hygiena KIT230094 Listeria Monocytogenes Detection LyoKit Instruction Manual

August 15, 2024
Hygiena

hygiena KIT230094 Listeria Monocytogenes Detection LyoKit

Specifications:

  • Product Name: Listeria monocytogenes Detection LyoKit
  • Product No.: KIT230092 / KIT230093 / KIT230094
  • Approvals: LICENSE NUMBER 070401
  • Revision: A, September 2023

Product Information

The Listeria monocytogenes Detection LyoKit is designed for the qualitative detection of Listeria monocytogenes.

Storage and Stability: The PCR strips must be stored in the provided aluminum bag with silica gel pads. Protect from light and moisture.

LyoKit Tube Profiles: Available in three tube profiles – white low profile tubes (0.1 mL), clear regular profile tubes (0.2 mL), and clear low profile tubes (0.1 mL).

Product Usage Instructions

  • Overview
     General Information

  • Ensure the PCR strips are stored as per the provided instructions to maintain stability.

  • Applicability
    This kit is designed for the qualitative detection of Listeria monocytogenes.

  • Kit Contents
    Refer to the kit contents list to ensure all components are included before starting the procedure.

Instructions

  1. Required Material
    Gather all necessary materials as listed in the instructions before proceeding.

  2. Precautions and Preparations
    Follow all precautions mentioned in the manual and prepare the workspace accordingly.

  3. Enrichment and DNA Extraction
    Follow the certified methods provided for enrichment and DNA extraction steps.

  4. Certified Methods
    Refer to the specified certified methods for accurate results.

  5. Procedure
    Follow the workflow, program setup, and data interpretation steps as outlined in the manual.

  6. Workflow
    Understand and follow the workflow described in the manual for efficient processing.

  7. Program Setup
    Set up the program as per the instructions provided for optimal results.

  8. Data Interpretation
    Interpret the data obtained following the guidelines given in the manual.

  9. Troubleshooting
    Refer to the troubleshooting section in case of any issues during the process.

  10. Support
    Contact support for any queries or assistance needed during product usage.

Additional Information

Refer to testing principles, trademarks, reference numbers, and change index sections for more details.

Frequently Asked Questions (FAQ):

  • Q: How should I store the PCR strips?
    A: Store the PCR strips in the provided aluminum bag with silica gel pads, protecting them from light and moisture.

  • Q: What are the available tube profiles for LyoKit?
    A: LyoKit is available in three tube profiles – white low profile tubes (0.1 mL), clear regular profile tubes (0.2 mL), and clear low profile tubes (0.1 mL).

Listeria
monocytogenes Detection LyoKit
PRODUCT INSTRUCTIONS

Documentation for the qualitative detection of Listeria monocytogenes
Product No. KIT230092 / KIT230093 / KIT230094

  • foodproof® Listeria monocytogenes Detection LyoKit
  • Product No.
  • LP: KIT230092
  • RP: KIT230093
  • DP: KIT230094
  • Kit for 96 reactions (lyophilized) for a maximum of 94 samples
  • Store kit at 2 to 8 °C
  • For testing of food and environmental samples
  • Approvals:
  • PRODUCT INSTRUCTIONS
    Revision A, September 2023

 OVERVIEW

  1. General Information
    Number of Reactions

    • The kit is designed for 96 reactions with a final reaction volume of 25 µL each. Up to 94 samples plus a positive and negative control can be analyzed per run.
    • Storage and Stability
    • Store all components at 2 to 8 °C. They are guaranteed to be stable through the expiration date printed on the label. Opening of the kit does not shorten the expiration date.
    • The PCR strips must be stored in the provided aluminum bag with silica gel pads. Protect from light and moisture.
    • LyoKit Tube Profiles
    • The LyoKit is available in three different tube profiles: white low profile tubes, 0.1 mL (LP), clear regular profile tubes, 0.2 mL (RP) and clear low profile tubes, 0.1 mL (DP).
    • The majority of real-time PCR cyclers use low profile tubes, 0.1 mL (LP). For the Dualo 32® R2 and a few other cyclers, please use clear low profile tubes, 0.1 mL (DP).
    • Refer to the cycler and tube compatibility chart for more information.
  2. Applicability

