THORLABS FPQFA Series Tunable Narrow Bandpass Fabry-Perot Filter User Guide
- July 3, 2024
- THORLABS
Table of Contents
- THORLABS FPQFA Series Tunable Narrow Bandpass Fabry-Perot Filter
- Product Information
- Product Usage Instructions
- Introduction
- Safety
- Installation
- Operation
- Maintenance and Cleaning
- Troubleshooting and Repair
- Disposal
- Thorlabs Worldwide Contacts
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
THORLABS FPQFA Series Tunable Narrow Bandpass Fabry-Perot Filter
Product Information
Specifications
FPQFA Specifications:
- Free Spectral Range: 30 GHz
- Finesse: >300
- Resolution: 80% Typ.
- On-resonance transmission Extinction (Excluding higher-order modes): 30 dB
- Cavity Length: 5 mm
- Piezoelectric Transducer Voltage: 0 to 150 V
- Voltage 1 FSR (Voltage < 50 V, 550 – 1300 nm)
- Electric Impedance
Item #: FPQFA-5
Mirror Substrate: UV Fused Silica
Mechanical Drawings
Product Usage Instructions
Safety Precautions
Take appropriate measures for eye safety. Never look directly into the laser
beam, not even at the output of the FPQFA.
Electrical Safety
- Risk of Electric Shock – Never touch the output of the piezoelectric driver.
- Never apply voltages below 0 V or above 150 V as it may damage the FPQFA.
Installation
Warranty Information:
This precision device is only serviceable if returned and properly packed into
the complete original packaging including the complete shipment plus the
cardboard insert that holds the enclosed devices. If necessary, ask for
replacement packaging. Refer servicing to qualified personnel.
FAQ
-
Q: What should I do if I accidentally apply a voltage above 150V?
A: If you accidentally apply a voltage above 150V, immediately disconnect the power source and do not attempt to use the product. Contact qualified personnel for assistance. -
Q: Can I clean the mirrors with regular glass cleaner?
A: No, it is recommended to use specialized cleaning solutions for optical components to avoid damaging the mirrors.
Introduction
Intended Use
The product may only be used by the instructions described in this manual. Any
other use will invalidate the warranty.
Explanation of Safety Warnings
Indicates information considered important, but not hazard-related, such as
possible damage to the product.
- Laser Radiation Warning
- Shock Warning
- The symbol on the product, the accessories or the packaging indicates that this device must not be treated as unsorted municipal waste but must be collected separately
Description
-
The FPQFA is a narrow tunable optical Fabry-Pérot filter with high transmission and large Free Spectral Range (FSR) used to examine the spectral fine-structure characteristics, especially for low light emitters. Due to the carefully designed cavity with high transmission and high extinction, the FPQFA is suitable to be used in a feedback loop to lock the transmission at a specific wavelength and reject other wavelengths including an optical pump beam.
-
The optical filter is based on a non-confocal cavity consisting of two high-reflectivity mirrors, one planar and one spherical. By varying the mirror separation with a piezoelectric transducer, the cavity acts as a very narrow tunable optical bandpass filter. Knowing the FSR of the FPQFA allows the time-base of an oscilloscope to be calibrated to facilitate quantitative measurements of a spectra. With a transmission of more than 80 %, 30-GHz FSR, and a resolution of better than 100 MHz, the emission spectra from low-light quantum emitters such as InGaAs quantum dots, Si and N vacancies in diamond can be studied efficiently.
Figure 1 Illustration of the non-confocal plano-concave cavity where the spherical mirror radius of curvature, R, is larger than the mirror spacing, d.
The non-confocal cavity in the FPQFA differs from the confocal cavity in the Thorlabs’ SA series of scanning Fabry-Pérot interferometers. In the latter, both mirrors are spherical with the same radius, which is also equal to the mirror separation. In the FPQFA, the cavity consists of one planar and one concave spherical mirror where the mirror radius is significantly larger than the mirror separation, see R and d in Figure 1. The benefit of a plano-concave cavity is that it is easier to achieve high performance at a large FSR i.e., smaller mirror distances, than with a confocal cavity while still having better spatial mode control than in a cavity with planar mirrors (plano- plano). A drawback is that the alignment of the cavity is slightly more sensitive than the confocal cavity but less sensitive than the plano-plano cavity.
