asecos UFA.025.030 Recirculating Air Filter Unit User Manual

June 3, 2024
asecos

EP.V.34660.01 | 05/2022

asecos logo b1

USER MANUAL

Recirculating air filter unit
UFA.025.030(.XL)

asecos UFA.025.030 Recirculating Air Filter Unit

1. NOTES • GUIDELINES • GUARANTEE
1.1. GENERAL SAFETY NOTES
  • Observe applicable statutes and regulations, and the notes in these operating instructions, when handling hazardous materials
  • Only use the recirculation filter attachment when it is in proper condition
  • Please follow the instructions of the safety experts, technical supervisory services (BG, GAA, pharmacy equipment) of the construction supervision, etc.
  • Observe accident prevention regulations and workplace ordinance
  • Check the recirculation filter attachment regularly for proper function (exhaust air volume flow and filter saturation by checking the indicators)
1.2. GUARANTEE

The guarantee for this product is agreed between you (the customer) and your dealer (the seller). As the manufacturer, asecos guarantees the products listed in the operating instructions for a period of 24 months from the date of delivery. All model safety equipment are subject to a compulsory annual inspection by specialised staff authorised by the manufacturer. Otherwise the customer’s guarantee claim against the manufacturer expires.

1.3. DEVELOPMENT AND MANUFACTURER

asecos GmbH, Weiherfeldsiedlung 16-18, 63584 Gründau, Germany
The type plate with the serial number is located on the outside of the unit’s housing

1.4. PURPOSE OF USE

The unit is used for technical ventilation from Ex zone 2 or no Ex zone of safety cabinets used for storing flammable liquids. The recirculation filter attachment is particularly low-noise and therefore ideal for use directly in the working environment.

Suitable for cabinet models:

UFA Model   Suitable cabinet groups
UFA.025.030  S-LINE / Q-LINE / V-LINE / UB-LINE / CS.195.105.F / E.195.105.F
UFA.025.030.XL  XL-LINE / S-CLASSIC-30

Set-up requirements

UFA.025.030 - Set-up requirements

UFA.025.030 - CAUTION

CAUTION:
The unit may only be installed in a non-hazardous area!
The integrated activated carbon filter removes 99.999% of the hydrocarbons from the exhaust air so that it can be discharged directly into the room air. The pollutant-containing exhaust air stream does not come into contact with electrical current-carrying components. The unit is therefore suitable for exhausting cabinets that are defined in EN 14470 Part 1 or DIN 12925 Part 1. An exhaust duct leading to the outside at a non-hazardous location can therefore be omitted. A CE certificate of conformity (also according to EC Directive 2014/34/EU – ATEX Directive) is available.

2. CONSTRUCTION AND FUNCTION
2.1. MATERIAL

UFA.025.030 - MATERIAL

INFO:
The housing is made of powder-coated fine sheet steel

2.2. ACCESSORIES
  • Activated carbon main filter incl. suspended matter pre-filter (order no. 40201), housing ABS/galvanised sheet steel, filling activated carbon (approx. 3500 gr).
  • Secondary filter (order no. 40202), ABS housing, ABEK activated carbon filling (140 g) incl. suspended matter prefilter
2.3. MODE OF OPERATION

UFA.025.030 - MODE OF OPERATION

  1. Fresh air
  2. ventilator
  3. activated carbon filter
  4. gases and vapours in cabinet

CAUTION:
Filter can only retain solvent vapours (hydrocarbons). Please consult us if other materials are to be stored.
Safe and faultless function is shown by the ventilation and filter monitoring mechanism which is fitted as standard. (3.2)

2.4. VENTILATION SETTING / MONITORING

The performance of the fan is sufficient to achieve an air exchange rate of at least 10 times for the cabinet models listed under point 1.3. The proper ventilation function is independently controlled by an electronic monitoring system. The air volumes are monitored in the exhaust air shaft by means of pressure differential measurement. An optical and acoustic alarm is triggered when the air exchange rate falls below the set value. The visual alarm only goes out when the setpoint is reached again, the acoustic alarm can be cancelled by pressing the “OK button” be interrupted.

2.5. POTENTIAL-FREE ALARM CONTACT

To use the potential-free alarm contact, the three-pole plug supplied must be used. The contact is designed for a maximum direct voltage of DC 30 V or an alternating voltage of AC 230 V. The maximum current carrying capacity is 10 A.
The plug is connected according to the following diagram:

Contact position
Normal operation: “L” and “1” are connected
Alarm or mains failure: “L” and “2” are connected

UFA.025.030 - POTENTIAL-FREE ALARM CONTACT

CAUTION:
The potential-free alarm contact must not be used to switch large loads. It is only intended to signal a malfunction of the recirculation filter attachment by means of a signal light, horn or similar.
The connection must be carried out by a qualified electrician!

