REAL system y-Series Rehabilitation Tool User Manual
- July 3, 2024
- REAL system
Table of Contents
- REAL system y-Series Rehabilitation Tool
- Product Information
- Product Usage Instructions
- FAQ
- DEVICE DESCRIPTON
- INDICATION FOR USE
- CONTRAINDICATIONS
- POTENTIAL ADVERSE EFFECTS/EVENTS
- OPERATOR PROFILE
- OPERATING PROCEDURE
- TECHNICAL SPECIFICATIONS
- SYMBOL GLOSSARY
- TECHNICAL INFORMATION
- Immunity to RF Wireless Communications Equipment
- FCC NOTE
- SPECIFICATION
- ABOUT COMPANY
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
REAL system y-Series Rehabilitation Tool
Product Information
Specifications
- Device Name: REAL y-Series
- Platform: Digital hardware and software medical device platform
- Technology: Virtual reality and full body tracking sensor
- Focus: Healthcare, physical, neurorehabilitation, and wellness needs
Product Usage Instructions
Components
The REAL y-Series consists of the following components:
- Headset (HMD)
- HMD Controller
- Large Sensor and Small Sensors (WTM and WSMs)
- Sensor Charger
- Router and Router Battery
- Sensor Bands
Headset (HMD)
The headset provides visual feedback of virtual reality experiences. To
use:
- Plug the headset power cord into the device to charge.
- Ensure the y-Series power cord is plugged into a wall outlet.
- Press the power button on top of the headset to turn it on/off.
- Adjust volume using the volume buttons located on the left side of the headset.
HMD Controller
The controller is for troubleshooting or administrative tasks. It is not
used during patient therapy. Use it to access settings or control volume.
Operating Configurations
The REAL y-Series is available in two configurations:
-
Online Configuration (Catalog Code: RLY1):
Operated online by connecting to a wireless network for assessments, therapies, and session history access. -
Offline Configuration (RLY1S and RLY1V):
Operated offline without internet for assessments and therapies. Session history access is not available.
FAQ
-
Can the headset be used wirelessly?
The headset needs to be connected to power for charging but can be used wirelessly once charged. -
How do I know if the sensors are fully charged?
The sensors are fully charged when they are placed back into the sensor charger, and the charging indicator shows a complete charge.
WARNING! CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE. OBSERVE ALL WARNINGS AND PRECAUTIONS NOTED THROUGHOUT THESE INSTRUCTIONS. FAILURE TO DO SO MAY RESULT IN COMPLICATIONS.
DEVICE DESCRIPTON
-
The REAL y-Series is a digital hardware and software medical device platform utilizing virtual reality and full body tracking sensor technology designed for use in healthcare and focusing on physical, neurorehabilitation, and/or wellness needs.
-
REAL y-Series consists of the following components:
- All-In-One Headset (HMD) with Software Experiences
- HMD Controller
- Large Sensor
- Small Sensors
- Sensor Charger (charging station)
- Tablet
- Router
- Router Battery
- REAL® Sensor Bands
-
Frequently used features and functions:
-
Headset (HMD)
-
The headset provides visual feedback of virtual reality experiences in concert with the REAL y-Series tablet and large and small sensors.
-
Plug headset power cord into headset to charge device, ensure y-Series power cord is plugged into wall outlet.
-
Press the power button to turn on/off headset. The power button is on top of the headset.
-
Press the volume buttons to adjust volume. The volume buttons are on the headset’s left side.
-
HMD Controller
-
Headset controller is to only be used for troubleshooting or administrative tasks. It is not used during patient therapy.
-
The controller is used to access headset settings or control volume.
-
Large Sensor and Small Sensors
-
Large and small sensors (WTM and WSMs) are equipped with mechanical and electrical components that measure motion and direction in physical space and then translate that information into a virtual environment.
-
The sensors are removed and placed back into the sensor charger (charging station) to activate or charge the devices.
-
The sensors are placed into the sensor bands that the user wears.
-
Sensor Charger
The sensor charger charges the sensors. Ensure the REAL y-Series power cord is plugged into a wall outlet. -
Tablet
-
The tablet is fitted with a touch screen that allows the user to:
-
Launch REAL TherapyView™ application on the tablet
-
Log in
-
Add or select patient (online configuration only)
-
Select a therapy experience, activity, or assessment
-
Initialize and sync to sensors
-
Select, start, modify, or end therapy session
-
Customize assessments and therapeutic and wellness activities with adjustable settings
-
View objective measures for a session
-
View session and measure history (online configuration only)
-
Log out
-
Plug tablet power cord into tablet to charge device, ensure y-Series power cord is plugged into wall outlet.
-
Power on/off the Tablet
Press the power button to power on/off the tablet. The power button is on the edge of the device. -
Router and Router Battery
-
Router allows the REAL y-Series components to communicate with each other.
-
Router battery provides power to the router when system is not connected to wall outlet.
-
Online Configuration Only: Router connects the REAL y-Series to internet.
-
Sensor Bands
Place or remove sensor bands on or from patient for upper or full body tracking.
