

Download and view the Infinity Orange operator’s manual by scanning this code or by clicking the button below.

CLICK HERE TO DOWNLOAD
WHO TO CALL FOR HELP

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-\(5\)

For supplies, contact your home care provider. Be sure to contact your home care provider at least days ahead of needing supplies so that you do not run out. Always keep
an extra days of supplies and formula on hand in case of an emergency.

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-\(7\)

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-\(8\)If your pump is alarming or if you have other pump-related questions, contact Moog’s Clinical Support. Be sure to read the message on the screen and write it down. You can silence the alarm by pressing the RUN/PAUSE button. You can also find troubleshooting assistance at moogmedical.com.

800.970.2337
Moog Clinical Support phone number

YOUR FEEDING PLAN

Rate
This is how fast the pump will deliver the formula.

Dose
This is how much formula will be fed.

Bag volume
This is how much formula you should add to your feeding bag at one time

Refill volume
If prescribed, this is how much formula you refill in the delivery set.

Refill interval
If prescribed, this is how often (in hours) you should refill the delivery set with formula.

Max hang time
The maximum amount of time the formula should be hanging in the bag, per the formula manufacturer.

INFINITY ORANGE PUMP & DELIVERY SETS

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-\(15\)

Pump Keypad

  • ON/OFF: press for 1.5 seconds to turn the pump on or off.
  • PRIME: press this button to prime the delivery set. See the priming section on the next page for details.
  • RUN/PAUSE: press to either start the pump or to pause it.
  • FOOD TYPE: press key to display and select the type of food to be delivered (formula or human milk).
  • VOLUME/TOTAL: press once to show volume delivered for the current feeding. Press again to see total volume delivered since it was last cleared.
  • RATE/DOSE: press to toggle between rate and dose settings.
  • CLEAR: press to reset whatever function is displayed to its minimum value. Press and hold for 3 seconds to clear rate, dose, and volume delivered all at once.
  • PLUS: press to increase the displayed feed function. Press and hold to increase rapidly.
  • MINUS: press to decrease the displayed feed function. Press and hold to decrease rapidly.

Note: PRIME, CLEAR,+ (plus) and – (minus) buttons only work when the pump is paused.

Video showing Infinity Orange pump overview

CLICK TO VIEW

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-\(16\)

Delivery Sets
Sets are available with an attached 100 mL bag or a screw cap lid that can be attached to any bottle with a 40 mm opening.

  1. Screw Cap Set (shown with bottle)
  2. 100 mL Bag Set
  3. Funnel / bag opening
  4. Cassette
  5. “ “ Symbol
  6. Tubing
  7. ENFit connector
  8. ENFit protective cover

Note: Any Infinity delivery set (both orange and teal colored) can be used with the Infinity Orange pump.

Video showing Infinity Orange delivery sets

CLICK TO VIEW

PREPARING THE DELIVERY SET FOR USE

If you are using the 100 mL delivery set

  1. Hold the bag upright. Slowly pour the formula or human milk into the bag. Pouring slowly allows the bag to open and helps the formula flow easily into the bag.
  2. Close the cap by aligning the cap on the funnel, and firmly pressing down.

If you are using the Screw Cap delivery set

  1. Fill a feeding bottle or container with 40 mm opening with formula or human milk. Firmly twist the Infinity Orange Screw Cap Set onto the bottle.
  2. Invert the bottle or container and place it in the suspension bag and secure the bag to an IV pole.

Video showing how to fill the 100 mL delivery set bag and the Screw Cap delivery set with formula or human milk

CLICK TO VIEW

PRIMING THE DELIVERY SET

After preparing the delivery set, but before using the pump, you must prime the delivery set. Priming the delivery set is the process of removing air from the tubing before feeding.
There are several different methods for priming the Infinity Orange delivery set. Select the method recommended by your healthcare provider or the method that best fits your needs.

Priming Options

  • Hold to Prime – page 6
  • Auto Prime – page 6
  • Manual priming for IV pole use – page 7
  • Manual priming for backpack use – page 7

Priming using the Hold to Prime feature
Hold to Prime can be used to prime the delivery set if it will be hung on an IV pole. Hold to Prime requires holding your finger on the button throughout the priming process, allowing you to stop priming whenever needed.

  1. Remove the protective cover from the delivery set, and make sure that the delivery set is not connected to the patient.

