Blue-White M1 Series Peristaltic Metering Pump Instruction Manual

August 14, 2024
Blue-White

Blue-White M1 Series Peristaltic Metering Pump

Specifications

  • Maximum Working Pressure (Excluding pump tubes): 100 PSI (6.89 bar)
  • Maximum Fluid Temperature: 185°F (85°C)
  • Maximum Viscosity: 5000 centipoise
  • Maximum Suction Lift: 30 ft. of water at sea level (14.7 atm psi)
  • Ambient Operating Temperature Range: 14°F to 115°F (-10°C to 46°C)
  • Ambient Storage Temperature Range: -40°F to 158°F (-40°C to 70°C)
  • Operating Voltage: 115V60Hz 1 PH (0.6A max.),220V50Hz 1 PH (0.3A max.), 230V60Hz 1 PH (0.3A max.), 230V50Hz 1 PH (0.3A max.), 240V50Hz 1 PH (0.3A max.)
  • Power Cord Options (removable): NEMA 5/15 (USA), CEE 7/VII (EU), AS 3112 (Australia/New Zealand), BS 1363/A (UK)
  • Motor: Brushless DC, 50W

Product Usage Instructions

Installation
Follow the steps below for proper installation of the Peristaltic Metering Pump:

  1. Mounting Location
    Select a suitable location for mounting the pump that meets the specified requirements.

  2. Pump Dimensions
    Refer to the provided dimensions to ensure adequate space for installation.

  3. Wall Mounting
    If wall mounting is required, use the provided Wall Mount Bracket for secure installation.

  4. Power Connections
    Connect the power cord based on the operating voltage requirement and plug type.

  5. Input / Output Connections
    Make the necessary connections for input and output according to the system requirements.

  6. M12 Connections
    If using M12 communications cables, ensure proper connection for communication purposes.

Pump Operation
Follow these steps for operating the pump efficiently:

  1. Priming the Pump

  2. Prime the pump before use as per instructions to ensure proper functionality.

  3. Manual Speed/Flow Adjust
    Adjust the speed or flow manually as needed for your application.

  4. 4-20mA Speed/Flow Control
    If applicable, utilize the 4-20mA input for speed and flow control.

Pump Maintenance
Maintain the pump regularly using the following guidelines:

  1. Routine Inspection and Maintenance
    Perform routine inspections and maintenance checks to keep the pump in optimal condition.

  2. Cleaning and Lubricating the Pump
    Clean and lubricate the pump as recommended for smooth operation.

Frequently Asked Questions (FAQ)

Q: How do I calibrate the pump tube?
A: Run the pump for one hour with a new tube installed prior to calibration for accurate results.

Q: Is there a specific identification number for re-ordering tubing?
A: Yes, the tubing has an identification number printed on it for easy re-ordering.

INTRODUCTION

Congratulations on purchasing Blue-White FLEXFLO® Series M1 peristaltic metering pump (a peristaltic metering pump is a type of positive displacement pump used for pumping a variety of fluids).

The FLEXFLO® Series M1 peristaltic metering pump is pre-configured for the tubing that is shipped with it.

NOTE:

  • For accurate calibration of pump tube, run the pump for one hour with new tube installed prior to calibration.
  • The tubing has an identification number printed on the tube for easy re-ordering.
  • The pump was pressure-tested at the factory with clean water before it was shipped. There may be trace amounts of clean water in the pre-installed tube assembly.