From a user’s perspective, the free spectral range of a non-confocal cavity is given by:
- ??? = ?/2? ,
where c is the speed of light and d is the cavity length. N.B. the factor 2 in
the denominator as for a plano-plano cavity instead of a factor 4 as in a
confocal cavity such as the Thorlabs’ SA series of scanning Fabry-Pérot
interferometers.
For discussions on the spatial mode structure, transmission spectrum, and
finesse of a Fabry-Pérot interferometer, please see the full web presentation
at www.thorlabs.com.
Technical Data
Specifications
FPQFA Specifications
Free Spectral Range| 30 GHz
Finesse| >300
Resolution| <100 MHz
On-Resonance Transmission| >80 %
Extinction (Excluding higher order modes)| Typ. 30 dB
Cavity Length| 5 mm
Piezoelectric Transducer Voltage| 0 to 150 V
Voltage 1 FSR
(Voltage < 50 V, 550 ≤ **l ≤ 1300 nm)**
| ≤22×(l/845) V
Electric Impedance| 2.2 µF
Item #| Wavelength Range| Mirror Substrate
---|---|---
FPQFA-5| 550 – 845 nm| UV Fused Silica
FPQFA-8| 845 – 1300 nm| UV Fused Silica
Mechanical Drawings
Safety
-
Warning
Take appropriate measures for eye safety. Never look directly into the laser beam, not even at the output of the FPQFA. -
Risk of Electric Shock
Never touch the output of the piezoelectric driver. -
Notice. Never apply voltages below 0 V or above 150 V as it may damage the FPQFA.
Installation
Warranty Information
This precision device is only serviceable if returned and properly packed into
the complete original packaging including the complete shipment plus the
cardboard insert that holds the enclosed devices. If necessary, ask for
replacement packaging. Refer servicing to qualified personnel.
Packing List
- FPQFA Tunable Narrow Bandwidth Fabry-Pérot Filter
- 2 SM05CP2 End Caps
Installation Instructions
- Depending on the application, insert the device in a fixed or kinematic mirror mount for Ø1″ optics.
- Remove the end caps.
- Connect the piezo cable to the driving electronics before turning it on.
- Follow the alignment instructions.
Operation
Warning
Take appropriate measures for eye safety. Never look directly into the laser
beam, not even at the output of the FPQFA.
General
Figure 3 The FPQFA used in two different applications: (a) As a frequency selective filter in a closed-loop feedback system. (b) As a frequency scanning filter to examine the light’s spectral properties.
The FPQFA Tunable Narrow Bandpass Fabry-Pérot Filter can be used in two major modes of operation, see Figure 3a and b:
- As an electrically tunable optical filter selecting a part of the incoming light’s spectrum given by the filter’s Lorentzian-shaped transmittance. In this case, the spectrally filtered light is collected by the down-stream optics for further use.
- As a frequency scanning filter in combination with a detector to measure the spectral structures of the incoming light.
Irrespective of which application the FPQFA will be used, it is very important to properly shape the illuminating light beam to match the fundamental mode supported by the cavity. This is also known as mode matching and will be discussed in the following sections.
Confined cavity mode
Figure 4 A plano-concave cavity illuminated with a mode matching the cavity fundamental mode.
The plano-concave cavity supports a fundamental mode whose waist is on the flat mirror, with a Gaussian intensity profile:
where the spot size ωcav is determined by the cavity length d and the radius of curvature of the curved mirror R by:
Waist Size vs Wavelength
Effective Waist Position vs Wavelength
For the FPQFA cavity, d = 5 mm, and R = 250 mm. The value of ωcav depends on the wavelength, with representative values plotted in Figure 5. If the light transmitted by the cavity should be collected by downstream optics, it is important to consider the impact of the spherical mirror. The spherical mirror will slightly affect the waist size and waist position as seen after the spherical mirror in Figure 4. This effective waist size is also plotted in Figure 5, and the position relative to the actual waist position (coinciding with the flat mirror) is plotted in Figure 6. A positive effective waist position means that the effective waist is located after the rear waist position in Figure 4.
It is important to recognize that the geometry of this confined cavity mode is completely determined by the cavity and the wavelength, independently of the light incident upon the cavity. In practice, we wish to “mode-match” the incoming light to the cavity by shaping the ingoing light to have the same spot size at the waist, positioning the waist of the ingoing light at the same position as the waist of the cavity mode, and orienting the ingoing light in the same direction as the cavity mode.
