Product Information

Specifications

  • Model: Swift 25.0-9800 Rev B
  • Manufacturer: Met One Instruments, Inc.
  • Address: 1600 NW Washington Blvd., Grants Pass, OR 97526
  • Telephone: 541-471-7111
  • Facsimile: 541-471-7116
  • Website: metone.com

Introduction

The Swift 25.0 Flow Calibrator is a product developed by Met One
Instruments, Inc. It is designed to provide accurate and timely
information for organizations and public authorities to make
informed decisions. The product is powered by Acoem, an
international group of companies committed to balancing progress
and preservation.

Unpacking

When you receive the Swift 25.0, please ensure that the shipment
contents are complete. To begin using the product, charge the Swift
25.0 using the provided charging cable.

Functions and Configurations

The Swift 25.0 offers various functions and configurations:

  1. Function Test: Allows you to test the functionality of the
    device.

  2. Inlet Cap: Covers the inlet of the device.

  3. Outlet Fitting: Connects the device to the outlet tubing.

  4. Electrical Connections: Provides electrical connections for the
    device.

  5. USB Driver Configuration: Configures the USB driver for data
    transfer.

  6. Configuration Software: Allows you to configure the device
    settings.

  7. Date and Time: Set the date and time of the device.

Operation

To operate the Swift 25.0:

  1. Installation: Follow the installation instructions provided in
    the manual.

  2. Outlet Tubing Adapter: Connect the outlet tubing to the device
    using the provided adapter.

  3. Bluetooth Antenna: Ensure that the Bluetooth antenna is
    properly connected for wireless communication.

  4. Temperature Probe: Connect the temperature probe to monitor
    temperature during operation.

FAQ

Q: Where can I find more information about Met One

Instruments?

A: You can visit the Met One Instruments website at metone.com for more information.

Q: Where can I find more information about Acoem?

A: For more information about Acoem, please visit their website
at acoem.com.

OPERATION MANUAL
Swift 25.0
Flow Calibrator Swift 25.0-9800 Rev B
Met One Instruments, Inc. 1600 NW Washington Blvd.
Grants Pass, OR 97526 Telephone: 541-471-7111 Facsimile: 541-471-7116
metone.com

Met One Instruments, Inc. is now part of the Acoem international group of companies.
Met One Instruments has been designing and manufacturing class-leading meteorological, ambient air sensing, and air quality monitoring instrumentation since its inception in 1989. Its line of robust industrial- grade meteorological equipment, air particulate monitoring equipment, and indoor air quality monitoring systems have set the standard for the industry. Headquartered in Grants Pass, OR, Met One Instruments, Inc. is fueled by a dedicated expert team who is diligently working to advance the technology required to ensure continued improvements in human and environmental health now and for generations to come.
Acoem is committed to helping organizations and public authorities find the right balance between progress and preservation — safeguarding businesses and assets and maximizing opportunities while conserving the planet’s resources. Headquartered in Limonest, France, Acoem delivers unrivaled inter-operable AI- powered sensors and ecosystems that empower our customers to make enlightened decisions based on accurate and timely information.
In 2021, Acoem acquired Met One Instruments, marking a pivotal moment when two industry leaders in the air quality monitoring sectors converged — creating a single, stronger and more future-focused provider of holistic environmental monitoring solutions. Now, Met One Instruments Powered by Acoem has opened new possibilities through an extensive offering of class leading, multi-parameter environmental monitoring and industrial reliability solutions. These integrated measurement systems, technologies, and services deliver comprehensive solutions for a range of applications, including environmental research, regulatory compliance, and industrial safety and hygiene.
For more information about Met One Instruments Powered by Acoem, please visit: metone.com
For more information about Acoem, please visit: acoem.com

Swift 25.0 Operation Manual – © Copyright 2023 Met One Instruments, Inc. All Rights Reserved worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language in any form without the express written permission of Met One Instruments, Inc.

Page 2

Swift 25.0-9800 Rev B

Table of Contents

1. INTRODUCTION ………………………………………………………………………. 7

1.1 About This Manual ………………………………………………………………………………. 7 1.2 Technical Service ……………………………………………………………………………….. 7 1.3 Swift 25.0 Flow Calibrator …………………………………………………………………….8 1.4 Swift 25.0 Specifications ………………………………………………………………………9
2. UNPACKING …………………………………………………………………………… 11

2.1 Shipment Contents……………………………………………………………………………… 11 2.2 Charging the Swift 25.0 ……………………………………………………………………… 12
3. FUNCTIONS and CONFIGURATIONS…………………………………………… 13

3.1 3.1.1 3.1.2 3.1.3 3.2
3.3 3.3.1 3.3.2 3.3.3 3.3.4 3.3.5 3.3.6 3.4
3.5
3.6
3.7
3.8 3.8.1

Multifunction Button…………………………………………………………………………… 13 Power …………………………………………………………………………………………………………… 13 Data Capture ……………………………………………………………………………………………….. 13 Bluetooth ……………………………………………………………………………………………………… 14
Function Test……………………………………………………………………………………. 14
User Interface Display ………………………………………………………………………… 14 Status Bar – Battery Life Icon ………………………………………………………………………… 16 Status Bar – Bluetooth Icon……………………………………………………………………………. 16 Flow Rate Measurement………………………………………………………………………………… 16 Barometric Pressure Measurement ………………………………………………………………… 16 Temperature Measurement ……………………………………………………………………………. 16 Relative Humidity Measurement ………………………………………………………………………17
Inlet Cap……………………………………………………………………………………………17
Outlet Fitting ……………………………………………………………………………………. 18
Electrical Connections ……………………………………………………………………….. 18
USB Driver Configuration…………………………………………………………………….. 19
Configuration Software………………………………………………………………………. 19 Date and time ……………………………………………………………………………………………….. 19

4. OPERATION …………………………………………………………………………..20

4.1 4.1.1 4.1.2 4.1.3 4.2
4.3
4.4

Installation ……………………………………………………………………………………….20 Outlet Tubing Adapter……………………………………………………………………………………22 Bluetooth Antenna …………………………………………………………………………………………23 Temperature Probe………………………………………………………………………………………..23
Data Retrieval …………………………………………………………………………………..24
Audit Sheet……………………………………………………………………………………….24
Storing the Swift 25.0 …………………………………………………………………………24