Available Models

Model Number Matrix

FLEXFLO® Model Number

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(1\)

What’s in the Box

  • Blue-White FLEXFLO® Series M1 peristaltic metering pump
  • Power Cord
  • Two (2) Tube assemblies (one installed, one spare)
  • Mounting Hardware Kit / Rear Bracket
  • Display Shield
  • Instruction Manual

See Accessory page for additional options (Suction / Discharge Tubing, Injection Valves, Suction Strainer, M12 communications cables, Wall Mount Bracket. )

ENGINEERING SPECIFICATIONS

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(2\)

  1. Refer to individual tube pressure and temperature ratings.

MATERIALS OF CONSTRUCTION

Wetted Components

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(3\)

Non-Wetted Components

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(4\)

FEATURES

  • Tube Failure Detection (TFD+) senses tube failure and shuts off the pump. No false triggering.
  • Heavy duty display shield protects pump controls.
  • Remote Start/Stop – one non-powered dry contact closure.
  • Compatible with Blue-White Flow Verification Sensor (FVS) system.
  • Outputs: 4-20mA, single 250V/3A (TFD or FVS), and a solid state “motor on” (0-60VDC sinking)
  • Self-priming, cannot vapor lock or lose prime.
  • Speed Control: Manual Speed Adjust and Scalable 4-20mA Input
  • Convenient M12 Connectors (IP67) for Input/Output Communications.(M12 cables ordered separately)

Agency Listings

This pump is ETL listed to conforms to the following: UL Standard 778 as a motor operated water pump. CSA Standard C22.2 as process control equipment

This pump complies to the Machinery Directive 2006/42/EC, BS, EN 60204-1, Low Voltage Directive 2014/35/EU BS EN 61010-1, EMC Directive 2014/30/EU, BS EN 50081-1/BS EN 50082-1.

This pump is certified to NSF/ANSI Standard 61- Drinking Water System Components – Health Effects

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(5\)

ENCLOSURE RATING

NEMA 4X
Constructed for either indoor or outdoor use to provide a degree of protection to personnel against incidental contact with enclosed equipment; to provide a degree of protection against falling dirt, rain, sleet, snow, windblown dust, splashing water, and hose-directed water; and that will be undamaged by external formation of ice on enclosure.

IP66
No ingress of dust; complete protection against contact. Water projected in powerful jets against enclosure from any direction shall have no harmful effects.

INSTALLATION

CAUTION

  • The pump should be serviced by qualified persons only. If equipment is used in a manner not specified in this manual, the protection provided by the equipment may be impaired.
  • Always wear protective clothing, face shield, safety glasses and gloves when working on or near your metering pump. Additional precautions should be taken depending on solution being pumped. Refer to MSDS precautions from your solution supplier.
  • All diagrams are strictly for guideline purposes only. Always consult an expert before installing metering pump on specialized systems. Metering pump should be serviced by qualified persons only.
  • Check system pressure and piping/tubing pressure limits before installing.
  • If using pump with chemicals that off-gas (i.e. sodium hypochlorite), do not leave chemicals in pump or between valves for excessive periods of non-use. Flush pump as necessary.
  • The pump should be supplied by an isolating transformer or RCD (operating current less or equal 30 mA).

Mounting Location

  1. The pump must be sheltered from environment (rain, snow, direct sunlight, etc.) Do not place pump on floor or in location with excessive chemical spills or vapors.
  2. Choose an area located near the chemical supply tank, chemical injection point, and electrical supply. Also, choose an area where the pump can be easily serviced.
  3. Find a secure and level surface and mount the pump close to the injection point. Keep the inlet (suction) and outlet (discharge) tubing as short as possible. Longer discharge tubing increases back pressure at pump head.

NOTE:

  • Mounting the pump lower than the chemical container will gravity-feed chemical into it. This “flooded suction” installation may minimize output variation by eliminating suction lift, and minimizing air and gasses entering the pump due to off-gassing and suction line leaks. A shut-off valve, pinch-clamp, or other means to halt gravity-feed to the pump must be installed during servicing.
  • Install a back flow prevention check valve at the discharge side of the pump to prevent the system fluid from flowing back through pump during tube replacement or during tube leak.
  • A pressure relief valve is recommended at the discharge side of the of pump to prevent premature wear and damage to the pump tube, in the event that the discharge line becomes blocked.
  • The pump does not require back pressure. Keep the discharge pressure as low as possible to maximize

Pump Dimensions

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(6\)

Wall Mounting

  1. Using #10 self-tapping screws, or appropriate hardware for the wall material, mount the bracket to a secure wall that is located no higher than 4 feet above ground for easy programing and service.Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG- \(7\)
    Note: If mounting bracket to shelf, table, or floor, we recommend using mounting holes on bracket feet and using hardware appropriate for base surface. See Accessories for Wall Shelf options.

