Directions for use

VOLDYNE 2500 VOLUMETRIC EXERCISER

INDICATIONS FOR USE

The Hudson RCI® Voldyne® Volumetric Exerciser is intended as an inspiratory deep breathing positive exerciser for adult and pediatric (above 5 yrs) patients. It is a single patient, multi-use device used in hospital or home care setting.

(1) Remove components from package, expand the tubing to its full length and attach open end of tubing to stem on front side of exerciser.
(2) Slide the green pointer to the prescribed volume level.
(3) Hold or stand exerciser in an upright position. Exhale normally. Then, place lips tightly around mouthpiece.
(4) Inhale slowly to raise the piston in the chamber. When inhaling, maintain the green flow chip in line with the smiley face flow range. Continue inhaling and try to raise the piston to prescribed volume level (top of piston indicates inspired volume).
(5) When inhalation is complete, remove mouthpiece, hold the breath as prescribed, and exhale normally. Allow piston to return to bottom of chamber. Rest and repeat exercises. Frequency of use and recommended inspiratory volumes should be at the direction of physician or therapist.

CLEANING INSTRUCTIONS
  • Remove the mouthpiece from the expandable tubing.
  • Dilute a ratio of 1 mL of liquid Dawn® dishwashing detergent into 1 L of warm tap water in a clean tub or bowl.
  • Immerse the mouthpiece in the soapy solution and soak for approximately 5 minutes.
  • After soaking but while the component is immersed, wipe the interior and exterior of the mouthpiece with a non-linting wipe for a minimum of 30 seconds. Rinse the mouthpiece under warm running tap water for a minimum of 1 minute or until all soap is removed.
  • Dry the mouthpiece using non-linting wipes.
  • Visually inspect the mouthpiece to ensure it is clean. Repeat cleaning steps if residual soil is present.
  • Reconnect the mouthpiece to the expandable tubing.

CAUTION/WARNING: After cleaning, visually inspect the corrugated tubing to ensure that the tubing is clean.

Predictive Nomogram-Inspiratory Capacity* (mL)**

| FEMALE|

HEIGHT

---|---|---
AGE| 58” 1.47m| 60” 1.52m| 62” 1.57m| 64” 1.63m| 66” 1.68m| 68” 1.73m| 70” 1.78m| 72” 1.83m| 74” 1.88m

20

| 1900| 2100| 2300| 2500| 2700| 2900| 3100| 3300| 3500†
25| 1850| 2050| 2250| 2450| 2650| 2850| 3050| 3250|

3450

30

| 1800| 2000| 2200| 2400| 2600| 2800| 3000| 3200| 2400
35| 1750| 1950| 2150| 2350| 2550| 2750| 2950| 3150|

3350

40

| 1700| 1900| 2100| 2300| 2500| 2700| 2900| 3100| 3300
45| 1650| 1850| 2050| 2250| 2450| 2650| 2850| 3050|

3250

50

| 1600| 1800| 2000| 2200| 2400| 2600| 2800| 3000| 3200
55| 1550| 1750| 1950| 2150| 2350| 2550| 2750| 2950|

3150

60

| 1500| 1700| 1900| 2100| 2300| 2500| 2700| 2900| 3100
65| 1450| 1650| 1850| 2050| 2250| 2450| 2650| 2850|

3050

70

| 1400| 1600| 1800| 2000| 2200| 2400| 2600| 2800| 3000
75| 1350| 1550| 1750| 1950| 2150| 2350| 2550| 2750|

2950

80

| 1300| 1500| 1700| 1900| 2100| 2300| 2500| 2700|

2900

| MALE|

HEIGHT

---|---|---
AGE| 58” 1.47m| 60” 1.52m| 62” 1.57m| 64” 1.63m| 66” 1.68m| 68” 1.73m| 70” 1.78m| 72” 1.83m| 74” 1.88m| 76” 1.93m| 78” 1.98m

20

| 2000| 2200| 2400| 2600| 2800| 3000| 3200| 3400| 3600| 3800| 4000†
25| 1950| 2150| 2350| 2550| 2750| 2950| 3150| 3350| 3550| 3750|

3950

30

| 1900| 2100| 2300| 2500| 2700| 2900| 3100| 3300| 3500| 3700| 3900
35| 1800| 2000| 2200| 2400| 2600| 2800| 3000| 3200| 3400| 3600|

3800

40

| 1750| 1950| 2150| 2350| 2550| 2750| 2950| 3150| 3350| 3550| 3750
45| 1700| 1900| 2100| 2300| 2500| 2700| 2900| 3100| 3300| 3500|

3700

50

| 1650| 1850| 2050| 2250| 2450| 2650| 2850| 3050| 3250| 3450| 3650
55| 1550| 1750| 1950| 2150| 2350| 2550| 2750| 2950| 3150| 3350|

3550

60

| 1500| 1700| 1900| 2100| 2300| 2500| 2700| 2900| 3100| 3300| 3500
65| 1400| 1600| 1800| 2000| 2200| 2400| 2600| 2800| 3000| 3200|

3400

70

| 1350| 1550| 1750| 1950| 2150| 2350| 2550| 2750| 2950| 3150| 3350
75| 1300| 1500| 1700| 1900| 2100| 2300| 2500| 2700| 2900| 3100|

3300

80

| 1250| 1450| 1650| 1850| 2050| 2250| 2450| 2650| 2850| 3050|

3250

  • Milliliter – Inspiratory capacity measured in milliliters rounded off to the nearest 50ml.
    ** Formula used in the above Nomogram published in The American Review of Respiratory Diseases, official journal of the American Thoracic Society, September 1979, Vol. 120, Number 3, by G. Polgar and V. Promadhat.
    † Inspiratory capacity measured in milliliters rounded off to the nearest 50 ml.

  Rx ONLY

Not made with natural rubber latex.

Do not use on multiple patients (Single Patient Use Only)

Warning: Single patient use: The use of products intended for single patient use on more than one patient creates a potential risk of serious injury and/or infection which may lead to death. Reprocessing of medical devices intended for single patient use only may result in degraded performance or a loss of functionality.

Hudson RCI, Teleflex and Voldyne are registered trademarks of Teleflex Incorporated or its affiliates.

All other trademarks are the properties of their respective owners.

Please scan here for instructional video.

For more information on incentive spirometry, please refer to the recommendations section of the 2011 AARC Clinical Practice Guidelines.