PCE INSTRUMENTS PCE-GM 60 Plus Gloss Meter Instruction Manual
- July 23, 2024
- PCE Instruments
Table of Contents
PCE INSTRUMENTS PCE-GM 60 Plus Gloss Meter
Specifications
- Product Name: Gloss Meter
- Models: PCE-GM 60 Plus, PCE-IGM 60, PCE-IGM 100, PCE-PGM 60, PCE-PGM 100
- Standards: ISO 2813, GB/T 9754, ASTM D 523, ASTM D 2457
- Display: 3.5-inch full-view display with a resolution of 480×320
- Power Source: Built-in rechargeable battery
Product Usage Instructions
Safety Instructions
The instrument is a safe device. Before operating, please read the safety
instructions and strictly comply with the following terms to avoid unexpected
damages. We shall not be liable for any losses by incorrect operation.
Battery| The device configures a built-in battery. Please use the original one. Do not use other batteries to prevent unnecessary damage.
Do not disassemble, extrude, or heat the battery.
When fully charged, please cut off the external power supply in time.
Please charge the battery biweekly if long time not use. Otherwise, it may cause damage to the battery.
For the first three times, please run out the power and charge fully to ensure the
battery reaches the optimal state.
---|---
External Power Source| Please use the original power adapter when charging.
Otherwise, it will shorten the battery life or cause an explosion.
Please cut off the external power source if
you don’t use it for a long time.
Gloss Meter| Do not use the device in an inflammable and explosive environment.
Do not disassemble the device which will cause damage and explosion.
Please stop using the machine when you smell something burning and send it back
to the repair center.
Introduction
The gloss meter is by IS0 2813(International Standard) and GB/T 9754(China
National Standard). It has the characteristics of easy-to-use, stable
performance, and precise measurement.
Advantages
- Large Screen (3.5 inches), high resolution (480*320), full-view display. b.Comply with ISO 2813、GB/T 9754、ASTM D 523、ASTM D 2457 standards.
- Aesthetic design perfectly combined with ergonomics structure.
- Three measurement angles (20°, 60°, 85°),can measure simultaneously(except the single-angle gloss meter).
- QC software with powerful extending functions (except certain models).
- Flexible choice of auto-calibrating when powered on, easy to operate.
- Display multiple sets of data, easy to compare.
- High hardware configuration with multiple innovative technologies.
- Built-in rechargeable battery, economical and environmental protection.
- Auto power-off function to save electricity consumption.
Cautions
- The gloss meter is a precise measuring instrument. Please avoid the dramatic changes in the external environment when measuring. Those changes including the flicker of surrounding light, and the rapid change of temperature and humidity will affect the measuring accuracy.
- Keep the instrument balanced. Make sure the measuring aperture clings to the test sample, and no shaking or shifting when measuring. Please prevent the gloss meter from fierce collision or crash. This device is not water-proof. Do not use it in an environment of high humidity or mist.
- Keep the device clean. Avoid dust, powder, or solid particles entering the measuring aperture and the devices.
- After using it, please turn it off. Keep the instrument and calibration board inthe instrument case.
- Keep the instrument in a cool, dry environment.
- Users can not make any changes on the device without permission. Since itmay affect measuring accuracy or even damage the device.
Cautions
- Avoid dramatic changes in the external environment during measurement.
- Ensure the gloss meter is balanced and stable during measurement; prevent collisions or exposure to high humidity.
- Keep the device clean and free from dust or particles.
- Turn off the device after use and store it in a cool, dry place.
External Structure Description
The gloss meter features a large 3.5-inch display and complies with various
standards for accurate measurements.
External Constructions
Button Function Instruction
- LCD Screen: Display measuring data and instrument operation navigation.
- Switch/Measure button: Long-press the button for 3 seconds to turn on or turn off the gloss meter. Short pressing the button to measure.
- Indicator Light : When turning on, it will show a green light. After startup, the light is off. Low power and charging status will show a red light. Fully charged will show a green light.
- RS-232/USB interface: This interface is common. The instrument automatically judges the connection status. USB interface is used to connect and transfer data to the PC. RS-232 interface is used to connect to the printer. USB cable can connect the power adapter and computer for charging the gloss meter (specification for the external adapter is: 5V=2A).
- Protective Cover (Calibration plate): It is to protect the measuring aperture. A built-in calibration plate is used for instrument calibrating.
Attention: The method to separate the protective cover from the instrument is shown in Figure 2. Hold the instrument with one hand, the other hand hold the protective cover, and then separate it according to the “Open” mark. You only need to separate it from one side. Do not separate it from two sides.
Operating Instructions
Turning On/Off
To turn on/off the device, follow the manufacturer’s instructions provided in
the manual. Long-press the switch button for 3 seconds to turn on the
instrument. The LCD screen will display the boot logo. After a few seconds, it
will enter into the measurement interface automatically, as shown in Figure 3.
Long-press the switch button for 3 seconds again, it will be turned off. The
instrument will enter the standby mode if without operation within five
minutes. And it will turn off atomically if without operation within one
minute in standby mode.
Calibration
The instrument comes with an auto-calibrating function. Follow the guidelines
for manual calibration if needed for specific applications.
Auto-Calibrating Description
The instrument has an auto-calibrating function. But for flexible use, it has
an optional designation to judge whether it will auto-calibrate at startup. If
the last power-off is done manually, it will calibrate automatically when
turned on next time. If the last power-off is automatic after 5 minutes of
backlight off, then it will not calibrate automatically when turned on next
time. When the operating environment changes (such as the rapid change in
temperature, altitude, and humidity), it must do calibration. To ensure
accuracy, please use the original standard plate for calibration. The dust on
the standard plate will affect calibration accuracy. Please clean the standard
plate and ensure it’s clean when calibrating. The Standard plate is a precise
optical component. Please avoid it from strong sunlight. Due to environmental
factors, the gloss value of the standard plate will change over time.
Therefore, it’s better to send it back to the factory or qualified local
National Institute of Metrology for calibration. (Once a year)
Attention:
- Each instrument has only a calibration plate. If using another calibration plate to pass the calibration, the measurement is also not accurate. Therefore, before calibrating, please check whether the serial number of the instrument is the same as that on the calibration plate.
- Before calibrating, please make sure the instrument and calibration plate are tightened. Otherwise, the calibration may fail and the measurement results may not be accurate.
Change Calibration Values
Users can change instrument calibration values through QC software.
Attention: Modify calibration value is better operated by the manufacturer or qualified metrological institutes. The calibration value needs to be modified only when it is different from the actual calibration plate value. Before modifying the standard value, please back up the original standard value.
Measurement
Place the measuring aperture on the test sample and ensure stability during
measurement to obtain accurate results. The measuring mode is basic (Figure
4). Basic mode is a basic sample test mode that will display the glossiness
value directly. It belongs to a single measurement. The results will be saved
automatically each time (except for certain models). It can display multiple
sets of test data simultaneously.
- “T005” at the upper left means the record number of the last measurement.
- “16:12” and “2015.10.23” are time and date.
- “T001-T005” is the record number of five measurements. (Some models only display three records.)
- “T102316” is the default name of the measuring record. It is made up of “T”+” month”+” day”, “T” means a basic record, and “102316” means the measuring record 4:00pm0 pm on 23rd October.
The last record will be shown in yellow.
Data Save
Utilize the data save feature as needed for recording measurements. It
defaults to save data automatically. Users can set a not-save function through
QC software. The instrument can save 1000 data. When it’s full, it will
prompt. If continue to measure, the new record will cover the last one. Users
can delete data or manage other operations through QC software. (Remarks: some
models don’t have this function.)
Connecting to PC
For connecting the gloss meter to a PC, refer to the user manual for detailed
instructions. Turn on the instrument, and connect the USB to the PC. Then the
instrument is charged and it will connect to the software automatically. Users
can control the measurements through QC software.
Remarks: Some models don’t have this function.
Print
If printing is required, follow the provided instructions for printing data
from the gloss meter.
Print
If the gloss meter connects miniature printer, it will print the testing data
when measuring.
Function Description of QC Software
When connecting to QC software, users can manage the following operations:
- Check Status (Instrument basic information, such as model, and serial number)
- Calibrate
- Change Calibration Value (It’s better to handle by manufacturer of qualified metrological institutes.)
- Data Management (Check record, delete record, export record, print report)
- Set Time and Data
- Set Language
- Select Angles (only for tri-angle gloss meter)
- Set Auto Save or Not-save Function.
Routine Maintenance
Regularly clean the device to prevent dust or particles from entering the measuring aperture. Store the instrument properly after each use.
-
The gloss meter is a precise instrument. Please operate and store it in a standard laboratory environment (Temperature: 20℃, Standardatmospheric pressure, Humidity: 50~70%RH). Please avoid using it in humid environments, strong electromagnetic interface environments, highlight intensity environments, and dusty environments.
-
A standard plate is a precise optical component. Avoid being damaged by sharp objects, avoid dirtying the board, and avoid exposing it to the sun. Regularly clean the standard plate by using a soft cloth with alcohol in one direction. Ensure there are no tiny particles or sundries on the soft cloth. Before calibrating, clean the standard plate to make accurate calibration.
Attention: Forbid to use acetone solvent! -
To ensure the accuracy of the instrument, it is better to send it back to the factory or qualified Local National Institute of Metrology for calibration. (Once a year)
-
It needs to modify the calibration value when the calibration value is different from the practical measurement value. View Chapter 2.2 for details.
-
The gloss meter is powered by a built-in battery. When a long time of not using this instrument, please charge it biweekly to protect the battery and extend the battery life.
-
The internal clean work of the instrument should be finished by manufacturer (suggest once a year). Do not use cleaning tools to clean the internal instrument. Otherwise, it will cause irreversible damage.
Technical Specifications
Refer to the user manual for detailed technical specifications of the gloss meter.
Measurement Angle | 20°/60°/85°/20°60°85° |
---|---|
Standard | ISO 2813, GB/T 9754, ASTM D 523, ASTM D 2457 |
Measuring Area(mm) | 20°: 10×10, 60°: 9×15, 85°: 5×36 |
Measuring Range | Multi-angle Models |
20°: 0-1000GU
60°: 0~1000GU
85°: 0~160GU
| 60°: 0~300GU| 60°: 0~200GU
Division Value| Some Models: 0.1 GU
Some Models: 1 GU
Range
| Multi-angle Models| Single Models| Some Single Models
0-10GU| 10-100GU| 100-1000GU| 0-10GU| 10-100GU| 100-300GU| 0-200GU
Repeatability| ±0.1GU| ±0.2GU| ±0.2GU%| ±0.1GU| ±0.2GU| ±0.2GU%| ±1GU
Reproducibility| ±0.2GU| ±0.5GU| ±0.5%GU| ±0.2GU| ±0.5GU| ±0.5%GU| ±1GU
Deviation| ±1.5, ±1.5%
Accuracy| Conform with JJG696 first-class gloss meter working requirements.
Chromaticity Corresponding| Corresponding with CIE 1931(2°) under CIE C light
source.
Measuring Time| 0.5s
Dimension| LWH : 160mm75mm90mm
Weight| 350g
Battery| 3200mAh Li-ion Battery, >5000 times (within 8 hours)
Display| TFT 3.5inch, resolution: 320*480
Interface| USB/RS-232
Storage| 1000
Software| GQC6 Quality Control Software with QC report printing function and
more extended
functions (except certain models).
Operation Temperature| 0~40℃ (32~104°F)
Storage Temperature| -20~50℃ (-4~122°F)
Humidity| <85% relative humidity, no condensation
Standard Accessories:| Power Adapter, USB cable, User Manual, GQC6 software
(except certain models), Wiping
Cloth, Calibration Plate
Optional Accessories:| Miniature Printer
Note| The specifications are subject to change without notice.
PCE Americas Inc.
- 1201 Jupiter Park Drive
- Suite 8
- Jupiter
- FL-33458
- USA
- From outside the US: +1
- Tel: 561-320-9162
- Fax: 561-320-9176
- info@pce-americas.com
PCE Instruments UK Ltd.
- Suite 1N-B, Trafford House
- Chester Rd
- Manchester M32 0RS
- United Kingdom
- From outside UK : +44
- Tel: (0) 2380 98703 0
- Fax: (0) 2380 98703 9
- info@pce-instruments.co.uk
FAQ
Q: Can I use the gloss meter in high-humidity environments?
A: It is not recommended to use the gloss meter in high-humidity environments
as it is not waterproof and may affect measurement accuracy.
Q: How do I know when to calibrate the gloss meter?
A: The gloss meter has an auto-calibrating function and will indicate when
calibration is needed based on usage patterns. Refer to the manual for
calibration instructions.