  2. Lower the pump so that the tab on the wall mount is inserted into the slot located on the back of the pump. The pump will now be secured to the mount bracket. To remove, pull tab forward and lift pump.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(8\)

POWER CONNECTIONS

WARNING

  • Risk of electric shock – cord connected models are supplied with a grounding conductor and grounding-type attachment plug. To reduce risk of electric shock, be certain that it is connected only to a properly grounded, grounding-type receptacle.
  • Electrical connections and grounding (earthing) must conform to local wiring codes.
  • Ensure to connect the pump to the proper supply voltage. Using the incorrect voltage will damage the pump and may result in injury. The voltage requirements is printed on the pump serial label.
    • Use the voltage for which the power cord is rated.
    • To prevent electronic noise interference, electronic signal wires and AC power wires must be kept separate. Do not bundle these cables together or run within the same conduit.
    • When there is a power interruption, the pump will restart (resume) in the same state as prior to power interrup-tion.
    • POWER: 115V60Hz (0.6A max.), 220V50Hz (0.3A max.), 230V60Hz (0.3A max.), 230V50Hz (0.3A max.), 240V50Hz (0.3A max.)
    • COVERS FOR USB CONNECTION AND M12 CONNECTIONS MUST BE IN PLACE WHEN NOT CONNECTED TO CABLES

NOTE: When in doubt regarding your electrical installation, contact a licensed electrician.

Pump (Right Side Panel)

IO Connection

WARNING

  • Risk of electric shock – All wiring must be insulated and rated 60V minimum.

COVERS FOR USB CONNECTION AND M12 CONNECTIONS MUST BE IN PLACE WHEN NOT CONNECTED TO CABLES (Failure to keep connecors in place may void warranty)

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(10\)

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(11\)

M12 cables not included with product. See Accessories for M12 cable options.
A-Type M12 connector with 5 position female sockets required.
Shielded cable should be used on all signal wires.

M12 Connector

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(12\)

M12 Input Connector I/O 1

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(13\)

M12 Input Connector I/O 2

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(14\)

M12 Output Connector I/O 3

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(15\)

TOUCHPAD LAYOUT

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(16\)

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(17\)

SET-UP AND PROGRAMMING

Remote Start/Stop

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(18\)

Item Number Directions for Normally Open Operation
1 Confirm that pump is in the OFF position
2 Press Remote Start/Stop button to activate
3 Light will turn Green for “Normally Open” Operation
4 To put pump is Ready state, press Start/Stop button
5 Display will show “SIG” when waiting for contact closure signal
6 Display will show pump speed (or RPM, 4-20ma) when signal is closed
Item Number Directions for Normally Closed Operation
--- ---
1 Confirm that pump is in the OFF position, and Remote Start/Stop light is

off
2| Press and hold Remote Start/Stop button for 5-7 seconds
3| Light will turn Red for “Normally Closed” Operation
4| To put pump is Ready state, press Start/Stop button
5| Display will show “SIG” when waiting for contact closure to Open
6| Display will show pump speed (or RPM, 4-20mA) when signal is Open

Programming FVS (Flow Verification Sensor)

Flow Verification System (FVS) uses pulse signal from flow meter/sensor. When FVS feature is activated, if running pump does not receive signal/pulse in the programmed delay time, the pump will stop and activate Alarm Output Relay. Display will read “FuS”. Contact Blue-White for sensor options.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(19\)

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(20\)

Activate by pressing FVS button once. Light On indicates activation. Press again to de-activate.

