PYROSTAR™ ES-F/Plate Multi-Test on the Microplate Reader – Kinetic
Turbidimetric Assay (KTA)
This is a guide to explain how to perform the Kinetic Turbidimetric method
with PYROSTAR™ ES-F/Plate on the Microplate Reader. Prior to performing the
assay, allow the reagent to come to room temperature. Ensure that the protocol
and parameters (Wavelength: 405 nm, Onset OD: 0.015, Intervals: 40 secs) are
set up on Toximaster® QC8 software. The temperature of the Microplate Reader
must be 37°± 1° C.
Step 1:
Prepare and set up the Toximaster® QC8 software for testing.
Step 2:
Reconstitute the Control Standard Endotoxin (CSE) with the LAL Reagent Water (LRW) to yield a concentration of 1000 EU/mL based on the Certificate of Analysis.
Step 3:
Vortex the reconstituted CSE for 2 minutes.
Step 4:
Label the appropriate endotoxin concentration for each dilution tube and add
the appropriate volume of LRW. Using the reconstituted CSE, prepare a 10-fold
dilution series for the appropriate range.
Dilutions may be prepared in different volumes if the same ratios are
maintained.
Dilution series for quantitative range: 10 EU/mL – 0.01 EU/mL
Step 5:
Add 50 µL of the endotoxin standards, controls, samples, etc., in the appropriate reaction wells.
Step 6:
Bring all contents to the bottom of PYROSTAR™ ES-F/Plate vial and reconstitute the LAL vial (volume shown on the label) with LRW. Swirl the LAL vial gently until the contents are dissolved, avoid making bubbles.
Step 7:
Using a repeater pipette, add 50 µL of reconstituted reagent to each reaction well.
Step 8:
Insert the microplate into the reader at 37°± 1° C and select “Start” on the Toximaster® software. The time course and well data generation will start immediately.
For more detailed information, refer to the PYROSTAR™ ES-F Plate package
insert.
Step 9:
After completion of the run, select “Stop”. Standard curves, group reports and audit trails are displayed for review.
Materials needed for Kinetic Turbidimetric Assay
Offered at FUJIFILM Wako
| Name of Products| Contents| Catalog Number
---|---|---|---
Depyrogenated Test Tubes| Borosilicate Test Tubes| 13 x 100 mm (50 test
tubes/pkg)| DL-13100
Endotoxin-free consumables| Aluminum Cap S| 14.7 x 18 mm (100 caps/pkg)|
293-28251
LAL Reagent Water, 30 mL| 20 x 30 mL glass bottle| LRW-2030
LAL Reagent Water, 100 mL| 12 x 100 mL glass bottle| LRW-12100
BioCleanTip Wako® 200 II| 200 µL/tip (100 tips/pkg)| 291-35021
BioCleanTip Wako® Extend S| 200 µL/tip (100 tips/pkg)| 294-35011
BioCleanTip Wako® 1000 II| 1000 µL/tip (100 tips/pkg)| 298-35031
BioCleanPlate Wako™| 96 wells/plate (50 plate/pkg)| 293-35221
LAL Reagents| PYROSTAR™ ES-F/Plate| Quantitative range:| WPEPK4-20015
4 multi-test vials (2.0 mL) + 1 vial CSE (500 ng/vial)| 10 EU/mL to 0.01 EU/mL
PYROSTAR™ ES-F/Plate| Quantitative range:| WPEPK4-50015
4 multi-test vials (5.2 mL) + 1 vial CSE (500 ng/vial)| 10 EU/mL to 0.01 EU/mL
Microplate Reader & Software| BioTek ELx808IU| 1BioTek ELx808IU unit| BIOT-
808IU
BioTek ELx808IU unit + Software| 1 Personal Computer/ Toximaster® QC8
Software| BIOT-35931
Part 11 software PC set| Toximaster® QC8 Software with 1 Personal Computer|
292-35931
Part 11 software only| Toximaster® QC8 Software| 294-35971
Not offered at FUJIFILM Wako
- Pipettes
- Test Tube Rack
- Vortex Mixer
Alpha Laboratories Ltd 40 Parham Drive Hampshire SO50 4NU UK
Tel: +44 23 8048 3000
Email: sales@alphalabs.co.uk
Web: www.alphalabs.co.uk
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>