3. ERECTION • COMMISSIONING
3.1. COMMISSIONING

UFA.025.030 - COMMISSIONING

ATTENTION:
Observe the instructions point 1.4 Intended use

3.2. SELF-TEST

UFA.025.030 - SELF-TEST

ATTENTION:
The unit is designed for continuous operation!

PLEASE NOTE:
As soon as the unit is connected to the mains, a self-test begins during which all safety-relevant components are checked. If there is a fault, the unit will not start. An error F7 is displayed alternating with the relevant faulty safety element F1-F4.

4. DISPLAY AND MENU
4.1. POWER INDICATOR

UFA.025.030 - POWER INDICATOR

  1. Display including error indication
  2. Control element with arrow keys and OK key

IN THE EVENT OF A MALFUNCTION:
Measures see 6. malfunctions

4.2. OVERVIEW OF THE MENU ICONS

Symbols in the active screen

UFA.025.030 - MENU ICONS 1       UFA.025.030 -
MENU ICONS 2

Volume flow OK     Filter saturation OK

Symbols in the main menu

UFA.025.030 - MENU ICONS 3       UFA.025.030 -
MENU ICONS 4     UFA.025.030 - MENU ICONS 5

Information           Settings submenu        Exit current display

Symbols in the Settings menu and sub-items

UFA.025.030 - MENU ICONS 6     UFA.025.030 - MENU
ICONS 7       UFA.025.030 - MENU ICONS 8

Set contrast           Carry out filter change   Set air monitoring parameters

UFA.025.030 - MENU ICONS 9      UFA.025.030 -
MENU ICONS 10

Replace sensor           Exit current display

4.3. SHOW SYSTEM INFORMATION

ATTENTION:
For a simplified service procedure, please have the system information ready.

UFA.025.030 - SHOW SYSTEM INFORMATION

NOTE
The system automatically switches to the normal display when there is no activity or no input.

4.4. SET DISPLAY CONTRAST

UFA.025.030 - SET DISPLAY CONTRAST

NOTE
The system automatically switches to the normal display when there is no activity or no input.

5. MALFUNCTIONS • ACTIONS
Malfunction Problem – Measure Note – be sure to observe!
**Pressure difference across the filter too low.
** (Fan defective or air outlet blocked). Please check the air outlet.

Please contact your specialist dealeror our service hotline 01805 – 92 20 92
| Pressure difference over the filter too high.
(filter too dirty, air flow rate too low).| Main filter change necessary! Please contact your specialist dealer or our service hotline 01805 – 92 20 92.
| The filter is almost completely saturated,
a timely change is necessary.| Attention: Possible accident inside the cabinet! Eliminate leakage professionally and change filter!
Please contact your specialist dealer or our service hotline 01805 – 92 20 92
| The filter is completely saturated and the fan has been switched off.
It is now imperative to change the filter at short notice.| Main filter and secondary filter change necessary!
Please contact your specialist dealer or our service hotline 01805 – 92 20 92
| The filter is close to its saturation limit,
a timely change is imperative.| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Filter runtime longer than 8040 operating hours (11 months)
The permissible operating time of the filter has almost expired. The filter must now be changed at short notice.| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Filter runtime longer than 8760 operating hours (12 months)
The permissible operating time of the filter has expired. The filter must now be changed at short notice.| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Filter sensor runtime longer than 48 months
A sensor replacement should be carried out in the course of the next safety inspection.| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Running time of the filter sensor longer than 54 months.
Sensor drift and thus safe operation can no longer be guaranteed. The sensor must be replaced.| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Differential pressure sensor error| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Error temperature/humidity sensor| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Filter sensor defective| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Fan defective (blocked)| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9
| Self test error
Display F7 alternating with the respective error F1 to F4| Please contact your responsible specialist dealer
or to our service hotline 01805 – 92 20 9

14 ct/called minute from the Deutsche Telekom AG fixed network, from mobile phones a maximum of 42 ct/minute.

6. REPAIRS

UFA.025.030 - ATTENTION ATTENTION
Repairs are only to be carried out by specialists specifically trained for this.
Given damage the appliance is to be repaired or replaced by the manufacturer.

7. FUNCTIONAL CHECK • MAINTENANCE • CARE

To ensure safety, the recirculation filter attachment should be serviced at least once a year and its function checked by a specialist. At the same time, the built-in filter unit must be replaced at least once a year, and immediately when the saturation limit is reached.
The functionality is checked electronically by the integrated ventilation and filter saturation monitoring, so that a function check of the recirculation filter attachment is carried out permanently during operation. The activated charcoal main filter of the recirculation filter attachment must be replaced with a new filter at short notice if the corresponding maintenance indicator (red screen) is shown. The filter can only be replaced by one of our trained service technicians and must be requested promptly on the service hotline. Filter replacement by asecos service personnel ensures that the contaminated activated carbon is fed into a defined disposal chain.

7.1. SENSORS AND MONITORING ELECTRONICS

The UFA contains sensitive sensors and complex monitoring electronics. The device must be designed as a “continuous 24/7” and was also designed for this purpose. The sensitive gas sensor system in particular is heavily stressed by continuous operation and, according to experience, can deliver unreliable measurement results after a certain period of operation (drift) after approx. 40,000 operating hours.
The UFA can then no longer be used safely (possible monitoring errors).
In this case, a visual warning is given. In this case, please contact the customer service for replacement of the sensor unit.

7.2. CLEANING

For external cleaning of the recirculating air filter attachment, please use common, mild household cleaners and a soft cloth. Disconnect your recirculating air filter attachment from the mains by unplugging the power cable before cleaning! Make sure that no moisture gets into the inside of the recirculating air filter attachment during cleaning.

CONTACT:
In the case of defects or complaints about our products (within and also after the warranty period), and for requesting safety checks or taking out a service contract, please contact our service hotline on: Tel: +44 1785 22 70-90 [email protected]

8. TECHNICAL DATA
    UFA.025.030   UFA.025.030.XL
External dimensions W x D x H  mm  308 x 555 x 256  308 x 555 x 256
Nominal voltage  85-265  85-265
Frequency  Hz  50/60  50/60
Weight  kg  17  19
Power consumption in operation  8,5  8,5
Maximum current consumption  0.13  0.13
Max. volume flow  m³/h  10  17
Min. volume flow  m³/h  10  17
Emission sound pressure level  dB  23.0  23.0
Potential-free alarm contact: maximum continous current  10  10
Potential-free alarm contact: maximun nominal voltage AC  230  230
Potential-free alarm contact: maximum nominal voltage DC  30  30
Potential-free alarm contact: single-phase motor load  185  185
Potential-free alarm contact: max. switching capacity (electromagnetic)  VA 
300  300
Potential-free alarm contact: max. switching capacity (ohmic)  VA  1500 

1500

  • including adapter

PLEASE NOTE
The data are approximate values and as such are not to be used in the sense of guaranteed properties. They correspond to the state of the art at the time of publication of the operating instructions. Further technical development expressly permits changes to these values without prior notice.

9. TECHNICAL DRAWING

[1] ventilator
[2] Sensor
[3] activated carbon filter
[4] Activated carbon secondary filter
[5] Potential-free alarm contact

UFA.025.030 - TECHNICAL DRAWING

a) Cleaned exhaust air
b) Fresh air

10. ATEX MARKING

UFA.025.030 - ATEX MARKING

[1] Marking of EX-protected units
[2] Equipment group according to RL 2014/34/EU
II Not underground operation
[3] Ignition protection types
Ex ic Intrinsic safety with the protection level “ic” – not capable of causing ignition in undisturbed operation
nA non-functioning
[4] Explosion groups according to EN 60079-0: 2019-09
IIB the group has a limiting gap width (MESG) 0.5-0.9 mm, suitable e.g. for use with ethylene
[5] Unit categories of the assemblies used
3/- Extraction from EX zone 2, installation exclusively in no EX zone
G Gas
[6] Temperature class
T4 max. surface temperature of the operating medium 135 °C
[7] Protection level EPL according to EN 60079-0: 2019-09
3/- G Gc Suitability for extraction from EX zone 2

asecos GmbH

Safety and Environmental Protection
Weiherfeldsiedlung 16-18
DE-63584 Gründau

  +49 6051 92200
  +49 6051 922010
  [email protected]

asecos bv

Veiligheid en milieubescherming
Tuinderij 15
NL-2451 GG Leimuiden

  +31 172 506476
  +31 172 506541
  [email protected]

asecos SARL

Sécurité et protection de l’environnement
1, rue Pierre Simon de Laplace
FR-57070 Metz

  +33 3 87 78 62 80
  +33 3 87 78 43 19
  [email protected]

asecos S.L.

Seguridad y Protección del
Medio Ambiente
CIM Vallès, C/ Calderí S/N
Oficinas 75 a 77
ES-08130 – Santa Perpètua de Mogoda
Barcelona

  +34 902 300385
  +34 902 300395
  [email protected]

asecos Ltd.

Safety and Environmental Protection
5 Eastgate Business Centre
Eastern Avenue
Burton on Trent, Staffordshire
DE13 0AT

  +44 7880 435436
  +49 6051 922010
  [email protected]

asecos

Safety and Environmental Protection Inc.
19109 West Catawba Avenue, Suite 200
Cornelius, NC 28031
USA

  +1 704 8973820
  +49 6051 922010
  [email protected]

asecos Schweiz AG

Sicherheit und Umweltschutz
Gewerbe Brunnmatt 5
CH-6264 Pfaffnau

  +41 62 754 04 57
  +41 62 754 04 58
  [email protected]

Read User Manual Online (PDF format)