NOTE: Alarm Delay Time range 1-20 seconds

Pump Lockout

Pump Lockout feature allows the user to “lock out” the ability to change pump controls, such as Pump Speed,Prime, FVS, Remote Start/Stop, Manual/4-20ma Modes. User will still be able to start and stop the pump.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(21\)

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(22\)

4-20mA Input

Selecting 4-20mA Input Mode

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(23\)

Directions

  1. Confirm that pump is in the OFF position
  2. Press 4-20mA Input button

Programming 4-20mA Input Mode

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(24\)

Directions

  1. Confirm that pump is in the OFF position.
  2. Press and hold 4-20mA Input button.
  3. MIN mA light will illuminate. Use arrow keys to toggle min mA value for the pump to operate at
  4. Press 4-20mA Input button for MIN % Speed light to illuminate
  5. Use arrow keys to toggle MIN % Speed value for the pump to operate.
  6. Press 4-20mA Input button for MAX mA light to illuminate.
  7. Use arrow keys to toggle MAX mA value for the pump to operate
  8. Press 4-20mA Input button for MAX % Speed light to illuminate.
  9. Use arrow keys to toggle MAX % Speed value for the pump to operate.
  10. Press and hold 4-20mA Input button to exit programming mode.

PUMP OPERATION

Priming The Pump

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(25\)

Need help calibrating your pump? Click on … or scan the QR code.

Manual Pump Operation

Pump speed can be adjusted manually using “Manual Adjust” Run Mode.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(27\)

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(28\)

4-20 Input Speed Control

Pump speed can be adjusted remotely using “4-20ma Input” Run Mode.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(29\)

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(30\)

OUTPUTS

  • MOTOR ON – Active when motor is running in Manual or 4-20ma Mode.
  • RELAY OUTPUT – Active when either TFD or FVS has triggered. Stop pump to reset. 4-2mA
  • OUTPUT – Non-scalable full-scale output proportional to pump speed.

ENHANCED TUBE FAILURE DETECTION (TFD+)

The pump is equipped with a Enhanced Tube Failure Detection (TFD+) system, which is designed to stop the pump and provide an output alarm in the event the pump tube should rupture, and a chemical enters the pump head.
This TFD+ system can detect the presence of many chemicals, including sodium hypochlorite (chlorine), hydrochloric (muriatic) acid, sodium hydroxide, oils- based polymers, water based polymers, and many others. The system will not be triggered by water (rain, condensation, etc.) or silicone oil (roller lubri- cant).

NOTE: If the system has detected a chemical, the pump tube must be replaced, and the pump head and roller assembly must be thoroughly cleaned. Failure to clean the roller assembly will void the warranty.

NOTE: If the TFD+ alarm is triggered, the pump will stop, and close an alarm output. See M12 Connector details to wire this alarm.

Confirming Chemical Detection

To determine if a chemical will be detected by the system:

  1. Remove the pump head cover, and the pump tube and roller assembly.

  2. Place a small amount of chemical in the bottom of the pump head that is enough to cover the sensors.

  3. Reinstall only the pump head cover.

  4. Turn on the pump by pressing the START button.
    NOTE: If the TFD+ system detects a chemical, the pump will stop after a two-second confirmation period. NOTE: If the TFD+ system does not detect a chemical, the pump will continue to operate after the confirmation period.

  5. Carefully clean the chemical out of the pump head. Ensure to remove all the chemical traces from the sensor probes.

  6. Replace the roller assembly and tubing.

  7. Reinstall the pump head cover.

  8. Press the START button to clear the alarm condition.

  9. Restart the pump.

PUMP MAINTENANCE

CAUTION
Always wear protective clothing, face shield, safety glasses and gloves when working on or near your metering pump. Additional precautions should be taken depending on the solution being pumped. Refer to MSDS precautions from your solution supplier.

Routine Inspection and Maintenance

The pump requires minimal maintenance. However, the pump and all the accessories should be checked weekly, especially when pumping chemicals. Inspect all the components for signs of leaking, swelling, cracking, discoloration, or corrosion. Immediately replace worn out or damaged components.