-
Operating Configurations
REAL y-Series are available in two configurations:
-
Online Configuration (Catalog Code: RLY1):
The REAL y-Series online configuration is operated online by connecting the system to a wireless network to utilize assessments, therapeutic and wellness activities, and access session history. Follow the instructions in Section 2 to connect the REAL y-Series to a wireless network. Online mode allows for patient history from session to session. -
Offline Configuration (RLY1S and RLY1V):
The REAL y-Series offline configuration is operated offline without the need to connect to a wireless network to utilize assessments and therapeutic and wellness activities. Access to session history is not available in this configuration. -
At full charge, the entire system is designed to operate for a minimum of 60 minutes. For optimal connectivity between system components, it is recommended that you keep your system plugged in and all components charging between sessions.
-
In the event of electromagnetic and/or RF disturbances, the performance of the REAL y-Series may be affected.
-
The REAL y-Series is a Type BF Applied Part.
-
There is no preventive inspection, calibration, or maintenance necessary for the REAL y-Series besides the initial set up procedure. During the three-year product lifespan of the REAL y-Series, the device will continue to perform safely without any routine maintenance. No parts within the REAL y-Series will require inspection nor maintenance by a service personnel to ensure basic safety during the three- year product lifespan. Circuit diagrams and calibration instructions are not provided because service or parts repair is not necessary.
-
Contact your local Penumbra representative or Penumbra Product Technical Support if the system no longer stays powered on and connected through the entire recommended duration of a therapy session.
-
REAL sensor bands are recommended for one hundred uses or until the sensor bands no longer hold the sensors or its orientation.
-
Supply mains are electrically isolated in medical equipment to maintain basic safety.
-
The full expected latency of the device, including movement detection, processing, and visual representation is 35 milliseconds or less. This value is considered minimal and sufficiently low enough so that movement can be quickly detected.
INDICATION FOR USE
The REAL y-Series is an immersive virtual reality and display system that interactively displays and tracks rehabilitation exercises using a combination of virtual environments and full presence tracked avatars for visual feedback.
CONTRAINDICATIONS
There are no known contraindications.
WARNINGS
- If a patient complains of motion sickness, dizziness, headache, eye strain, or fatigue when using the device, stop use of device immediately.
- Use caution when using this device if a patient has a history of vestibular issues or motion sickness.
PRECAUTIONS
- Ensure a safe environment for the patient while performing activities with the device (i.e. remove any surrounding obstacles and ensure that the patient is unlikely to trip or fall). Ensure a medical professional is with the patient at all times to prevent any injury.
- Be aware of the patient’s limitations in range of motion and avoid device or program use that could lead to excessive gestures that could injure a patient.
- Incorrect placement of the sensors on the patient may result in the avatar appearing incorrectly or distorted on the headset and tablet.
- Damage (mechanical and electrical) may result if the tablet, headset, sensors, and/or sensor charger are dropped or struck against another object. Device is not intended for continued use if dropped from higher than 1 meter.
- Surface temperature around the headset exhaust may reach 46 °C if operating above nominal room temperature.
- Sensors will transmit inaccurate position data if used near metal including, but not limited to, wheelchairs, walkers, utility carts, smartwatches and mobile devices.
- Headset tracking can be lost or compromised if there is irregular or dim lighting in the room, the front facing camera sensors on the headset are obstructed or facing highly reflective surfaces, there is excessive motion in front of the patient, or a lack of distinct visual features.
- Third conductor of the AC cord is a functional earth. To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
- At no time should liquid products be allowed near any device component.
- No modification of this equipment is allowed.
- Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
- Portable RF communications equipment (including peripherals such as antenna cables, external antennas, smart watches, and mobiles devices) should be used no closer than 30 cm (12 inches) to any part of REAL y-Series, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
- Use of REAL y-Series adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
- Accessories such as power adapters and cords should not be replaced by the end user and should only be replaced by Penumbra. Any changes or replacements of accessories will likely impact compliance of REAL y-Series.
- Use of system with online configuration should be in a secure information technology environment. Https communication channels must be open.
POTENTIAL ADVERSE EFFECTS/EVENTS
Visual stimulation through head-mounted displays has a small possibility of provoking an epileptic seizure. Should this occur, stop using the device immediately. Other possible complications include, but are not limited to, the following:
- claustrophobia
- discomfort or pain in the head or eyes
- disorientation/vertigo/dizziness
- drowsiness
- eye strain
- falls or fractures
- headache/migraine
- insomnia
- light-headedness
- motion sickness
- nausea
- pain
- seizure
- repetitive strain injury
- vision problems
- skin irritation
Should any of the above occur, stop using the device immediately.
OPERATOR PROFILE
Operators of the REAL y-Series should be trained in rehabilitation therapy.
Follow facility guidelines for use and access to account login credentials (if
applicable). For systems with online configuration, the same account login
credentials shall not be used by more than one REAL y-Series at any given
time.
Note: These rehabilitation exercises and assessments are intended to be
conducted in a clinical environment and prescribed and supervised by a medical
professional trained in rehabilitation therapy. Rehabilitation therapy and
technique decisions will vary based on the clinical judgement of the treating
medical professional. A medical professional must be present at all times to
provide direct supervision throughout the course of therapy.