Swift 25.0-9800 Rev B

Page 3

5. SWIFT UTILITY SOFTWARE ………………………………………………………25

5.1 5.1.1 5.1.2 5.2
5.3
5.4 5.4.1 5.4.2 5.5 5.5.1 5.5.2 5.5.3 5.5.4 5.6 5.6.1 5.6.2 5.7
5.8 5.8.1 5.8.2 5.9
5.10

Menu Bar …………………………………………………………………………………………. 25 File Tab …………………………………………………………………………………………………………25 About ……………………………………………………………………………………………………………25
Establishing a Connection ……………………………………………………………………26
Swift Utility Software Interface ……………………………………………………………. 27
Device Info and Settings ……………………………………………………………………… 28 Get Button …………………………………………………………………………………………………….28 Set Button……………………………………………………………………………………………………..28
User Settings ……………………………………………………………………………………. 28 Temperature Units …………………………………………………………………………………………28 Pressure Units………………………………………………………………………………………………. 29 Flow Units ……………………………………………………………………………………………………. 29 Advanced ……………………………………………………………………………………………………. 29
Download Data ………………………………………………………………………………….30 All Button …………………………………………………………………………………………………….. 30 From Button…………………………………………………………………………………………………. 30
Clear Memory …………………………………………………………………………………… 31
Date and Time ………………………………………………………………………………….. 31 Set Clock ……………………………………………………………………………………………………… 31 Set From ……………………………………………………………………………………………………….32
Zero Flow …………………………………………………………………………………………33
History …………………………………………………………………………………………….33

6. MAINTENANCE and TROUBLESHOOTING …………………………………… 34

6.1
6.2 6.2.1 6.2.2 6.3
6.4 6.4.1 6.4.2 6.4.3 6.4.4 6.4.5

Recommended Periodic Maintenance Table …………………………………………….34
Inlet Screen Cleaning ………………………………………………………………………….34 Disassembly ………………………………………………………………………………………………… 35 Inspection and Cleaning ……………………………………………………………………………….. 35
Basic Problem and Cause/Solution Table ………………………………………………..36
Coin Cell Battery Replacement …………………………………………………………….. 37 Back Panel Removal ………………………………………………………………………………………37 Remove Main Battery …………………………………………………………………………………….38 Remove Battery Tray……………………………………………………………………………………..38 Replace the Coin Cell Battery ……………………………………………………………………….. 39 Reassembly …………………………………………………………………………………………………. 39

7. SPARE PARTS and ACCESSORIES…………………………………………….. 40

7.1 Consumables, Replacement Parts, and Accessories ………………………………….40

8. SWIFT BLUETOOTH APPLICATION ……………………………………………. 41

9. FLOW AUDIT SHEET ……………………………………………………………….. 41

Page 4

Swift 25.0-9800 Rev B

Table of Figures
Figure 1-1 Swift 25.0 Flow Calibrator ………………………………………………………………………………8 Figure 2-1 Standard Components for the Swift 25.0 ……………………………………………………… 12 Figure 3-1 Swift 25.0 Front View ………………………………………………………………………………….. 13 Figure 3-2 Swift 25.0 User Interface …………………………………………………………………………….. 15 Figure 3-3 Inlet Cap Screen …………………………………………………………………………………………..17 Figure 3-4 Swift 25.0 Bottom View………………………………………………………………………………. 18 Figure 4-1 Flow Measurements on UUTs with Standard 1.25″ O.D. Inlet …………………………… 21 Figure 4-2 Connection to a SASS with Tubing Adaptors ……………………………………………….. 22 Figure 4-3 Bluetooth Antenna …………………………………………………………………………………….. 23 Figure 4-4 Temperature Probe ……………………………………………………………………………………. 23 Figure 5-1 File Tab………………………………………………………………………………………………………. 25 Figure 5-2 About Screen …………………………………………………………………………………………….. 25 Figure 5-3 Start Screen ………………………………………………………………………………………………. 26 Figure 5-4 Main Screen ………………………………………………………………………………………………..27 Figure 5-5 Device and Settings……………………………………………………………………………………. 28 Figure 5-6 Get Button…………………………………………………………………………………………………. 28 Figure 5-7 Get and Set Buttons……………………………………………………………………………………. 28 Figure 5-8 Temperature Units ……………………………………………………………………………………… 28 Figure 5-9 Pressure Units ……………………………………………………………………………………………. 29 Figure 5-10 Flow Units ………………………………………………………………………………………………… 29 Figure 5-11 Advanced …………………………………………………………………………………………………. 29 Figure 5-12 Download Data Options ……………………………………………………………………………. 30 Figure 5-13 Clear Memory Prompt ………………………………………………………………………………… 31 Figure 5-14 Clear Memory ……………………………………………………………………………………………. 31 Figure 5-15 PC Time Prompt…………………………………………………………………………………………. 31 Figure 5-16 Date and Time Settings ……………………………………………………………………………… 31 Figure 5-17 Select Time Prompt …………………………………………………………………………………… 32 Figure 5-18 Date Dropdown Menu ……………………………………………………………………………….. 32 Figure 5-19 Zero Tare Prompt ……………………………………………………………………………………… 33 Figure 5-20 Zero Flow ………………………………………………………………………………………………… 33 Figure 5-21 History Panel…………………………………………………………………………………………….. 33 Figure 6-1 Inlet Assembly…………………………………………………………………………………………….. 34 Figure 6-2 Inlet Assembly ……………………………………………………………………………………………. 35 Figure 6-3 Back Panel Removal …………………………………………………………………………………….37 Figure 6-4 Battery Removal………………………………………………………………………………………… 38

Swift 25.0-9800 Rev B

Page 5

Figure 6-5 Battery Tray Removal ………………………………………………………………………………… 38 Figure 6-6 Coin Cell Battery ……………………………………………………………………………………….. 39 Figure 6-7 Reassembly ……………………………………………………………………………………………….. 39
Table of Tables
Table 1-1 Swift 25.0 Specifications …………………………………………………………………………………9 Table 6-1 Preferred Maintenance Period………………………………………………………………………. 34 Table 6-2 Trouble Shooting Cause and Solutions …………………………………………………………. 36 Table 7-1 Parts and Accessories ………………………………………………………………………………….. 40

Page 6

Swift 25.0-9800 Rev B

1. INTRODUCTION

1.1 About This Manual
This document is organized with the most important information toward the front of the manual. All users should read and understand the sections on setup, operation, and field audits. Toward the back are sections that provide in-depth information on subjects such as diagnostics and accessories. These sections should be consulted as needed.