Cracking, crazing, or discoloration during the first week of operation are signs of a severe chemical attack. If this occurs, perform the following steps:

  1. Immediately remove the chemical from the pump.
  2. Determine which parts are being attacked.
  3. Replace the damaged parts with parts that have been manufactured using more suitable materials.
  4. After servicing, operate the pump to verify normal operation.

NOTE: The manufacturer does not assume responsibility for damage to a pump that has been caused by a chemical attack.

Cleaning and Lubricating the Pump

  • The pump will require occasional cleaning, depending on type of service, operating conditions, and run time.
  • When changing the pump tube assembly, the pump head chamber, the roller assembly, and the pump head cover should be wiped of any dirt and debris.
  • Clean the motor shaft with a clean towel, and then apply a small amount of grease to the shaft. This will help prevent the rotor from sticking to the motor shaft.
  • Periodically, or when necessary, grease the pump head cover bearing. Apply a small amount of grease (Aeroshell aviation grease #5 or equivalent).
  • 100% silicone lubrication may be used on the roller assembly.
  • Periodically clean the injection fitting /check valve assembly, especially since injecting fluids, like sodium hypochlorite, can calcify. These lime deposits and other buildups can clog the fitting, increase back pressure, and interfere with the check valve operation.
  • Periodically clean the suction strainer.

Not Sure … Watch Instructional Video
Click on … or scan QR code.

TUBE REPLACEMENT

CAUTION

  • Prior to service, pump clean water through the pump and suction / discharge line to remove any chemical.
  • Always wear protective clothing, face shield, safety glasses and gloves when working on or near your metering pump. Additional precautions should be taken depending on solution being pumped. Refer to MSDS precautions from your solution supplier.
  • Use the provided Tube Installation Tool to leverage tubing into the pump head. Do not use your fingers.
  • Use extreme caution when replacing the pump tube. Do not place your fingers near the rollers.

The pump tube assembly will eventually become non-operational if it is not regularly inspected and replaced. The tube life is affected by many factors, such as the type of chemical being pumped, the amount of back pressure, the motor revolutions per minute (RPM), and temperature.

Tube Removal

CAUTION

Safety first. Remove the pressure. Relieve (remove) the system pressure on the discharge and suction side of the pump. Failure to do so will cause the solution to squirt when disconnecting the tube connections.

  1. Disconnect the system plumbing from the pump tube adapters.
  2. Press the Start/Stop key to stop the pump.
  3. Remove the three black thumb screws from the front of the pump head cover by unscrewing counterclockwise. Remove the pump head cover by pulling straight out.
  4. Set the motor speed to 10%. Press the Start/Stop key to start the pump.
  5.  With the pump running, pull the inlet (suction) fitting out of the pump head. Guide the tube counterclockwise away from the rollers. Pull the outlet (discharge)fitting out of the pump head.Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG- \(34\)
  6. Press the Start/Stop key to stop the pump.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(35\)

Tube Installation

NOTE: Thoroughly clean the pump head and rotor. The rotor can be removed by pulling it straight out. After cleaning, push the rotor back on the shaft.

  1. Set the motor speed to 10%. Press the Start/Stop key to start the pump.

  2. Insert the inlet (suction) side of the pump tube fitting into the lower retaining slot of the pump head. Carefully guide the pump tube into the pump head.

  3. Stretch the tube slightly and insert the outlet (discharge) fitting into the upper retaining slot of the pump head.
    Pull the tube installation tube out of the pump head.Blue-White-M1-Series-
Peristaltic-Metering-Pump-FIG- \(37\)

  4. Place the clear cover onto the pump head. Secure the cover with the provided three thumb screws. The pump is now ready for operation.Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG- \(38\)

Updating The Firmware

REQUIREMENTS: Download & Install the M1 Programmer file by visiting the Firmware Update section of the M1 product page at www.blue- white.com as well as download the Firmware Update File.