OPERATING PROCEDURE
GETTING STARTED AND CHARGING COMPONENTS
-
Remove REAL y-Series case from the shipping container and inspect for damaged or missing components.
-
Connect the REAL y-Series case to its power cord (AC adapter power cord). Plug the power cord into a grounded electrical outlet, making sure that it is the same voltage as indicated on the unit nameplate. Ensure the power receptacle is connected to a supply mains with protective earth. Ensure the case lid is left open when charging components.
-
Ensure sensors with white sides facing forward are in their corresponding size slots on the sensor charger. LED lights on sensors will show the following:
-
Ensure the headset is connected to its power cord (USB-C cord). LED light on top of headset will show blinking red, blinking green or solid green to indicate charging status.
-
Ensure the tablet is connected to its power cord (USB-C cord).
INTERNET CONNECTIVITY (Online Configuration Only)
Note: Prior to first time use, the REAL y-Series must be configured and
connected to the local internet. To continue using the REAL y-Series over its
lifespan, user may be prompted/required to complete over-the-air software
updates.
See Get Connected Guide section in the REAL y-series Quick Start Guide for
more detailed instructions. Continue to Section 3 when the system has secure
internet connectivity.
Connecting to Wi-Fi
For a connection to Wi-Fi, it is recommended that steps provided below are
performed by the facility’s IT team. For remote technical support please
contact the Penumbra Product Technical Support team 1.855.732.5797 (select
Option 2) or realsupport@penumbrainc.com from Mon–Fri, 6 am – 6 pm PST.
- Turn on tablet by pressing and holding down power button for approximately 5 seconds.
- Launch the TherapyView application if it is not already open.
- Press “OK” on the REAL Service prompt.
- On the login page, click on the “Network Setup” button in the bottom right corner.
- Connect using the desired wireless network name and password.
- Press the button at the top left of the screen to return to the TherapyView application home screen.
SOFTWARE UPDATES
To receive over-the-air (OTA) software updates, REAL y-Series must be
connected to a Wi-Fi network.
- Online Configuration
1. Turn on the tablet and headset (HMD).
2. Ensure the tablet is connected to Wi-Fi*.
3. If required, follow prompts on the tablet.
Note: To connect your REAL y-Series online system to the Internet, refer to “Internet Connectivity” section.
- Offline Configuration
Contact your local Penumbra representative or Penumbra Product Technical Support team for software updates.
START-UP SYSTEM FOR PATIENT USE
-
Unplug power cords from tablet and headset when ready to use and sufficiently charged.
-
If the tablet is not turned on, turn on.
-
Turn on the headset by pressing and holding the power button for approximately 5 seconds.
-
Place the headset on patient’s head in the sequence numbered below. Size of headset strap can be adjusted for fit. Top of head pad may be removed temporarily for better fit. Patient can immediately begin visually interacting with the environment.
-
Note: Patient can keep eyeglasses on.
-
Note: To improve focus, you can use the interpupillary distance (IPD) slider on the bottom of the headset. Refer to the image below.
-
-
Confirm audio can be heard but does not block out the patient’s ability to communicate; adjust volume on headset as needed.
-
Remove sensor bands from reusable packaging (sold separately) and note the catalog number on the product label. Each patient should have their own sensor bands.
-
Remove all sensors from sensor charger.
-
REAL y-Series has two sensor tracking options, upper and full body (default).
- Sensor Band Catalog Code BND supports only upper body tracking.
- Sensor Band Catalog Code BND1 supports both upper and full body tracking.
-
Place small sensors onto bands by sliding them into elasticized loops. For the hand sensor bands, tighten elasticized loop using the buckle. Place large sensor into pocket of shoulder band. Refer to catalog number for correct orientation.
-
Once sensors are placed in the bands, put each band onto its corresponding body part. See images below. Refer to catalog number for correct placement. Connect hook and loop fasteners of shoulder band if desired. Adjust bands for comfortable fit, if necessary. Ensure the waist band is worn at the sacrum of the back and the sensor is firmly secured and is centered on the back.
Note: Ensure the elbow sensor is placed right above the elbow and opposite the bicep.
Note: Ensure the knee sensor band is placed right below the knee cap. Wrap the extra sensor band over the sensor to secure the sensor firmly.
-
Launch the TherapyView application if it is not already open.
-
Online Configurations
-
Select the “Sign In” button and enter your TherapyView credentials.
-
Add a new patient or select existing patient from patient list.
-
Enter or edit patient information, avatar settings, and sensor tracking to match your patient as needed.
Note: Patient information and avatar settings will be saved to the patient’s TherapyView profile for subsequent sessions. Session history will also be available for therapist review. -
Offline Configurations
-
Select “Start Now” button.
-
Enter patient information, avatar settings and sensor tracking to match your patient as needed.
-
Select “Start Session.” Note: Patient information and avatar settings are not saved.
-
Once a patient is selected or created, initiate the session by pressing “Start Session.”