Mode matching
The two most common situations that arise are that of a free-space incident
beam, and that of light emerging directly from an optical fiber. We will treat
both cases in turn.
Free-space beam
In most cases, the incident beam has a Gaussian profile with the intensity distribution
Note that ω0 is the radius at which the intensity has dropped from the on-axis intensity by a factor of 1⁄?2; 86 % of the optical power is contained within this radius. It is conventional to refer to ω0 as the “radius” of the beam and 2ω0 as its “diameter”. It is straightforward to measure ω0 with a knife edge, a beam profiler, or a CMOS camera. After propagation through the lens with focal length f, the light is focused to a spot size
a distance
from the lens, as shown in Figure 7. The equations tell us that a lens with a smaller focal length f will focus a beam more tightly. The parameter z0 is called the Rayleigh range and is the distance at which the spot size of the beam has increased a factor of √2 from the waist size. It is calculated as
For z0 ≫ f, the expression for the spot size reduces to
and the distance to
To mode-match from a collimated beam of radius ω0 to a cavity mode spot size ωcav, one should use a lens with a focal length
For example, a collimated beam from a TC06 Triplet Fiber Optic Collimator entering the lens with radius ω0 = 0.56 mm and requiring a focused spot size of ωcav = 84 µm at λ= 633 nm, gives a focal length of f = 233 mm. To mode- match an incident beam of the same radius ω0 to a cavity mode radius ωcav = 120 µm at λ= 1300 nm, a focal length of f = 163 mm is needed. Lenses with focal lengths of f = 250 mm in the first case, and f = 150 mm or 175 mm in the latter example would be reasonable choices. In practice, the incident beam radius ω0 is usually not known precisely, and satisfactory results will be obtained with a focal length of in the vicinity of the calculated value. For best results, it is recommended to verify the waist spot size and position by measurement as described earlier. If the setup cannot accommodate the lens due to space constraints, increasing the incident beam waist ω0 will also increase the required focal length.
Optical fiber
The other situation that occurs frequently is that the light to be injected into the cavity emerges from a single-mode fibre. In this case, an adjustable fibre collimator, such as Thorlabs CFC5A or PAF2-7A, can shape the optical mode to the desired shape. The light emerges from the fiber with a spot size given by
where θis the numerical aperture of the fibre (typically around θ= 0.13). As an example, for fibre 780HP with θ= 0.13 at a wavelength of 845 nm, one finds ω0 = 2.06 µm. Note that this is slightly smaller than half of the given mode field diameter of 5.0 ± 0.5 µm specified for this fibre.
If the distance d1 from the fibre tip to the lens in the collimator is taken to be
The light emerging from the collimator will be refocused to a spot size ω1 at a distance
from the lens, as shown in Figure 8. In practice, one chooses a collimator with a focal length that will give a convenient distance of d2 to the cavity. For example, with ω0 = 2.07 μm and ω1 = ωcav = 97 μm, one finds
A collimator with a focal length of f = 4.6 mm, such as Thorlabs CFC5A, would give d1 = 4.69 mm and d2 = 221 mm. In a real setup, one would adjust the collimator (and thus d1) to get the desired waist size and position of the emerging beam.
Alignment
After selecting the proper lens according to the previous section, it is
necessary to have enough degrees of freedom in the alignment setup to adjust
the incoming focused beam to achieve proper mode-matching. There are several
possible solutions to achieve the necessary degrees of freedom of which two
are presented here.
Figure 9 Example of an alignment setup with the five necessary degrees of freedom for mode-matching (only three degrees of freedom are shown in the figure), where the FPQFA is used as a tunable optical filter.
Correct mode-matching requires spatially overlapping the injected light with the fundamental mode of the cavity. This requires five degrees of freedom, three for translation (x, y, and z) and two for angular alignment (around x- and y-axes).