  1. Connect the Pump to a computer via USB-A / USB-C cable (Pump must not be powered before starting.) (Be sure to replace USB cover after updating firmware)

  2. Power up the Pump

  3. Open the M1 programer

  4. Select “Enable” USB in the Communication settings window

  5. Click on Connect
    Watch this Video for further assistance.Blue-White-M1-Series-Peristaltic-
Metering-Pump-FIG- \(40\)

  6. Select “Load Hex File”Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG- \(41\)

  7. Select “.HEX” file (See Blue-White website for latest version)

  8. Click on “Erase-Program-Verify”Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG- \(43\)

  9. Once Hex file is loaded successfully

  10. Click on “Disconnect”. Your firmware has now been updated.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(44\)

REPLACEMENT PARTS LIST

Item Description Part Quantity
1 Spacer, back 90011-014 1
  Roller assembly complete (rotor) 71000-350
2 Roller assembly complete (rotor)

71000-159

| 1
 | Roller assembly complete (rotor)| 71000-255|
 | Tube assembly, 1/4” OD, Flex-A-Thane®| A1-1|
 | Tube assembly, 7/16” OD, Flex-A-Thane® Tube assembly, 1/4” OD, Flex-A-Prene®| A1-3
A1-4|
3| A1-6
| 2
Tube assembly, 7/16” OD, Flex-A-Prene®| A1-7|
Tube assembly, 7/16” OD, Flex-A-Chem®| A1-8
|
4                                         Pump head cover, Acrylic| A1-SXX-C| 1
5             Thumb screw with 9/64” key drive, maximum torque 6-8 in. lbs.| 90011-160| 3
6                               Tube nut, compression, for 3/8” tubing| C-330-6| 2
7                               Cover M1 Polycarbonate (not pictured)| 90002-684| 1

NOTES:
Designate “T” Tube Compression, or “M” 1/2” MNPT connection types when ordering.

EXPLODED VIEW

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(45\)

ACCESSORIES

The following accessories are available for the M1 FLEXFLO® Peristaltic Metering Pump. Please visit Bluewhite.com for more information. All accessories are sold separately.

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(46\)

KIT-M12

  • Kit contains: Two M12 cables. 10 foot length.
    Other cable lengths:

  • KIT-M12-2-15 15 foot length.

  • KIT-M12-2-30 30 foot length.

NOTE: THIS DIAGRAM IS FOR THE PUMP’S M12 PORT

KIT-M12-3

  • Kit contains: Three M12 cables. 10 foot length.
    Other cable lengths:

  • KIT-M12-3-15 15 foot length.

  • KIT-M12-3-30 30 foot length.

CABLE-UAC
Kit contains: One 3’ USB-A to USB-C cable.

POWER CORDS – DETACHABLE

  • 90010-663 115V/60Hz NEMA 5/15
  • 90010-664 220V/50Hz CEE 7/V11
  • 90010-665 230V/50Hz BS 1363/A
  • 90010-666 240V/50Hz AS 3112
  • 90010-696 230V/60Hz NEMA 6/15
  • 90010-821 115V/60Hz NEMA 5/15 (Lockable)

SUCTION AND DISCHARGE TUBING

  • C-334-6 Tubing, Suction, clear PVC 3/8” O.D. x 5’ length
  • C-334-6-10 Tubing, Suction, clear PVC 3/8” O.D. x 10’ length
  • C-334-6-100 Tubing, Suction, clear PVC 3/8” O.D. x 100’ length
  • C-335-6 Tubing, Discharge, opaque PE 3/8” O.D. x 5’ length
  • C-335-6-10 Tubing, Discharge, opaque PE 3/8” O.D. x 10’ length
  • C-335-6-100 Tubing, Discharge, opaque PE 3/8” O.D. x 100’ length

KIT-S07
Kit contains: One 7 gallon tank, One 3/8” suction tube, One 3/8” discharge tube, One foot valve and strainer and One mounting bracket with screws

KIT-S15
Kit contains: One 15 gallon tank, One 3/8” suction tube, One 3/8” discharge tube,One foot valve and strainer and One mounting bracket with screws

KIT-S30
Kit contains: One 30 gallon tank, One 3/8” suction tube, One 3/8” discharge tube,One foot valve and strainer and One mounting bracket with screws

KIT-PSM
Kit contains: One HDPE Bracket, (4) 3/8” x 2-3/4” long dia anchor bolts.