  2. Press the ON/OFF button for 1.5 seconds to turn the pump on. Seat the cassette in the pump and close the pump door.

  3. Press and hold the PRIME button. The alarm will sound once, and the pump will prime the set for as long as the button is pressed. The display will read “TO STOP LET GO.”

  4. After air is removed from the downstream tubing, release the PRIME button.
    The delivery set is now fully primed.

  5. Both the pump and delivery set can be hung on an IV pole.

Video showing how to prime the set using the Hold to Prime feature on the pump

CLICK TO VIEW
Priming using the Auto Prime feature

Auto Prime can also be used to prime the delivery set if it will be hung on an IV pole. When enabled, auto prime is the easiest priming method as it allows you to prime the set with a touch of a button.

  1. Remove the protective cover from the delivery set, and make sure that the delivery set is not connected to the patient.

  2. Press the ON/OFF button for 1.5 seconds to turn the pump on. Seat the cassette in the pump and close the pump door.

  3. Press and hold the PRIME button for 1.5 seconds. Alarm will sound once, and the pump will begin to prime the set. The display will read “TO STOP PUSH PRIM.” When auto prime is finished, the pump will stop and an alarm will sound.

  4. Remove any remaining air in the tubing by pressing and holding the PRIME button again for 1.5 seconds.
    After air is removed from the downstream tubing, press the PRIME button again to stop the pump. The delivery set is now fully primed.

  5. Both the pump and delivery set can be hung on an IV pole.

Video showing how to prime the set using the Auto Prime feature on the pump

CLICK TO VIEW
Manual priming for IV pole use
Manually priming the delivery set is the fastest way to prepare the set for use. Manual priming can be used when the pump and set will be hung on an IV pole for use. For use with the 100 mL delivery set only.

  1. Remove the protective cover from the delivery set, and make sure that the delivery set is not connected to the patient.

  2. Grasp the bag with one hand while holding the delivery set cassette in the other hand.

  3. Gently pinch the teal colored tubing below the “ “ symbol. Hold this position until air is removed from the downstream tubing. Gently squeeze the bag at the same time to assist fluid flow.
    After air is removed from the downstream tubing, release the “ “ symbol. The delivery set is now fully primed.

  4. Seat the cassette in the pump and close the pump door.

  5. Both the pump and delivery set can be hung on an IV pole.

Video showing how to manually prime the set for IV pole use

CLICK TO VIEW

Note: This video shows Infinity delivery sets being primed, but the steps are the same for Infinity Orange delivery sets.

Manual priming for backpack use
Using the pump and delivery set with a backpack requires the removal of all air in the delivery set bag and tubing. This can be achieved using the manual prime method. For use with the 100 mL delivery set only.

  1. Remove the protective cover from the delivery set, and make sure that the delivery set is not connected to the patient.

  2. Turn the bag upside down so that the tubing port is at the top. Grasp the bag with one hand while holding the delivery set cassette in the other hand.

  3. Gently pinch the teal colored tubing below the “ “ symbol. Hold this position until air is removed from the bag and downstream tubing. Gently squeeze the bag at the same time to assist fluid flow. Tilt the bag as needed to evacuate air from the tubing port.
    After air is removed from the downstream tubing, release the “ “ symbol. The delivery set is now fully primed.

  4. Seat the cassette in the pump and close the pump door.

  5. Both the pump and delivery set can be placed into a backpack for use.

Video showing how to manually prime the set for backpack use

CLICK TO VIEW

Note: This video shows Infinity delivery sets being primed, but the steps are the same for Infinity Orange delivery sets.

PROGRAMMING THE PUMP

After the delivery set is prepared, primed, and either hung on an IV pole or placed into a backpack, follow these basic instructions to program the Infinity Orange pump. For more detailed instructions, refer to the operator’s manual or watch the videos on the moogmedical.com website showing how to program the pump.

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-57

NOTE: If pump is turned off and turned back on again, the volume delivered counter resets to 0 and food type resets to formula. Rate, dose, and food type must be confirmed before restarting the pump with a new feeding.

NOTE: Any changes to pump settings (except food type) during a feeding cycle will cause pump to start a new feeding; it will not start where it left off. Rate, dose, and food type must be confirmed by viewing each parameter before restarting the pump.