    • The kit described in these Product Instructions has been developed for real-time PCR instruments with a FAM and a VIC/Yakima Yellow or HEX detection channel. The performance of the kit was tested with the following real-time PCR instruments: LightCycler® 480, LightCycler® 96 (Roche Diagnostics), Mx3005P® (Agilent Technologies), Applied BiosystemsTM 7500 Fast (Thermo
    • Scientific), and CFX96TM (Bio-Rad).
    • When testing the samples on a LightCycler 480 instrument, it is recommended to carry out the DNA extraction with a wash step (Using foodproof® StarPrep® Two Kit, KIT230177, Extraction
    • Procedure A: STANDARD protocol).
    • The kit must not be used in diagnostic procedures.
  3. Kit Contents
    A schematic representation of the foodproof Listeria monocytogenes Detection LyoKit with all its components.

    • LP: KIT230092
    • RP: KIT230093
    • DP: KIT230094

| Component| Details
---|---|---
1| Microplate|

  • 12 x 8-tube strips, prefilled with lyophilized ready-to-use PCR mix.
  • Three different tube profiles are available: white low profile tubes (LP), clear regular profile tubes (RP) and clear low profile tubes (DP).*

2| 12 x 8-cap strips| For use in real-time PCR after addition of samples.
3| 2 x H2O PCR-grade (colorless cap)| 1 mL nuclease-free, for use as a PCR run negative control.


4

| Control Template (purple cap)| 250 µL, contains a stabilized solution of DNA for use as a PCR run positive control.

  • Tube profile and instrument compatibility chart is available online

INSTRUCTIONS

  1. Required Material
    Most of the required equipment and reagents are available through Hygiena®. Please contact us for further information.

    • Use a real-time PCR cycler suitable for detection of respective probes as well as for using low or regular profile strip tubes.
    • In case the strip tubes don’t fit in the instrument, the samples should be transferred to appropriate PCR vessels after resuspension of the lyophilized PCR mix.
    • Material
    • Nuclease-free, aerosol-resistant pipette filter tips.
    • PCR strip or plate centrifuges
    • Without vortex: Mini microcentrifuge for 4 x 8-strips
    • With vortex: Multispin MSC-6000 for 4 x 8-strips
    • With vortex: CVP-2 for 12 x 8-strips and plates
    • DNA Extraction Kits
    • foodproof StarPrep Two Kit (KIT230177) or
    • foodproof StarPrep Two 8-Strip Kit (KIT230186) or
    • foodproof Magnetic Preparation Kit II (Kit230181)

Precautions and Preparations
The kit provides all reagents required for the PCR. However, in order to achieve reliable results, the entire assay procedure must be performed under nuclease-free conditions. Follow the instructions below to avoid nucleases, carry-over or cross-contamination:

  • Keep the kit components separate from other reagents in the laboratory.
  • Use nuclease-free labware (e.g., pipettes, pipette tips, reaction vials).
  • Wear gloves when performing the assay.
  • To avoid cross-contamination of samples and reagents, use fresh aerosol barrier pipette tips.
  • To avoid carry-over contamination, transfer the required solutions for one experiment into a fresh tube, rather than directly pipetting from stock solutions.
  • Physically separate the workplaces for DNA preparation, PCR setup and PCR to minimize the risk of carry-over contamination. Use a PCR hood for all pipetting steps.
  • Sample Material: Use any sample material suitable for PCR in terms of purity, concentration and absence of inhibitors.
  • DNA Extraction: We provide sample preparation kits suitable for all kind of food and other samples.
  • Positive Control: Always run a positive control with the samples. Use the provided control DNA (Control Template) or a positive sample preparation control.
  • Negative Control: Always run a negative control with the samples. To prepare a negative control, replace the template DNA with PCR-grade water. Include a negative control during sample preparation to monitor reaction purity and cross-contamination. This extraction control can be used as an additional negative control reaction.
  • Confi rmation: If required, positive results may be confi rmed by appropriate methods (e.g., reference method).
  • Waste Disposal: All contaminated and potentially infectious material, like enrichment cultures or food samples, should be autoclaved before disposal and eliminated according to local rules and  regulations. For more information, e.g., proper disposal of unused chemicals, please refer to the appropriate safety data sheet (SDS).
  • Keep the PCR mix away from light and moisture.
  • For more information, please refer to the appropriate safety data sheet (SDS). The SDS is available online at www.hygiena.com/sds.