TROUBLESHOOTING

Error Code Explanation Troubleshooting
E01 Motor Over Current Check that tube is installed correctly
E02 Over Voltage Check power supply output voltage
E03 Under Voltage Check power supply output voltage
E04 Temperature exceeds 75°C at control Check ambient conditions,

restart pump once cooled to ambient temperature
E05| Inverter Error| Contact Blue-White Industries 714-893-8529 customerservice@blue- white.com
E06| No Motor Connection| Contact Blue-White Industries 714-893-8529 customerservice@blue- white.com
E08| Motor Stall| Check that tube is properly installed
E10| Capacitor bank charging error| Contact Blue-White Industries 714-893-8529 customerservice@blue- white.com
E17| Communication error at display| Contact Blue-White Industries 714-893-8529 customerservice@blue- white.com

WARRANTY

LIMITED WARRANTY
Your new FLEXFLO pump is a quality product and is warrantied for 60 months from date of purchase (proof of purchase is required). The pump will be repaired or replaced at our discretion. Failure must have occurred due to defect in material or workmanship and not as a result of operation of the product other than in normal operation as defined in the pump manual. Warranty status is determined by the pump’s serial label and the sales invoice or receipt. The serial label must be on the pump and legible. The warranty status of the pump will be verified by Blue-White or a factory authorized service center.

Pump Head and roller assembly is warrantied against damage from chemical attack when proper TFD+ (Tube Failure Detection) system instructions and maintenance procedures are followed.

WHAT IS NOT COVERED

  • Pump Tube Assemblies and rubber components – They are perishable and require periodic replacement.
  • Pump removal, or re-installation, and any related labor charge.
  • Freight to the factory, or service center.
  • Pumps that have been tampered with, or in pieces.
  • Damage to the pump that results from misuse, carelessness such as chemical spills on the enclosure, abuse, lack of maintenance, or alteration which is out of our control.
  • Pumps damaged by faulty wiring, power surges or acts of nature.

PROCEDURE FOR IN WARRANTY REPAIR
Contact the factory to obtain a RMA (Return Material Authorization) number. Carefully pack the pump to be repaired. It is recommended to include foot strainer and injection/check valve fitting since these devices may be clogged and part of the problem. Please enclose a brief description of the problem as well as the original invoice or sales receipt, or copy showing the date of purchase. Prepay all shipping costs. COD shipments will not be accepted. Warranty service must be performed by the factory or an authorized service center. Damage caused by improper packaging is the responsibility of the sender. When In-Warranty repair or replacement is completed, the factory pays for return shipping to the dealer or customer.

PRODUCT USE WARNING
Blue-White products are manufactured to meet the highest quality standards in the industry. Each product instruction manual includes a description of the associated product warranty and provides the user with important safety information. Purchasers, installers, and operators of Blue-White products should take the time to inform themselves about the safe operation of these products. In addition, Customers are expected to do their own due diligence regarding which products and materials are best suited for their intended applications. Blue-White is pleased to assist in this effort but does not guarantee the suitability of any particular product for any specific application as Blue-White does not have the same degree of familiarity with the application that the customer/end user has. While Blue-White will honor all of its product warranties according to their terms and conditions, Blue- White shall only be obligated to repair or replace its defective parts or products in accordance with the associated product warranties. BLUE-WHITE SHALL NOT BE LIABLE EITHER IN TORT OR IN CONTRACT FOR ANY LOSS OR DAMAGE WHETHER DIRECT, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL, ARISING OUT OF OR RELATED TO THE FAILURE OF ANY OF ITS PARTS OR PRODUCTS OR OF THEIR NONSUITABILITY FOR A GIVEN PURPOSE OR APPLICATION.