CLEANING

Notes about cleaning

WARNING: To avoid electrical shock, never clean pump or Infinity Orange AC Adapter/ Charger with charger plugged into an outlet or pump turned on.

  • Make sure the Infinity Orange AC Adapter/Charger is completely dry before plugging into an electrical outlet.
  • The pump and delivery set must be dry before use to ensure optimal pump performance.
  • If needed, a dry or slightly damp cloth may be used to clean the outside of the power adapter while it is disconnected from a power outlet.
  • Avoid using harsh cleaners or disinfectants. They may cause damage to the pump surfaces.

To clean the pump
The Infinity Orange pump may be cleaned by rinsing under a stream of warm water. Standard dish soap and a non-abrasive sponge or cloth can also be used for deeper cleaning.

MOOG-Infinity-Orange-Small-Volume-Enteral-Feeding-
Pump-\(51\)

Use a cotton swab to clean the pathways around where the cassette is placed into the pump (areas shown in orange in the image above).

Note: Refer to the operator’s manual for a list of acceptable and unacceptable cleaning agents, including cautions and warnings.

Video showing how to clean the pump

CLICK TO VIEW

ALARMS & MESSAGES

Following is a list of alarms and messages seen on the pump, along with some information on why it was displayed and how to fix the situation.

ER01 – ER99
The pump has failed a self-test. Try turning the pump off and then back on.

NO FLOW OUT
The pump detected a blockage in the tubing between the pump and patient. Check for kinks or blockages. Check for cleanliness around the pump sensors.

LOAD SET
The run button was pressed without a set being properly loaded. Pause the pump, ensure that a delivery set is properly loaded, close the door, and try again.

LOW BATT
There is 1 hour or less of battery life remaining. Plug in the AC Adapter/Charger to recharge the battery.

NO FOOD
The pump detected air in the tubing. Check if the food bag is empty or if there are air bubbles in the tubing. If necessary, refill the food bag and prime the set to remove air from the tubing.

PUSH RUN TO FEED
The pump has been in pause mode for 2 minutes. Program and use the pump, or press the ON/OFF button to turn it off.

NO FLOW IN
The pump detected a blockage in the tubing between the pump and food bag. Check for kinks or blockages. Check for cleanliness around the pump sensors.

SHUT DOOR
The pump door was opened while the pump was running. Pause the pump. Check to make sure the door is latched securely.

Other messages
The pump will occasionally display other messages such as DOSE DONE, NEXT DOSE, CHEK, etc. These messages are used to provide updates on feeding updates and pump performance. See the “Alarms, Messages, and Indications” chapter of the operator’s manual for detailed information. You may also refer to the Troubleshooting section on the following page for additional assistance.

View the Infinity Orange Operator’s Manual for additional information on the various alarms, messages, and indications that can be seen on the Infinity Orange pump.

CLICK HERE TO DOWNLOAD
GROWING WITH INFINITY
You may see Infinity pumps that are similar to your Infinity Orange, but are a different color. There are a few significant differences between the pumps, making each ideal for use at different stages of a patient’s life. Here is a quick comparison.

Infinity Orange Small Volume Enteral Feeding Pump
Designed specifically for patients receiving small volume feedings feeding at lower rates (< 300 mL/hr), especially those being fed human milk, The Infinity Orange features several built-in safety features making it ideal for use in hospitals.

Infinity Enteral Feeding Pump
As a patient begins to require larger doses above 100 mL and is expected to be on long-term enteral feedings, moving to the Infinity pump could be considered. The Infinity pump offers a higher maximum rate (600 mL/hr), and allows for convenient infinite dose programming.

View a video highlighting the differences between the pumps and various delivery sets.

CLICK TO VIEW

TROUBLESHOOTING

ER01 ER99 Alarms

Was pump dooropen while pump was turned on or while running?| YES/UNSURE| Turn pump off. Unload and reload delivery set cassette. Make sure door is closed securely, then turn pump back on.
NO| If alarm continues, contact Moog Customer Service at 800.970.2337 for assistance.
LOAD SET Alarm

Is door closed securely?| NO/UNSURE| Turn pump off. Unload and reload delivery set cassette. Make sure door is closed securely, then turn pump back on.
YES|
Is door cracked or tab on side of door broken?| YES| If alarm continues, contact Moog Customer Service at 800.970.2337 to order a new door.
NO| If alarm continues, contact Moog Customer Service at 800.970.2337 for assistance.
NO FOOD Alarm – If re-priming of the pump delivery set is required, it is advisable to disconnect the enteral adapter from the patient’s feeding tube while re-priming.