MANAGING A THERAPEUTIC OR WELLNESS SESSION
-
On the TherapyView selection screen, choose a therapy, experience, activity or assessment by selecting one of the following options available at the top of the screen.
-
Body Target: Filter activities by area of focus, motor goals, and/or cognitive goals
-
Choose desired areas of focus by tapping the parts of the avatar’s body, ‘Motor Goals’, and/or ‘Cognitive Goals’ options.
-
Choose the desired activity from the list by selecting the activity card or the “i” icon and select “Run.”
-
Exercises: Filter by exercise to find the desired activity
-
Choose your preferred exercise by tapping on the exercise to the left of the screen.
-
Choose the desired activity from the list by selecting the activity card or the “i” icon and pressing “Run.”
-
ADLs/IADLs: Filter activity by ADLs/IADLs
Choose the desired activity from the list by selecting the “i” icon and selecting “Run.” -
Assessment: View all available assessments
Choose the desired assessment from the list by selecting the assessment card or the “i” icon and select “Run.” -
Experiences: View all available REAL y-Series experiences
Choose the desired experience by selecting the experience card or the “i” icon and select “Run.”
-
-
Have the patient sit in a neutral position away from metal components, facing forward with hands on thighs in an upright position and knees in a right-angled position to the floor. Press the button on the tablet screen to map the sensors.
-
Confirm patient’s avatar in virtual environment corresponds to patient’s physical movement. If the avatar looks correct, press “Continue”. If not, remap the sensors by selecting “Remap Sensors.”
-
Press “Start” to begin an activity. You may also use the TherapyView selection screen to select a new activity, experience, or assessment. (See Section 6 for more details about therapeutic and wellness experiences).
-
Always remain with the patient throughout therapy session to provide direct supervision.
-
Monitor patient view on tablet; select and begin new experiences or activities as desired.
Note: After 20 minutes of inactivity in TherapyView, the user will be automatically logged out of the TherapyView application. To start a new session, exit out of TherapyView completely and return to step 12 in Section -
Note: If sensors lose synchronization or headset loses tracking, reset the entire system. Headset should be turned off and on using the power button. TherapyView application should be relaunched on the tablet. Sensors should be reset by placing them in the powered charging station (while the REAL y-Series is plugged in) and then removing them.
-
When planned therapy session is complete, press “End” on tablet.
-
Assess patient progress by reviewing objective measures from activities and/or assessments. Activity measures including such as utilization data, sit-to-stand transfers, full body measures, balance, assessment measures, lower body isotonic exercise measures, and functional range of motion.
Online Configuration Only: Patient information and avatar settings will be saved to the patient’s profile for subsequent sessions. Previous patient sessions are available for review in addition to the most recent session.
SYSTEM REMOVAL
- Remove sensors and bands.
- Remove headset from patient’s head.
- Power off headset by pressing and holding onto the power button for approximately 5 seconds.
- Log out of the TherapyView application.
- All system components (or patient contacting components) can be wiped down with an institutional approved sanitizing wipe* as needed between sessions.
- Ensure REAL y-Series Case is plugged in.
- Connect tablet and headset to their respective power cords. Red or green LEDs should be noted on the headset.
- Return all sensors to sensor charger. Red or green LEDs should be noted on sensors when placed correctly in the charging station.
- Sufficiently charge all components after each use and prior to next use for a minimum of 60 minutes.
Note: Standard operation is to power cycle and relaunch the system between therapy sessions to re-establish system connections.
- Do not use petroleum-based compounds, acids, caustics, or chlorinated solvents to clean or lubricate any parts. Use water-based solvents for cleaning. Handle the lenses on the headset carefully to avoid smears or scratches. Wipe lenses with a soft cloth for cleaning. Ultraviolet C (UVC) cleaners may be used to clean plastic components of the headset and sensors. Ensure headset lenses are covered if using a UVC cleaner. Please check with the UVC cleaner manufacturer for effectiveness and possible damage to the sensors’ plastic. Damage to the plastic components due to UV light exposure is not covered by warranty.
REAL y-Series Travel Case Instructions:
REAL y-Series is housed in a travel case that may be locked with the attached
combination lock that secures the zipper. The combination lock should be
turned to the red dot position at all times. If the combination lock is turned
away from the red dot, turn it to the red dot position.
- How to reset the combination lock:
1. Set all combination dials to the following: (0-0-0 default).
2. Find the hole located to the right of the dials. Use a paper clip or similar tool to press down on the reset button until an audible “click” is heard.
3. Set personal combination by turning the dials to display the desired set of numbers, e.g. 2-8-7.
4. Push the slide button located on the left of the dials towards the direction of the arrow and the reset button will push back up. An audible “click” will be heard.
5. Remember the personal combination. If reset is necessary, set the dials to the previous set combination and repeat steps 2 to 4 above.
- How to use the combination lock:
1. To unlock: Turn the dials to the correct combination. Push the slide button on the left of the dial towards the direction of the arrow to unlock.
2. To lock: Put the loop portion of the zipper into the slots of the lock, then turn the dials randomly to conceal the personal combination to lock.
SOFTWARE DESCRIPTION
REAL y-Series contains a variety of activities that incorporate clinically
recognized, existing therapeutic, functional, and wellness exercises to
facilitate motor learning, cognitive function, memory, and relaxation.
Activities are grouped into experiences that offer a similar environment,
typically for one or more activities. REAL y-Series also offers in-VR
assessments that reflect clinically recognized assessments. Settings for
activities involve parameters such as turning on and off avatar features and
environmental factors. While using the REAL y-Series, the clinician remains
responsible for the patient’s safety and the appropriateness of individual
exercises, including range of motion (ROM) attempts and any other limb or
joint limitations unique to that patient.
-
REAL Home with Assessments
-
Functional Reach™
Assess a patient’s functional reach in a standing position. Inspired by standard functional reach assessment.- Instructions:
- Setup: Ask the patient to stand with their back straight and feet hip-distance apart. Place feet at the position indicated in blue in the headset.
- Calibration: Ask the patient to extend either the right or left arm at the position indicated by a shadow arm and hold for 5 seconds.
- Assessment: A ruler and butterfly/bubble will appear. Ask your patient to reach for the butterfly/bubble by stretching their arm as far as they can safely without taking a step.
- Measures: A reach measure will automatically be detected when the patient retracts their arm. Press “Stop” on the TherapyView screen to accept the reach measurements. You can also reject the result by tapping “Tap to Reject” option on TherapyView.
-
Static Stance™
Assess a patient’s standing balance capabilities in various standing stances.- Instructions:
- Setup: Choose desired stance when prompted on TherapyView
- Assessment: Ask the patient to stand and follow the different stances shown by a prompt in the headset. Place feet at the position indicated in the headset. Press “Start” to begin the assessment. Hold stance for a desired amount of time. Once complete, press “Stop” on the tablet to register the measurements.
- Measures: Hold time and sway-related measures will be generated after pressing stop.
-
Happy Valley™
Happy Valley is designed to support physical rehabilitation, cognitive function, and wellness immersing the patient in a delightful world of penguins. Happy Valley includes a total of 18 therapeutic and wellness activities.
Activity | Catch & Glow™ | Hide & Seek™ | Mad Tavern™ |
---|---|---|---|
Goal | Exercise visual scanning, gaze stability and cervical ROM. | ||
How to | Use the blue line | Look for the penguin | Use your neck |
perform the | pointer to help a | with the propeller | movement to hover |
activity | penguin catch |
fireflies in a specific
| hat on using the blue
line pointer.
| the blue line pointer
over each character
| pattern.| | to wake them up.
- **Hot Air Balloon Activities
Activity| Bumper Band™| Balloon Pilot™| Summit
Rescue™
---|---|---|---
Goal| Focus on trunk mobility and core control.
How to perform the activity**| Move your torso in all directions to bump
the small balloons.| Move torso in all directions to move the balloon to hit
the treesandknock off the apples to feed the pigs.| Move torso forward and
backward to move the balloon to rescue the characters.
- **Sunrise Activities
Activity** | Ice Cave™ | Sunrise™ | Harvest™ |
---|---|---|---|
Goal | Focus on trunk mobility and core control. | ||
How to | Melt the ice by lifting | Lift the sun by | Grow vegetables by |
perform the | and setting the sun. | raising your arms or | lifting and |
setting
activity| Lift the sun by| lower legs. Set it by| the sun. Lift the sun
| raising your arms or| lowering your arms| by raising your arms
| lower legs. Set it by| or lower legs to| or lower legs. Set it
| lowering your arms| starting position.| by lowering your
| or lower legs to| | arms or lower legs
| starting position.| | to starting position.
- **Bird Forest Activities
Activity** | Free Birds™ | Nest Hop™ | Bird Match™ |
---|---|---|---|
Goal | Work on dynamic reaching, functional reach, and wrist ROM. | ||
How to perform the activity | Use either hand to pick up and place | ||
individual birds into nests with unique objects. | Use either hand to pick up | ||
and place a bird into a series of nests. | Pick up a brightly colored bird, |
locate the nest with a matching ribbon, and place the bird in nest.
- **Sports Park Activities
**
Activity| Chuckleball™| Chuckleball Arena™| Flying
Fish™
---|---|---|---
Goal| Focus on functional movement, dynamic reaching, reactive movement
and reflexes.
How to perform the activity| Use hands, head, or legs to block blue
fish.| Block blue fish with hands, head, or legs to defend your goal and hit
fish into the opposing goal.| Block blue fish with hands, head, or legs and
dodge red spiky fish.
- **Creative Canvas Activities
Activity** | Paint by Number™ | Free Stamp™ | Free Paint™ |
---|---|---|---|
Goal | Exercise visual scanning, gaze stability and cervical ROM. | ||
How to perform the activity | Use either hand or the blue line pointer to | ||
control a paint brush and match the number to the color. | Use either hand to |
stamp selected images on
virtual canvas backgrounds.
| Use either hand to freely express artistic creativity on virtual backgrounds.
Choose Blank Canvas or Coloring Book modes.
- **Island Antics
**
Island Antics experience is designed for patients who need to work on upper
body reconditioning and motor functions. It transports patients to a city of
antics where they get to be a celebrated hero tasked to help island citizens
using multi-directional movements to complete tasks.
Activity | Save the Seagull™ | Citizen Crossing™ |
---|---|---|
Goal | Use and work on trunk control, functional reach, and cross body | |
motions. | Engineered to improve trunk control, functional reach, and dynamic |
balance.
How to perform the activity| Grab the small UFOs like a frisbee, follow
the path in front of you, and release.| Use your hand to pick the citizen up
and guide it along the path connecting the two buildings.
Activity | Leaks & Levers™ | Coconut Chuck™ |
---|---|---|
Goal | Use and work on your shoulder flexion and extension. | Exercise |
functional reach and trunk movement.
How to perform the activity| Follow the path to move the lever, wheel,
or pulley using your hands.| Use your hands to reach out and pull the handle
in front of you to shoot the coconut on to the boats.
- **Mindful Market
**
Welcome to Mindful Market, where patients can emphasize high-level cognitive
skills, performing instrumental ADLs (Activities of Daily Living), and stimuli
tolerance, in an engaging, lively marketplace. Recover executive functions,
working memory, sequencing and sorting abilities, and resilience to
stimulation by volunteering at multiple small businesses in the charming
market square.
Activity | Sandwich Shop™ | Harvest Helper™ | Stamp Stand™ |
---|---|---|---|
Goal | Help build sequencing and working memory skills. | ||
How to | Choose the correct | Lift your handsand | Choose the correct |
perform the | ingredients to make | catch the object | quantity of stamp |
activity | the customer’s | thrown and place | sheets to match the |
desired sandwich. | the object on the matching pallet. | customer’s |
designated budget.
- **Music in Motion
**
Music in Motion is an entertaining music-themed VR rehabilitation experience
designed for upper & lower extremity and cognitive rehabilitation.
Activity| Song Safari™| Lean into the Music™| Reach
for the Rhythm™
---|---|---|---
Goal| Focus on neck and cervical movements.| Use cervical ROM, trunk and
core control.| Utilize shoulder ROM & trunk control.
How to perform the activity| Spot candy critters by holding the blue
line pointer on the critter hiding in the environment.| Lean from side to side
andmove your head up and down to steer the bird to hit the targets.| Use your
right or left hand to hit the blue or red targets respectively.
Activity | Turn with the Tempo™ | Step to the Beat™ |
---|---|---|
Goal | Engage wrist and shoulder ROM. | Exercise knee & hip ROM. |
How to perform the activity | Match the angle of the music note using | |
your right or left hand for the blue or red notes respectively. | Use your |
right and left feet to step on the blue and red targets respectively.
- **Pleasant Cove
**
Pleasant Cove focuses on the mental wellness component of therapy for patients
recovering cognitive skills while providing options for the therapist to
address rehabilitation goals through more directed, stress-relieving
experiences.
Activity | Bountiful Birdseed™ | Playful Percussion™ |
---|---|---|
Goal | Work on sustained attention and wrist ROM. | Focus on sequencing, |
psychomotor skills, and working memory.
How to perform the activity| Shake the seeds into the pink circle to
feed the bird. Extend your hand.| Using the mallets, follow the pink arrow and
hit each note to complete a song.
Activity | Green Thumb Gardening™ | ADL Cards™ |
---|---|---|
Goal | Use object recognition, shoulder, and wrist ROM. | Focus on |
sequencing, object identification, and memory.
How to perform the activity| Grow flowers by picking up the equipment
around you.| Practice ADLs by interacting with ADL-based cards. Choose from
different modes: Item Find, Sequencing, or Sorting.
- **REAL Pinball
** Aim for your highest scores in this fast-paced, pinball experience that supports range of motion, processing speeds, muscle tone & endurance, control of voluntary movement functions, neuromusculoskeletal skills, and psychomotor skills.
Activity| Carnival Blast™| Alien Arrival™| Magical
Garden™
---|---|---|---
Goal| Engineered to improve upper or lower body range of motion.
How to perform the activity| Use arms or legs to hit and block ball
using different control methods such as kayak paddle, pong, rackets, foot
flippers or gaze control.
- **Virtual Athletic Club
**
Virtual Athletic club is a high paced and engaging experience that motivates
patients to focus on upper body compound exercises and reactive movement.
Activities in this futuristic athletic club are designed to challenge and
improve upper extremity strength and endurance.
Activity | Bionic Bow™ | Paddle Pong™ | Power Punch™ |
---|---|---|---|
Goal | Focus on upper body ROM & strengthening in an archery activity. | ||
Work on upper body ROM and reactive movements. | Work on upper body endurance |
and strengthening in a futuristicboxing activity.
How to perform the activity| Use one hand to grasp the bow and hold it
upright. Use the other hand to hold the bowstring. Aim the bow at the target
in front of you, then release the bowstring to shoot the target.| Use one or
both hands to hit the balls at the launch pad. Whenthe ball is successfully
hit, additional ballsare added to the mix.| Move your arms to punch the
targets coming towards you. If a red or green target with arrows appear, twist
your torso in direction of the arrows and punch the target with the matching
glove color.
- **REAL Life
**
Practice activities of daily living (ADLs) in this supportive, virtual
environment. Clinicians can utilize REAL Life to rebuild patient confidence
and practice performing ADLs, such as bathing. Immersed in a calm and relaxing
environment, patients are encouraged to engage in activities that promote
independent living.
Activity | Shower Sanctuary™ |
---|---|
Goal | Designed to teach or practice bathing exercises. |
How to perform the activity | Turn on the shower, pick up the shower |
head. Once wet, turn off the shower and pick the soap from the right side of the tub and lather it over your body. Next, reach for the shampoo on the left side of the tub. Squeeze the bottle and rub the shampoo in your hair. Turn on the shower to rinse the soap. Pick up the towel to dry yourself.
- **Under the Sea™
**
Focus on lower body strength and endurance in a tranquil underwater
environment.
Activity | Undersea Garden™ | Coral Revival™ |
---|---|---|
Goal | Visually scan to interact with a peaceful underwater environment. |
Revitalize coral reef by completing a variety of exercises.
How to perform the activity| Gaze at different sea creatures and objects
to interact with them.| Aim the cannon at bleached coral while performing a
movement.
Complete the exercise by returning to the starting position to shoot and heal the targeted coral.
Activity | Treasure Hunt™ | Ruin Reveal™ |
---|---|---|
Goal | Direct a submarine and retrieve underwater treasures by performing | |
various exercises. | Discover an ancient underwater kingdom by working on |
diverse exercises.
How to perform the activity| Move the submarine scooper using head
movements.
Perform the exercise and hold to activate the scooper.
Complete the exercise by returning to the starting position to collect the treasure.
| Aim the cannon at the ancient walls while performing a movement. Complete the exercise by returning to the starting position to shatter the walls.
- **Float
**
Engage in intentional movements and mindfulness in a light-hearted
environment.
Activity | Warm- Up™ | Gifting™ | Meditation™ |
---|---|---|---|
Goal | Gently warm up by mirroring simple movements led by a glowing | ||
orb. | Gift Floaties by choosing different fashion accessories. | Follow a |
guided meditation session to practice mindfulness.
How to perform the activity| Follow the avatar/orb through warm-up
exercises using your arms and/or neck.| Move your head to select a Floatie,
and then use your hands to choose an accessory.| Find a comfortable
positionandfollow an instructor-led guided meditation session.
Activity | Breathing™ | Petting™ | Tai Chi™ |
---|---|---|---|
Goal | Follow a Floatie through guided breathwork training. | Use | |
functional reach to pet the Floatie. | Practice slow intentional movements with |
a Floatie.
How to perform the activity| Find a comfortable position and follow a
Floatie-led breathwork session.| Use your hands to pet a Floatie at
specificareas indicated on the screen.| Use your arms
to mirror the Floatie through slow intentional tai chi movements.
TECHNICAL SPECIFICATIONS
Sensor Accuracy* | ± 2 cm at a max distance of 75 cm |
---|---|
Sensor Precision* | 2 cm or less |
Latency | ≤35 milliseconds |
Operating Temperature | 15°C to 30°C (59°F to 86°F) |
Operating Pressure | 102 kPa or less |
Operating Relative Humidity | 30% to 90% |
Operating Elevation | 2,500 meters or less |
Radio Module | Output power (EIRP*): 6.31 mW (8 dBm) max |
Frequency Band: ISM (Industrial, Scientific, and Medical) Typical Center frequency: 2.44 GHz
Channel: 77 channels Bandwidth: 2 MHz per channel
Modulation: GFSK (Gaussian frequency-shift keying) Data flow: Bi-directional
*EIRP = Equivalent Isotropically Radiated Power
*REAL y-Series is calibrated appropriately to detect movement in virtual reality space in relation to real space accurately and precisely. Sensors will compute and display position at an accuracy of a 2 cm radius with respect to real space at a max distance of 75 cm relative to the headset. Sensors will also reproducibly compute position at a maximum deviation of a 2 cm radius for repeated movements at a max distance of 75 cm relative to the headset. Please note that accuracy and precision specifications contain limitations and are dependent on certain factors such as the amount of metal near the system. For example, if the patient is in a metal wheelchair and cannot move to a non-metal chair, reduction in accuracy and precision may occur.
SYMBOL GLOSSARY
TECHNICAL INFORMATION
REAL y-Series is intended for use in the electromagnetic environment specified below.
The customer or the user of REAL y-Series should assure that it is used in such an environment.
Emissions Test| Compliance|
RF emissions CISPR 11| ****
Group 1
| REAL y-Series uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11| Class B| REAL y-Series is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage Fluctuations/ Flicker emissions| Complies
REAL y-Series is intended for use in the electromagnetic environment specified
below.
The customer or the user of REAL y-Series should assure that it is used in such an environment.
Immunity test| IEC 60601
test level
| Compliance level| Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ±8 kV contact
±15 kV air
| ±8 kV contact
±15 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4| ±2 kV for power supply lines
±1 kV for input/output lines
| ±2 kV for
power supply lines
±1 kV for input/output lines
| Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
| ±1 kV differential mode
±2 kV common mode
| ±1 kV differential mode
±2 kV common mode
| Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short interruptions and voltage variations on power supply input
lines IEC 61000-4-11| Voltage Dips 30% reduction, 25/30 periods
At 0°
| Voltage Dips 30% reduction,
25/30 periods At 0°
| Mains power quality should be that of a typical commercial or hospital environment. If the user of the EQUIPMENT requires continued operation during power mains interruptions,
it is recommended that REAL y-Series be powered
from an uninterruptible power supply or a battery.
Voltage Dips > 95% reduction, 0.5 period
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
| Voltage Dips > 95% reduction, 0.5 period
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
Voltage Dips > 95% reduction, 1 period At 0°| Voltage Dips > 95% reduction, 1
period At 0°
Voltage Interruptions > 95% reduction, 250/300 periods| Voltage Interruptions
95% reduction, 250/300 periods
(50/60 Hz)
magnetic field IEC 61000-4-8
| 30 A/m| 30 A/m| Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or hospital
environment.
REAL y-Series is intended for use in the electromagnetic environment specified
below.
The customer or the user of REAL y-Series should assure that it is used in such an environment.
Immunity test| IEC 60601
test level
| Compliance level| Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6| 3 Vrms
150 kHz to 80 MHz (6 Vrms in ISM radio
Bands within 150kHz – 80MHz)
| 3 Vrms| Portable and mobile RF communications equipment should be used no closer to any part of REAL y-Series, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz d = 2.3√P 800 MHz to 2.7 GHz
Radiated RF IEC 61000-4-3| 10 V/m
80 MHz to 2.7 GHz
| 10 V/m
| | | where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
| | | Field strengths from fixed RF transmitters, as determined by an
electromagnetic site surveya, should be less than the compliance level in each
frequency rangeb.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which REAL y-Series is used exceeds the applicable RF compliance level above, REAL y-Series should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating REAL y-Series.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and REAL y-Series
REAL y-Series is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of REAL
y-Series can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and REAL y-Series as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
| Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2√P
| 80 MHz to 800 MHz
d = 1.2√P
| 800 MHz to 2.7 GHz
d = 2.3√P
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Immunity to RF Wireless Communications Equipment
Test Frequency (MHz)| Band a)
(MHz)
| Service a)| Modulation b)| Maximum Power (W)| Distance
(m)
| IMMUNITY TEST LEVEL
(V/m)
---|---|---|---|---|---|---
385| 380–390| TETRA 400| Pulse modulation b) 18 Hz| 1.8| 0.3| 27
450| 430–470| GMRS 460,
FRS 460
| FM c)
±5 kHz deviation
1 kHz sine
| 2| 0.3| 28
710| 704–787| LTE Band 13, 17| Pulse modulation b) 217 Hz| 0.2| 0.3| 9
745
780
810| 800–960| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
| Pulse modulation b) 18 Hz| 2| 0.3| 28
870
930
1720| 1700–1990| GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3,
4, 25; UMTS
| Pulse modulation b) 217 Hz| 2| 0.3| 28
1845
1970
2450| 2400–2570| Bluetooth, WLAN,
802.11 b/g/n, RFID 2450, LTE Band 7
| Pulse modulation b) 217 Hz| 2| 0.3| 28
5240| 5100–5800| WLAN 802.11
a/n
| Pulse modulation b) 217 Hz| 0.2| 0.3| 9
5500
5785
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
FCC NOTE
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
This device contains license-exempt transmitter(s)/receiver(s) that comply
with Innovation, Science and Economic Development Canada’s license-exempt
RSS(s). Operation is subject to the following two conditions:
- This device may not cause interference.
- This device must accept any interference, including interference that may cause undesired operation of the device.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a residential setting. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
SPECIFICATION
-
Highest Clock Frequency:
- HMD: 32 MHz
- WSM: 32 MHz
- WTM: 24.576 MHz
-
Frequency Range:
- 2402 MHz
- 2479 MHz
-
Transmitting Frequency and Modulation:
Gaussian Frequency-shift Keying (GFSK) modulation. 2 Mbps modulation for all transmitter frequencies. -
Antenna Make, Model, and Gain:
** Device| Antenna Make| Antenna Model| Antenna Gain
---|---|---|---
WSM| **
Johanson
| ****
P/N 2450AT43B100E
| Peak Gain 1.3 dBi
Average Gain -0.5 dBi
WTM
HMD| Penumbra, Inc.| P/N 17107| Peak Gain 0.7 dBi
Average Gain -2.6 dBi
- Power Output and Data Rate:
Device| Power Output| Data Rate
---|---|---
WSM| Programmed by the firmware to +8 dBm.| GFSK modulation, 2 Mbps data rate.
WTM| Programmed by the firmware to +4 dBm.
HMD| Programmed by the firmware to +8 dBm.
ABOUT COMPANY
- Manufacturer: Penumbra Inc.
- One Penumbra Place Alameda, CA 94502 USA
- Email: askreal@realsystem.com
- Tel: 1.855.REAL-SYS 1.855.732.5797