1.2 Technical Service

Should support still be required after consulting the documentation, contact one of the expert Met One Instruments, Inc. Technical Service representatives during normal business hours of 7:00 a.m. to 4:00 p.m. Pacific Time, Monday through Friday. In addition, technical information and service bulletins are often posted on our website. Please contact us and obtain a Return Authorization (RA) number before sending any equipment back to the factory. This allows us to track and schedule service work and to expedite customer service.

Contact Information:

Tel: + 541 471 7111 Fax: + 541 471 7115 Web:http://www.metone.com Email: service@metone.com

Address:

Met One Instruments, Inc. 1600 Washington Blvd Grants Pass, Oregon 97526 U.S.A.

Please have the instrument serial number available when contacting the manufacturer. On most models manufactured by Met One Instruments, it will be located on a silver product label on the unit and printed on the calibration certificate. The serial number will begin with a letter and be followed by a unique five-digit number such as U15915.

Swift 25.0-9800 Rev B

Page 7

1.3 Swift 25.0 Flow Calibrator
The Met One Instruments, Inc. Model Swift 25.0 is a multi-function Flow Calibrator specifically designed for accurate flow measurements from 2.0 to 25.0 liters per minute (LPM). The device’s easy-to-read display shows Flow Rate, Flow Temperature and Ambient Pressure. Flow measurements are traceable to NIST standards, and a certificate of calibration is provided with each unit. The Swift 25.0 comes complete with a portable carrying case, flow adapters and a USB-C charging cable.

Page 8

Figure 1-1 Swift 25.0 Flow Calibrator
Swift 25.0-9800 Rev B

1.4 Swift 25.0 Specifications

Table 1-1 Swift 25.0 Specifications

PARAMETER

SPECIFICATION

Flow Range

2 to 25 LPM

Flow Accuracy

±1% of reading to -10° C

Display Resolution (Flow)

0.01 LPM

Flow Units

LPM (Liters Per Minuet) or SLPM (Standard Liters Per Minuet)

Barometric Pressure Range 400 – 900 mmHg (533 ­ 1200 mbar)

Barometric Pressure Accuracy

± 1.125 mmHg (± 1.5 mbar) @ 25° C 750 mmHg ± 1.875 mmHg (± 2.5 mbar) error band between -20° C to 85° C

Standard Conditions

25° C and 101.3 kPa

Display Resolution (Pressure) 0.1 mmHg (0.133 mbar)

Pressure Units

mbar, mmHg, kPa, inHg or Pa

Pressure Drop

7 mmHg @ 25 LPM

Temperature Accuracy

± 0.2° C

Temperature Probe Accuracy ± 0.2° C

Temperature Resolution

0.01° C/°F

Temperature Units

° C or ° F

Relative Humidity Range

0 to 100% RH non-condensing

Relative Humidity Accuracy ±3.5% RH

Data Storage Capacity

200 records, snapshot method utilizing power button

Electrical
DC Power Charging Serial Battery Battery Life
Software
Driver Software Bluetooth App

5 VDC, 500 mA max USB-C USB-C Internal 3.7 V, 2000mAh (7.4 Wh) Lithium-ion battery 35 hours
CP210X (MS Windows only) Swift Utility software (MS Windows only) In development, instrument is Bluetooth ready

Interface
Display

Black on White, Backlit, Sunlight Readable, 320 x 154 Resolution

Swift 25.0-9800 Rev B

Page 9

Display Parameters
Physical
Weight Size Outlet Fitting Size Case Size
Environmental
Operating Temperature Range Operating Humidity Water Resistance

Flow Rate: LPM or SLPM Barometric Pressure (ambient): mbar, mmHg, kPa, inHg or Pa Temperature: ° C or ° F, in flow or external if probe is plugged in Relative Humidity (ambient): % Average Flow Rate: units will follow Flow units. Battery Level and Charge Status Bluetooth connection
19.8 oz. (561 g) 4 in. wide x 6.1 in. height x 1.35 in. depth (10.6 cm x 15.5 cm x 3.4 cm) 1.25″ I.D (31.75 mm) slip on with O-ring seal 13″ x 11.4″ x 5.5″ (33 cm x 29 cm x 14 cm)
-20° C to +50° C
0 to 100% Non-condensing Enclosure is water resistant. Connections are IP67 rated.
Specifications may be subject to change without notice.

Page 10

Swift 25.0-9800 Rev B

2. UNPACKING

Before operating the Swift 25.0 for the first time, it is recommended to fully charge the flow calibrator using the included USB cable.

Use the following information to correctly assemble, configure, and deploy the Swift 25.0 Flow Calibrator. There are no special precautions or handling concerns except for the normal level of care required for handling scientific equipment. Refer to the instructions and diagrams on the following pages.
When unpacking the Swift 25.0 Flow Calibrator and accessories, compare them to the check off list below to make sure all items are present. Any damage incurred to the equipment during shipping is the responsibility of the carrier. If any damage to the shipment is noticed before unpacking, a claim must be filed with the commercial carrier immediately. Follow any special unpacking instructions provided by the carrier as all items are carefully removed from the containers and each component inspected. It is recommended to document and photograph all damaged packages and items before, during, and after unpacking them. Contact Met One Instruments (see Section 1.2 of this manual) to arrange for any replacement items needed.

2.1 Shipment Contents
The normal instrument configuration is supplied with the following standard components and accessories:
Swift 25.0 Flow Calibrator Check Off List
One Operator’s Manual (Swift 25.0-9800) One Calibration Certificate. NIST One Ambient Temperature Probe (83954) One Temperature Probe Cable Clip (480235) One Bluetooth Antenna, for use with mobile app (680792) One Flow Adapter: for 3/8″ I.D. tubing, SASS Application (83953) One Black EDPM Tubing- 4′ or 1.22 m, 1/4″ I.D x 3/8″ O.D. or6.35 mm I.D x 9.53 mm O.D.
(960057)
One USB Cable (502116) One Carrying Case (83959) Software & Driver Available at metone.com

The standard components for the Swift 25.0 are shown in Figure 2-1 Standard Components for the Swift 25.0 and Figure 3-1 Swift 25.0 Front View.

Swift 25.0-9800 Rev B

Page 11

Figure 2-1 Standard Components for the Swift 25.0
Please keep all the special shipping items (box, foam packing material, etc.) used to ship the Swift 25.0. They should be reused if the Swift 25.0 is to be transported (changing site locations, returning to the factory, etc.). Contact Met One Instruments (see Section 1.2 of this manual) for replacement packing materials if necessary.
2.2 Charging the Swift 25.0
The battery is charged using the included USB cable. 1. Insert the USB-C end (small oval plug) into the Swift 25.0 on the bottom left side of the unit. 2. The USB-A end (rectangle plug) can connect to standard USB port or to a standard 110V wall outlet by using a USB wall adapter such as a cell phone charging box (not supplied).
See Section 3.3.1 Status Bar – Battery Life Icon for more information.

Page 12

Swift 25.0-9800 Rev B

3. FUNCTIONS and CONFIGURATIONS
Figure 3-1 Swift 25.0 Front View shows the Multifunction Button, User Interface Display, Inlet Cap and Outlet Fitting.

Inlet Cap

User Interface Display

Multifunction/po wer Button

Outlet Fitting

Figure 3-1 Swift 25.0 Front View

3.1 Multifunction Button
The Multifunction Button shown in Figure 3-1 performs the functions in the sections below.

3.1.1 Power · Press the power button once to turn the power on. · Press and hold down the power button for five seconds to turn the power off. IMPORTANT: Be sure to allow at least ten seconds for the Swift 25.0 to complete the power-on cycle and initialization before mating it to a flow inlet.

3.1.2 Data Capture
· Press the power button once to save all displayed data along with the date and time while the instrument is powered on. A Page Icon will appear at the top of the display indicating a screenshot is being saved. When saved, the page icon will disappear.
See Section 3.8 Configuration Software for data retrieval.
NOTE: Data is stored in a rotational buffer of 200 records with first in first out, meaning when the 200 records are full, the first record is deleted when the next record is stored. Data is retained when the unit is powered off.

Swift 25.0-9800 Rev B

Page 13

3.1.3 Bluetooth
· When the Swift 25.0 is first powered on, and the Bluetooth antenna installed, a Bluetooth signal will begin broadcasting. The signal broadcast is indicated by the blinking Bluetooth icon found in the status bar (see Section 3.3.2 Status Bar Bluetooth Icon).
· When the Bluetooth icon is solid (no longer blinking), a Bluetooth connection has been established with a mobile device.
· The Bluetooth signal will only broadcast for six minutes if a connection is not made. After six minutes the signal will end, and the icon will disappear from the status bar.
· To start the Bluetooth signal broadcast again, press the power button twice in quick succession. The Icon will begin to blink for another six minutes or until a connection is made with a mobile device.
3.2 Function Test
Before operating the Swift 25.0 for the first time, it is recommended to fully charge the flow calibrator using the included USB cable.
Complete the following steps to verify proper operation.
1. Press the Power button to turn on the power.
2. Observe the Met One Instruments, “Powered by Acoem” startup screen for three seconds, then the Product Model and Firmware screen for three seconds, the Operate screen will then appear.
3. Hold the Power button for five seconds to power down the flow calibrator.
4. Unit is ready for use.
3.3 User Interface Display
The user interface display is shown in Figure 3-2. The display’s back light is always on with one fixed level of intensity. The display is direct sunlight readable. The user interface contains the following readings in order from top to bottom:
· The status bar contains the battery life indicator, Bluetooth connection status and screenshot Icon.
· Flow rate in liters per minute (LPM) or standard liters per minute (SLPM) · Barometric pressure in mbar, mmHg, kPa, inHg, or Pa · Temperature in °F or °C. · Relative humidity (RH) in %. · Average Flow. Averaged over 10 seconds.
NOTE: The Engineering units of measurement can be changed for user preference using the Swift Utility software. (Reference Swift Utility software manual.) Example: Fahrenheit to Celsius

Page 14

Swift 25.0-9800 Rev B

Battery Status Icon
Flow Value

Bluetooth Icon

Temperature value
Average Flow value

Figure 3-2 Swift 25.0 User Interface

Status Bar
Barometric Pressure Value
Relative Humidity
value

Swift 25.0-9800 Rev B

Page 15

3.3.1 Status Bar – Battery Life Icon
The battery life icon is located in the status bar at the top of the display. Charging is indicated by three strobing bars within the battery icon. A completely discharged battery will fully charge in approximately eight hours. Battery life is roughly measured in 33% increments with each bar. As battery life depletes, the bars will disappear, one at a time.
NOTE: It is recommended that the battery be fully charged before operating the Swift 25.0 for the first time.

3.3.2 Status Bar – Bluetooth Icon
The Bluetooth icon is located in the status bar, next to the battery life icon, at the top of the display. It is an antenna symbol that will flash when advertising a signal. The broadcast name is “Swift 25”. Connect the Swift 25.0 to a mobile device using the Swift Bluetooth mobile application. The application will be available for both Apple and Android devices. Refer to the Swift Bluetooth application manual for more information.

3.3.3 Flow Rate Measurement
Flow rate is reported on the first line of the display, below the status bar (See Figure 3-2). Ambient temperature, barometric pressure, and relative humidity are used are used to calculate the flow rate. Default flow rate units are liters per minute (LPM). Displayed flow rate measurement units can be changed from volumetric liters per minute (LPM) to standard liters per minute (SLPM) using the Swift Utility software or Swift Bluetooth app described in their respective manuals. The manuals and Swift Utility software can be downloaded at metone.com under the support tab. If assistance is needed, please contact the service department.

3.3.4 Barometric Pressure Measurement
Barometric pressure readings reported on the second line of the display. The reading comes from an independent pressure sensor, not in the flow path. Default pressure units are mmHg. Displayed pressure units can be changed to mbar, inHg, mmHg, Pa, or kPa using the Swift Utility software or Swift mobile application.

3.3.5 Temperature Measurement
Temperature readings are reported on the third line of the display. Temperature readings are measured either from the flow chamber or from the external probe when attached. Airflow must pass through the unit to obtain an accurate temperature reading when no probe is attached. The measurement source will automatically switch to the external probe when plugged in. Flow chamber temperature readings may be affected by external factors.

Page 16

Swift 25.0-9800 Rev B

Default temperature units are Celsius (°C). Displayed temperature units can be changed to Fahrenheit (°F) using the Swift Utility software or Swift mobile application.
3.3.6 Relative Humidity Measurement Relative Humidity is displayed on the fourth line of the display. The reading comes from an independent sensor, not in the flow path. The units are displayed as a percentage (%).
3.4 Inlet Cap
Figure 3-3 shows the Swift 25.0 Inlet Cap assembly. It is designed to allow airflow through the inlet but prevent rain and debris from damaging the internal sensors. The inlet cap is a two-piece assembly which houses a mesh screen to prevent debris and bugs from entering the device. See Section 6.2 Inlet Screen Cleaning for disassembly and maintenance.

Figure 3-3 Inlet Cap Screen
Caution: Do not carry the instrument by the Inlet Cap or the Outlet Fitting. They are held in place by the friction of O-rings and could come loose.

Swift 25.0-9800 Rev B

Page 17

3.5 Outlet Fitting
Figure 3-1 shows the location of the Swift 25.0 outlet fitting which is designed to fit on the down tube of the ambient air monitor equipment with 1.25″ (31.75 mm) outer diameter (O.D.) inlets, the standard size of Met One Instruments, Inc. BAM monitors. The flow calibrator can be used on FRM and FEM ambient air samplers including these Met One Instruments, Inc products:
· BAM 1020 · BAM 1022 · E-SEQ-FRM · E-FRM-DC · E-BAM Plus · E-BAM · SASS and SuperSASS · For a complete list, contact
sales@metone.com
3.6 Electrical Connections
Figure 3-4 shows the bottom view which includes connection points and features.
· Bluetooth antenna connector.
· RH sensor vent. This vent has a special membrane to allow moisture to enter and not allow water or dust to enter the instrument.
Important: Do not poke the membrane, it can easily be damaged allowing water or dust to enter the instrument.
· USB-C Port. For charging the internal battery, editing Engineering Units, and downloading data.
· Temperature Probe Connector.
Bluetooth Antenna Connector

RH Sensor Vent

Figure 3-4 Swift 25.0 Bottom View

USB C Connector
Temperature Probe
Connector

Page 18

Swift 25.0-9800 Rev B

3.7 USB Driver Configuration
A Silicon Labs CP210x Driver for the USB connection must be installed before connecting to the USB Type C port on the Swift 25.0. This is compatible with a Windows operating system only. The USB driver can be downloaded from our website at https://metone.com/software/. This only needs to be done once for each computer that is used with the flow calibrator.
It is recommended that the unit be fully charged before operating the Swift 25.0 for the first time. The battery is charged using the included USB cable. Insert the smaller USB C into the Swift 25.0 on the left side of the unit. The USB “A” end of the cable can connect to any USB port or connect to a standard outlet by using a USB wall adapter.
3.8 Configuration Software
The Swift 25.0 features Utility software to change the temperature and pressure units, Download Snapshot data, and Tare or zero the flow sensor. The Swift Utility software can be downloaded from our website at https://metone.com/software/. Use the setup wizard to install the software on a computer. Connect the flow calibrator to the computer with the USB cable. Open the Swift Utility software and select “GET” to load the device information.
Once the flow calibrator information is displayed, the temperature and pressure Engineering units can be changed using the drop-down menus. Temperature can be displayed in C (Celsius) or F (Fahrenheit). Pressure can be displayed in mbar, mmHg, inHg, Pa, or kPa. Press “Set” to change the displayed parameter.
3.8.1 Date and time
Important: Set the instrument to local Date, Time, and Engineering units prior to taking flow measurements. See Section 5.8 Date and Time For more information.

Swift 25.0-9800 Rev B

Page 19

4. OPERATION
The Swift 25.0 should be powered on before mating to a unit for measuring flow rate. The Swift 25.0 performs a zero-flow measurement when first powered on. No air should be passing through the flow calibrator before energizing the unit. The Swift 25.0 is ready to begin sampling after a short boot up, approximately ten seconds. When the home screen is displayed, the instrument is ready for use. The Swift 25.0 will provide readings for the following:
· Battery status.
· Bluetooth connection status.
· Flow rate in liters per minute (LPM) or standard liters per minute (SLPM).
· Barometric Pressure in mbar, mmHg, kPa, inHg, or Pa.
· Temperature in °F or °C., Relative Humidity (RH) in %.
· Average Flow. (Averaged over 10 second intervals).
Readings are updated on the display once per second. Refer to Section 3.3 User Interface Display for more details about the displayed readings.
4.1 Installation
Install the Swift 25.0 by placing the outlet adaptor onto the flow inlet of the unit being tested. As was mentioned in Section 3.5 Outlet Fitting, the adaptor will fit an inlet with an outer diameter of 1.25″. There are two O-rings inside the outlet adaptor to seal the flow path. When pressing the flow calibrator on the inlet, two points of resistance will be felt from the O-rings. When the second point of resistance is passed and can go no further, the meter is fully seated on the inlet.

Page 20

Swift 25.0-9800 Rev B

Figure 4-1 Flow Measurements on UUTs with Standard 1.25″ O.D. Inlet

Swift 25.0-9800 Rev B

Page 21

4.1.1 Outlet Tubing Adapter
When connecting to a SASS or other instrument that require the use of tubing for flow calibrations, use the outlet tubing adaptor and supplied tubing from the accessory kit as shown in Figure 4-2. Refer to the owner’s manual of the specific instrument when performing a measurement, calibration, or audit.

Figure 4-2 Connection to a SASS with Tubing Adaptors

Page 22

Swift 25.0-9800 Rev B

4.1.2 Bluetooth Antenna 1. Connect the Bluetooth antenna by gently mating the threaded collar of the antenna to the gold-colored connection point on the bottom of the Swift 25.0 (shown in Figure 3-4). 2. Gently tighten the threaded collar untill it is finger tight.
Figure 4-3 Bluetooth Antenna
4.1.3 Temperature Probe NOTE: When connecting or disconnecting the external temperature probe, the probe or internal temperature sensor reading will be displayed after a one second delay.
1. Connect the temperature probe by aligning the four pins in the blue connector with the holes of the silver connector, at the bottom of the Swift 25.0 (shown in Figure 3-4).
2. Gently press the blue connector into the silver connector until it stops. 3. Twist the threaded collar of the blue connector clockwise until it is finger tight.

Figure 4-4 Temperature Probe
Swift 25.0-9800 Rev B

Page 23

4.2 Data Retrieval
Utilize the Swift Utility software to retrieve captured data. See Section 5.6 for more information.
4.3 Audit Sheet
Section 9 FLOW AUDIT SHEET contains an example audit sheet from a Met One Instruments BAM 1020. Refer to the owner’s manual of the specific instrument when performing an audit.
4.4 Storing the Swift 25.0
It is recommended to store the Swift 25.0 and its accessories in the Hard Case for storage and transport. Leave the inlet cap and outlet fitting attached to the Swift 25.0.

Page 24

Swift 25.0-9800 Rev B

5. SWIFT UTILITY SOFTWARE

The Swift Utility Software supports all three Swift models: Swift 6.0, Swift 20.0, and Swift 25.0. The Swift 25.0 supports more features than the other two models, so if the software detects a Swift 25.0 is connected, then those extra options will become available. The further sections in this document will expand on these features.
5.1 Menu Bar
The menu bar can be found in the upper right corner of the main screen, directly below the Acoem “A” symbol and Swift software title. The bar includes About and File tabs.

5.1.1 File Tab

The file tab includes a dropdown menu for language options and Quit.

5.1.1.1 Language

Pressing the language button will show the available language options. A checkbox indicates which

language is currently selected. To change the language shown by the user interface, simply click on

the desired language shown in the dropdown. When a new language is selected, the main screen will refresh and reflect the chosen language.

Figure 5-1 File Tab

5.1.1.2 Quit The Quit option in the file tab will close the SWIFT program.

5.1.2 About
Pressing the about tab on the menu bar will open the software’s information page. The information page includes the Acoem Logo, Met One Instruments company name, software name, software version and a clickable link leading to the Metone.com web page.

Figure 5-2 About Screen
Swift 25.0-9800 Rev B

Page 25

5.2 Establishing a Connection
Use the setup wizard to install the software on your computer. Connect the flow calibrator to the computer with the USB cable. Open the Swift Setup Software and select the GET button, outlined in green at the bottom left side of the start screen, to load the device information (See Figure 5-3).

Figure 5-3 Start Screen

Page 26

Swift 25.0-9800 Rev B

5.3 Swift Utility Software Interface
The Swift 25.0 features setup software to change the temperature and pressure units, download snapshot data, and zero/tare the flow sensor. The Swift Setup Software can be downloaded from our website at https://metone.com/software/.

Measurement Units Sections:5.5.1, 5.5.2,
5.5.3

Clear Memory Section: 5.7

Device Info Section:

Download Data
Section: 5.6

Set Date/Time Section: 5.8

Zero/Tare Section: 5.9

Advanced Section: 5.5.4

Figure 5-4 Main Screen

History Section: 5.10

Swift 25.0-9800 Rev B

Page 27

5.4
·
·

Device Info and Settings
The Swift Utility software provides the device information including the model number, software version, serial number, and options for measurement units and current comport connection. A list of units is available in the dropdown menu for each measurement. The units of measurement can be configured to user preferences.

Figure 5-5 Device and Settings

5.4.1 Get Button
As was explained in Section 5.2, the GET button will establish a connection to the Swift 25.0. When the button is selected, device info, serial number and units of measurement will populate in the information fields. This is confirmation that the instrument is linked to the software. Changes can now be made, and data downloaded.

Figure 5-6 Get Button

5.4.2 Set Button
The set button will appear after the Swift 25.0 is linked with the utility software. When settings are changed, click on the SET button to send those changes to the unit. The changes will be seen on the devices’ display.

Figure 5-7 Get and Set Buttons
5.5 User Settings
Changing user settings is an easy process. Simply click on the unit drop-down menu to view the available options. Click on the desired option to set the unit of measurement.
5.5.1 Temperature Units Temperature unit options are C (Celsius) or F (Fahrenheit) illustrated by Figure 5-8.

Page 28

Swift 25.0-9800 Rev B

Figure 5-8 Temperature Units

5.5.2

Pressure Units

Pressure unit options are mbar (Millibar), mmHg (millimeters of mercury), inHg (Inches of Mercury), Pa (Pascal), kPa (Kilopascals) illustrated by Figure 5-9.

Figure 5-9 Pressure Units
5.5.3 Flow Units Flow Type options are Actual (LPM) and Mass (SLPM) illustrated by Figure 5-10.

5.5.4 Advanced
Figure 5-11 Advanced

Figure 5-10 Flow Units
The advanced option is for customers utilizing many COM ports. The advanced settings panel is hidden by default. When hidden, the software will automatically begin to scan all comports searching for a SWIFT device. To show the advanced settings, press the “Advanced” button to enable. When enabled, the software will use the selected COM port to communicate to the current device connected. When disabled, the software will search all available COM ports for a Swift device. This advanced option was added for customers using many COM ports and just want to specify the port instead of waiting for the software to find the device.

Swift 25.0-9800 Rev B

Page 29

5.6 Download Data
· The Swift 25.0 has the capability of capturing and saving data as stated in Section 3.1.2 Data Capture. The Swift Utility software facilitates downloading saved data in three forms: NEW, ALL, and FROM.
· When data has been downloaded, the information will be saved in a basic text format. The text can be copied into an Excel spread sheet for easier viewing.

An example of raw data is illustrated below.
Time, Flow(LPM),AT(C),BP(mbar),RH(%) 2023-06-22 09:17:04,00.02,+23.16,1013.8,043,000
2023-06-22 09:17:11,00.00,+23.18,1013.8,044,000 2023-06-22 09:17:24,00.17,+23.27,1013.8,046,000 2023-06-22 09:17:32,03.70,+23.34,1013.8,045,000

Figure 5-12 Download Data Options

5.6.1 All Button
Clicking the “All” button will download all data including every data point stored within the Swift 25.0 memory.

5.6.2 From Button
The From function allows the user to download data from a specified date and time to present. For example, if there was data stored from November 1st to December 10th and the last ten days of data are required. The “FROM” date would be selected for 2023-12-01. The downloaded data will be from December 1st to present.

Page 30

Swift 25.0-9800 Rev B

5.7 Clear Memory
· Pressing the “CLEAR” button will clear all the data in the current Swift 25.0 device.
· A prompt will appear asking the user if they are sure they want to clear all the current data in the device. Pressing Yes will delete all data in the current Swift 25.0 device. Pressing No will cancel this action and return the user back to the main screen.
Figure 5-14 Clear Memory
Figure 5-13 Clear Memory Prompt
5.8 Date and Time
The SWIFT Utility software provides two options for setting the device’s date and time, “SET CLOCK” and “SET FROM”. The date and time are used to identify when data was captured by the end user.
5.8.1 Set Clock “SET CLOCK” allows the user to configure the date and time of the flow calibrator using the computer’s current date and time. Pressing the “SET CLOCK” button will open a prompt to confirm if the Swift 25.0 device’s clock should be set to the current PC time. Pressing Yes will set the clock. Pressing No will cancel this action and return the user to the main screen.

Figure 5-15 PC Time Prompt

Figure 5-16 Date and Time Settings

Swift 25.0-9800 Rev B

Page 31

5.8.2 Set From
“SET FROM” allows the user to configure the date and time of the flow calibrator by selecting the desired date and time.
· To edit the date, click on the small calendar icon drop down menu to select the desired date.
· To edit the time, click on hours, minutes, or seconds fields to highlight them. When the numbers are highlighted, use the keyboard numbers to type in the desired values. hours, minutes, and seconds need to be edited individually.
· Click on the “SET FROM” button to save the changes. A prompt will appear, confirming if the date and time should be set to the current selection. Pressing Yes will apply the chosen date and time. Pressing No will cancel this action and return the user to the main screen.

Figure 5-18 Date Dropdown Menu

Figure 5-17 Select Time Prompt

Page 32

Swift 25.0-9800 Rev B

5.9 Zero Flow
The “ZERO” button will perform a zero-flow reset of the internal flow sensor. Be sure that the flow calibrator is not mounted to a running particle monitor and that no flow is passing though the inlet. If there is air flowing though the inlet, the zero tare will create an offset, adversely affecting the accuracy of the device.
· Click on the “ZERO” button to zero the flow, resetting the internal flow sensor offset.
· A prompt will appear, confirming that no flow is passing though the unit and that a zero tare should be performed. Pressing Yes will apply the chosen date and time. Pressing No will cancel this action and return the user to the main screen.

Figure 5-20 Zero Flow

Figure 5-19 Zero Tare Prompt
5.10 History
The history panel tracks all actions made in the software. This panel will show if an action was successful or unsuccessful.

Figure 5-21 History Panel
Swift 25.0-9800 Rev B

Page 33

6. MAINTENANCE and TROUBLESHOOTING

This section provides information regarding routine maintenance, and field troubleshooting procedures.

6.1 Recommended Periodic Maintenance Table

The table below shows the recommended interval for the regular maintenance, field checks, and service tasks.

Table 6-1 Preferred Maintenance Period

Maintenance Item

Period

Factory Calibration

12 months

Clean Inlet Screen

As Required

Coin cell battery (replaced during factory calibration)

12 months

Request a factory calibration by calling Met One Instruments at + 541 471 7111 or email service at service@metone.com See Section 1.2 Technical Service for more information.

6.2 Inlet Screen Cleaning
The Swift 25.0 inlet assembly is made up of two parts, the inlet cap, and the inlet screen holder. It is recommended to occasionally inspect and clean the inlet screen. The time interval between cleanings varies depending on the local particulate levels and time of year.

Figure 6-1 Inlet Assembly (Inlet Cap – left, Inlet Screen Holder – right)

Page 34

Swift 25.0-9800 Rev B

6.2.1 Disassembly
· To remove the inlet cap assembly from the Swift 25.0, hold the cap with a firm grip and pull in the opposite direction of the flow outlet.
· Separate the inlet cap from the inlet screen holder by gripping the cap with one hand. With the other hand, grip the portion of the inlet with the two grooves and pull the two pieces apart. It may be necessary to use a rocking motion while pulling.
o The inlet cap may separate from the inlet screen holder leaving it still attached to the flow calibrator. If this happens, set the inlet cap aside, grip the inlet screen holder, and pull in the same direction as the cap, to remove it from the flow calibrator. You may need to twist the inlet while pulling.

Figure 6-2 Inlet Assembly Inlet Cap ­ left, inlet screen holder ­ right.
6.2.2 Inspection and Cleaning 1. Inspect the screen for any obstructions or debris. 2. Use compressed air to blow away anything stuck to the screen. a. A medium bristled nylon or natural bristled brush may be used to remove anything left on the screen. 3. Use compressed air again to blow away any leftover debris removed by the brush. 4. Reassemble the inlet assembly by pressing the inlet cap back on the inlet screen holder. A snap should be felt when the O-ring is properly seated. 5. Install the inlet assembly by inserting the grooved portion of inlet screen holder into the flow inlet. Two snaps should be felt as the grooves lock in with the inlet O-rings.

Swift 25.0-9800 Rev B

Page 35

6.3 Basic Problem and Cause/Solution Table

The following table contains information on some of the more common problems that may be encountered, and some steps to identify and remedy the problems. Met One Instruments welcomes customer suggestions for new items to be included in this section for future manual revisions. If the solution cannot be found in the following table, then contact one of our expert service technicians for help in resolving the problem.

Problem:

Table 6-2 Trouble Shooting Cause and Solutions
The flow calibrator will not turn on

Cause/Solution:

· The battery needs to be charged.
o Connect the USB charging cable to the Swift 25.0 and appropriate power supply.

Problem: Cause/Solution:

The reported flow is significantly lower than expected
· The flow calibrator was powered on while connected to a unit to test with flow engaged. The Swift 25.0 performs a zero-flow measurement when first powered on. o Remove the Swift 25.0 from the UUT and cycle power.
· The flow calibrator was powered on in windy ambient conditions. o Cover the inlet while power cycling the unit.

Problem: Cause/Solution:

The temperature does not match expected ambient conditions
· Airflow must pass through the unit to obtain an accurate temperature reading. o Ensure the Swift 25.0 is connected to equipment with air flow.

Problem: Cause/Solution:

The time stamp data is not correct after setting with utility software · Internal coin cell is dead and needs to be replaced.