Figure 9 illustrates a setup where the cavity is fixed, and the incident beam is shaped to match the cavity mode and aligned to the cavity. This setup is suitable when the light transmitted through the cavity is collected by downstream optics for further processing since the direction of that light is set by the cavity and hence fixed. Two steering mirrors provide control of the position and angle of the incoming beam, and the axial position of the waist is controlled by moving the focusing lens. Of all the degrees of freedom, the axial position (z) is the least critical. A characteristic length scale for it is provided by the distance
at which the spot size of the confined mode increases due to diffraction by a factor of √2, indicating that the axial position needs to be correct to only a few millimetres. This distance is also known as the Rayleigh range. If space permits, an optional beam splitter can help establish the correct lens position. When the waist is correctly positioned on the flat mirror of the cavity, the reflected beam will be collimated which can be evaluated by looking at the reflection from the beam splitter. Also, an aperture placed before the steering mirrors can help to ensure that the incident beam is perpendicular to the flat cavity mirror. Aligning the mirrors such that light reflected by the cavity passes back through the aperture ensures that the injected beam is perpendicular to the cavity input coupling mirror. In a fibre system, the same goals may be achieved by using a circulator and optimizing the reflected power.
After this first rough alignment, when placing a detector after the cavity and displaying the signal on an oscilloscope, one can expect to see a transmission signal as the cavity length is scanned with the piezoelectric actuator, see Figure 10a. To optimize the alignment of the two steering mirrors, the suggested procedure is as follows:
- With the mirror furthest from the cavity, turn the knob controlling the horizontal motion a small amount in the clockwise direction, then optimize the transmitted power with the knob controlling the horizontal motion on the other mirror.
- If the transmission is better than where it started, return to the first mirror, and turn the knob again in the clockwise direction and repeat.
- If the transmission is worse than where it started, turn the knob on the first mirror in the counterclockwise direction and repeat.
- Continue the process until the horizontal degree of freedom is optimized.
- Repeat the procedure in the vertical motion.
- Iterate optimizing the horizontal and vertical motion as necessary until the best transmission is found.
At this point, there should be very little power coupling into higher-order modes (Figure 10d). If the mode-matching is still not satisfactory, it may be necessary to translate the lens axially or change the focal length.
Figure 11 Example of an alignment setup with the five necessary degrees of freedom for mode-matching (only three degrees of freedom are shown in the figure), where the FPQFA is used as a frequency scanning filter.
There are many alternatives to the setup illustrated in Figure 9 in which the cavity is fixed, and two steering mirrors are used for alignment. One popular scheme has the cavity itself on a kinematic mirror mount (we recommend the Polaris-K1E or KS1 mount), providing two degrees of freedom, in which case only one steering mirror is needed to provide two more degrees of freedom, see Figure 11. This setup has the benefit of easier alignment compared to the previous example since the coupling between beam angular and position alignment is less. The steering mirror mainly affects the horizontal and vertical spot position in the cavity and the cavity tip-tilt mainly affects the injected beam’s direction in the cavity. The drawback is that the transmitted beam after the cavity will change direction during the alignment. Hence, this setup is most suitable when a detector or camera is placed directly after the cavity.
FPQFA Mounting
When mounting the FPQFA in a setup, it is recommended that the incoming light enters the device through the entry port located on the side marked with the red ring, see Figure 12a. Also, the axial focus position of the shaped incoming beam should coincide with the focus alignment groove on the housing. This groove is aligned with the internal planar cavity mirror. To facilitate alignment for the user, when the FPQFA is mounted in a Thorlabs mirror holder for Ø1”-optics (such as the Polaris-K1E or KS1), the focus alignment groove coincides with the mirror holder front face, see Figure 12b.
If needed, the FPQFA can also be operated in reverse with the incoming light entering the device through the exit port if the focus position is aligned to the focus alignment groove.
Driving Electronics
Risk of Electric Shock
Never touch the output of the piezoelectric driver.
NOTICE
Never apply voltages below 0 V or above 150 V as it may damage the FPQFA.
Depending on the mode of operation, the requirements of the electronics used to drive the FPQFA tunable optical filter are different. If the FPQFA is used in a closed feedback loop, the drive electronics must be able to generate the voltage and current to follow the PID correction signal. In case the FPQFA is used as a frequency scanning device, the drive electronics need to be able to generate a sawtooth or triangle wave of high enough voltage and current to drive the capacitive load, such as the Thorlabs SA201B.
Piezo Displacement vs. Voltage
Figure 13 Piezoelectric actuator displacement without load versus applied voltage for increasing and decreasing voltage.
Common to all piezoelectrical actuators is that they are non-linear and exhibit hysteresis, including the actuator used in the FPQFA. The typical displacement of the piezoelectrical actuator when applying a static voltage across it is displayed in Figure 13. Due to that the displacement is not linear with the applied voltage, the required voltage to scan the filter 1 FSR (a distance ?⁄2) is not only wavelength dependent but also dependent on the applied voltage. Since the displacement per Volt decreases with increasing voltage, the required voltage increases with increasing voltage. An expression for a rough estimate of the required voltage for 1 FSR at voltages below 50 V is
where λis the wavelength in nm.
The maximum current requirement when applying a time-varying voltage over a capacitive load is:
where C is the capacitance of the piezoelectric element and ??(?)⁄???? is the maximum slope of the signal (slew rate).
If the FPQFA is used in a high-speed closed-loop control system, the current requirement will depend on the characteristics of the control system, such as bandwidth, latency, etc. Depending on the requirements, an amplifier with a higher output current, such as the BPA100, may be required.
An expression for the maximum current requirement when applying a sinusoidal signal is:
where f is the frequency, C is the piezoelectric actuator capacitance, and Vpp is the peak-to-peak voltage.
When applying a triangular wave with the SA201B control box for Scanning Fabry-Pérot Interferometers (maximum output current 25 mA), the maximum voltage slope is 11 V/ms, which can be sufficient for applications when the FPQFA is used as a frequency scanning device.
Maintenance and Cleaning
- There are no user-serviceable parts on the FPQFA, and the device should only be sent back to Thorlabs for fault analysis and repair after consultation with Thorlabs personnel.
- If needed, the outside of the FPQFA housing can be cleaned with a damp lint-free cloth. Make sure not to let any water enter the entry and exit ports.
- The backside of the mirrors which is visible through the entry and exit ports can only be cleaned with clean compressed air. Never touch the backside of the mirrors with some physical object.
Troubleshooting and Repair
Below are a few checks to help in troubleshooting problems that may arise. Please contact your local Thorlabs Technical Support office with any questions.
Problem
|
Suggested Solutions
---|---
There is no transmission signal on the oscilloscope.
| Verify that the FPQFA is connected to the driving electronics.
Verify that the driving electronics are on and generating the right output.
Verify that the transmitted or reflected beam is not obstructed and hits the
detector properly.
Slightly move the beam in the horizontal and vertical direction with the first
steering mirror until a signal as in Figure 10 a is achieved.
Increase the voltage scan range to make sure the frequency scan is at least 1
– 2 FSRs, see FPQFA Specifications Table in Section 1.4.1.
Verify that the cable is still soldered to the piezoelectrical actuator by
applying a 100-Hz 10-V sawtooth wave and listening for a “buzzing” sound.
Disposal
Thorlabs verifies our compliance with the WEEE (Waste Electrical and Electronic Equipment) directive of the European Community and the corresponding national laws. Accordingly, all end users in the EC may return “end of life” Annex I category electrical and electronic equipment sold after August 13, 2005, to Thorlabs, without incurring disposal charges. Eligible units are marked with the crossed-out “wheelie bin” logo (see right), were sold to and are currently owned by a company or institute within the EC and are not dissembled or contaminated. Contact Thorlabs for more information. Waste treatment is your responsibility. “End of life” units must be returned to Thorlabs or handed to a company specializing in waste recovery. Do not dispose of the unit in a litter bin or at a public waste disposal site. It is the user’s responsibility to delete all private data stored on the device before disposal.
Thorlabs Worldwide Contacts
For technical support or sales inquiries, please visit us at www.thorlabs.com/contact for our most up- to-date contact information.
Corporate Headquarters
- Thorlabs, Inc. 43 Sparta Ave Newton, New Jersey 07860 United States
- sales@thorlabs.com
- techsupport@thorlabs.com.
EU Importer
- Thorlabs GmbH Münchner Weg 1 D-85232 Bergkirchen Germany
- sales.de@thorlabs.com
- europe@thorlabs.com.
Product Manufacturer
- Thorlabs Sweden AB Bergfotsgatan 10 431 35 Mölndal Sweden
- scandinavia@thorlabs.com
- techsupport.se@thorlabs.com.
UK Importer
- Thorlabs Ltd. 204 Lancaster Way Business Park Ely CB6 3NX United Kingdom
- sales.uk@thorlabs.com
- techsupport.uk@thorlabs.com.
References
- Thorlabs, Inc. - Your Source for Fiber Optics, Laser Diodes, Optical Instrumentation and Polarization Measurement & Control
- Thorlabs, Inc. - Your Source for Fiber Optics, Laser Diodes, Optical Instrumentation and Polarization Measurement & Control
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>