Is bag of pump delivery set empty?| YES| Refill bag, re-prime delivery set and resume feeding.
NO|
Is air present in the tubing?| YES| Has air been removed from bag of pump delivery set?| NO| Manually prime tubing by turning bag upside down, gently squeezing lower half of bag while pinching the teal colored tubing just below the “” drop symbol. Hold this position until all air is removed from bag and tubing.
YES| Press and hold the PRIME button to move the air in the tubing past pump delivery set cassette and resume feeding.
NOTE: If your pump is equipped with an auto prime option, refer to the Operator’s Manual for details.
NO|
Is formula blenderized or aggressively mixed?| YES| Let formula sit for 10-15 minutes before preparing pump delivery set for feeding. If foam is present in tubing, flush by priming pump delivery set either manually or with the pump and resume feeding.
NO|
Are air bubbles trapped inside the cassette?| YES| Remove tubing from pump and manually prime tubing by turning bag upside down, gently squeezing lower half of bag while pinching the teal colored tubing just below the drop symbol. Hold this position until all air has passed through the teal segment of the tubing.
NO|
Has the appropriate food type setting on the pump been selected for this feeding?| YES| Pause pump and change food type by pressing FOOD TYPE key and either + or to change between formula and human milk.
NO| If alarm continues, contact Moog Customer Service at 800.970.2337 for assistance.
NO FLOW IN or NO FLOW OUT Alarm
NO FLOW IN: Occlusion between delivery set and pump.
NO FLOW OUT: Occlusion between pump and patient.

Is delivery set tubing pinched or kinked?| YES| Check delivery set for pinches or kinks in tubing and resume feeding.
NO|
Is tab on inside of door broken?| YES| Contact Moog Customer Service at 800.970.2337 to order a new door.
NO|
Is pressure sensor region of cassette receptacle clean?| NO| Clean pressure sensor area with a cotton swab, soft cloth, or dampened sponge; or wash entire pump under running water. Do not use abrasive materials or harsh chemicals.
YES|
Are there visible signs of damage to pressure sensors area?| YES| Contact Moog Customer Service at 800.970.2337 to return pump for service
NO| Disconnect feeding set from patient. Press the RUN button. If pump delivers food without alarm, check patient’s feeding tube for obstruction.
If alarm continues, remove set from pump, gently massage the “” drop symbol, then manually prime set until occlusion is cleared and formula advances through tubing. Replace set and resume feeding.
If unable to advance formula through tubing or alarm continues, contact Moog Customer Service at 800.970.2337 to return pump for assistance.
*Note: Do not use the PRIME button to troubleshoot or resolve these alarms.
Charging Problems

Is A/C adapter charger plugged into wall outlet and pump properly?| NO/UNSURE| Check connections for both wall plug and pump adapter. Green indicator light on charger wall plug should be illuminated and pump display should have plug symbol with scrolling bars between E and F of fuel gauge. When pump is fully charged, bars will stop scrolling.
YES|
Is wall outlet functioning properly?| NO/UNSURE| Plug another device into outlet to verify outlet is functioning properly.
YES|
Does charger adapter appear to have any damage?| YES| Contact Moog Customer Service at 800.970.2337 to order a new charger
NO|
Does charger port appear to have any damage including bent or missing pins?| YES| Contact Moog Customer Service at 800.970.2337 to return pump for service.
NO|
Is Battery Symbol and E and F of fuel gauge flashing?| YES| Contact Moog Customer Service at 800.970.2337 to return pump for service.
NO

Speak live with a Moog Clinical Representative for pump questions and troubleshooting guidance 24 hours a day, 7 days a week.

Clinical & Customer Support
800.970.2337
Visit moogmedical.com/enteral/orange for additional information about Infinity Orange enteral feeding pumps.

Infinity® is a registered trademark of Zevex, Inc. Moog® is a registered trademark of Moog, Inc. ENFit® is a registered trademark of GEDSA. © 2024 Moog, Inc. Moog Medical,
4314 Zevex Park Lane, Salt Lake City, UT 84123 USA. LS-66638 Rev. A

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals