Elitech RC-51H Multi-use Temperature and Humidity User Manual

June 7, 2024
Elitech

Elitech RC-51H Multi-use Temperature

 **and Humidity User Manual**

Elitech RC-51H Multi-use Temperature and Humidity User
Manual

Product Overview

This temperature and humidity data logger is mainly used in the fields or places of medicine, food, life science, flowers , breeding industry, ice chest, container, shady cabinet, medical cabinet, refrigerator, laboratory and greenhouse, etc. RC-51 H is plug and play and It can directly generate the data report, with no need to install data management software. The data can still be read in case battery runs out.

Structure Description

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Product
Overview

  1. Transparent cap
  2. USB port
  3. LCD screen
  4. Seal ring
  5. Button & Bi-color indicator (red and green)
  6. Sensor
  7. Product label

LCD screen

Elitech RC-51H Multi-use Temperature and Humidity User Manual - LCD
Screen

  • A Battery indicator
  • B Mean kinetic temperature
  • C Start recording indicator
  • D Stop recording indicator
  • E Cyclic recording indicator
  • F Computer connection indicator
  • G Temperature unit(°CRF)
  • H Humidity unit or Progress percentage
  • I Timing indicator
  • J Average value indicator
  • K Number of records
  • L Combined indicator

Product label

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Product
label

  • a Model
  • b Firmware version
  • C Certification information
  • d Barcode
  • e Serial number

Picture is for reference only, please take real object as standard.

Technical Specifications

  • Recording Options: Multi-Use

  • Temperature Range: -30 to 70

  • humidity Range: 10%95%

  • Temperature & humidity Accuracy: ±0.5(-20/+40);±1.0(other range) ±3%RH (2520%90%RH)±5%RH(other range)

  • Data Storage Capacity 32,000 readings:

  • Software: PDF /Elitech Log Win or Mac (latest version)

  • Connection Interface: USB 2.0, A-Type

  • Shelf Life/Battery: 2 years1/ER14250 button cell

  • Recording Interval: 15 minutes(standard)

  • Startup Mode: Button or software

  • Stop Mode: Button, software or stop when full

  • Weight: 60g

  • Certifications: EN12830, CE, RoHS

  • Validation Certificate: Hardcopy

  • Report Generation: Automatic PDF report

  • Temperature & 0.1(Temperature)
    humidity Resolution 0.1%RHcHimidity

  • Password Protection: Optional on request

  • Reprogrammable: With free Elitech Win or MAC software

  • Alarm Configuration: Optional, up to 5 points, Humidity only supports upper and lower limit alarm

  • Dimensions: 131mmx24mmx7mm(LXD)

1.Depending on optimal storage conditions (±15 to +23/45% to 75% RH)

Parameter Instruction

Users can reconfigure the parameters by data management software per actual needs. The the original parameters and

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Parameter
Instruction

Operating Instructions

This data logger can be stopped by software. Users can stop the logger by clicking the stop button in the data management software.

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Operating
Instructions

View data
When the data logger is inserted into the USB port of the computer, the data report will be created automatically. The red and green indicators flash in turn when the document is being created, and the LCD screen shows the progress of PDF Report creation. The red and green indicators light at the same time immediately after the document is created, then users can view the data report . The document creation will last for no more than 4 minutes.

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Rotate the
transparent cap

(1) Rotate the transparent cap in the direction of the arrow and remove it.

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Insert the
data logger into the computer

(2) Insert the data logger into the computer and view the data report.

Description of the indicator flashing status

Elitech RC-51H Multi-use Temperature and Humidity User Manual -
Description of the indicator flashing status
Description of the menus
Elitech RC-51H Multi-use Temperature and Humidity User Manual - Description
of the menus
Description of the combined indicators and other status
Elitech RC-51H Multi-use Temperature and Humidity User Manual - Description
of the combined indicators and other status

Note:

  1. Menu 1 appears only when the corresponding function is selected.
  2. ” ►” should be in a slate of blinking .•
  3. The display in the combined indicator area. The same as below.

Replace battery

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Replace
battery

Report

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Report 1st
page

The First Page

Elitech RC-51H Multi-use Temperature and Humidity User Manual - Report other
page

Other Page

  1. Basic information
  2. Description of the usage
  3. Configuration information
  4. Alarm threshold and related statistics
  5. Statistical information
  6. Temperature and humidity graph
  7. Temperature and humidity data details
  8. A Document creation time (record stop lime)
  9. B Alarm (Alarm status as shown in the figure above)
  10. C Stop mode that has been set.
  11. D Alarm status of the temperature alarm zone
  12. E Total times of exceeding the temperature alarm threshold
  13. F Total time of exceeding the temperature alarm threshold
  14. G Alarm delay and alarm type
  15. H Alarm threshold and temperature alarm zones
  16. I Actual stop mode (different from the item C)
  17. J Vertical coordinate unit of the data graph
  18. K Alarm threshold line (corresponding to the item L)
  19. L Alarm threshold
  20. M Record data curve (black indicates temperature, deep green indicates humidity)
  21. N Document name (serial number & description of usage ID)
  22. O Record time range in the current page
  23. P Records when date changes (date & temperature and humidity)
  24. Q Records when the date is not changed (time & temperature and humidity)

Attention: The data above is only used as explanation of the report. Please refer to the actual document for specific configuration and information

What’s included

1 temperature and humidity data logger
1 Er14250 battery
1 user manual

The product should avoid contact with chemicals like unstable chemical solvents or other organic compounds. Particularly, it should not be stored in or exposed to environments that contain high concentration of ketene, acetone, ethanol, isopropanol, toluene and etc. Please do not use the product in the soil for a long time.

Elitech Technology, Inc.
1551 McCarthy Blvd, Suite 112, Milpitas, CA 95035 USA Tel: (+1)408-844-4070 Sales: sales@elitechus.com Support: support@elitechus.com Website: www.elitechus.com Software Download: elitechus.com/download/software

Elitech (UK) Limited
2 Chandlers Mews, London, E14 8LA UK Tel: (+44)203-645-1002 Sales: sales@elitech.uk.com Support: service@elitech.uk.com Website: www.elitech.uk.com Software Download: elitechonline.co.uk/software

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>