Page 36

Swift 25.0-9800 Rev B

6.4 Coin Cell Battery Replacement
The Swift 25.0 Flow Calibrator uses a CR1220 coin cell battery to keep the correct date and time when screen shot data is stored. If the captured data does not display the correct date and time the coin cell battery should be replaced. Use the instructions below to replace the CR1220 coin cell battery.
6.4.1 Back Panel Removal · Turn the Swift 25.0 off and remove the 6 screws from the back panel.
· A plastic screwdriver or fine pick may be needed to gently lift the back panel.
NOTE: Be careful not to dislodge the rubber O-ring gasket seated around the perimeter under the battery cover.

Figure 6-3 Back Panel Removal

Swift 25.0-9800 Rev B

Page 37

6.4.2 Remove Main Battery · Disconnect the battery pack from the PCB. Remove battery pack.

Figure 6-4 Battery Removal
6.4.3 Remove Battery Tray · Unfasten the battery tray by removing the 2 panhead screws with a Phillips screwdriver and lift the tray out exposing the PCBA board underneath.

Page 38

Figure 6-5 Battery Tray Removal
Swift 25.0-9800 Rev B

6.4.4 Replace the Coin Cell Battery
· Release the Battery by pressing the retainer spring to the outside edge of the battery. The battery should pop out. You may need to use a fingernail or a pencil eraser to push the retainer spring out of the way.
· Install a new coin cell battery (positive side facing out) by sliding one end the battery under the two small tabs, opposite to the retanier spring. Push down on the battery allowing the spring to lock in place with a faint click

Figure 6-6 Coin Cell Battery

6.4.5 Reassembly
1. Screw the battery tray back in place using the two panhead screws.
2. Connect the battery pack connector to the PCBA.
3. Set the battery pack into the battery tray and route the battery cable nicely on the side of the tray.
4. Put the back panel back in place and secure with the six screws.
o In order to power the flow calibrator back on after the main battery has been unplugged, external power must be applied.
5. Connect the USB-C cable connector to the port at the bottom of the Swift 25.0 and the other end of the cable to a power source.
6. Press the power button and allow the flow calibrator to fully power on.
o The date and time will need to be reset.
Refer to Section 5.8 Date and Time for setting the local date and time.

Figure 6-7 Reassembly

Swift 25.0-9800 Rev B

Page 39

7. SPARE PARTS and ACCESSORIES

7.1 Consumables, Replacement Parts, and Accessories

The following parts are available from Met One Instruments for maintenance, replacement, service, and upgrades. If unsure about a part, please contact the Service department and provide the serial number of the Swift 25.0.

Table 7-1 Parts and Accessories Description

Part Number

USB CABLE, TYPE A MALE TO TYPE C MALE, 3 FEET

502116

SWIFT 25.0 CASE

83959

INLET ASSEMBLY

83985

OUTLET ASSEMBLY

83986

OUTLET FITTING O-RINGS

720124

TUBING, EPDM RUBBER, 1/4″ I.D. x 3/8″ O.D., 4 FEET

960057

ADAPTOR BARB

83953

AMBIENT TEMPERATURE PROBE

83954

BLUETOOTH ANTENNA

80792

COIN CELL BATTERY

390099

Page 40

Swift 25.0-9800 Rev B

8. SWIFT BLUETOOTH APPLICATION
See the Swift Bluetooth application manual for more information.
9. FLOW AUDIT SHEET
This section (next page) contains an example Audit Sheet for a Met One Instruments BAM 1020. Refer to the owner’s manual of the specific instrument when performing an audit.

Swift 25.0-9800 Rev B

Page 41

BAM 1020 Audit Sheet

Model: BAM 1020 Serial Number:

Audit Date:

Audited By:

Flow Reference Standard Used:

Model:

Temperature Standard Used: Model:

Barometric Pressure Standard Used:

Model:

Leak Check Value:

as found:

lpm

Ambient Temperature:
Barometric Pressure:
Flow Rate (Actual Volumetric):
Flow Rate (EPA Standard):

as found:
as found:
as found:
as found:

BAM
C mmHg
lpm slpm

Flow Audits Serial No: Serial No: Serial No:

as left:

Ref. Std.
C

as left:

mmHg

as left:

lpm

as left:

slpm

as left:

Calibration Date:
Calibration Date:
Calibration Date:

lpm

BAM
C

Ref. Std.

C

N/A

mmHg

mmHg N/A

lpm

lpm

N/A

slpm

slpm

N/A

Pump muffler unclogged: Sample nozzle clean:
Tape support vane clean: Capstan shaft clean:
Rubber pinch rollers clean: Chassis ground wire installed:

as found as found as found as found as found
as found

Mechanical Audits

as left as left as left as left as left
as left

PM10 particle trap clean: as found

PM10 drip jar empty:

as found

PM10 bug screen clear:

as found

PM2.5 particle trap clean: as found

Inlet tube water-tight seal OK: as found

Inlet tube perpendicular to BAM:

as found

as left as left as left as left as left
as left

N/A N/A N/A N/A

Analog Voltage Output Audit

N/A

Set Out

Measured

1.000 Volts

Volts

Volts

2.500 Volts

Volts

Volts

5.000 Volts

Volts

Volts

Membrane Audit Zero Count (I1)
Span Count (I2)
Measured Mass
Percent Error

Flow Control Range

Flow Set Point

BAM Flow

15.0 LPM

16.70 LPM 18.3 LPM

Parameter Expected Found

Setup and Calibration Values Parameter Expected Found

Page 42

Swift 25.0-9800 Rev B

Parameter Expected Found

Clock Time/Date Inlet Type
Conc Units
Beta Count
BAM Sample
Cycle Mode
Standard Temp
Background Span Membrane
Span Check K
Usw
Error
1 2 3

FRH Setpoint
Low Power
Pressure Units
Time Stamp
Tape Pressure Alarm
Conc Error Alarm
Maintenance Alarm
Station ID MET Average

Conc Range Analog Output Conc Offset Analog Range 1
Analog Range 2
RS-232 baud
Flow Control232
RS-485 baud
Protocol Type Modbus Port Modbus Address

Last 6 Errors in BAM 1020 Error Log

Date

Time

Error

4

5

6

Date

Time

Swift 25.0-9800 Rev B

Page 43

Audit Notes:

Page 44