CHEMICAL RESISTANCE WARNING
Blue-White offers a wide variety of wetted parts. Purchasers, installers, and operators of Blue-White products must be well informed and aware of the precautions to be taken when injecting or measuring various chemicals, especially those considered to be irritants, contaminants or hazardous. Customers are expected to do their own due diligence regarding which products and materials are best suited for their applications, particularly as it may relate to the potential effects of certain chemicals on Blue-White products and the potential for adverse chemical interactions. Blue-White tests its products with water only. The chemical resistance information included in this instruction manual was supplied to Blue-White by reputable sources, but Blue- White is not able to vouch for the accuracy or completeness thereof. While Blue-White will honor all of its product warranties according to their terms and conditions, Blue-White shall only be obligated to repair or replace its defective parts or products in accordance with the associated product warranties. BLUE-WHITE SHALL NOT BE LIABLE EITHER IN TORT OR IN CONTRACT FOR ANY LOSS OR DAMAGE, WHETHER DIRECT, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL, ARISING OUT OF OR RELATED TO THE USE OF CHEMICALS IN CONNECTION WITH ANY BLUE- WHITE PRODUCTS.

APPENDIX A: ACRONYMS

  • °C Celsius
  • °F Fahrenheit
  • AC Alternating current
  • bar Unit of pressure
  • CIP Clean-in-place
  • cm Centimeters
  • COD Cash on Delivery
  • D Depth
  • DC Direct current
  • EEE Electrical and electronic equipment
  • EP Ethylene propylene
  • ETL Electrical Testing Labs/Intertek
  • EU European Union
  • FDA Food and Drug Administration
  • FKM Fluoroelastomer
  • FVS Flow Verification Sensor
  • GF Glass fiber
  • GPD Gallons per day
  • GPH Gallons per hour
  • H Height
  • Hz Hertz
  • ID Inside diameter
  • IO Input/Output
  • Kg Kilogram
  • lb. Pound
  • LLDPE Linear low-density polyethylene
  • LPH Liters per hour
  • mA Milliampere
  • min Minute
  • mL Milliliters
  • MSDS Material Safety Data Sheet
  • N.C. Normally Close
  • N.O. Normally Open
  • NPT National Pipe Thread
  • NSF National Sanitation Foundation
  • OD Outside diameter
  • P.N. Part Number
  • PBT Polybutylene Terephthalate
  • PE Polyethylene
  • PSI Pounds per Square Inch
  • PVC Polyvinyl chloride
  • PVDF Polyvinylidene fluoride
  • RCD Residual-current device
  • Rev. Revision
  • RMA Return Material Authorization
  • RPM Revolutions per minute
  • SIP Steam-in-place
  • SS Solid state
  • TFD+ Enhanced Tube Failure Detection
  • TFE/P Tetrafluoroethylene propylene
  • UL Underwriters Laboratories
  • US United States
  • V Volt
  • W Watt
  • W Width
  • WEEE Waste Electrical and Electronic Equipment

Model Number Matrix

FLEXFLO® Model Number

Blue-White-M1-Series-Peristaltic-Metering-Pump-FIG-
\(52\)

Users of electrical and electronic equipment (EEE) with the WEEE marking per Annex IV of the WEEE Directive must not dispose of end of life EEE as unsorted municipal waste, but use the collection framework available to them for the return, recycle, recovery of WEEE and minimize any potential effects of EEE on the environment and human health due to the presence of hazardous substances.

The WEEE marking applies only to countries within the European Union (EU) and Norway. Appliances are labeled in accordance with European Directive 2002/96/EC.

Contact your local waste recovery agency for a Designated Collection Facility in your area.

5300 Business Drive
Huntington Beach, CA 92649
USA

TEL: 714-893-8529

FAX: 714-894-9492
www.blue-white.com
sales@blue-white.com
customerservice@blue-white.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals