Elitech Multi Use Temperature And Humidity Data Logger User Manual
Overview
RC-61/GSP-6 is a temperature and humidity data logger with two external probes that allow various probe combination methods. It features a large LCD Screen, audible-visual alarm, auto shortened interval for alarms and other functions; its built-in magnets are also easy for mounting during usages. It can be used to record the temperature/humidity of medicines, chemicals, and other goods during storage, transportation and in each phase of the cold chain including cooler bags, cooling cabinets, medicine cabinets, refrigerators and laboratories.
- LED Indicator
- LCD Screen
- Button
- USB Port
- Temperature-Humidity-Combined Probe (TH)
- Temperature Probe (T)
- Glycol Bottle Probe (optional)
Specifications
Model
|
RC-61/GSP-6
---|---
Temperature Measurement Range| -40″C~+BS”C (-40″F~18S”F)
Temperature Accuracy| TH Probe: ±0.3″C/±0.6″F (-20″C~+40″C), ±0.S”C/±0.9″F
(others)
T Probe: ±0.S”C/±0.9″F (-20″C-+40″C), ±1″C/±1.8″F (others)
Humidity Measurement Range| 0%RH-100%RH
Humidity Accuracy| ±3%RH (25″C, 20%RH-80%RH), ±5%RH (others)
Resolution| 0.1″C/”F; 0.1%RH
Memory| Maximum 16,000 points
Logging Interval| 10 seconds to 24 hours
Data Interface| USB
Start Mode| Press button; Use software
Stop Mode| Press button; Auto-stop; Use software
Software| ElitechLog, for mac □ S & Windows system
Report Format| PDF/EXCEL/TXT* via ElitechLog software
External Probe| Temperature-humidity combined probe, temperature probe;
glycol bottle probe (optional)**
Power| ER14505 battery/USB
Shelf Life| 2 years
Certification| EN12830, CE, RoHS
Dimensions| 118×61.Sx19 mm
Weight| 100g
*TXT for Windows ONLY. •• The glycol bottle contains 8ml propylene glycol.
Operation
1. Activate Logger
-
Open the battery cover, gently press the battery to hold it in position.
-
Pull out the battery insulator strip.
-
Then re-install the battery cover.
2. Install Probe
Please install the probes to the corresponding jacks ofT and H, details
are shown below:
3. Install Software
Please download and install the free ElitechLog software (macOS and Windows)
from Elitech US: www.elitechustore.com/pages/download
or Elitech UK: www.elitechonline.co.ul</software or Elitech BR:
www.elitechbrasil.com.br.
4. Configure Parameters
First, connect the data logger to the computer via USB cable, wait until the
!;l icon shows on the LCD, then configure via:
ElitechLog Software: If you do not need to change the default parameters (in
Appendix); please click Quick Reset under Summary menu to synchronize local
time before usage; – If you need to change the parameters, please click the
Parameter menu, enter your preferred values, and click the Save Parameter
button
to complete the configuration.
Warning! For first time user or o~er battery replacement:
To ovoid time or time zone errors, please make sure you click Quick Reset or
Save Parameter before usage to configure your loco/ time into the logger.
Note: The parameter of Interval Shortened is disabled by default. If you
set it to Enable. it will auto shorted the fogging interval to once per
minute if it exceeds temperature/humidity limit(s).
5. Start Logging
Press Button: Press and hold the ► button for S seconds until the icon shows
on the LCD, indicating the logger starts logging.
Note: If the ► icon keeps flashing, it means the logger configured with
start delay; it wi/1 start fogging ofter the set delay time elapses.
6. Stop Logging
Press Button: Press and hold the button for S seconds until the ■ icon shows
on the LCD, indicating the logger stops logging.
Auto Stop: When the logging points reach the maximum memory, the logger will
stop automatically.
Use Software: Connect the logger to your computer; open ElitechLog software,
click Summary menu and Stop Logging button.
Note: Default stop is via Press Button, if set as disabled, the button stop
function will be invalid; please open the EfitechLog software and click Stop
Logging button to stop it.
7. Download Data
Connect the data logger to your computer via USB cable, and wait until the !;;I icon shows on the LCD, then download data via: ElitechLog Software: The logger will auto-upload data to ElitechLog, then please click Export to select your desired file format to export. If data failed for auto-upload, please manually click Download and then repeat above operation.
8. Reuse the Logger
To reuse a logger, please stopped it first; then connect it to your computer
and use ElitechLog software to save or export the data.
Next, reconfigure the logger by repeating the operations in 4, Configure
Parameters*, After finished, follow 5. Start Logging to restart the logger for
new logging.
Status Indication
1. LCD Screen
- Battery Level
- topped
- Logging
- Circular Logging
- Over Limit Alarm
- Connected to PC
- Max./Min./MKT/Average Values
- High/Low Temperature Limit
- High/Low Temperature/Humidity Limit
- Current Time
- month-Day
- Logging Points
2. LCD Interface
_ Temperature (Humidity); Logging Points
Maximum, Current Time
Minimum, Current Date
Hight Alarm limit
Low Alarm Limit
Average
Probe Not Connected
_
3. Buttons-LCD-LED Indication
• To enable buzzer function, please open ElitechLog software and go to Parameter menu-> Buzzer-> Enable.
Battery Replacement
-
Open the battery cover, remove the old battery.
-
Install a new ER14505 battery into the battery compartment. Please note the negative cathode is installed to the spring end. l:I1
-
Close the battery cover.
What’s Included
- Data Logger x 1
- Temperature-Humidity-Combined Probe x 1
- ER14505 Battery x 1
- Temperature Probe x 1
- USB Cable x 1
- User Manual x1
- Calibration Certificate x1
Warning
Please store your logger at room temperature.
Please pull out the battery insulator strip in the battery compartment before
using.
If you use the logger for the first time, please use ElitechLog software to
synchronize system time and configure parameters.
Do not remove the battery if the logger is recording.
The LCD screen will be auto off after 15 seconds of inactivity (by default).
Press the button again to turn on the screen.
Any parameter configuration on Elitech Log software will delete oil logged
doto inside the logger. Please save doto before you apply any new
configurations.
To ensure the humidity occurocy. please avoid contact with unstable chemical
solvents or compounds. especially avoid long-term storage or exposure ta the
environments with high concentrations of ketene, acetone, ethanol,
isapropanai, toluene etc.
Do not use the Jagger far Jang-distance transport if the battery icon is less
than half as ~ .
The glycol filled battle probe can be considered as a thermal buffer that con
simulate the actual temperature variations inside, which is suitable far
vaccine, medical or similar scenarios.
Default Parameters
Model
|
RC-61
|
CSP-6
---|---|---
Logging Interval| 15 minutes| 15 minutes
Start Mode| Press Button| Press Button
Start Delay| 0| 0
Stop Mode| Use Software| Use Software
Repeat Start/Circular Logging| Disable| Disable
Time Zone| |
Temperature Unit| ·c| ·c
Low/High Temperature Limit| -30″[/6 □”[| -3 □ “[/60″[
Calibration Temperature| o·c| o·c
Low/High Humidity Limit| 10%RH/9 □ %RH| 1 □ %RH/90%RH
Calibration Humidity| □ %RH| □ %RH
Button Tone/Audible Alarm| Disable| Disable
Display Time| 15 seconds| 15 seconds
Sensor Type| Temp (Probe T) + Hurni (Probe H)| Temp (Probe T) + Hurni
(Probe H)