Enrichment and DNA Extraction

  • The foodproof Listeria monocytogenes Detection LyoKit is intended for the rapid detection of Listeria monocytogenes isolated by foodproof DNA extraction methods from enrichment cultures of all relevant kind of foods and primary production stage (PPS) samples that are potentially contaminated with Listeria monocytogenes. For DNA extraction, please use the kits mentioned in 2.1 Required Material.
  • Certified Methods
    The performance of the foodproof Listeria monocytogenes Detection LyoKit in combination with several foodproof DNA extraction procedures, including the foodproof StarPrep Two Kit (Procedure A and B) and the foodproof StarPrep Two 8-Strip Kit (Procedure A), has been approved in an AOAC RI Performance Tested MethodsSM program (License no. 070401) and in a NordVal International method extension (Certificate no. 025). For the AOAC-RI PTM validation study, the following food categories were tested: raw meat, processed meat, seafood, milk and dairy, and fruits/juices. The following categories were tested for the NordVal International validation study: composite foods/ready-to-eat and ready-to-reheat, meat products, milk and dairy products, vegetables, seafood and fishery products, and environmental samples. For further information about the tested matrices, enrichment protocols and the DNA extraction procedures please refer to ANNEX 1 and ANNEX 2 at the end of the manual.

Procedure

This protocol describes how to perform the analysis of DNA extracts by real- time PCR.

Workflow

  1. PLACE STRIPS IN RACK
    Take needed number of PCR tube strips out of aluminum bag. Important: close bag tightly afterwards. Place strips in a suitable PCR tube rack.
    If needed, gently tap the tubes to move the lyophilized
    pellets to the bottom of all tubes.

  2.  DECAP
    Open strips carefully direct before fi lling and discard caps.
    Important: do not leave open longer than necessary.

  3. ADD SAMPLES AND CONTROLS
    Pipette 25 µL of samples, negative control (colorless cap) or
    Control Template (purple cap) into respective wells.
    If using less volume, add PCR-grade H2O to reach 25 µL.
    To reduce the risk of cross-contamination, prepare only one strip at a time.

  4. SEAL
    Seal the tubes with the provided 8-cap strips tightly.

  5. MIX
    Resuspend pellet after sealing by mixing thoroughly.
    Alternatively, resuspend pellet by pipetting up and down multiple times in step 3.

  6. CENTRIFUGE
    Briefl y spin strips, e.g., 5 sec at 500 – 1,000 x g, in a suitable centrifuge.

  7. START REAL-TIME PCR RUN
    Cycle samples as described in the program setup (2.4.2). Place tubes in a vertical, balanced order into the cycler, e.g., two strips can be placed in the fi rst and last column.

Program Setup
Program your real-time PCR instrument before setting up the PCR reactions. Select the following channels:

FAM (L. monocytogenes), and VIC (Internal Control).
As an alternative to VIC, HEX can be used. For the PikoReal® 24, Yakima Yellow has to be selected.

  • Pre-incubation: 1 cycle Step 1: 37 °C for 4 min Step 2: 95 °C for 5 min
  • Amplifi cation: 50 cycles Step 1: 95 °C for 5* sec
  • Step 2**: 60 °C for 60 sec
  • *Use 15 sec for the LightCycler 96 **Fluorescence detection
  • For some real-time PCR instruments the probe quencher as well as the usage of a passive reference dye has to be specified. This kit contains probes with a non-fluorescent (“dark”) quencher and no passive reference dye.
  • For users of the Agilent Mx3005P instrument: Click “Instrument” and “Filter Set Gain Settings” to open the Filter Set Gain Settings dialog box in which the gain settings may be viewed and modified.
  • For FAM the Filter Set Gain Setting has to be modified to “x1”.
  • Data Interpretation
    • Verify results of positive (Control Template) and negative control (H2O) before interpreting sample results. Always compare samples to positive and negative controls. Review data from each channel and interpret results as described in the table
FAM VIC Result Interpretation
+ + or – Positive for L. monocytogenes
+ Negative for L. monocytogenes
Invalid

Troubleshooting

Problem Possible Cause Recommendation

Squashed or crooked tubes, or open / dislodged tube lids after run, or the cycler does not open or close properly.| Wrong tube format.|

  • Choose the correct tube format for your cycler. Tube profile and instrument compatibility chart is available online.
  • If necessary, the samples can be transferred to appropriate PCR vessels after resuspension of the lyophilized PCR mix.

Wrong placement of tubes.| Place tubes into the cycler in a vertical and balanced order, as described in the instructions for the PCR instrument.
No signal increase is observed, even with positive controls.| Incorrect detection channel has been chosen.| Set channel settings for respective dyes accordingly.
Pipetting errors.|

  • Check for correct reaction setup and repeat the PCR run.
  • Always run a positive control along with your samples.

No data acquisition programmed.| Check the cycle programs.
A sample shows no signals, including the internal control. Positive and negative control have proper signals.| Inhibitory effects of the sample material (e.g., caused by insufficient purification).| Use the recommended DNA extraction kit.

Dilute samples or pipette a lower amount of sample DNA (e.g., 20 µL PCR-grade water and 5 µL sample instead of 25 µL sample).

Negative control samples are positive.| Carry-over contamination.|

  • Exchange all critical solutions and reagents for DNA/RNA extraction.
  • Repeat the complete experiment with fresh batches of all reagents.
  • Always handle samples, kit components and consumables in accordance with commonly accepted practices to prevent carry-over contamination.
  • Add positive controls after sample and negative control reaction vessels have been sealed.

Fluorescence intensity is too low.| Inappropriate storage of kit components.| Store lyophilized PCR mix at 2 to 8 °C, protected from light and moisture.
Low initial amount of target DNA.| If possible, increase the amount of sample DNA. Depending on the chosen DNA isolation method, inhibitory effects may occur.

Problem Possible Cause Recommendation
Strong decrease of fluorescence baseline. Resuspension of lyophilized PCR mix
not complete.
  • Always resuspend lyophilized PCR mix thoroughly.
  • Use the recommended vortex centrifuge with the correct settings.

Fluorescence intensity varies or changes abruptly during the run.| Insufficient centrifugation of the PCR strips, e.g., resuspended PCR mix is still in the upper part of the vessel or bubbles trapped in the mix.|

  • Always centrifuge PCR strips.
  • Use the centrifuge models and settings recommended in this manual.
  • Avoid the introduction of air bubbles during pipetting.

Outer surface of the vessel or the seal is dirty (e.g., by direct skin contact).|

  • Always wear gloves when handling the vessels and seal.
  • Do not mark vessels on the outside of the tubes or directly on top of the reaction mix.

Pellets are difficult to dissolve.| The lyophilized PCR mix started to rehydrate.|

  • Store the lyophilized PCR mix always in the aluminum bag with the silica gel pads. Make sure that the lids are tightly closed.
  • Remove strips from the aluminum bag only shortly before PCR setup.
  • Open strip shortly before filling.

Support
If have questions or experience any problems with our products, please contact us:

www.hygiena.com/support

Our aim is to provide you with a solution as quickly and effectively as possible. We would also like you to contact us if you have any suggestio n case  you would like to use our product for a different application. We highly value your feedback.

ADDITIONAL INFORMATION

Testing Principle
The foodproof kit provides all necessary reagents and a control template for reliable interpretations of results. To ensure maximum reliability of the kit and to prevent misinterpretation of negative results due to inhibition of the amplifi cation, an Internal Control (IC) is included. A hydrolysis probe was designed to bind specifi cally the IC, allowing detection in the respective channel, whereas the target DNA is detected in another channel. In case of a negative result due to inhibition of the amplifi cation by the sample DNA of interest, the amplifi cation of the IC is suppressed as well, whereas a negative result for the sample DNA of interest and amplifi cation of the IC clearly indicates the absence of parameter in the sample. The real-time PCR kit minimizes contamination risk and contains all reagents (except for template DNA) needed for the detection of target DNA. Primers and probes provide specifi c detection of target DNA in food and environmental samples, including primary production stage samples. The described performance of the kit is guaranteed for use only on the real-time PCR instruments listed in 1.2 Applicability. For other instruments, please contact us.
Step-by-Step Procedure

  1. Using the kit‘s sequence-specifi c primers in a polymerase chain reaction (PCR), the PCR instrument and the supplied reagents amplify fragments of specifi c sequences for target DNA.

  2.  The PCR instrument detects these amplifi ed fragments in real time through
    fluorescence generated by cleavage of the hybridized probe due to the 5′ nuclease activity of the Taq DNA polymerase. The probe is labeled at the 5′ end with a reporter fl uorophore and at the 3′ end with a quencher.

  3. During the annealing/elongation phase of each PCR cycle, the probe hybridizes to
    an internal sequence of the amplicon and is cleaved by the 5′ nuclease activity of the Taq DNA polymerase. This cleavage of the probe separates the reporter dye from the quencher dye, increasing the reporter dye signal.

  4. The PCR instrument measures the emitted fl uorescence of the reporter dye.

Prevention of Carry-Over Contamination
The heat-labile Uracil-DNA N-Glycosylase (UNG) is suitable for preventing carry-over contamination between PCRs. This technique relies on the incorporation of deoxyuridine triphosphate (dUTP) during all amplifi cation reactions, and the pretreatment of all successive PCR mixtures with the heat- labile UNG. The UNG cleaves DNA at any site where a deoxyuridine residue has been incorporated. The resulting abasic sites are hydrolyzed due to the high temperatures during the initial denaturation step, and can no longer serve as PCR templates. The heat-labile UNG is inactivated during the initial denaturation step. Native DNA (e.g., the isolated target genomic DNA) does not contain uracil and is therefore not degraded by this procedure. Since dTTP is replaced with dUTP and UNG is included in this kit, decontamination can be achieved with the provided reagents.

  • Trademarks foodproof®, microproof®, vetproof®, ShortPrep®, StarPrep®, RoboPrep® and LyoKit® are trademarks of Hygiena Diagnostics GmbH.
  • Hygiena® is a registered trademark of Hygiena.
  • Other brand or product names are trademarks of their respective holders.

Reference Number

  • The reference number and original Hygiena Diagnostics GmbH article numbers:
  • R 602 23 -1 (KIT230092), R 602 23 -2 (KIT230093), and R 602 23 -3 (KIT230094).

Change Index

  • Version 1, September 2014:
  • First version of the package insert.
  • Version 2, March 2017:
  • License Notice changed. Introduction of vortex centrifuges into the PCR Setup Procedure.
  • Version 3, December 2019:
  • Introduction of the following: ANNEX 1: AOAC-RI #070401 Extension Table and additional information in the Applicability Statement; R 602 23-3, new tube format for other real-time PCR instruments; AOAC-RI logo.
  • Version 4, February 2022:
  • Rebranding, new document layout and content, new order number. Introduction of the following: NordVal International logo; ANNEX 2: NordVal #025 Extension Table and additional information in
  • the Applicability Statement.
  • Revision A, September 2023:
  • Additional rebranding, layout, images and tracking.
  • R 602 23 20 -> INS-KIT230092-93-94-REVA

ANNEX 1

  • AOAC-RI #070401 Extension Table for the foodproof Listeria  monocytogenes Detection LyoKit
  • The following table shows the recommended enrichment time for different food matrices with enrichment in Half Fraser broth in combination with different foodproof DNA extraction procedures that have been validated for the AOAC RI PTM extension of the foodproof Listeria monocytogenes Detection LyoKit.
  • For further information regarding the DNA extraction procedures below, please refer to the appropriate Hygiena Diagnostics GmbH package inserts on www.hygiena.com.

DNA Extraction| Enrichment time in Half Fraser broth at 30 °C| Enrichment time per matrix
---|---|---
foodproof ShortPrep II Kit| 24 h – 48 h| minced meat 48 h melon 24 h raw fish 48 h sausage 48 h Harzer cheese 48 h
foodproof StarPrep Two Kit Extraction Procedure A STANDARD| 24 h| minced meat 24 h melon 24 h raw fish 24 h sausage 24 h Harzer cheese 24 h

foodproof StarPrep Two Kit Extraction Procedure B RAPID

| 24 h – 48 h| minced meat 48 h melon 24 h raw fish 48 h sausage 48 h Harzer cheese 48 h
foodproof StarPrep Two 8-Strip Kit| 48 h| minced meat 48 h melon 48 h raw fish 48 h sausage 48 h Harzer cheese 48 h
foodproof Magnetic Preparation Kit II| 24 h – 48 h| minced meat 48h melon 24 h raw fish 24 h sausage 48 h Harzer cheese 48 h

Tested food categories : Raw meats, processed meats, seafood, milk and dairy, fruit/ juices
Sample size: 25 g / 225 mL broth
Reference method: ISO 11290:1996/Amd 1:2004

ANNEX 2

  • NordVal #025 Extension Table for the foodproof Listeria monocytogenes Detection LyoKit
  • The following table shows the recommended enrichment time for the tested food categories and environmental samples in Actero Listeria Enrichment Media (ALEM) and Half Fraser broth in combination with different foodproof DNA extraction procedures that have been validated for the NordVal International method extension of the foodproof Listeria monocytogenes Detection LyoKit (Certificate No. 025).
  • For further information regarding the DNA extraction procedures below, please refer to the appropriate Hygiena Diagnostics GmbH package inserts on: www.hygiena.com.

DNA Extraction| Enrichment time in ALEM at 36 °C| Enrichment time in Half Fraser broth at 30 °C| DNA extract for PCR
---|---|---|---
foodproof StarPrep Two Kit Extraction Procedure A STANDARD| 22 ± 2 h| 25 ± 1 h| 5 µL + 20 µL PCR- grade water
foodproof StarPrep Two Kit Extraction

Procedure B RAPID

| —| 48 ± 2 h| 5 µL + 20 µL PCR- grade water
foodproof StarPrep Two 8-Strip Kit Extraction Procedure A STANDARD| 22 ± 2 h| 25 ± 1 h| 5 µL + 20 µL PCR- grade water

Tested categories:

  • Composite foods/ready-to-eat and ready-to-reheat
  • Meat products
  • Milk and dairy products
  • Vegetables
  • Seafood and fishery products
  • Environmental samples

Sample size:

  • 25 g + 150 mL Actero Listeria Enrichment Media or 225 mL Half Fraser broth
  • 1 swab + 10 mL Actero Listeria Enrichment Media or 10 mL Half Fraser broth
  • 1 sponge + 100 mL Actero Listeria Enrichment Media or 100 mL Half Fraser broth
  • 1 wipe + 225 mL Actero Listeria Enrichment Media or 225 mL Half Fraser broth
  • For sampling after cleaning process, premoisten:
    • 1 swab + 1 mL broth universal neutralizing (+ 9 mL ALEM or Half Fraser)
    • 1 sponge + 10 mL broth universal neutralizing (+ 90 mL ALEM or Half Fraser)
    • 1 wipe + BPW + 10 % neutralizing agent (+ 225 mL ALEM or Half Fraser)

Reference method : ISO 11290-1/A1 (May 2017)

Hygiena® Camarillo, CA 93012 USA diagnostics.support@hygiena.com
Manufactured by Hygiena Diagnostics GmbH Hermannswerder 17 14473 Potsdam Germany
www.hygiena.com